DYNATRON ION IONTOPHORESIS ELECTRODE

K060814 · Dynatronics Corp. · EGJ · Jun 14, 2006 · Physical Medicine

Device Facts

Record IDK060814
Device NameDYNATRON ION IONTOPHORESIS ELECTRODE
ApplicantDynatronics Corp.
Product CodeEGJ · Physical Medicine
Decision DateJun 14, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5525
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Dynatron® Ion™ iontophoresis electrode is designed for clinical use to introduce soluble salts and other drugs into the body as an alternative to hypodermic injection.

Device Story

Dynatron® Ion™ iontophoresis electrode consists of an active drug delivery electrode and a return electrode. Device used in clinical settings by clinicians to deliver soluble salts and drugs into the body via iontophoresis, serving as an alternative to hypodermic injection. Electrodes are designed for single patient use. Device facilitates transdermal drug delivery; clinical benefit is non-invasive administration of medication.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Iontophoresis electrode system comprising an active drug delivery electrode and a return electrode. Designed for single patient use. No specific materials or energy source specifications provided.

Indications for Use

Indicated for the introduction of soluble salts and other drugs into the body as an alternative to hypodermic injection. Intended for clinical use.

Regulatory Classification

Identification

An iontophoresis device is a device that is intended to use a direct current to introduce ions of soluble salts or other drugs into the body and induce sweating for use in the diagnosis of cystic fibrosis or for other uses if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug. When used in the diagnosis of cystic fibrosis, the sweat is collected and its composition and weight are determined. Additionally, an iontophoresis device intended for any other purposes is a prescription device that is intended to use a current to introduce ions of drugs or non-drug solutions into the body for medical purposes other than those specified in paragraph (a) of this section, meaning that the device is not intended for use in diagnosis of cystic fibrosis, or a specific drug is not specified in the labeling of the iontophoresis device.

Special Controls

*Classification.* Class II (special controls). The device is classified as class II. The special controls for this device are:(i) The following performance testing must be conducted: (A) Testing using a drug approved for iontophoretic delivery, or a solution if identified in the labeling, to demonstrate safe use of the device as intended; (B) Testing of the ability of the device to maintain a safe pH level; and (C) If used in the ear, testing of the device to demonstrate mechanical safety. (ii) Labeling must include adequate instructions for use, including sufficient information for the health care provider to determine the device characteristics that affect delivery of the drug or solution and to select appropriate drug or solution dosing information for administration by iontophoresis. This includes the following: (A) A description and/or graphical representation of the electrical output; (B) A description of the electrode materials and pH buffer; (C) When intended for general drug delivery, language referring the user to drug labeling approved for iontophoretic delivery to determine if the drug they intend to deliver is specifically approved for use with that type of device and to obtain relevant dosing information; and (D) A detailed summary of the device-related and procedure-related complications pertinent to use of the device, and appropriate warnings and contraindications, including the following warning: *Warning:* Potential systemic adverse effects may result from use of this device. Drugs or solutions delivered with this device have the potential to reach the blood stream and cause systemic effects. Carefully read all labeling of the drug or solution used with this device to understand all potential adverse effects and to ensure appropriate dosing information. If systemic manifestations occur, refer to the drug or solution labeling for appropriate action.(iii) Appropriate analysis/testing must demonstrate electromagnetic compatibility, electrical safety, thermal safety, and mechanical safety. (iv) Appropriate software verification, validation, and hazard analysis must be performed. (v) The elements of the device that may contact the patient must be demonstrated to be biocompatible. (vi) The elements of the device that may contact the patient must be assessed for sterility, for devices labeled as sterile. (vii) Performance data must support the shelf life of the elements of the device that may be affected by aging by demonstrating continued package integrity and device functionality over the stated shelf life.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K0608/4 JUN 1 4 2006 ## SUMMARY OF SAFETY AND EFFECTIVENESS #### SUBMITTED BY: DYNATRONICS CORPORATION 7030 Park Centre Drive Salt Lake City UT 84121 Phone: (800) 874-6251; (801) 568-7000; Fax: (801) 568-7711 #### 1. DEVICE NAME (Trade/common, and classification): Dynatron® Ion™ iontophoresis electrode | Classification: | Class III | |-------------------|-----------| | Regulation Nos .: | 890.5525 | | Product Codes: | EGJ | #### 2. PREDICATE DEVICES: North Coast Buffered Iontophoresis Electrode – cleared under K052019 #### 3. PERFORMANCE STANDARDS: NA #### 4. DESCRIPTION: The Dynatron® Ion™ iontophoresis electrode consists of an active drug delivery electrode and a return electrode. Electrodes are designed for single patient use. ## 5. INTENDED USE/INDICATIONS FOR USE: The Dynatron® Ion™ iontophoresis electrode is designed for clinical use to introduce soluble salts and other drugs into the body as an alternative to hypodermic injection. The Intended Use/Indications For Use stated herein are consistent with the cleared indications for the predicate device. {1}------------------------------------------------ # 6. SUBSTANTIAL EQUIVALENCE (SE) RATIONALE: The Dynatron® Ion™ iontophoresis electrode shares the same or similar basic characteristics, features and intended use as the predicate device and, therefore, is substantially equivalent to the North Coast Buffered Iontophoresis Electrode (applicable 'K' number listed above). - 7. SAFETY AND EFFECTIVENESS: There are no substantive differences between the product defined in this 510(k) submission and the predicate device. They are similar to هئي the technologies that are currently used in other similar medical devices. The product is safe and effective when used as indicated in specific applications under a clinician's supervision/therapy program. Signed: Kelvyn Cullimore, Jr., President and CEO DYNATRONICS CORPORATION h 20, 2006 Dated: {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 1 4 2006 Dynatronics Corporation % Mr. Kelvyn Cullimore, Jr., President and CEO 7030 Park Centre Drive Salt Lake City, Utah 84121 Re: K060814 Trade/Device Name: Dynatron® Ion™ iontophoresis electrode Regulation Number: 21 CFR 890.5525 Regulation Name: Iontophoresis devices Regulatory Class: III Product Code: EGJ Dated: May 24, 2006 Received: May 26, 2006 Dear Mr. Cullimore: ~--- We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act), as long as you comply with all of the Act's requirements relating to drugs labeled or promoted with the device as described below. You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if {3}------------------------------------------------ Page 2 - Mr. Kelvyn Cullimore, Jr. applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Our substantially equivalent decision does not apply to the drugs that you will label or promote for use with your device. Therefore, you may neither label nor promote your device for use with specific drugs, nor package drugs with your device prior to FDA having approved the drugs for iontophoretic administration. For information on the requirements for marketing new drugs, you may contact: > Director Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland As you are aware, there are concerns relating to the fact that no drug is currently labeled for administration via an iontophoresis device. The Agency currently is evaluating this public health concern regarding the safety and effectiveness of this route of administration of drugs, and in the near future will inform manufacturers of certain additional steps the Agency believes are necessary to address this concern. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation, please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification," (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (240) 276-3150, or at its {4}------------------------------------------------ Page 3 - Mr. Kelvyn Cullimore, Jr. Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): 长060 814 Device Name: Dynatron® Ion™ iontophoresis electrode Indications for Use: The Dynatron® Ion™ iontophoresis electrode is indicated to introduce soluble salts and other drugs into the body as an alternative to hypodermic injection Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Oaubare Bneehup for mxm (Division Sign-Off) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ Division of General, Restorative, and Neurological Devices **510(k) Number** K060814 Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...