VANTAGE INDIRECT OPHTHALMOSCOPE

K060822 · Keeler Instruments, Inc. · HLI · May 23, 2006 · Ophthalmic

Device Facts

Record IDK060822
Device NameVANTAGE INDIRECT OPHTHALMOSCOPE
ApplicantKeeler Instruments, Inc.
Product CodeHLI · Ophthalmic
Decision DateMay 23, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 886.1570
Device ClassClass 2

Indications for Use

The Keeler Vamage Plus Indirect Ophthalmoscope is intended to examine the Cornea, aqueous, lens, vitreous and retina of the eye.

Device Story

Vantage Plus is a head-mounted binocular indirect ophthalmoscope used by trained clinicians to visualize ocular structures (cornea, aqueous, lens, vitreous, retina). Device comprises an illumination element (6V, 5W Xenon lamp, condensing lens, IR filter, user-selectable diagnostic filters, variable aperture stops, projection lens, illumination mirror) and a viewing element (adjustable +2 dioptre lenses and directional mirrors). Clinician wears the device on a headband; directs light into the patient's eye; views the fundus through ophthalmic lenses. Integrated aperture selector and viewing mirror adjustments improve ease of use compared to predicate. Device aids in diagnostic assessment of ocular health. Power provided by external wall pack, smart pack, or wireless battery pack.

Clinical Evidence

Bench testing only. Testing included light hazard assessment per BS EN ISO 15004:1998, electrical safety per BS EN 60601-1 and 60601-1-2, and verification of light output and aperture sizes. Risk management conducted per BS EN ISO 14971:2001.

Technological Characteristics

Head-mounted binocular indirect ophthalmoscope. Illumination: 6V, 5W Xenon filament lamp. Optics: +2 dioptre viewing lenses, adjustable directional mirrors. Filters: Permanent IR blocker, selectable blue, red-free, and diffuser filters. Aperture sizes: 60mm, 45mm, 22mm. Power: External wall pack, smart pack, or wireless battery pack. Materials: Self-extinguishing (HB or V0 rated). Compliance: BS EN 60601-1, 60601-1-2, BS EN ISO 15004:1998, BS EN ISO 14971:2001.

Indications for Use

Indicated for use by trained personnel to examine the cornea, aqueous, lens, vitreous, and retina of the eye.

Regulatory Classification

Identification

An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.

Special Controls

*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### Special 510K Summary Keeler Instruments Inc #### Submitter Contact Information: Company Name/Address: | | 456, Parkway, Broomall, PA 19008 | | MAY 23 2006 | |--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|-------------| | Company Phone No: | (610) 353 4350 | | | | Company Fax No: | (610) 353 7814 | | | | Contact Person: | Mr. Eugene VanArsdale (Marketing Director). | | | | Manufacturing Location: | Keeler Ltd Clewer Hill Road, Windsor, Berkshire. SL4 4AA. United Kingdom. | | | | Company Phone No: | (44) 1753 857177 | | | | Company Fax No: | (44) 1753 830247 | | | | Contact Person: | Mr. Neil Atkins (Development Manager). | | | | Device Information: | | | | | Device Trade Name: | Vantage Plus Binocular Indirect Ophthalmoscope | | | | Common Name: | Ophthalmoscope | | | | Class: | II | | | | Classification Panel: | 86 | | | | Product Code: | HLI (AC) and HLJ (DC) | | | | Regulation Number: | 886.1570 | | | | Substantial Equivalence: | | | | | Predicate Device: | Vantage Binocular Indirect Ophthalmoscope. | | | | 510(k) No: | K942104 | | | | Manufacturer: | Keeler Ltd (FDA Registration No 1000391004) (Address as above). | | | | Distributor: | Keeler Instruments Inc (FDA Registration No 2519784) (Address as above). | | | | Device Description: | The Vantage Plus is a head mounted Binocular Indirect ophthalmoscope designed to be used by trained personnel for illuminating and viewing parts of the eye such as the Cornea and retina when used in conjunction with an ophthalmic lens. The instrument has two main elements i) The illumination element ii) the viewing element. The illumination element consists of a 6Volt, 5Watt Xenon filament lamp, a condensing lens, IR filtration, User filters for diagnostic purposes, variable stop sizes for increasing/decreasing light aperture sizes, a projection lens and an illumination mirror used by the user to direct the light. The binocular Viewing element consists of low powered Viewing Lenses (+2 dioptre) and | | | The binocular Viewing element consists of low powered Viewing Lenses (+2 dioptre) and directional mirrors which are adjustable to obtain optimal views of the eye. The unit has adjustable illumination levels and is powered by the same power sources to that used on the Vantage (Re: 510(k) K942104). To improve the ease of use of the instrument, the Vantage Plus combines the Aperture selector control and the Viewing mirror adjustments to achieve optimal instrument setting with minimum fuss. This feature is not present on the Vantage (Re: 510(k) K942104). # Dia # 그 | 강 #### D K0608zz {1}------------------------------------------------ ### Comparison of Technological Characteristics: - : | | | | Comparison of Indirect Ophthalmoscope Technological Characteristics | | |-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------| | Criteria | | Predicate Device<br>Keeler Vantage<br>(Re:510(k) K942104) | Keeler Vantage Plus (see note 4) | | | Type | | Binocular (Headband Mounted) | Binocular (Headband Mounted) | | | Illumination | | 6V, 10 Watt Halogen Filament<br>Bulb | 6V, 5 Watt Xenon Filament Bulb | | | Safety Filter IR Blocker | | Selectable | Permanent | | | Light Output (See note 1) | | 954 Lux | 913 Lux | | | Blue Filter | | Selectable | Selectable | | | Red Free Filter | | Selectable | Selectable | | | Diffuser | | Selectable | Selectable | | | Light aperture sizes (See note 2) | | Large (60mm) | Large (60mm) +/-5% | | | | | Medium (45mm) | Medium (45mm) +/-5% | | | | | Small (22mm) | Small (22mm)+/-5% | | | | | Inter Pupillary Distance Adjustment | | 52mm to 76mm | | Lens Power Viewing Optics | | +2 Dioptre | +2 Dioptre | | | Attachments | | Hi-Mag<br>Teaching Mirror | Hi-Mag<br>Teaching Mirror | | | Power Sources (See Note 2) | | Wall Pack<br>Smart Pack<br>Wireless battery Pack | Wall Pack<br>Smart Pack<br>Wireless Battery Pack | | | | Note 1: Average light output from a selection of bulbs for a patch size of 60mm.<br>Note 2: Measurements taken from 440mm from the front of the instrument. | | Note 3: All Power supplies comply with relevant parts of BS EN 60601-1 and 60601-1-2.<br>Note 4: All testing has been carried out on a fully functional prototype. | | | | Quality approvals: | Keeler is an ISO 9001:2000 and BS EN 13485:2003 compliant company and are obliged to ensure<br>that our operating and design practices fully comply. As part of the design process, all necessary risk<br>analysis, risk management, design verification and validation are conducted to ensure that the safety<br>and effectiveness of product designs either meet or surpass the product specification and comply too<br>the relevant standards. | | | | Risk Management: | Completed as per BS EN ISO 14971:2001 | | | | | Light Hazard: | The Vantage Plus has been tested in accordance with BS EN ISO 15004:1998 and is within the limits<br>specified for Optical radiation Hazard (Section 6.3) (See separate report under Technical<br>Documentation). | | | | | Electrical: | The Vantage Plus is using the same Power Sources as the Vantage detailed in 510(k) K942104. All<br>power sources are compliant with BS EN 60601-1 and 60601-1-2 and Keeler has not received any<br>adverse reports on these products (See separate reports under Technical Documentation). | | | | | Explosion: | Low probability. All measures have been taken to reduce the risk of explosion by using non-<br>explosive materials, ensuring the product meets all of the relevant electrical safety requirements and<br>detailed user instructions defining the requirements for safe usage of the instrument. | | | | | Fire Hazard: | Low Probability. All measures have been taken to use self extinguishing materials which are either<br>HB or VO rated. The system is illuminated using a 5watt Xenon lamp and all materials used in the<br>vicinity are specifically designed to operate in high temperature environments safely. | | | | | Surface Temperature: | Complies with BS EN ISO 60601-1 for temperature of external surfaces and controls. Caution in use instructions indicates 'Do not remove bulb whilst hot'. | | | | {2}------------------------------------------------ Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, forming a human-like figure. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. JUL 3 2008 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Keeler Instruments, Inc. c/o Mr. Eugene VanArsdale Marketing Director 456 Parkway Broomall, PA 19008 Re: K060822 Trade/Device Name: Vantage Plus Binocular Indirect Ophthalmoscope Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: II Product Code: HLI and HLJ Dated: May 4, 2006 Received: May 8, 2006 Dear Mr. VanArsdale: This letter corrects our substantially equivalent letter of May 23, 2006. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Malvina B. Edwards, ud Malvina B. Eydelman, M.I Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Copy ### Indications for Use 510(k) Number (if known): Device Name: Vantage Plus Binocular Indirect Ophthalmoscope Indications for Use: i The Keeler Vamage Plus Indirect Ophthalmoscope is intended to examine the Cornea, aqueous, lens, vitreous and retina of the eye. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Oft) Division of Ophthalmic Ear, Nose and Throat Devises 510(k) Number K060822 Page 1 of 1 22
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...