PDS PLUS ANTIBACTERIAL SUTURE

K061037 · ETHICON, Inc. · NEW · Jul 14, 2006 · General, Plastic Surgery

Device Facts

Record IDK061037
Device NamePDS PLUS ANTIBACTERIAL SUTURE
ApplicantETHICON, Inc.
Product CodeNEW · General, Plastic Surgery
Decision DateJul 14, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4840
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

PDS* Plus Antibacterial monofilament synthetic absorbable sutures are indicated for use in soft tissue approximation, including use in pediatric cardiovascular tissue where growth is expected to occur and ophthalmic surgery (other than contact with cornea and sclera). PDS Plus is not indicated in adult cardiovascular tissue, microsurgery, and neural tissue. These sutures are particularly useful where the combination of an absorbable suture and extended wound support (up to six weeks) is desirable.

Device Story

PDS* Plus is a sterile, monofilament synthetic absorbable suture composed of poly(p-dioxanone). It incorporates Irgacare MP (triclosan), a broad-spectrum antibacterial agent, at concentrations up to 2360 ug/m. Used by surgeons for soft tissue approximation; provides extended wound support for up to six weeks. The device is absorbed by the body over time. It is intended for clinical use in surgical settings.

Clinical Evidence

Bench testing only. Non-clinical laboratory testing demonstrated conformance to USP Monograph for absorbable surgical sutures. In-vivo and in-vitro testing confirmed the device performed as intended.

Technological Characteristics

Monofilament synthetic absorbable suture; material: poly(p-dioxanone); antibacterial agent: Irgacare MP (triclosan) <= 2360 ug/m; sterile; conforms to USP Monograph for absorbable surgical sutures.

Indications for Use

Indicated for soft tissue approximation in pediatric cardiovascular tissue (where growth expected) and ophthalmic surgery (excluding cornea/sclera). Contraindicated for adult cardiovascular tissue, microsurgery, and neural tissue.

Regulatory Classification

Identification

An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.

Special Controls

*Classification.* Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K061037 page 1 of 2 ## Summary of Safety and Effectiveness Information supporting claims of substantial equivalence, as defined under the Federal Food, Drug and Cosmetic Act, respecting safety and effectiveness is summarized below. #### Submitted by: Brvan A. Lisa Sr. RA Associate, Regulatory Affairs Ethicon, Inc., A Johnson & Johnson Company Route 22 West, PO Box 151 Somerville, NJ 08876 ### Name/Classification of Device: Class II in 21 CFR § 878.4493, Absorbable poly(glycolide/l-lactide) surgical suture (GAM) #### Trade Name: PDS* Plus (Polydioxanone) Antibacterial Suture #### Predicate Devices: PDS II* (Polydioxanone) Suture (N18331) VICRYL* Plus Antibacterial Suture (K032420) MONOCRYL* Plus Antibacterial Suture (K050845) #### Statement of Intended Use: PDS* Plus Antibacterial monofilament synthetic absorbable sutures are indicated for use in soft tissue approximation, including use in pediatric cardiovascular tissue where growth is expected to occur and ophthalmic surgery (other than contact with cornea and sclera). PDS Plus is not indicated in adult cardiovascular tissue, microsurgery, and neural tissue. These sutures are particularly useful where the combination of an absorbable suture and extended wound support (up to six weeks) is desirable. #### Device Description: PDS* Plus (Polydioxanone) Antibacterial monofilament synthetic absorbable suture is prepared from a polyester, poly (p-dioxanone). The suture is available dyed (D&C Violet No. 2) or undyed (natural). The suture contains Irgacare MP** (triclosan), a broad-spectrum antibacterial agent, at no more than 2360 ug/m." {1}------------------------------------------------ ## Summary of Technological Characteristics of New Device to Predicate Devic The modified device has similar technological characteristics as the predicate devices. Like currently marketed PDS* II Suture, it is a sterile, monofilament synthetic absorbable suture that conforms to the USP Monograph for absorbable surgical sutures, except for diameter. Like the currently marketed Coated VICRYL Plus Antibacterial suture and MONOCRYL Plus Antibacterial suture, the modified device contains Irgacare** MP, an antibacterial agent. #### Performance Data: Non-clinical laboratory testing was performed demonstrating that the device conformed to the USP Monograph for absorbable surgical sutures. Additionally, in-vivo/in-vitro testing was provided showing that the device performed as intended and as claimed. #### Conclusions: Based on the 510(k) summaries and 510(k) statements (21 CFR 807) and the information provided herein, we conclude that the modified device is substantially equivalent to the Predicate Devices under the Federal Food, Drug, and Cosmetic Act. * Trademark of Ethicon, Inc. **Trademark of Ciba Specialty Chemicals Corporation {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ethicon, Inc., a Johnson & Johnson Co. % Mr. Bryan A. Lisa Senior Regulatory Affairs Associate Route 22 West, P.O. Box 151 Somerville, New Jersey 08876 JUL 1 4 2006 Re: K061037 Trade/Device Name: PDS* Plus (Polydioxanone) Antibacterial Suture Regulation Number: 21 CFR 878.4840 Regulation Name: Absorbable polydioxanone surgical suture Regulatory Class: II Product Code: NEW Dated: July 5, 2006 Received: July 6, 2006 Dear Mr. Lisa: We have reviewed your Section 510(k) premarket notification of intent to market the device w f nave a showe and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can may of Saily of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 -- Mr. Bryan A. Lisa This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chubary Buckup Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page 1 of 1 ## Indications for Use # 510(k) Number (if known): Device Name: PDS* Plus (Polydioxanone) Antibacterial Suture Indications for Use: PDS* Plus Antibacterial monofilament synthetic absorbable sutures are indicated for use in soft tissue approximation, including use in pediatric cardiovascular tissue where growth is expected to occur and ophthalmic surgery (other than contact with cornea and sclera). PDS Plus is not indicated in adult cardiovascular tissue, microsurgery, and neural tissue. These sutures are particularly useful where the combination of an absorbable suture and extended wound support (up to six weeks) is desirable. Prescription Use X__ _ _ _ _ _ _ _ _ _ _ AND/OR _ Over-The-Counter Use __ _ _ _ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Electric Bulletin Division of Gen eral. Restorative. and Neurological Devices 510(k) Number K(661053)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...