K060879 · Siemens Medical Solutions USA, Inc. · ITX · Apr 19, 2006 · Radiology
Device Facts
Record ID
K060879
Device Name
MODEL EC9-4 TRANSDUCER
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
ITX · Radiology
Decision Date
Apr 19, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1570
Device Class
Class 2
Attributes
Pediatric
Indications for Use
EC9-4 is a convex array ultrasound transducer intended for fetal, small organ, neonatal cephalic, transvaginal and transrectal applications on the ACUSON Antares™ ultrasound system. - 1. ACUSON Antares™ ultrasound system is intended for use in the following applications: General Radiology, Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Transesophageal, Pelvic, Neonatal/Adult Cephalic, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
Device Story
EC9-4 is a convex array ultrasound transducer designed for use with the ACUSON Antares ultrasound system. It captures acoustic signals from internal anatomical structures during transrectal, transvaginal, fetal, neonatal, and small parts examinations. The transducer converts these signals into electrical data processed by the Antares system to generate real-time diagnostic images. Operated by clinicians in clinical settings, the device facilitates anatomical measurements and clinical diagnosis. The system adheres to ALARA principles for acoustic output. The output is displayed on the system monitor, allowing healthcare providers to visualize internal structures, assess pathology, and guide clinical decision-making. The device benefits patients by providing non-invasive, real-time diagnostic imaging for various clinical applications.
Clinical Evidence
No clinical data. Bench testing only, including compliance with safety and EMC standards (EN 60601-1, EN 60601-1-1, EN 60601-1-2, EN 60601-2-37) and biological evaluation (ISO 10993-1).
Technological Characteristics
Convex array ultrasound transducer. Materials comply with ISO 10993-1. Operates with ACUSON Antares system. Adheres to AIUM/NEMA UD-2 and UD-3 standards for acoustic output. Safety compliance: UL 60601-1, CSA C22.2 No. 601-1, EN 60601-1 series.
Indications for Use
Indicated for transrectal, transvaginal, fetal, neonatal, and small parts imaging applications in patients requiring diagnostic ultrasound evaluation.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
{0}------------------------------------------------
EC9-4 ultrasound transducer Special 510 (k): Device Modification
## SECTION 11
# 510(k) Summary of Safety and Effectiveness
| Sponsor: | Siemens Medical Solutions USA, Inc.,<br>Ultrasound Division<br>1230 Shorebird Way<br>P.O. Box 7393<br>Mountain View, California 94039-7393 | APR 19 2006 |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Contact Person: | Patrick Lynch<br>Sr. Regulatory Affairs Specialist<br>Telephone: (425) 557-1825<br>Fax: (425) 391-9198 | |
| Submission Date: | March 22, 2006 | |
| Device Name: | EC9-4 Ultrasound Transducer | |
| Common Name: | Diagnostic Ultrasound Transducer | |
| Classification: | | |
| Regulatory Class: | II | |
| Review Category: | Tier II | |
| Classification Panel: | Radiology | |
| 21 CFR 892.1550 | | |
| Diagnostic Ultrasound Transducer | FR # | Product Code |
|----------------------------------|----------|--------------|
| | 892.1570 | 90-ITX |
#### Predicate Devices:
The EC9-4 transducer has already been cleared on the following Siemens Medical Solutions' ultrasound platform:
· ACUSON Antares™ ultrasound platform (#K050034)
#### Device Description:
Presently cleared EC9-4 endocavity curved array ultrasound transducer is primarily indicated for transrectal and transvaginal, fetal and neonatal and small parts applications on the ACUSON Antares ultrasound imaging platform.
{1}------------------------------------------------
### Intended Use:
EC9-4 is a convex array ultrasound transducer intended for fetal, small organ, neonatal cephalic, transvaginal and transrectal applications on the ACUSON Antares™ ultrasound system.
- 1. ACUSON Antares™ ultrasound system is intended for use in the following applications:
General Radiology, Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Transesophageal, Pelvic, Neonatal/Adult Cephalic, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
The Antares ultrasound system has been designed to conform to the following product safety standards:
- UL 60601-1, Safety Requirements for Medical Equipment .
- CSA C22.2 No. 601-1, Safety Requirements for Medical Equipment .
- AIUM/NEMA UD-2, 1998, Acoustic Output Measurement Standard for Diagnostic . Ultrasound
- AIUM/NEMA UD-3, 1998, Standard for Real Time Display of Thermal and . Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
- . 93/42/EEC Medical Device Directive
- Safety and EMC Requirements for Medical Equipment .
- . EN 60601-1
- . EN 60601-1-1
- . EN 60601-1-2
- . EN 60601-2-37
- ISO 10993-1, Biological Evaluation of Medical Devices Part 1: Evaluation and . Testing
The systems' acoustic output is in accordance with ALARA principle (as low as reasonably achievable)
## Technological Comparison to Predicate Devices:
EC9-4 ultrasound transducer is substantially equivalent in its technology and functionality to the EC9-4 transducer which is already cleared under the following 510(k) premarket notification number: K050034 (ACUSON Antares ultrasound system),
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three tail feathers, representing the three branches of government. The eagle is surrounded by a circle of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
APR 1 9 2006
Mr. Patrick Lynch Senior Regulatory Affairs Specialist Siemens Medical Solutions USA, Inc. Ultrasound Division Headquarters 1230 Shorebird Way P.O. Box 7393 MOUNTAIN VIEW CA 94039-7393
Re: K060879
Trade Name: EC9-4 Ultrasound Transducer Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: ITX Dated: March 30, 2006 Received: March 31, 2006
Dear Mr. Lynch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act). You may, thereforc, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the ACUSON Antares™ Ultrasound System, as described in your premarket notification:
Transducer Model Number
1-09-4
ling Public . I
{3}------------------------------------------------
Page 2 - Mr. Lynch
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions rcgarding the content of this letter, please contact Andrew Kang. M.D. at (301) 594-1212.
Sincerely yours,
1/ January 3, 1942
Nancy C. Brogdon Director. Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
Siemens Medical Solutions USA, Inc. Ultrasound Division
EC9-4 ultrasound transduccr Special 510(k): Device Modification
510(k) Number (if known): K060879
Device Name: EC9-4 ultrasound transducer
Indications for Use:
The EC9-4 endocavity curved array ultrasound transducer is primarily indicated for transrectal and transvaginal, fetal and neonatal and small parts applications on the ACUSON Antares ultrasound imaging platform.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
S. C. Dunston
Sign Off)
reproductive, Abdominal
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.