← Product Code [MTJ](/productcode/MTJ) · K060987

# OSCERA7 SYNTHETIC ABSORBABLE BONE WAX, MODEL OS31 (K060987)

_Cp Medical · MTJ · Nov 2, 2006 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K060987

## Device Facts

- **Applicant:** Cp Medical
- **Product Code:** [MTJ](/productcode/MTJ.md)
- **Decision Date:** Nov 2, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Indications for Use

The Oscera7™ Synthetic Absorbable Bone Wax is indicated in the control of bleeding from cut or damaged bones by acting as a pressure tamponade, or mechanical barrier .

## Device Story

Oscera7™ Synthetic Absorbable Bone Wax is a kneadable, biocompatible material used to achieve hemostasis in surgical settings. It functions as a mechanical barrier (tamponade) when pressed into the pores of cut or damaged bone. The material is synthetic and absorbable, composed of a fatty acid salt, isopropyl hexadecanoate, an alkylene oxide copolymer, and β-tricalcium phosphate. Post-application, these components are dispersed and absorbed by the body. It is intended for use by clinicians to manage bone bleeding.

## Clinical Evidence

Verification testing included bench and animal studies. Testing confirmed hemostatic capability, bioresorbability, and biocompatibility, demonstrating that design inputs were achieved and the device is safe and effective.

## Technological Characteristics

Kneadable, synthetic, absorbable material. Composition: fatty acid salt, isopropyl hexadecanoate, alkylene oxide copolymer, and β-tricalcium phosphate. Mechanical tamponade principle of operation.

## Predicate Devices

- CP Medical Bone Wax ([K024372](/device/K024372.md))
- Ethicon Bone Wax (Pre-amendment)
- Orthostat Bone Putty ([K043260](/device/K043260.md))
- Autosuture Bone Wax ([K971680](/device/K971680.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K060987 page 1/2

Image /page/0/Picture/1 description: The image shows the logo for CP Medical. The text "CP Medical" is displayed in a bold, outlined font. Below the company name, it says "Portland, OR USA" in a smaller font.

NOV - 2 2006

## 510(k) SUMMARY OF SAFETY & EFFECTIVENESS

Device Name: Oscera7TM Synthetic Absorbable Bone Wax

Device Model Number: Classification Name:

Device Classification:

Predicate Devices:

Bone Wax (MTJ)

Unclassified

CP Medical 803 NE 25th Ave.

OS31

CP Medical Bone Wax, K024372 Ethicon Bone Wax, (Pre-amendment) Orthostat Bone Putty, K043260 Autosuture Bone Wax, K971680

Manufacturer:

Establishment Registration Number:

Official Contact:

3032563 Betsy Cortelloni Regulatory Affairs Manager Theragenics Corporation® 5203 Bristol Industrial Way

Portland, OR 97232 USA

Buford, GA 30518 Phone: 770-271-0233 Fax: 770-831-4369

Intended Use: The Oscera7™ Synthetic Absorbable Bone Wax is indicated in the control of bleeding from cut or damaged bones by acting as a pressure tamponade, or mechanical barrier .

Device Description: The Oscera7™ Synthetic Absorbable Bone Wax is a kneadable, biocompatible, synthetic absorbable material indicated for control of bleeding from bone surfaces. Hemostasis is accomplished by the physical mechanism of tamponade as it is pressed into the pores of the cut or damaged bone surface.

Post-application, the component materials, which include a fatty acid salt, isopropyl hexadecanoate, an alkylene oxide copolymer and ß-tricalcium phosphate, are dispersed and absorbed within the body.

Substantial Equivalence Comparison: Oscera7™ has the same intended use control of bleeding - as each of the predicates listed. Oscera7™ is absorbable, as 

{1}------------------------------------------------

are the Orthostat and Autosuture bone waxes. All of the materials used in the Oscera7™ formulation have been used individually in the predicate products.

Design Verification: Verification testing of Oscera7™ included both bench and animal testing. The testing demonstrated the hemostatic capability, bioresorbability, and biocompatibility, thereby establishing compliance to the functional requirements of the device.

Conclusion: The results of verification testing confirmed that design inputs were achieved and the cumulative test results demonstrated the functionality, safety and effectiveness of Oscera7™, as well as its substantial equivalence to the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows a circular logo. The logo contains the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" arranged in a circular fashion around an emblem. The emblem is a stylized depiction of an eagle or bird-like figure, with three curved lines representing the body and wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

CP Medical % Theragenics Corporation Ms. Betsy Cortelloni Regulatory Affairs Manager 5203 Bristol Industrial Way Buford. Georgia 30518

NOV - 2 2006

Re: K060987

Trade/Device Name: Oscera7" Synthetic Absorbable Bone Wax Product Code: MTJ Dated: September 20, 2006 Received: September 21, 2006

Dear Ms. Cortelloni:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have becn reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally 

{3}------------------------------------------------

Page 2 - Ms. Betsy Cortelloni

marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## INDICATIONS FOR USE FORM

Page 1 of

510(K) number (if known): K060987

Device Name: Oscera7™ Synthetic Absorbable Bone Wax

Indications for Use:

The Oscera7™ Synthetic Absorbable Bone Wax is indicated in the control of bleeding from cut or damaged bones by acting as a pressure tamponade, or mechanical barrier .

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Division of General. Restorative, and Neurological

---

**Source:** [https://fda-staging.innolitics.com/device/K060987](https://fda-staging.innolitics.com/device/K060987)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).