LINGLOCK

K061042 · Sybron Dental Specialties · DYW · Jan 19, 2007 · Dental

Device Facts

Record IDK061042
Device NameLINGLOCK
ApplicantSybron Dental Specialties
Product CodeDYW · Dental
Decision DateJan 19, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5470
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use of LingLock is to stabilize the lower dental arch following orthodontic treatment. The LingLock retainer enables patients to floss their teeth interdentally in a regular manner.

Device Story

LingLock is an interlock retainer system for post-orthodontic stabilization of the lower dental arch. The device consists of male and female polycrystalline alumina (aluminum oxide) ceramic attachments bonded to the lingual surfaces of the lower six anterior teeth. An application tool, comprising a polypropylene holder and stainless steel guide strip, ensures precise positioning of the attachments across tooth contact points. Orthodontic adhesive is applied to the ceramic components, light-cured, and the application tool is removed. The engaged ceramic attachments prevent tooth shifting while maintaining an interdental gap to facilitate regular flossing. The device is intended for use by dental professionals in a clinical setting.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on technological characteristics and intended use comparisons to legally marketed predicate devices.

Technological Characteristics

Material: Polycrystalline alumina (aluminum oxide) attachments; polypropylene attachment holder; stainless steel guide strip. Principle: Mechanical interlock via bonded ceramic attachments. Form factor: Male/female ceramic components for lingual tooth surface. Connectivity: None. Sterilization: Not specified.

Indications for Use

Indicated for patients requiring stabilization of the lower dental arch following orthodontic treatment. The device is bonded to the lingual surface of the lower six anterior teeth.

Regulatory Classification

Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a logo for Sybron Dental Specialties. The logo features the letters 'sds' in a stylized font, with a curved line above the letters. Below the letters, the words 'SYBRON DENTAL SPECIALTIES' are written in a simple, sans-serif font. The logo is black and white. K061042 ## Section III - 510(k) Summary of Safety and Effectiveness ## Submitter: JAN 1 9 2007 Sybron Dental Specialties, Inc. 100 Bayview Circle, Suite 6000 Newport Beach, California 92660 (949) 255-8766 - Phone (949) 255-8763 - Facsimile Colleen Boswell - Contact Person Date Summary Prepared: April 2006 #### Device Name: - Trade Name LingLock . - Common Name Lingual Orthodontic Retainer . - Classification Name Retainer, Screw Expansion, Orthodontic, per 21 CFR § 872.5410 . #### Devices for Which Substantial Equivalence is Claimed: - . Ormco Corporation, Bondable Lingual Retainer - 3M Unitek, Clarity Ceramic Brackets . #### Device Description: The device is an interlock retainer composed of polycrystalline alumina (aluminum oxide) that is bonded to the lingual surface of the lower anterior teeth and is intended to stabilize the lower dental arch following orthodontic treatment. LingLock is comprised of a set of ceramic attachments (male/female) bonded to the teeth across the contact points of the lower six anterior front teeth; the male component on one side and the female component on the other. When engaged, the ceramic attachments prevent the teeth from shifting out of position, yet have a slight gap which enables patients to floss their teeth interdentally in a regular manner. The LingLock retainer has an application tool consisting of a ceramic attachment holder (polypropylene) and a guide strip (stainless steel) which ensures an exact positioning of the ceramic attachments both in relation to each other and the teeth to be retained. An orthodontic adhesive is placed on the bonding surface of the ceramic attachments and the guide strip is guided in between two neighbouring teeth. The ceramic attachment holder is brought in contact with the enamel at the lingual surface of the teeth and the orthodontic adhesive is light cured. After curing, the guide strip is removed by pulling it forward in a slight upward rotational movement. The ceramic attachment holder is then removed. {1}------------------------------------------------ ## Intended Use of the Device: The intended use of LingLock is to stabilize the lower dental arch following orthodontic treatment. The LingLock retainer enables patients to floss their teeth interdentally in a regular manner. ## Substantial Equivalence: LingLock is substantially equivalent to other legally marketed devices in the United States. Linglock functions in a manner similar to and is intended for the same use as the Bondable Lingual Retainer currently manufactured by Ormco Corporation and is composed of the same material and has the same bonding surface as the ceramic bracket manufactured by 3M Unitek. {2}------------------------------------------------ Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Colleen Boswell Vice President, Regulatory Affairs Sybron Dental Specialist, Incorporated 100 Bayview Circle, Suite 6000 Newport Branch, California 92660-8915 JAN 1 9 2007 Rc: K061042 Trade/Device Name: LingLock Regulation Number: 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: II Product Code: DYW Dated: December 29, 2006 Received: January 4, 2007 Dear Ms. Boswell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Ms. Boswell Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Suen Kumno Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use Koblo 42 510(k) Number (if known): Device Name: LingLock Indications For Use: LingLock is an interlock retainer composed of polycrystalline alumina that is bonded to the lingual surface of the lower anterior teeth and is intended to stabilize the lower dental arch following orthodontic treatment. The LingLock retainer enables patients to floss their teeth interdentally in a regular manner. Suer Uunm 1 CE SB XOL 4 Coquor, Dental De K061042 Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...