ELEMENT BLOOD GLUCOSE TEST SYSTEM

K060793 · Home Diagnostics, Inc. · NBW · Jul 25, 2006 · Clinical Chemistry

Device Facts

Record IDK060793
Device NameELEMENT BLOOD GLUCOSE TEST SYSTEM
ApplicantHome Diagnostics, Inc.
Product CodeNBW · Clinical Chemistry
Decision DateJul 25, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2

Indications for Use

The Element Blood Glucose Test System is intended for the quantitative determination of glucose in human whole blood taken from the finger or forearm. The system is intended to be used to assist the patient and healthcare professional in the management of diabetes. For at home use and for use by healthcare professionals in physician’s offices and convalescent care bedside testing facilities.

Device Story

Element Blood Glucose Test System measures glucose in human whole blood samples (finger or forearm). System uses electrochemical biosensor technology with glucose oxidase reagent on test strips. Meter applies voltage to electrodes; measures resulting current proportional to glucose concentration; displays plasma-equivalent glucose value. Designed for use by patients at home or healthcare professionals in clinical settings. Modifications from predicate include reduced test time (5 seconds), reduced memory capacity, and removal of certain user-accessible features (morning averages, date/time display, user-accessible serial port). Output assists in diabetes management.

Clinical Evidence

No clinical data provided in the document; substantial equivalence determination based on bench testing and comparison to predicate technology.

Technological Characteristics

Electrochemical glucose monitoring system; consists of meter and test strips; measures glucose in capillary whole blood; portable form factor for point-of-care or home use.

Indications for Use

Indicated for quantitative glucose determination in human whole blood (finger or forearm) to assist patients and healthcare professionals in diabetes management.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K060793 B. Purpose for Submission: Reduction in test time and software modifications to eliminate or modify some features available in previously cleared meter C. Measurand: Glucose D. Type of Test: Quantitative, Glucose Oxidase, Electrochemical Biosensor E. Applicant: Home Diagnostics, Inc. F. Proprietary and Established Names: Element Blood Glucose Test System G. Regulatory Information: 1. Regulation section: 21 CFR 862.1345, Glucose Test Systems 2. Classification: Class II 3. Product code: NBW – System, Test, Blood Glucose, Over the Counter CGA – Glucose Oxidase, Glucose {1} 4. Panel: 75 Clinical Chemistry H. Intended Use: 1. Intended use(s): See Indication for Use below 2. Indication(s) for use: The Element Blood Glucose Test System is intended for the quantitative determination of glucose in human whole blood taken from the finger or forearm. The system is intended to be used to assist the patient and healthcare professional in the management of diabetes. 3. Special conditions for use statement(s): For at home use and for use by healthcare professionals in physician’s offices and convalescent care bedside testing facilities. 4. Special instrument requirements: Element Blood Glucose meter I. Device Description: The Element Blood Glucose Test System is comprised of the Element Blood Glucose Meter, Element Test Strips, and Element Control Solutions (low and high). J. Substantial Equivalence Information: 1. Predicate device name(s): TrueTrack Smart System 2. Predicate 510(k) number(s): K030703, K032657 {2} 3 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Quantitative measurement of glucose in human whole blood from finger or forearm by people with diabetes and healthcare professionals | Same | | Reagent | Glucose oxidase | Same | | Sample size | 1 ul | Same | | Measurement range | 20-600 mg/dL | Same | | Test strips | Identical, new name | | | Control solutions | Identical, new name | | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Meter memory display | 50 blood or control results | 365 blood results, 1 control result | | Calculate and display morning averages | Not available | Available | | Test time | 5 seconds | 10 seconds | | Units of measure | Factory set | Programmed by user | | Date/time in display | Not available | Available | | Data upload capability | Serial port not accessible to user | Serial port accessible to user | K. Standard/Guidance Document Referenced (if applicable): None referenced. L. Test Principle: Glucose measurement is based on the electrical potential caused by the reaction of glucose in the sample with reagents contained on the test strip's electrodes. The glucose in the sample is oxidized by glucose oxidase, producing gluconolactone and the reduced form of an electron mediator. The amount of reduced mediator is proportional to the amount of glucose present in the sample. The reduced electron mediator is then oxidized at the surface of the measurement electrodes when a specified voltage is applied across the electrodes by the meter. The resulting current is measured and converted to a glucose concentration by the meter and displayed as a plasma equivalent value {3} M. Performance Characteristics (if/when applicable): The only performance characteristics applicable to this submission are precision and accuracy, to demonstrate that measurements at 5 seconds (Element meter) are equivalent to measurements at 10 seconds (TrueTrack meter). The remaining performance characteristics are unaffected by the current modifications and remain the same as those presented in the predicate 510(k)s. 1. Analytical performance: a. Precision/Reproducibility: Pooled blood at 3 levels (70, 98 and 215 mg/dL) were tested on both meters using 3 lots of test strips. The overall % CV for 5 replicates of each level run on the Element meter was compared to the same results on the TrueTrack meter. Acceptance criteria = at 70 mg/dL, mean CV must be ≤ 9.8%; at 98 mg/dL and 215 mg/dL, mean CV must be ≤ 8.3%. Results of this study showed that the average within vial % CV was < 6.0% at all levels for each strip lot, and the between vial % CV was < 5.0%. b. Linearity/assay reportable range: The linearity/assay reportable range was cleared with the TrueTrack Smart System (K030703 and K032657). c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability, stability and expected values cleared with the TrueTrack Smart System (K030703 and K032657). d. Detection limit: The detection limit was cleared with the TrueTrack Smart System (K030703 and K032657). e. Analytical specificity: Analytical specificity was cleared with the TrueTrack Smart System (K030703 and K032657). f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Pooled blood adjusted to 9 glucose levels, ranging from 30-535 mg/dL, was run with 3 test strip lots on both meters. Accuracy was demonstrated by the comparison of Element regression statistics against True Track using the same {4} samples. Acceptance criteria = slope of regression line must be $1.00 \pm 0.10$; y intercept must be $0.0 \pm 10.0$; and the correlation coefficients must be $\geq 0.950$. Each of the 3 strip lots must meet the criteria. The following results were obtained: | | Acceptance limits | Lot 1619 | Lot 1620 | Lot 1621 | | --- | --- | --- | --- | --- | | Slope | 0.90 to 1.10 | 1.00 | 0.99 | 1.02 | | Y Intercept | 10 to +10 | 2.6 | 1.4 | 0.5 | | R² | 0.95 to 1.00 | 0.999 | 0.999 | 0.999 | b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Performance in the hands of lay users was cleared with the TrueTrack Smart System (K030703 and K032657). 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Expected values/ reference range provided with TrueTrack Smart System (K030703 and K032657). N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...