K060672 · One Stick, LLC · FPA · Mar 31, 2006 · General Hospital
Device Facts
Record ID
K060672
Device Name
ONE STICK Y EXTENSION SET AND VOLUMETER
Applicant
One Stick, LLC
Product Code
FPA · General Hospital
Decision Date
Mar 31, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The One Stick™ Volumeters™ and Y-Set families are indicated for the collection and handling of blood specimens prior to further testing.
Device Story
Blood collection assistance device; facilitates drawing samples from patients with suboptimal blood flow (e.g., elderly, pediatric, prone to vessel collapse/hemolysis). Device consists of male luer fitting for venipuncture attachment, tubing, Y-connector with Roberts clamp for flow control, and two needleless access ports connected to a vented volume collection tube. Used in clinical settings; operated by healthcare professionals. Device acts as an intermediary between patient and collection tube to manage flow and prevent vessel collapse. Benefits include improved sample collection success in difficult venous access cases. Components include standard market-available Y-sets and BD-manufactured syringe barrels.
Clinical Evidence
Bench testing only. Engineering reports dated July 13th and August 17th, 2005, confirmed that devices built to specifications function as expected. No clinical data provided.
Technological Characteristics
Sterile, single-patient use. Components: male luer fitting, tubing, Y-connector, Roberts clamp, needleless access devices, vented volume collection tube. Materials: Class VI polypropylene (vent cap), silicone (O-ring), Gore expanded PTFE supported membrane. Manufactured per cGMP standards. Sizes: 3cc and 5cc.
Indications for Use
Indicated for the collection and handling of blood specimens prior to further testing in patients, including elderly and pediatric populations, who may experience suboptimal blood flow, vessel collapse, or hemolysis during standard venipuncture procedures.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Medegen Medical Manufacturing Services Intravascular Administration Set (K051499)
Submission Summary (Full Text)
{0}------------------------------------------------
MAR 3 1 2006
g
One Stick 510(k) Application Section E
K060672 P.1 of 3
## 510(K) SUMMARY
### FOI RELEASABLE
Pursuant to § 513(i)(3)(A) of the Food, Drug, and Cosmetic Act, One Stick is r ursuall to g 515(1)(2)(11) of the Peop, Stag, Stag, Stag,
required to submit with this Premarket Notification ". . adequate summary of any information respecting safety and effectiveness or state that such information will be made available upon request of any person." One Stick chooses to submit a summary of information respecting safety and effectiveness.
| Classification Name: | | Intravascular Administration Set and<br>Extension Set<br>(includes needle less access devices/systems<br>and blood flow regulators) | |
|--------------------------------------------------------|-------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Common/Usual Name: | | Intravascular Administration Set | |
| Proprietary Name: | | One Stick™ Y Extentsion Set and Volumeter™ | |
| Device Classification: | | Class II | |
| Name | | Number | 21 CFR Ref. |
| Intravascular Admin Set<br>Subcutaneous, Intravascular | | FPA | 8 880.5440 |
| Owner/Operator: | One Stick LLC<br>P.O. Box 3575<br>Lubbock, Texas 79452-3575<br>(806) 842-3999 | | |
| Contact Person: | Tracy Magee, Vice President | | |
{1}------------------------------------------------
#### DESCRIPTION OF DEVICE
One Stick" Volumeters™ and Y-Set family of sets are supplied as sterile devices, and are intended for single patient use only. They are available in 3cc and 5cc sizes and are manufactured per cGMP Standards. The Y sets are comprised of standard components currently on the market. The Y sets are currently covered by 510(k) K051499. The barrel of the Volumeter™ is currently marketed by Becton Dickinson in their 3ml syringe. The Vent cap is molded with Class VI polypropylene, silicone O-ring and a Gore expanded PTFE supported membrane.
#### INDICATIONS FOR USE
The One Stick™ Volumeters™ and Y-Set families are indicated for the collection and handling of blood specimens prior to further testing.
## DESCRIPTIVE AND TECHNOLOGICAL CHARACTERISTICS OF PROPOSED AND PREDICATE DEVICES
The One Stick™ Device is a blood collection assistance device for drawing samples from patients with suboptimal blood flow for standard procedures (Elderly and pediatric patients, those patients that may experience blood vessel collapse or excessive hemolysis when conventional Vacutainers are used directly). The device consists of a male luer fitting for attaching to a venipuncture, a section of tubing connecting the luer to a Y-Connector with a Roberts clamp for interrupting flow, two additional sections of tubing attached to needless access devices. This Y device is then attached to a vented volume collection tube.
The Y set or device was approved under 510(k) K051499 to Medegen Medical Manufacturing Services. In our case these devices are being produced by Medegen for One Stick. The barrel of the Volumeter is produced by BD and is molded from their standard tooling.
## PERFORMANCE CHARACTERISTICS
The results of these test show that when the devices are built to specifications they all function as expected. Detailed results can be found in engineering reports dated July 13th and August 17th 2005.
{2}------------------------------------------------
One Stick 510(k) Application K464672 123 SF 3 510(K) A Section E
#### CONCLUSION
One Stick believes that the One Stick" Blood collection assistance device is substantially equivalent to the currently marketed Intravascular Administration Set by Medegen Medical Manufacturing Services 510(k) K051499. A comparison of the descriptive characteristics of these products demonstrate the One Stick Blood collection assistance device is equivalent in its indications for use, while being very similar in design and materials. In addition, One Stick has presented laboratory testing and biocompatibility information. The information presented provides assurance that the One Stick Blood collection assistance device will meet the minimum requirements that are considered acceptable for its intended use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" are arranged in a circular pattern around the caduceus. The seal is black and white.
MAR 3 1 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
One Stick Limited Liability Company C/O Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, Limited Liability Company 1394 25th Street, NW Buffalo, Minnesota 55313
Re: K060672
Trade/Device Name: One Stick™ Y Extension Set and Volumeter™ Regulation Number: 880.5440 Regulation Name: Intravascular administration set Regulatory Class: II Product Code: FPA Dated: March 10, 2006 Received: March 14, 2006
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent [(for the indications for use stated in the enclosure)] to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page-2 - Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements mount that a regulations and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration r ou intine (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and fisms (21 cert forth in the quality systems (QS) regulation (21 CFR Part 820); and if requirelates the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 1110 loter warket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), 11 you desire t the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Inter of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
C. Liu, Ph.D.
Chiu Lin, Ph.D. Director Division of Dental, Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
Kφb
er Use
# Indications for Use
510(k) Number (if known):
One Stick Tim YExtension Device Name:
Indications For Use:
FOR 7th Collections AND Ax بي جي جي هم INDICATED OF White -17-11-17 Blours SPECIMIANS Priva 70 FURTHER TESTING
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
Ten Service General H. : C: Lichial :
K06672
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.