VITROS CHEMISTRY PRODUCTS CALIBRATOR KIT 99, VITROS CHEMISTRY PRODUCTS AAT/HPT PERFORMANCE VERIFIER I, II, AND III
Applicant
Ortho-Clinical Diagnostics, Inc.
Product Code
JIX · Clinical Chemistry
Decision Date
Apr 24, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Indications for Use
VITROS Chemistry Products Calibrator Kit 99 is used to calibrate VITROS 5, 1 FS Chemistry systems for the quantitative measurement of proteins in body fluids. VITROS Chemistry Products AAT/HPT Performance Verifiers I, II, III are assayed controls used to monitor performance of protein measurements in body fluids with VITROS 5, 1 FS Chemistry systems.
Device Story
VITROS Chemistry Products Calibrator Kit 99 and AAT/HPT Performance Verifiers are processed human serum-based products containing inorganic salts, buffers, and preservatives. Calibrator Kit 99 provides a five-level set for system calibration; Performance Verifiers provide a three-level set for monitoring measurement performance. Used exclusively with VITROS 5, 1 FS Chemistry systems, these products support the quantitative immunoturbidimetric assay of alpha-1 antitrypsin and haptoglobin. The system uses these materials to establish and verify the accuracy of protein measurements in clinical laboratory settings. By ensuring proper calibration and control, the device enables healthcare providers to obtain reliable quantitative results for patient protein levels, facilitating clinical assessment of associated disease states.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Processed human serum matrix with added inorganic salts, buffers, and preservatives. Liquid, ready-to-use format. Five-level calibrator set; three-level control set. Standardized to BAM-IRMM-LGC ERM-DA 470 reference material. Designed for use with VITROS 5, 1 FS Chemistry systems (random access, fully automated clinical chemistry analyzer).
Indications for Use
Indicated for use as calibrators and assayed controls for the quantitative measurement of alpha-1 antitrypsin (AAT) and haptoglobin (HPT) in body fluids using VITROS 5, 1 FS Chemistry systems.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
VITROS Chemistry Products AAT Performance Verifiers I, II, and III (K052819)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k060940
B. Purpose for Submission:
Addition of Haptoglobin to an existing device.
C. Measurand:
Alpha-1 Antitrypsin
Haptoglobin
D. Type of Test:
Quantitative immunoturbidimetric assay
E. Applicant:
Ortho-Clinical Diagnostics, Inc.
F. Proprietary and Established Names:
VITROS Chemistry Products Calibrator Kit 99
VITROS Chemistry Products AAT/HPT Performance Verifiers I, II, and III
G. Regulatory Information:
1. Regulation section:
21CFR§ 862.1660, Quality Control Material (Assayed and Unassayed)
21CFR§ 862.1150, Calibrator
2. Classification:
Calibrator - Class II
Quality control material - Class I
1. Product code:
JJX, Single (Specified) Analyte Controls (Assayed and Unassayed)
JIX, Calibrator, Multi-Analyte Mixture
4. Panel:
Chemistry (75)
H. Intended Use:
1. Intended use(s):
VITROS Chemistry Products Calibrator Kit 99 is used to calibrate VITROS 5, 1 FS Chemistry systems for the quantitative measurement of proteins in body fluids.
VITROS Chemistry Products AAT/HPT Performance Verifiers I, II, III are assayed controls used to monitor performance of protein measurements in body fluids with VITROS 5, 1 FS Chemistry systems.
2. Indication(s) for use:
Same as above
3. Special conditions for use statement(s):
For prescription use only.
4. Special instrument requirements:
For use in VITROS 5, 1 FS Chemistry systems (k031924).
I. Device Description:
{1}
The VITROS Chemistry Products Calibrator Kit 99 is a five level set of fluids used to calibrate VITROS 5, 1 FS Chemistry systems for the quantitative measurement of $\alpha 1$ -antitrypsin (AAT), and haptoglobin (HPT) using VITROS Chemistry Products AAT and HPT Reagents. VITROS Chemistry Products Calibrator 99 is prepared from processed human serum to which inorganic salts, buffers, and preservatives have been added.
The VITROS Chemistry Products AAT/HPT Performance Verifiers I, II, and III is a three level set of assayed liquid controls used to monitor the performance of $\alpha 1$ -antitrypsin (AAT), and haptoglobin (HPT) measurements, using VITROS Chemistry Products AAT and HPT Reagents with the VITROS 5, 1 FS Chemistry systems. The VITROS AAT/HPT Performance Verifiers are prepared from processed human serum to which inorganic salts, buffers, and preservatives have been added.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
VITROS Chemistry Products Calibrator Kit 99
VITROS Chemistry Products AAT Performance Verifiers I, II, and III
2. Predicate $510(\mathbf{k})$ number(s):
K052819
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | VITROS Calibrator Kit 99 (New) | VITROS Calibrator Kit 99 (Current) |
| Standardization | BAM-IRMM-LGC (Institute for reference Methods and Materials/Laboratory of the Gov’t Chemist) ERM-DA 470 | Same |
| Sample type | Serum | Same |
| Number of levels | Five levels | Same |
| Matrix | Prepared from processed human serum to which salts, buffers and preservatives have been added | Same |
| Format | Liquid ready to use. | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | VITROS Chemistry Products Calibrator Kit 99 is used to calibrate VITROS 5, | VITROS Chemistry Products Calibrator Kit 99 is used to calibrate |
| | VITROS 5, 1 FS | VITROS 5, 1 FS |
{2}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | VITROS AAT/HPT Performance Verifiers I, II, III | VITROS 5, 1 FS Chemistry systems for the quantitative measurement of α1-antitrypsin (AAT). |
| Analytes | α1-antitrypsin (AAT), and haptoglobin (HPT) | α1-antitrypsin (AAT), |
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | VITROS AAT/HPT Performance Verifiers I, II, III | VITROS AAT Performance Verifiers I, II, III |
| Standardization | BAM-IRMM-LGC (Institute for reference Methods and Materials/Laboratory of the Gov’t Chemist) ERM-DA 470 | Same |
| Sample type | Serum | Same |
| Number of levels | Three levels | Same |
| Matrix | Prepared from processed human serum to which salts, buffers and preservatives have been added | Same |
| Format | Liquid ready to use. | Same |
| Differences | | |
| Intended Use | VITROS Chemistry Products Calibrator Kit 99 is used to calibrate VITROS 5, 1 FS Chemistry systems for the quantitative measurement of proteins in body fluids. | VITROS Chemistry Products Calibrator Kit 99 is used to calibrate VITROS 5, 1 FS Chemistry systems for the quantitative measurement of α1-antitrypsin (AAT). |
| Analytes | α1-antitrypsin (AAT), and haptoglobin (HPT) | α1-antitrypsin (AAT), |
# K. Standard/Guidance Document Referenced (if applicable):
None referenced.
# L. Test Principle:
The VITROS Chemistry Products Calibrator Kit 99 (New) and VITROS Chemistry Products AAT/HPT Performance Verifiers I, II, III are to be used with VITROS 5, 1 FS Chemistry systems (k031924). The VITROS 5, 1 FS Chemistry systems is a random access, fully automated clinical chemistry analyzer, intended for use in the in vitro quantitative measurement of a variety
{3}
of analytes in certain biological fluids. The analyzer operates in conjunction with reagents, calibrators and controls designed for use with the system.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
**a. Precision/Reproducibility:**
Not applicable.
The new device is for the addition of the haptoglobin analyte to Calibrator Kit 99 and Performance Verifiers I, II, III. Haptoglobin immunological test system (§866.5460) is a Class II exempt from premarket notification procedures. Refer to predicate device (k052819) for precision/reproducibility studies of AAT.
**b. Linearity/assay reportable range:**
Not applicable.
The new device is for the addition of the haptoglobin analyte to Calibrator Kit 99 and Performance Verifiers I, II, III. Haptoglobin immunological test system (§866.5460) is a Class II exempt from premarket notification procedures. Refer to predicate device (k052819) for linearity studies of AAT.
**c. Traceability, Stability, Expected values (controls, calibrators, or methods):**
The values assigned to the VITROS Chemistry Products Calibrator Kit 99 for AAT/HPT are traceable to BAM-IRMM-LGC (Bundesanstalt für Materialforschung und-prüfung/Institute for Reference Methods and Materials/Laboratory of the Gov't Chemist) ERM-DA 470 Reference Material. A five level set of calibrators are prepared through VITROS 5, 1 FS Chemistry System on-analyzer dilution of ERM-DA470 and are used to calibrate the Manufacturer's Selected Measurement Procedure, a VITROS 5, 1 FS Chemistry System analyzer using VITROS Chemistry Products AAT or HPT Reagents, respectively. The calibration standards and the VITROS chemistry System are used to support value assignment for working calibrators, VITROS Chemistry Products Calibrator Kit 99 (Master Lot).
The Master Lot and the Manufacturer's Selected Measurement Procedure, a VITROS 5, 1 FS Chemistry System analyzer using VITROS Chemistry Products AAT or HPT Reagents, respectively transfers the trueness of the working calibrators to new product lots of the VITROS Chemistry Products AAT/HPT Reagent and VITROS Chemistry Products Calibrator Kit 99.
Calibrator Kit Stability – Unopened kit are stable until expiration date printed on the label when stored refrigerated at 2°-8°C. Opened vials are stable at ≤3 days when stored capped between 2°-8°C.
Performance Verifiers Stability – Unopened kit are stable until expiration date printed on the label when stored refrigerated at 2°-8°C. The stability of opened vials stored capped between 2°-8°C is up to 4 weeks.
4
{4}
d. Detection limit:
Not applicable
The new device is for the addition of the haptoglobin analyte to Calibrator Kit 99 and Performance Verifiers I, II, III. Haptoglobin immunological test system (§866.5460) is a Class II exempt from premarket notification procedures. Refer to predicate device (k052819) for determination of detection limit.
e. Analytical specificity:
Not applicable.
The new device is for the addition of the haptoglobin analyte to Calibrator Kit 99 and Performance Verifiers I, II, III. Haptoglobin immunological test system (§866.5460) is a Class II exempt from premarket notification procedures. Refer to predicate device (k052819) for analytical specificity studies.
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
The new device is for the addition of the haptoglobin analyte to Calibrator Kit 99 and Performance Verifiers I, II, III. Haptoglobin immunological test system (§866.5460) is a Class II exempt from premarket notification procedures. Refer to predicate device (k052819) for method comparison with predicate device.
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Not applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.