← Product Code [DZE](/productcode/DZE) · K061003

# SFB & CFB IMPLANTS (K061003)

_Nobel Biocare AB · DZE · Jun 30, 2006 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K061003

## Device Facts

- **Applicant:** Nobel Biocare AB
- **Product Code:** [DZE](/productcode/DZE.md)
- **Decision Date:** Jun 30, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3640
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Nobel Biocare's SFB and CFB implants are endosseous implant intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. Nobel Biocare's SFB and CFB implants are indicated for single or multiple unit restorations in splinted or non-splinted applications. Nobel Biocare's SFB and CFB implants may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied.

## Device Story

SFB and CFB implants are threaded, root-form dental implants; manufactured from commercially pure titanium; feature TiUnite surface treatment and Groovy groove design. Used by dental surgeons in clinical settings for implantation into upper or lower jaw bone. Supports prosthetic devices (artificial teeth) to restore esthetics and chewing function. Placed via single or two-stage surgical procedures; supports immediate loading if initial stability criteria are met. Output is a stable foundation for dental prosthetics, improving patient oral function.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Commercially pure titanium; threaded root-form design; TiUnite surface treatment; Groovy groove design feature. Non-powered, mechanical device.

## Regulatory Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

## Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

## Predicate Devices

- TiUnite Implants ([K050705](/device/K050705.md))
- Groovy Implants ([K050258](/device/K050258.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K661003

## 1.4 510(k) Summary of Safety and Effectiveness

JUN 300 2006

| Submitted by:                       | Herbert Crane<br>Director Regulatory Affairs                               |
|-------------------------------------|----------------------------------------------------------------------------|
| Address:                            | Nobel Biocare USA LLC<br>22715 Savi Ranch Parkway<br>Yorba Linda, CA 92887 |
| Telephone:                          | (714) 282-4800, ext. 5074                                                  |
| Facsimile:                          | (714) 282-9023                                                             |
| Date of Submission:                 | April 10, 2006                                                             |
| Classification Name:                | Endosseous Dental Implant (21 CFR 872.3640)                                |
| Trade or Proprietary or Model Name: | SFB & CFB Implants                                                         |
| Legally Marketed Device(s):         | TiUnite Implants (K050705)<br>Groovy Implants (K050258)                    |

## Device Description:

Nobel Biocare TiUnite® Implants are threaded, root-form dental implants intended for use in the upper and/or lower jaw to support prosthetic devices, such as artificial teeth, in order to restore patient esthetics and chewing function to partially or fully edentulous patients. The SFB and CFB implants are manufactured from commercially pure titanium. The implants utilize the TiUnite surface treatment and have the Groovy groove design feature.

Nobel Biocare SFB and CFB implants may be placed in the oral cavity using either a single stage surgical procedure or a two stage surgical procedure. If a single stage procedure is used, the implants may be immediately loaded following insertion where good initial stability can be obtained.

## Indications for Use:

Nobel Biocare's SFB and CFB implants are endosseous implant intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. Nobel Biocare's SFB and CFB implants are indicated for single or multiple unit restorations in splinted applications. Nobel Biocare's SFB and CFB implants may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## JUN 3 0 2006

Nobel Biocare AB C/O Mr. Herbert Crane Director of Regulatory Affairs Nobel Biocare USA. LLC 22715 Savi Ranch Parkway Yorba Linda, California 92887

Rc: K061003

Trade/Device Name: SFB & CFB Implants Regulation Number: 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: April 10, 2006 Received: April 11, 2006

Dear Mr. Crane:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice. labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Mr. Crane

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also. please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on vour responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours

Silon Kuno

Chiu-Li, Ph.D.

Chiu Lin. Ph.D. Child Liller Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

K061003 510(k) Number (if known):

Device Name: SFB & CFB Implants

Indications For Use:

Nobel Biocare's SFB and CFB implants are endosseous implant intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. Nobel Biocare's SFB and CFB implants are indicated for single or multiple unit restorations in splinted or non-splinted applications. Nobel Biocare's SFB and CFB implants may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied.

× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Shashi Panwar

!

of Anasthesiology, General Hospital in Contici. Dental Devices

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K061003](https://fda-staging.innolitics.com/device/K061003)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).