MODIFICATION TO: TAMPER EVIDENT ULTRASAFE PASSIVE DELIVERY SYSTEM, ULTRASAFE PASSIVE DELIVERY SYSTEM (X-SERIES)
K060743 · Safety Syringes, Inc. · MEG · Apr 28, 2006 · General Hospital
Device Facts
Record ID
K060743
Device Name
MODIFICATION TO: TAMPER EVIDENT ULTRASAFE PASSIVE DELIVERY SYSTEM, ULTRASAFE PASSIVE DELIVERY SYSTEM (X-SERIES)
Applicant
Safety Syringes, Inc.
Product Code
MEG · General Hospital
Decision Date
Apr 28, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Single use devices that are indicated for use as an accessory with prefilled ISO Standard glass syringes to aid in the protection of healthcare professionals, patients who self-inject doctor prescribed medications, and individuals that assist self-injecting patients, from accidental needlesticks. The intended patient population is unrestricted and may include children and adults, and parenteral methods of administration.
Device Story
Antineedlestick accessory for pre-filled ISO Standard glass syringes (0.5 mL to 5.0 mL); passive needle guard mechanism; protects users from accidental needlesticks. Used by healthcare professionals, patients, and caregivers in clinical or home settings. No changes to design, materials, or technology from predicate; modification limited to expanded indications for use. Device provides safety benefits to non-healthcare professional users; usability testing confirms non-healthcare users can follow instructions.
Clinical Evidence
No clinical data. Usability testing performed to evaluate the ability of non-healthcare personnel to understand and follow the Instructions for Use, confirming the device performs as intended for the expanded user population.
Technological Characteristics
Passive needle guard accessory for ISO Standard glass syringes. Fits syringe sizes 0.5 mL, 1.0 mL, 1.0 mL long, 1.5 mL, 2.25 mL, 3.00 mL, and 5.0 mL. Mechanical design; no energy source or software.
Indications for Use
Indicated for healthcare professionals, patients self-injecting doctor-prescribed medications, and caregivers assisting self-injecting patients. Unrestricted patient population (children and adults) requiring parenteral administration.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Tamper Evident UltraSafe® Passive™ Delivery System
UltraSafe® Passive™ Delivery System (X-Series)
Submission Summary (Full Text)
{0}------------------------------------------------
#### 510(k) Summary
# Special 510 (k) Summary of Safety and Effectiveness
#### Company and Contact Information
Safety Svringes, Inc. 1939 Palomar Oaks Way, Suite A Carlsbad, CA 92011
Voice 760.918.9908 Fax 760.918.0565
Thomas L. Hall Sr. Director, Regulatory Affairs & Quality Assurance
### Device Identification
Trade Names:
- 1. Tamper Evident UltraSafe® Passive™ Delivery System
2. UltraSafe® Passive™ Delivery System (X-Series) Regulation Number - 21 CFR 880.5860 Regulation Name - Syringe, Piston (accessory) Classification - Class II Product Code - MEG
### Predicate Devices
- 1. Tamper Evident UltraSafe® Passive™ Delivery System
- 2. UltraSafe® Passive™ Delivery System (X-Series) (Subsequent to the submission of the premarket notification for this device, the trade name was changed from UltraSafe® Passive™ X-Series Needle Guard)
### Description of Modified Devices
The Safety Syringes UltraSafe Delivery System/Needle Guard devices are antineedlestick accessories for pre-filled ISO Standard glass syringes. ISO Standard glass syringes are supplied in multiple sizes. These devices are designed to fit syringe sizes 0.5 mL, 1.0 mL, 1.0 mL long, 1.5 mL, 2.25 mL, 3.00 mL, and 5.0 mL.
The only change from the currently legally marketed devices, and the subject of this premarket notification is the expansion of the Indications for Use, that is currently limited to healthcare professionals, to include physician prescribed medication selfinjecting patients and individuals who assist self-injecting patients. There are no changes to the design of these devices, materials used to produce these devices, or to their fundamental technology.
{1}------------------------------------------------
## Indications for Use
Single use devices that are indicated for use as an accessory with prefilled ISO Standard glass syringes to aid in the protection of healthcare professionals, patients who self-inject doctor prescribed medications, and individuals that assist selfiniecting patients, from accidental needlesticks.
The intended patient population is unrestricted and may include children and adults, and parenteral methods of administration.
### Compliance with Design Controls
Design Control activities for this device have been performed in conformance with the design control procedure requirements as specified in 21 CFR part 820.30.
## Risk Management
Risks associated with each product family have been identified, evaluated and, after mitigation, the risks fall into the "Broadly Acceptable Region" following ISO 14971.
## Conclusion
The expansion of the indications for use to include patients who self-inject doctor prescribed medications and individuals who assist self-injecting patients provides needed protection to an expanding population of users who can benefit with the use of anti-needlestick devices. The ability of non-healthcare personnel to understand and follow the Instructions for Use has been tested and the results of that testing indicates that the devices will perform as intended in the hands of typical nonhealthcare users.
This labeling change brings the safety benefits to a new user group. No issues have arisen to indicate that this labeling change poses unreasonable risks to nonhealthcare professional users.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three wing-like shapes, representing health, services, and people. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the eagle.
APR 2 8 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Thomas L. Hall Senior Director, Regulatory Affairs and Quality Assurance Safety Syringes, Incorporated 1939 Palomar Oaks Way, Suite A Carlsbad, California 92011
Re: K060743
Trade/Device Name: Tamper Evident UltraSafe® Passive™ Delivery System UltraSafe® Passive™ Delivery System (X-Series) Regulation Number: 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG Dated: April 12, 2006 Received: April 14, 2006
Dear Mr. Hall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federa! Register.
{3}------------------------------------------------
Page 2 - Mr. Hall
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number:
K060743
Device Name:
UltraSafe® Passive™ Delivery System
Indications for Use:
Single use devices that are indicated for use as an accessory with prefilled ISO Standard glass syringes to aid in the protection of healthcare professionals, patients who self-inject doctor prescribed medications, and individuals that assist self-injecting patients, from accidental needlesticks.
The intended patient population is unrestricted and may include children and adults, and parenteral methods of administration.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Antony D. ins
Ancolliesiology, General Hospital, : Connol Dental Devices
大46φ743
000 03
{5}------------------------------------------------
# Indications for Use
510(k) Number:
#### K060743 .
Device Name:
Tamper Evident UltraSafe® Passive™ Delivery System
Indications for Use:
Single use devices that are indicated for use as an accessory with prefilled ISO Standard glass syringes to aid in the protection of healthcare professionals, patients who self-inject doctor prescribed medications, and individuals that assist self-injecting patients, from accidental needlesticks.
The intended patient population is unrestricted and may include children and adults, and parenteral methods of administration.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Ching veng
ാത്രം (1) അടിമൃത്രം (General Hospital) In Com.of Dental Devices
K06243
000 02
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.