K060810 · Ind Diagnostic, Inc. · DKZ · Aug 7, 2006 · Clinical Toxicology
Device Facts
Record ID
K060810
Device Name
ONE STEP DRUG TEST - AMP/MET/MOR/THC/COC
Applicant
Ind Diagnostic, Inc.
Product Code
DKZ · Clinical Toxicology
Decision Date
Aug 7, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3100
Device Class
Class 2
Indications for Use
The One Step Drugs of Abuse Test is a prescription lateral flow competitive immunoassay panel or single test intended for professional central laboratory use by trained users. It provides qualitative visual screening results for Marijuana, Cocaine, Morphine, Amphetamine and Methamphetamine at cutoff concentrations of 50 ng/mL Marijuana, 300 ng/mL Morphine, 1000 ng/mL Morphine, 1000 ng/mL Amphetamine and 1000 ng/mL Methamphetamine in urine in combinations (Multi-Panel) or separately (Single). It is not intended for over the counter sales to the lay person. For in vitro diagnostic use only. As with all qualitative tests, this assay provides only a preliminary result. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas chromatography/mass spectroscopy (GC/MS) is the recommended confirmatory method
Device Story
Lateral flow competitive immunoassay; detects drugs of abuse in urine. Input: urine sample (dip strip or cassette well). Principle: drug conjugate competes with drug in urine for limited antibody binding sites; antibody-colloid gold conjugate moves via capillary action. Output: visual pink line on test band (negative) or absence of line (positive). Used in professional central laboratories by trained personnel. Provides preliminary screening results; requires confirmatory testing (GC/MS) for clinical decision-making. Benefits: rapid, qualitative identification of drug presence to guide further diagnostic or clinical action.
Indicated for professional laboratory use for qualitative screening of Marijuana, Cocaine, Morphine, Amphetamine, and Methamphetamine in human urine. Not for OTC or point-of-care use.
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k060810
B. Purpose for Submission:
New 510(k)
C. Measurand:
Amphetamine, Methamphetamine, Cannabinoids, Cocaine, and Morphine in urine
D. Type of Test:
Qualitative visually read immunochromatographic assay
E. Applicant:
IND Diagnostics, Inc.
F. Proprietary and Established Names:
One Step Amphetamine, Methamphetamine, Cocaine, Morphine, and Marijuana Tests
G. Regulatory Information:
1. Regulation section:
862.3100, Enzyme Immunoassay, Amphetamine
862.3870, Enzyme Immunoassay, Cannabinoids
862.3250, Enzyme Immunoassay, Cocaine and Cocaine Metabolites
862.3610, Thin Layer Chromatography, Methamphetamine
862.3640, Liquid Chromatography, Morphine
2. Classification:
Class II
3. Product code:
DKZ, LDJ, DIO, DJC, and DPK, respectively
{1}
4. Panel: 91 (Toxicology)
H. Intended Use:
1. Intended use(s):
See indications for use.
2. Indication(s) for use:
The One Step Drugs of Abuse Test is a prescription lateral flow competitive immunoassay panel or single test intended for professional central laboratory use by trained users. It provides qualitative visual screening results for Marijuana, Cocaine, Morphine, Amphetamine and Methamphetamine at cutoff concentrations of 50 ng/mL Marijuana, 300 ng/mL Cocaine, 300 ng/mL Morphine, 1000 ng/mL Amphetamine and 1000 ng/mL Methamphetamine in urine in combinations (Multi-Panel) or separately (Single). It is not intended for over the counter sales to the lay person.
For in vitro diagnostic use only.
As with all qualitative tests, this assay provides only a preliminary result. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used. To obtain a confirmed analytical result, a more specific alternative chemical method is needed. Gas chromatography/Mass spectrometry is the recommended confirmatory method.
3. Special conditions for use statement(s):
The assay is for prescription use. The assay is not designated for use in point-of-care settings.
4. Special instrument requirements:
Not applicable. The device is a visually read single-use device.
I. Device Description:
The product may be sold as a single test or as part of a panel, and may be packaged in either a strip or a cassette format. Operators dip the test strip into the urine, or add drops of urine to the cassette test well to initiate the test reaction. Results are visually read.
The sponsor indicates the device has no human source material.
J. Substantial Equivalence Information:
2
{2}
1. Predicate device name(s):
Princeton Biomeditech Accusign DOA 5
2. Predicate 510(k) number(s):
k962353
3. Comparison with predicate:
Both devices have the same intended use, utilize the same test methodology and have the same cutoff concentrations. The reagent formulations vary between the two devices, as do the manufacturers.
## K. Standard/Guidance Document Referenced (if applicable):
The sponsor did not reference any standards in this submission.
## L. Test Principle:
The test is based on the principle of a competitive inhibition immunoassay, in which a chemically labeled drug (drug conjugate) competes with the drug which may be present in urine, for limited antibody binding sites. The test device consists of a membrane strip, which is pre-coated with DOA-BSA conjugate on the test band region, and a colored anti-DOA monoclonal antibody-colloid gold conjugate pad, which is placed at the end of the membrane.
In the absence of drug in the urine, the colored antibody-colloid gold conjugate moves with the sample by capillary action along the membrane until it reaches the immobilized drug conjugate in the test band region. At this point, the antibody-colloid gold conjugate reacts with the pre-coated drug conjugate and forms a visible pink colored line as the antibodies form complexes with the drug conjugate. Therefore, formation of a visible pink color line on the test band region is interpreted as a negative test result.
When the drug is present in the urine, the drug/metabolite antigen will compete with the drug conjugate coated in the test band region for the limited antibody sites. When a sufficient concentration of drug is present, it will fill the limited antibody binding sites, and will prevent attachment of the colored antibody-colloid gold conjugate to drug conjugates pre-coated in the test band region. Therefore, absence of the pink color band on the test region indicates a positive result for the specified drug. A control line indicates that an adequate volume of sample was added and that the test membrane is intact.
## M. Performance Characteristics (if/when applicable):
1. Analytical performance:
{3}
a. Precision/Reproducibility:
Precision of the test was characterized at the sponsor's facility by two trained technicians. Replicates of drug-free urine were tested, as well as drug-free urine spiked to the cutoff concentration, 25% above the cutoff concentration, and 25% below the cutoff concentration. Testing was performed on strip, cassette, and multi-panel test formats. Results are presented below.
Marijuana
| Marijuana | 0 ng / mL | 37.5 ng / mL | 50 ng / mL | 62.5 ng / mL |
| --- | --- | --- | --- | --- |
| Negative | 60 | 54 | 0 | 0 |
| Positive | 0 | 6 | 60 | 60 |
Cocaine
| Benzoylecgonine | 0 ng / mL | 225 ng /mL | 300 ng / mL | 375 ng / mL |
| --- | --- | --- | --- | --- |
| Negative | 60 | 54 | 0 | 0 |
| Positive | 0 | 6 | 60 | 60 |
Methamphetamine
| D-MET | 0 ng / mL | 750 ng / mL | 1000 ng / mL | 1250 ng / mL |
| --- | --- | --- | --- | --- |
| Negative | 60 | 54 | 0 | 0 |
| Positive | 0 | 4 | 60 | 60 |
Amphetamine
| D-AMP | 0 ng / mL | 750 ng / mL | 1000 ng / mL | 1250 ng / mL |
| --- | --- | --- | --- | --- |
| Negative | 60 | 54 | 0 | 0 |
| Positive | 0 | 6 | 60 | 60 |
{4}
Morphine
| Morphine | 0 ng / mL | 225 ng /mL | 300 ng / mL | 375 ng / mL |
| --- | --- | --- | --- | --- |
| Negative | 60 | 54 | 0 | 0 |
| Positive | 0 | 6 | 60 | 60 |
b. Linearity/assay reportable range:
Not applicable. The assay is intended for qualitative use.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Controls are required with this assay but are not specifically identified in the labeling. The sponsor did not indicate any degree of traceability for their devices. Users are instructed to follow federal, state, and local guidelines concerning QC practices.
d. Detection limit:
Sensitivity of a qualitative assay may be characterized by validating performance around the claimed cutoff concentration of the assay, and demonstrating the lowest concentration of drug that is capable of or consistently producing a positive result. This information appears in the precision section, above.
e. Analytical specificity:
An extensive list of compounds were tested for positive interference. Each compound was dissolved in drug-free control urine to a concentration of 100 µg / mL. Samples were adjusted to 3 pH levels (5, 7, 9) and 3 specific gravity levels (low, normal, high). All test results were negative.
Cross-reactivity of structurally related compounds was estimated by testing various amounts of drug or drug metabolite added to a drug-free urine control. Results are expressed as the lowest concentration required to produce a positive result.
{5}
Cross-reactivity of One-Step Marijuana
| Compounds | Concentration (ng/ml) |
| --- | --- |
| 11-nor-Δ-9-Tetrahydrocannabinol-COOH | 50 |
| 11-nor--Δ-8-Tetrahydrocannabinol-COOH | 440 |
| Δ-9-Tetrahydrocannabinol | 20,000 |
| Δ-8-Tetrahydrocannabinol | 17,500 |
| Canabinol | 50,000 |
| Cannabidiol | 100,000 |
Cross-reactivity of One-Step Cocaine
| Compounds | Concentration (ng/ml) |
| --- | --- |
| Cocaine | 300 |
| Benzoylecogonine | 300 |
Cross-reactivity of One-Step Morphine test
| Compounds | Concentration (ng/ml) |
| --- | --- |
| Morphine | 300 |
| Codeine | 300 |
| Ethyl morphine | 300 |
| Hydrocodone | 400 |
| Hydromorphone | 400 |
| Levophanol | 5,000 |
| Meperidine | 70,000 |
| Morphine-3-glucuronide | 500 |
| Naporphine | 2,000 |
| Norcodeine | 30,000 |
| Oxycodone | 13,125 |
| Oxymorphone | 50,000 |
| Thebaine | 26,250 |
6
{6}
Cross Reactivity of One Step Amphetamine
| Compounds | Concentration (ng/ml) |
| --- | --- |
| D-Amphetamine | 1000 |
| D/L-Amphetamine | 2000 |
| 3,4-Methylenedioxyamphetamine | 2500 |
| L-Amphetamine | 30,000 |
| D-(+)-Methamphetamine | 43,750 |
| D,L 3,4-Methylenedioxymethamphetamine | 43,750 |
Cross-Reactivity of One Step Methamphetamine
| Compounds | Concentration (ng/ml) |
| --- | --- |
| D-(+)-Methamphetamine | 1000 |
| 3,4-Methylenedioxyethylamphetamine(MDEA) | 1000 |
| d-Amphetamine | 50,000 |
| l-Amphetamine | 50,000 |
| 3,4-Methylenedioxyamphetamine (MDA) | 50,000 |
There is the possibility that other substances and/or factors not listed above may interfere with the test and cause false results, e.g., technical or procedural errors.
f. Assay cut-off:
The identified cutoff concentration(s) of the assay(s) are consistent with other cleared DOA assays.
2. Comparison studies:
a. Method comparison with predicate device:
Clinical samples were purchased from a laboratory in order to characterize performance of the One-Step assays against GC/MS. Samples were selected in order to obtain adequate distribution of the targeted drug around the claimed cutoff concentration. Results of the studies are presented below:
{7}
Comparison of One-step THC Test with GC/MS
| THC Metabolite | Low Negative (< 50% cutoff) | Near Cutoff Negative (between -50% + cutoff) | Near Cutoff Positive (between cutoff + 50%) | High Positive (> 50% cutoff) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 0 | 11 | 22 |
| Negative | 31 | 8 | 3 | 0 |
Comparison of One-step Cocaine Test with GC/MS
| Benzoylec-gonine | Low Negative (< 50% cutoff) | Near Cutoff Negative (between -50% + cutoff) | Near Cutoff Positive (between cutoff + 50%) | High Positive (> 50% cutoff) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 0 | 5 | 30 |
| Negative | 30 | 8 | 2 | 0 |
Comparison of One-step Methamphetamine Test with GC/MS
| D-Meth | Low Negative (< 50% cutoff) | Near Cutoff Negative (between -50% + cutoff) | Near Cutoff Positive (between cutoff + 50%) | High Positive (> 50% cutoff) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 1 | 11 | 23 |
| Negative | 31 | 9 | 0 | 0 |
Comparison of One-step Amphetamine Test with GC/MS
| D-Amph | Low Negative (< 50% cutoff) | Near Cutoff Negative (between -50% + cutoff) | Near Cutoff Positive (between cutoff + 50%) | High Positive (> 50% cutoff) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 2 | 9 | 26 |
| Negative | 31 | 7 | 0 | 0 |
{8}
Comparison of One-step Morphine Test with GC/MS
| Morphine | Low Negative (< 50% cutoff) | Near Cutoff Negative (between -50% + cutoff) | Near Cutoff Positive (between cutoff + 50%) | High Positive (> 50 % cutoff) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 1 | 7 | 28 |
| Negative | 31 | 7 | 1 | 0 |
b. Matrix comparison:
Not applicable. The assay is intended for only one sample matrix.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable. Clinical studies are not typically submitted for this device type and matrix.
b. Clinical specificity:
Not applicable. Clinical studies are not typically submitted for this device type and matrix.
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
Validation of the clinical appropriateness of the cutoff is not typically submitted for this device type and matrix.
5. Expected values/Reference range:
Not applicable. No elicit drugs should be present in urine.
O. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
{9}
P. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.