ARTHROCARE WANDS

K060823 · Arthrocare Corp. · GEI · Apr 10, 2006 · General, Plastic Surgery

Device Facts

Record IDK060823
Device NameARTHROCARE WANDS
ApplicantArthrocare Corp.
Product CodeGEI · General, Plastic Surgery
Decision DateApr 10, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Wands are intended for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in orthopedic, arthroscopic, spinal, and neurological procedures

Device Story

ArthroCare Wands are bipolar, single-use, high-frequency electrosurgical devices. Used in orthopedic, arthroscopic, spinal, and neurosurgical procedures. Device functions by delivering high-frequency energy to soft tissue to perform resection, ablation, and coagulation, and to achieve hemostasis of blood vessels. Operated by surgeons in clinical settings. Output is controlled by the surgeon to modify tissue during surgery, directly impacting clinical outcomes by enabling precise tissue management.

Clinical Evidence

Bench testing only.

Technological Characteristics

Bipolar, single-use, high-frequency electrosurgical device. Materials and dimensional specifications modified from predicate. Sterilization parameters remain consistent with predicate.

Indications for Use

Indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in patients undergoing orthopedic, arthroscopic, spinal, and neurological procedures.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for "ArthroCare Corporation". The logo consists of a stylized graphic to the left of the text. The graphic is composed of several curved black lines that create a circular shape. To the right of the graphic is the word "ArthroCare" in a bold, sans-serif font, with a trademark symbol next to it. Below "ArthroCare" is the word "Corporation" in a smaller, less bold font. APR 1 0 2006 510(k) Summary KO6082 ArthroCare Corporation ArthroCare Wands Page 1 of 2 General Information Submitter Name/Address: ArthroCare Corporation 680 Vaqueros Avenue Sunnyvale, CA 94085-2936 Establishment Registration Number: Contact Person: 2951580 March 24, 2006 Valerie Defiesta-Ng Director, Regulatory Affairs Date Prepared: Device Description Trade Name: Generic/Common Name: Classification Name: ArthroCare® Wands Electrosurgical Device and Accessories Electrosurgical Cutting and Coagulation Device and Accessories (21 CFR 878.4400) ### Predicate Device ArthroCare 8000S Coblator Surgery Systems K053297 ### Product Description The ArthroCare Wands are bipolar, single use, high frequency electrosurgical devices designed for orthopedic, arthroscopic, and spinal, and neurosurgical procedures. ### Intended Use The Wands are intended for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in orthopedic, arthroscopic, spinal, and neurological procedures ### Substantial Equivalence This Special 510(k) proposes modifications in dimensional specifications, material, and labeling for the ArthroCare Wands, which were previously cleared under K053297 on December 6, 2005. The indications for use, technology, principle of operation, performance specifications, and sterilization parameters of the Wands remain the same as in the predicate cleared 510(k). {1}------------------------------------------------ 060823 # ್ನಿ ಲ್ಲಿ ### Summary of Safety and Effectiveness The modified ArthroCare Wands, as described in this submission, are substantially equivalent to the predicate Wands. The proposed modification in dimensional specifications, materials, and labeling are not substantial changes or modifications, and do not significantly affect the safety or efficacy of the devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three stripes forming its wings, symbolizing strength and service. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 1 0 2006 ArthroCare Corporation c/o Ms. Valerie Defiesta-NG Director, Regulatory Affairs 680 Vaqueros Avenue Sunnyvale, California 94085-2936 Re: K060823 Trade/Device Name: ArthroCare Wands Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI, GXI Dated: March 24, 2006 Received: March 27, 2006 Dear Ms. Defiesta-NG: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ #### Page 2 - Ms. Valerie Defiesta-NG forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, elmalle Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Statement KO60823 510(k) Number: Device Name: ArthroCare Wands Indications for Use: The Wands are intended for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in orthopedic, arthroscopic, spinal, and neurological procedures. Prescription Use (21 CFR 801 Subpart D) AND/OR X Over-the-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of C Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number ko60823
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...