The Comprehensive® Primary Shoulder Stems are intended for hemi or total shoulder replacement: 1) Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis. 2) Rheumatoid arthritis. 3) Revision where other devices or treatments have failed. 4) Correction of functional deformity. 5) Fractures of the proximal humerus where other methods of treatment are deemed inadequate. 6) Difficult clinical management problems, including cuff tear arthropathy, where other methods of treatment may not be suitable or may be inadequate. Humeral/glenoid components with a porous coated surface coating are indicated for either cemented or uncemented biological fixation applications. (Metal backed glenoid components offer optional screw fixation.) The Comprehensive® Primary Shoulder Stems are intended for use only with the Bio-Modular® humeral heads and glenoid components and the Versa-Dial™ humeral head components. The devices are single-use implants.
Device Story
Comprehensive® Primary Shoulder Stems are titanium alloy humeral implants for hemi or total shoulder arthroplasty; designed for cemented or uncemented biological fixation via plasma-spray titanium porous coating. Stems are used exclusively with Biomet Bio-Modular® and Versa-Dial™ humeral heads and glenoid components. Implants are single-use; intended for surgical implantation by orthopedic surgeons in clinical/hospital settings. Device restores joint function and addresses degenerative disease, fractures, or revision needs. Substantial equivalence established through identical material composition to Bio-Modular® system and identical taper geometry to Comprehensive® Humeral Fracture Stems.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and design comparison to legally marketed predicate devices.
Technological Characteristics
Humeral stems constructed of titanium alloy with plasma-spray titanium porous coating. Designed for cemented or uncemented fixation. Dimensions and taper geometry match predicate devices. Single-use implant. No software or electronic components.
Indications for Use
Indicated for patients requiring hemi or total shoulder replacement due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, proximal humerus fractures, or difficult clinical management cases like cuff tear arthropathy, or revision after failed prior treatment.
Regulatory Classification
Identification
A shoulder joint metal/polymer/metal nonconstrained or semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits movement in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys such as cobalt-chromium-molybdenum (Co-Cr-Mo) and titanium-aluminum-vanadium (Ti-6Al-4V) alloys, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, or a combination of an articulating ultra-high molecular weight bearing surface fixed in a metal shell made of alloys such as Co-Cr-Mo and Ti-6Al-4V. The humeral component and glenoid backing have a porous coating made of, in the case of Co-Cr-Mo components, beads of the same alloy or commercially pure titanium powder, and in the case of Ti-6Al-4V components, beads or fibers of commercially pure titanium or Ti-6Al-4V alloy, or commercially pure titanium powder. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance: Shoulder Joint Metal/Polymer/Metal Nonconstrained or Semi-Constrained Porous-Coated Uncemented Prosthesis.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "BIOMET" in a stylized, bold font. Above the word "BIOMET" is the date "MAY 30 2006". The text is black and the background is white. The word "BIOMET" is in all caps.
b0692
## 510(k) Summary
| Preparation Date: | March 13, 2006 |
|--------------------|---------------------------------------|
| Applicant/Sponsor: | Biomet Manufacturing Corp. |
| Contact Person: | Susan Alexander |
| Proprietary Name: | Comprehensive® Primary Shoulder Stems |
| Common Name: | Shoulder Prosthesis |
## Classification Name:
- Prosthesis, Shoulder, Non-Constrained, Metal/Polymer, Cemented (21 CFR §888.3650) .
- . Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer, Cemented (21 CFR §888.3660)
- Shoulder Joint, Metal/Polymer/Metal, Non-Constrained or Semi-Constrained, Porous Coated, . Uncemented Prosthesis (21 CFR §888.3670)
- . Shoulder Joint, Humeral, (Hemi-Shoulder), Metallic, Uncemented Prosthesis (21 CFR §888.3690)
## Legally Marketed Devices To Which Substantial Equivalence Is Claimed:
- Bio-Modular® Shoulder System; Biomet Manufacturing; Corp. (K030710) .
- Comprehensive® Humeral Fracture Stems; Biomet Manufacturing Corp. (K023063) .
- . Integrated™ Shoulder System (Kirschner Shoulders with Titanium Plasma Spray); Biomet, Inc. (K961260)
#### Device Description:
The Comprehensive® Primary Shoulder Stems are humeral stems comprised of titanium alloy. The stems are available in various lengths and sizes. Portions of the devices are coated with plasma-spray titanium porous coating. The taper geometry of the new devices is the exactly the same as the predicate Comprehensive® Humeral Fracture Stems (K023063). The new devices are comprised of the exact same material as the predicate Bio-Modular® Shoulder System stems (K030710). The Comprehensive® Primary Shoulder Stems are designed to be used with Biomet's Bio-Modular® and Versa-Dial™ humeral heads and can be used in hemi or total shoulder arthroplasty.
Intended Use: The Comprehensive® Primary Shoulder Stems are indicated for:
- Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis. 1)
- 2) Rheumatoid arthritis.
- Revision where other devices or treatments have failed. 3)
- 4) Correction of functional deformity.
- 5) Fractures of the proximal humerus where other methods of treatment are deemed inadequate.
- 6) Difficult clinical management problems, including cuff tear arthropathy, where other methods of treatment may not be suitable or may be inadequate.
Humeral/glenoid components with a porous coated surface coating are indicated for either cemented or uncemented biological fixation applications. (Metal backed glenoid components offer optional screw fixation.)
5 - 1
MAHING ADDRESS -190. Box 587
그
SHIPPING ADDRESS S6 E. Rell In ac
미
く)がいいとこと ラフ・4.2(バア (いんぷり)
FAX -574.267.8137
1: MAIL bromet@houset.com
{1}------------------------------------------------
K060692 p²/₂
The Comprehensive® Primary Shoulder Stems are intended for use only with the Bio-Modular® humeral heads and glenoid components and the Versa-Dial™ humeral head components.
The devices are single-use implants.
Summary of Technologies: The technological characteristics (material, design, sizing, indications) of the Comprehensive® Primary Shoulder Stems are similar or identical to the predicate devices.
Non-Clinical Testing: Non-clinical laboratory testing was performed to determine substantial equivalence. The results indicated that the device was functional within its intended use.
Clinical Testing: None provided as a basis for substantial equivalence.
Unless otherwise indicated, all trademarks are property of Biomet.
{2}------------------------------------------------
## 510 (K)・ROUTE SLIP TRADITIONAL
| 510 (k) NUMBER | K060692 | PANEL OR | DIVISION DGRND | BRANCH |
|-----------------------------------------------------|---------------------------------------------------------|---------------|--------------------|--------|
| ELECTRONIC SUBMISSION | N | | | |
| TRADE NAME | COMPREHENSIVE PRIMARY SHOULDER STEMS | | | |
| COMMON NAME | SHOULDER PROSTHESIS | | | |
| PRODUCT CODE | | | | |
| APPLICANT | BIOMET MANUFACTURING CORP. | | | |
| SHORT NAME | BIOMETE | | | |
| CONTACT | SUSAN ALEXANDER | | | |
| DIVISION | | | | |
| ADDRESS | PO BOX 587<br>56. E. BELL DRIVE<br>WARSAW, IN 465810587 | | | |
| PHONE NO. | (574) 267-6639 | FAX NO. | (574) 372-1683 | |
| MANUFACTURER | BIOMET MANUFACTURING CORP.<br>STERIS ISOMEDIX SERVICES | REG NO. | 1450662<br>1450662 | |
| DATE ON SUBMISSION | 13-MAR-2006 | DATE DUE POS | 29-APR-2006 | |
| DATE RECEIVED IN ODE | 15-MAR-2006 | 75th DAY | 29-MAY-2006 | |
| DECISION | | DECISION DATE | | |
| Is this 510(k) identified as a Class III device | YES<br>YES | | NO | |
| Is this 510(k) the result of additional information | YES<br>YES | | NO | |
CD Enclosed
Ream SE
gss 5/24/06
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 30 2006
Biomet Manufacturing Corp. c/o Ms. Susan Alexander Regulatory Specialist P.O. Box 587 Warsaw, Indiana 46581-0587
Re: K060692
Trade/Device Name: Comprehensive Primary Shoulder Stems Regulation Number: 21 CFR 888.3670 Regulation Name: Shoulder joint metal/polymer/metal non-constrained or semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: MBF, KWT, KWS, HSD Dated: March 13, 2006 Received: March 15, 2006
Dear Ms. Alexander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
## Page 2 - Ms. Susan Alexander
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Akut Lemmer wo
← Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K060692
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Comprehensive® Primary Shoulder Stems
Indications For Use:
The Comprehensive® Primary Shoulder Stems are intended for heml or total shoulder replacement:
- 1) Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis.
- 2) Rheumatoid arthritis.
- 3) Revision where other devices or treatments have failed.
- 4) Correction of functional deformity.
- 5) Fractures of the proximal humerus where other methods of treatment are deemed inadequate.
- 6) Difficult clinical management problems, Including cuff tear arthropathy, where other methods of treatment may not be suitable or may be inadequate.
Humeral/glenold components with a porous coated surface coating are Indicated for either cemented or uncemented biological fixation applications. (Metal backed glenold components offer optional screw fixation.)
The Comprehensive® Primary Shoulder Stems are intended for use only with the Blo-Modular@ humeral heads and glenoid components and the Versa-Dial™ humeral head components.
The devices are single-use implants.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | <img alt="signature" src="signature.png"/> |
|------------------------------------------------------------|--------------------------------------------|
| Division of General, Restorative, and Neurological Devices | Page 1 of 1 |
| K060692 | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.