K061022 · Science For Biomaterials · MQV · May 24, 2006 · Orthopedic
Device Facts
Record ID
K061022
Device Name
BIO 1-KIT BIOSORB FILLED CARTRIDGE
Applicant
Science For Biomaterials
Product Code
MQV · Orthopedic
Decision Date
May 24, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
BIOSORB ® Resorbable Void Filler is a resorbable calcium salt bone void filler intended to fill bony voids or gaps of the skeletal system (i.e. the extremities, spine and pelvis,) caused by trauma or surgery, that are not intrinsic to the stability of the bony structure. BIOSORB ® Resorbable Void Filler does not possess sufficient mechanical strength to support reduction of a defect prior to soft and hard tissue ingrowth. Rigid fixation techniques are recommended as needed to assure stabilization of the defect in all plans. The BIO 1-KIT ® BIOSORB® filled cartridge is intended for use as a piston syringe system for the aspiration of autogenous blood and/or bone marrow. This cartridge provides the surgeon with a convenient way to mix autologous blood or bone marrow with the BIOSORB ® Resorbable Void Filler and deliver the material to the orthopedic surgical site.
Device Story
BIO 1-KIT is a sterile, single-use orthopedic delivery system comprising a pre-filled syringe of synthetic beta-tricalcium phosphate (β-TCP) porous ceramic pellets, a connector, and a bone marrow aspiration needle. Used by surgeons in clinical settings to aspirate autogenous blood or bone marrow, mix it with the β-TCP pellets within the syringe under aseptic conditions, and deliver the resulting composite to bone defects. The β-TCP implant acts as an osseo-conductive scaffold, resorbing over 6-12 months as it is replaced by host bone and soft tissue. The system minimizes air exposure during mixing, preserving cell and protein viability in the marrow. It provides a convenient, closed-system method for graft preparation and site delivery, facilitating bone defect repair.
Clinical Evidence
Clinical evidence for the BIOSORB bone void filler component was established via human clinical trials under K021963, demonstrating biocompatibility, osseointegration, and resorption/replacement by osseous tissue. Bench testing for the BIO 1-KIT cartridge included infra-red spectrometry to assess polypropylene syringe compatibility and particle release, alongside functional testing of aspiration and impregnation performance.
Technological Characteristics
Materials: Synthetic beta-tricalcium phosphate (β-TCP) pellets (ASTM F 1088-2004) and polypropylene syringe. Structure: Macroporous (45% and 70% porosity) interconnected structure. Form factor: Syringe-based delivery system with connector and aspiration needle. Sterilization: Sterile, single-use. Connectivity: None.
Indications for Use
Indicated for filling bony voids or gaps in the skeletal system (extremities, spine, pelvis) caused by trauma or surgery that are not intrinsic to structural stability. Contraindicated for use where mechanical strength is required to support defect reduction prior to tissue ingrowth.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
JU(K) Fremarket Notification
BIO 1-KIT ® BIOSORB® filled cartridge
CIENCE FOR ATERIALS
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
### 1. GENERAL INFORMATION
MAY 2 4 2006
| Trade Name | BIO 1-KIT ® BIOSORB® filled cartridge |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Bone void filler and bone graft delivery syringe |
| Classification Name | Resorbable Calcium Salt Bone Void Filler device |
| Class | II |
| Product Code | MQV |
| CFR section | 21CFR 888.3045 |
| Device panel | Orthopedic |
| Legally marketed predicate devices | BIOSORB® Resorbable Void Filler K021963 (SBM)<br>VITOSS-Filled cartridge K032130 (Orthovita Inc) |
| Submitter | SCIENCE FOR BIOMATERIALS<br>Sciences et Bio Matériaux<br>ZI du Monge<br>F 65100 LOURDES - FRANCE<br>Owner operation Number : 9063735<br>Establishment Registration Number:3004549189 |
| Contact | Denis CLEMENT, General Manager<br>Phone : +33 (0)5 62 42 21 01<br>Fax : +33 (0)5 62 42 21 00<br>e-mail : denis.clement@sgm-fr.com<br>Regulatory contact: Idée Consulting (FRANCE)<br>Isabelle DRUBAIX e-mail : idrubaix@nordnet.fr |
# 2. DEVICE DESCRIPTION
BIOSORB ® bone void filler (BIO1 range of products) (K021963) is an osseo-conductive macroporous implant made of synthetic beta tri Calcium Phosphate (βTCP) indicated for Bone Void Filling.
BIOSORB ® bone void filler presents a multidirectional interconnected porosity structure (45% and 70%), similar to that of the human cancellous bone. BIOSORB ® bone void filler implant slowly resorbs during the remodeling and bone defect repair process and is progressively replaced with bone and soft tissues. The progressive resorption within 6-12 months prevents premature resorption and failed filling.
The impregnation of bone substitute with marrow is largely practiced by the surgeons. Usually this operation is done by immersion of the bone substitute in a container containing the osseous marrow. In order to facilitate this operation and to limit the contact of the mixture with the ambient air at the origin of the degradation of the cells
00000 7
{1}------------------------------------------------
I BIOSORB filled
and proteins contained in osseous marrow, S.B.M has developed a system allowing the and procession of autologous osseous marrow, the impregnation of the calcium phosphate pellets and the impaction of the mixture in the cavity to be filled. This system allows the impregnation of porous pellets with osseous marrow under aseptic conditions.
BIO 1-Kit® comprises :
- a pre-filled syringe of BIOSORB® (β-ΤCP porous ceramic pellets) A
- a connector with an adapter and a needle 公
- a bone marrow aspiration needle A
BIO 1-Kit® is delivered sterile and intended for a single use.
BIO 1-Kit® is available either with BIOSORB® granules (1.5 and 3.0 mm in diameter) or BIOSORB® blocks (4x4x4 mm) and in quantities ranging from 5 cc to 45 cc.
### 3. INTENDED USE
BIOSORB ® Resorbable Void Filler is a resorbable calcium salt bone void filler intended to fill bony voids or gaps of the skeletal system (i.e. the extremities, spine and pelvis,) caused by trauma or surgery, that are not intrinsic to the stability of the bony structure.
BIOSORB ® Resorbable Void Filler does not possess sufficient mechanical strength to support reduction of a defect prior to soft and hard tissue ingrowth. Rigid fixation techniques are recommended as needed to assure stabilization of the defect in all plans.
The BIO 1-KIT ® BIOSORB® filled cartridge is intended for use as a piston syringe system for the aspiration of autogenous blood and/or bone marrow. This cartridge provides the surgeon with a convenient way to mix autologous blood or bone marrow with the BIOSORB ® Resorbable Void Filler and deliver the material to the orthopedic surgical site.
### 4. PERFORMANCE DATA
BIOSORB® (K021963) conforms to the recognized consensus standard specification ASTM F 1088-2004 Standard Specification for Beta-Tricalcium Phosphate for Surgical Implantation. FDA has recognized the use of this consensus standard as verification of material characteristics and biocompatibility for surgical application (Recognition List Number: 011 Effective Date: 09/01/2004).
Moreover, the biocompatibility of βTCP implants is well documented. As a biomaterial β TCP has consistently proven to be non toxic, non allergenic, biocompatible and elicits no inflammation. No adverse system effects have been reported.
A wide variety of tests (cytotoxicity, sensitization, genotoxicity and mutagenicity, systemic toxicity, irritation, systemic tolerance, Pyrogenicity) was performed on
# 000008
{2}------------------------------------------------
BIOSORB® (refer to K021963). Testing performed on BIOSORB" shows an excellent biocompatibility with no significant adverse observations of any kind.
A human clinical trial (K021963) has been performed to investigate the safety and effectiveness of BIOSORB ® bone void filler and more especially the achievement of a stable osseous fusion in consolidation of bone defects, the absence of inflammatory or septic response and the resorption of the BIOSORB® implant and its replacement by osseous tissues. Clinical and radiological data demonstrate the biocompatibility, the osseo integration and the resorption of BIOSORB ® bone void filler.
In addition the following tests were performed on the BIO 1-KIT ® BIOSORB® filled cartridge:
- A The compatibility and the absence of particles release from the polypropylene syringe has been assessed by infra-red spectrometry
- A Testing of aspiration and impregnation has been performed
# 5. SUBSTANTIAL EQUIVALENCE
BIO 1-KIT ® BIOSORB® filled cartridge is substantially equivalent to its predicate device VITOSS-Filled cartridge K032130 (Orthovita Inc) in terms of intended use, material, design and function. Any minor differences between these two devices do not raise new questions of safety and effectiveness.
Summary preparation date: March 31, 2006
0000 ్లో ప్రారా
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the Department of Health and Human Services. The logo features a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH AS." are arranged in a semi-circle around the left side of the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 4 2006
Science for Bio Materials % Mr. Denis Clement General Manager ZI du Monge F 65100 Lourdes - France
Re: K061022
Trade/Device Name: BIO 1-KIT® BIOSORB® filled cartridge Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: April 7, 2006 Received: April 24, 2006
Dear Mr. Clement:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact-the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Hebert Leemans
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and .Radiological Health
Enclosure
{5}------------------------------------------------
Page 3 - Mr. Denis Clement
# Indications for Use
510(k) Number (if known):
Device Name:
Indications For Use:
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of
cc: HFZ-401 DMC
:
{6}------------------------------------------------
### 061022 510(k) Number (if known):
Device Name: BIO 1-KIT ® BIOSORB® filled cartridge
## Indications for Use:
BIOSORB ® Resorbable Void Filler is a resorbable calcium salt bone void filler intended to fill bony voids or gaps of the skeletal system (i.e. the extremities, spine and pelvis,) caused by trauma or surgery, that are not intrinsic to the stability of the bony structure.
BIOSORB ® Resorbable Void Filler does not possess sufficient mechanical strength to support reduction of a defect prior to soft and hard tissue ingrowth. Rigid fixation techniques are recommended as needed to assure stabilization of the defect in all plans.
The BIO 1-KIT ® BIOSORB® filled cartridge is intended for use as a piston syringe system for the aspiration of autogenous blood and/or bone marrow. This cartridge provides the surgeon with a convenient way to mix autologous blood or bone marrow with the BIOSORB ® Resorbable Void Filler and deliver the material to the orthopedic surgical site.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**Division Sign-Off**
Division of General, Restorative,
and Neurological Devices
510(k) Number K061022 000002
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.