← Product Code [JIX](/productcode/JIX) · K061056

# RANDOX LIQUID PROTEIN CALIBRATOR (K061056)

_Randox Laboratories, Ltd. · JIX · Jul 20, 2006 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K061056

## Device Facts

- **Applicant:** Randox Laboratories, Ltd.
- **Product Code:** [JIX](/productcode/JIX.md)
- **Decision Date:** Jul 20, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1150
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Randox Laboratories Limited Liquid Protein Calibrator (for neat sample assays) is derived from normal human serum obtained from volunteer donors. It has been developed for the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, IgA, IgG, IgM, Prealbumin, Transferrin and Haptoglobin assays (all neat sample assavs). Assignment was performed at Randox Laboratories by immunoturbidimetry with reference to International Reference Material CRM 470. The constituent concentrations of these Calibrators are present at levels 1, 2, 3, 4 and 5. This calibrator can be used on Abbott Spectrum, Abbott aeroset, Abbott Architect ഗ്രാമം Architect i2000sr, Ace Analyser, Bayer Advia 1650, Advia 2400, Advia 1200, Dade Dimension RXL, Dimension AR, Hitachi 704, Hitachi 717, Hitachi 911, Hitachi 917, Hitachi 912, Hitachi 747, Kone progress, Olympus AU800, AU600, AU400, AU2700, AU5400, Selectra vitalab, Synchron CX4 Synchron CX5 Synchron CX7, Synchron LX20, ILAB300, ILAB900, ILAB1800, ILAB600, RX Daytona, RX Imola, Cobas Mira, Cobas Mira S, Cobas Mira plus systems The Randox Laboratories Limited Liquid Protein Calibrators, should only be used by suitably qualified laboratory personnel under appropriate laboratory conditions. For in vitro diagnostic use.

## Device Story

Liquid protein calibrators derived from normal human serum; used to calibrate clinical chemistry and immunoassay systems for specific protein analytes (ASO, C3, C4, CRP, Ferritin, Haptoglobin, IgA, IgG, IgM, Prealbumin, Transferrin). Supplied in 5 levels; assigned values via immunoturbidimetry against CRM 470 reference material. Used by laboratory personnel to ensure accuracy of patient sample testing on various automated analyzers. Benefits include standardized calibration across multiple instrument platforms, ensuring reliable diagnostic results for protein analytes.

## Clinical Evidence

No clinical data. Bench testing only; performance established via assignment of constituent concentrations using immunoturbidimetry against International Reference Material CRM 470.

## Technological Characteristics

Liquid calibrator derived from normal human serum. Multi-level (1-5) concentration format. Principle of operation: immunoturbidimetric assignment against CRM 470. Compatible with various automated clinical chemistry analyzer platforms.

## Regulatory Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k061056

B. Purpose for Submission:
New Devices

C. Measurand:
Protein analytes: ASO, Complement C3, Complement C4, CRP, Ferritin, IgA, IgG, IgM, Prealbumin, Transferrin, Haptoglobin

D. Type of Test:
Calibrators

E. Applicant:
Randox Laboratories Ltd.

F. Proprietary and Established Names:
Randox Liquid Protein Calibrators (For Neat Sample Assays)

G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1150, Calibrator
2. Classification:
Class II
3. Product code:
JIX, Calibrator, Multi-Analyte Mixture
4. Panel:
Chemistry 75

H. Intended Use:
1. Intended use(s):
Liquid Protein Calibrators are intended for in vitro diagnostic use in the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgA, IgG, IgM, Prealbumin, and Transferrin assays on Clinical Chemistry and Immunoassay systems, Abbott Spectrum, Abbott Aeroset, Abbott Architect i2000, Architect i2000sr, Ace analyser, Bayer Advia 1650, Advia 2400, Advia 1200, Dade Dimension RXL, Dimension AR, Hitachi 704, Hitachi 717, Hitachi 911, Hitachi 917, Hitachi 912, Hitachi 747, Kone progress, AU800, AU600, AU400, AU2700, AU5400, Selectra Vitalab, Synchron CX4, Synchron CX5, Synchron CX7, Synchron LX20, ILAB300, ILAB900, ILAB1800, ILAB600, RX Daytona, RX Imola, Cobas Mira, Cobas Mira S, Cobas Mira Plus systems.
2. Indication(s) for use:
The Randox Laboratories Limited Liquid Protein Calibrator (for neat sample assays) is derived from normal human serum obtained from volunteer donors. It has been developed for the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, IgA, IgG, IgM, Prealbumin, Transferrin and Haptoglobin assays (all neat sample assays). Assignment was performed at Randox Laboratories by immunoturbidimetry with reference to International Reference Material CRM 470. The constituent

{1}

concentrations of these Calibrators are present at levels 1, 2, 3, 4 and 5.

This calibrator can be used on Abbott Spectrum, Abbott Aeroset, Abbott Architect i2000 and Architect i2000sr, Ace analyser, Bayer Advia 1650, Advia 2400, Advia 1200, Dade Dimension RXL, Dimension AR, Hitachi 704, Hitachi 717, Hitachi 911, Hitachi 917, Hitachi 912, Hitachi 747, Kone progress, Olympus AU800, AU600, AU400, AU2700, AU5400, Selectra Vitalab, Synchron CX4, Synchron CX5, Synchron CX7, Synchron LX20, ILAB300, ILAB900, ILAB1800, ILAB600, RX Daytona, RX Imola, Cobas Mira, Cobas Mira S, Cobas Mira Plus systems.

The Randox Laboratories Limited Liquid Protein Calibrators, should only be used by suitably qualified laboratory personnel under appropriate laboratory conditions.

For in vitro diagnostic use.

3. Special conditions for use statement(s):

For prescription use only.

4. Special instrument requirements:

Abbott Spectrum, Abbott Aeroset, Abbott Architect i2000 and Architect i2000sr, Ace analyser, Bayer Advia 1650, Advia 2400, Advia 1200, Dade Dimension RXL, Dimension AR, Hitachi 704, Hitachi 717, Hitachi 911, Hitachi 917, Hitachi 912, Hitachi 747, Kone progress, Olympus AU800, AU600, AU400, AU2700, AU5400, Selectra Vitalab, Synchron CX4, Synchron CX5, Synchron CX7, Synchron LX20, ILAB300, ILAB900, ILAB1800, ILAB600, RX Daytona, RX Imola, Cobas Mira, Cobas Mira S, Cobas Mira Plus systems.

I. Device Description:

The liquid protein calibrators are supplied at 5 levels: 1, 2, 3, and 5. Target values are supplied for the following analytes in Calibrator 1, 2, 3, and 5: Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgA, IgG, IgM, Prealbumin, and Transferrin. Target values are supplied for the following analytes in Calibrator 4: ASO, Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgA, IgG, IgM, Prealbumin, and Transferrin.

J. Substantial Equivalence Information:

1. Predicate device name(s):

Bayer Specific Protein Reference Serum

Randox Liquid Protein Calibrator

2. Predicate 510(k) number:

k033791 (Haptoglobin, IgA, IgG, IgM, Transferrin, C3, C4)

k031608 (IgA, IgG, IgM, ASO, Ferritin, Prealbumin, CRP, C3, C4)

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | New Device | Predicate Device  |
|  Calibrator material source | Human serum | Same  |

{2}

|  Differences  |   |   |
| --- | --- | --- |
|  Item | New Device | Predicate  |
|  Indication for Use | The Randox Laboratories Limited Liquid Protein Calibrators (for neat sample assays) is derived from normal human serum obtained from volunteer donors. It has been developed for the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, IgA, IgG, IgM, Prealbumin, Transferrin and Haptoglobin assays (all neat sample assays).

Assignment was performed at Randox Laboratories by immunoturbidimetry with reference to International Reference Material CRM 470. The constituent concentrations of these Calibrators are present at levels 1, 2, 3, 4 and 5.

This calibrator can be used on Abbott Spectrum, Abbott aeroset, Abbott Architect i2000, Abbott Architect i2000sr, Ace analyser, Bayer Advia 1650, Advia 2400, Advia 1200, Dade Dimension RXL, Dimension AR, Hitachi 704, Hitachi 717, Hitachi 911, Hitachi 917, Hitachi 912, Hitachi 747, Kone progress, Olympus AU800, AU600, AU400, AU2700, AU5400, Selectra Vitalab, | Same Indication for Use, except for the equipment listings:
(1) Bayer Predicate Device: For calibration of the ADVIA® Chemistry Systems, Technicon RA-500®, Technicon RA-1000®, Technicon® RA-XT™, Technicon RA-2000®, and opeRA® chemistry system specific protein methods.
(2) Randox Predicate Device: For Bayer ADVIA 1650 analyzer only.  |

{3}

|  Differences  |   |   |
| --- | --- | --- |
|  Item | New Device | Predicate  |
|   | Synchron CX4, Synchron CX5, Synchron CX7, Synchron LX20, ILAB300, ILAB900, ILAB1800, ILAB600, RX Daytona, RX Imola, Cobas Mira, Cobas Mira S, Cobas Mira Plus systems. |   |
|   |  The Randox Laboratories Limited Liquid Protein Calibrators, should only be used by suitably qualified laboratory personnel under appropriate laboratory conditions.
For in vitro diagnostic use.  |   |
|  Calibrator level(s) | One level value for ASO and five level values for C3, C4, CRP, Ferritin, Haptoglobin, IgA, IgG, IgM, Prealbumin, and Transferrin | The Bayer Predicate Device had one level value for C3, C4, IgA, IgG, IgM, Transferrin, Haptoglobin.
The Randox Predicate Device had one level value for ASO, CRP, Ferritin, C3, C4, Prealbumin and five level values for IgA, IgG, IgM, Transferrin  |

K. Standard/Guidance Document Referenced (if applicable):
None referenced.

L. Test Principle:
Not Applicable

M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not Applicable
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability of the calibrators (except for ASO and Ferritin) was via the

{4}

reference material BCR Reference Material Proteins, CRM 470; NIBSC 94-572. The reference materials for ASO and Ferritin were ‘1st IRP AST91’ and ‘NIBSC 94/572’ respectively.

## Stability

The expiration date claim was 12 months.

Open vial stability - Data included control recovery from day 0, 7, 14, 21, 30. The results ranged from - 0.8% to 4.3% which were within the acceptable criteria of ± 5% versus fresh calibrators. On day 0 of the stability study, analytes are calibrated using a freshly opened vial calibrator set. (After the required volume is removed, vials are re-capped and stored at 2-8°C). Data are recorded. The controls are then tested and controls must recover within their specified insert ranges for each analyte. On each subsequent time-points, the analytes are calibrated using previously stored open vial calibrators. Control recovery for those calibrated on open calibrator set must be within 5% to the recoveries of the control when calibrated on the freshly opened calibrator set.

Real time stability - Data included testing from day 0, 13, and 27 days at 37°C compared to material stored at recommended temperature. The results were within the acceptance criteria of ≤ 5%. On day 0 of the real time stability study, sufficient volume of the calibrator sets (enough to cover all time points) are placed at three storage temperatures (+30°, 2-8°, -80 °C). The +30 °C stability is only assessed until week 8. At 4 and 8 week time-points a set of calibrators from each of the three stored temperatures is tested. Control recoveries are run and all control material must recover within the analyte specific ranges. After week 8, only the 2-8° and -80 °C calibrator sets are tested at weeks 8, 26, 30, 52, 56, 78, 90, 104, 108.

d. Detection limit:
Not Applicable.

e. Analytical specificity:
No Applicable.

f. Assay cut-off:
Not Applicable.

2. Comparison studies:
a. Method comparison with predicate device:
Not Applicable
b. Matrix comparison:
Not Applicable.

3. Clinical studies:
a. Clinical sensitivity and specificity:
Not Applicable
b. Other clinical supportive data (when a is not applicable):
Not Applicable.

4. Clinical cut-off:
Not applicable

{5}

5. Expected values/Reference range:
Not Applicable

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

6

---

**Source:** [https://fda-staging.innolitics.com/device/K061056](https://fda-staging.innolitics.com/device/K061056)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
