K060971 · Omi Manufacturing Pty., Ltd. · MEG · Oct 24, 2006 · General Hospital
Device Facts
Record ID
K060971
Device Name
OMI RETRACTABLE SAFTEY SYRINGE, VARIOUS
Applicant
Omi Manufacturing Pty., Ltd.
Product Code
MEG · General Hospital
Decision Date
Oct 24, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OMI Safety Insulin Syringes are intended for the subcutaneous injection of insulin. In addition, the syringe is designed to aid in prevention of needlestick injuries. This device is a safety hypodermic syringe for intramuscular and subcutaneous injection of medication to patients. This device aids in prevention of needlestick injuries. EO sterilized, non-pyrogenic, and latex free. The function of the OMI Retractable Safety Syringe is to provide a safe, accurate and reliable method of injecting medication into a patient.
Device Story
OMI Retractable Safety Syringe is a sterile, single-use, latex-free, non-pyrogenic piston syringe with an integral needle. Available in 1ml, 3ml, and 10ml sizes. Operates like a conventional hypodermic syringe; features a user-activated spring-loaded mechanism that retracts the contaminated needle into the plunger immediately following injection. Used by healthcare professionals for subcutaneous and intramuscular medication administration. Retraction mechanism serves to contain the needle, reducing risk of accidental needlestick injuries to the operator.
Clinical Evidence
Bench testing only. Compliance with ISO 6009, ISO 7864, ISO 7886-1, ISO 7886-4, and ISO 10993 standards demonstrated performance and safety.
Technological Characteristics
Piston syringe with integral needle; spring-loaded retraction mechanism. Materials: Latex-free. Sterilization: Ethylene Oxide (ISO 11135). Standards: ISO 6009 (color coding), ISO 7864 (needles), ISO 7886-1/4 (syringes), ISO 10993 (biocompatibility).
Indications for Use
Indicated for intramuscular and subcutaneous injection of medication (including insulin) to patients. Aids in the prevention of needlestick injuries.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
K060971
001 2 4 2006
### EXHIBIT 2 510(k) Summary
- 1. Submitter Name: OMI Manufacturing Pty Ltd
- 2. Address: 1/12 Booran Dr.
Slacks Creek, Queensland,
- Australia 4127
- 3. Phone:
- 4. Fax: 61 7 32094765
- 5. Email: projects(@omiltd.com
- 6. Contact: Graham McNicol Project Manager
61 7 34517000
- 7. Establishment Registration Number: 3003462738
- 8. Date summary prepared: March 1, 2006
- 9. Device Trade or Proprietary Name: OMI Retractable Safety Syringe
- 10. Device Common or usual name: Safety Syringe
- 11. Device Classification Name: Piston Syringe/Syringe Anti-stick
- 12. Device Classification Code: MEG
- 13. Device Class: class 11 as per regulation 21 CFR 880.5860
- 14. Compliance (performance Standard): None established under section 513-514 Complies with voluntary standards and FDA guidance documents: ISO6009: Hypodermic needles for single use color coding for identification. ISO7864 Sterile hypodermic needles for single use
-
- ISO7886-1 Sterile Hypodermic syringes for single use
- ISO7886-4 Syringes with reuse Prevention Features
- ISO10993: Biological evaluation of medical devices
- EN550/ISO11135 Sterilization of Medical Devices- Validation and routine control Of Ethylene Oxide Sterilization
- 15: Predicate Device: Vanishpoint Syringe (K946219) from Retractable Technologies Inc.
- 16: Device Description: The OMI Retractable Safety Syringe is a Sterile, Non Toxic, Non Pyrogenic, Latex Free, Single Use, automatically activated anti-stick piston syringe with integral needle. Available in 1ml, 3ml and 10ml sizes with various needle sizes.
The OMI Retractable Safety Syringe works like a conventional hypodermic syringe except for its ability to retract the contaminated needle inside the plunger immediately after patient injection. Needle retraction is activated by the syringe user.
The primary intended use is to administer safe and accurate subcutaneous and intramuscular injections. The secondary intended use is to retract and contain the contaminated needle after injection for the purpose of aiding in the prevention of accidental needle stick injuries.
{1}------------------------------------------------
KCC9.71
- 17. Indications for Use: This device is a safety hypodermic syringe for intramuscular and subcutaneous injection of medication to patients. This device aids in prevention of needlestick injuries. EO sterilized, non-pyrogenic, and latex free. The function of the OMI Retractable Safety Syringe is to provide a safe, accurate and reliable method of injecting medication into a patient.
### 18. Summary Comparing Technological Characteristics With Predicate Devices:
The OMI Manufacturing Pty Ltd makes a Substantial Equivalence claim of the OMI Retractable Safety Syringe to Retractable Technologies, Inc., Pop-n-Lock Syringe (currently marketed as the Vanishpoint syringe), 510(k) # k946219. Both syringes are similar and in some cases the same, with regards to parts, design, material, operating procedure and intended use. Both the OMI and Vanishpoint syringes consist of a syringe barrel, syringe plunger, single lumen hypodermic needle, needle hub and stem. Both have a spring retracting mechanism which is user activated once medication has been injected. Both are plunger syringes which are supplied with permanently attached needles. Both are supplied sterile, single-use and disposable.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## OCT 2 4 2006
OMI Manufacturing Pty., Limited C/O Mr. Daniel Kamm Kamm & Associates P.O. Box 7007 Deerfield, Illinois 60015
Re: K060971
Trade/Device Name: OMI Retractable Safety Syringe and OMI Insulin Safety Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG Dated: October 16, 2006 Received: October 19, 2006
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 -- Mr. Kamm
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susi Runne
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K060971
Device Name: OMI Insulin Safety Syringe
Indications For Use:
The OMI Safety Insulin Syringes are intended for the subcutaneous injection of insulin. In addition, the syringe is designed to aid in prevention of needlestick injuries.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Antar Ume
n of Anesthesiology, General Hospital, on Control, Dental Devices
K46971
Page 1 of 2
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): צו 609 7/
Device Name: OMI Retractable Safety Syringe
Indications For Use:
This device is a safety hypodermic syringe for intramuscular and subcutaneous injection of medication to patients. This device aids in prevention of needlestick injuries. EO sterilized, non-pyrogenic, and latex free.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ell for ADK 10/25/2022
in of Anesthesiology, General Hospital, con Control, Dental Devices
Page 1 of 1
1124
Page 7 of 323
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.