← Product Code [KPS](/productcode/KPS) · K060631

# BIOGRAPH 64 AND BIOGRAPH 40 (K060631)

_Siemens Medical Solutions USA, Inc. · KPS · Mar 17, 2006 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K060631

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [KPS](/productcode/KPS.md)
- **Decision Date:** Mar 17, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1200
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The Siemens PET/CT scanners are combined Positron Emission Tomography (PET) and X-Ray Computed Tomography (CT) scanners. The Biograph 64 and 40 are intended to be utilized by appropriately trained health care professionals to: - Image and measure the distribution of injected positron emitting radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body; and - Produce cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plan taken at different angles or spiral planes taken at different angles. The Biograph 64 and 40 PET/CTs provide registration and fusion of highresolution metabolic and anatomic information. The PET component utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for fused PET and CT images. Additionally, the system maintains independent functionality of the PET and CT devices, allowing for single modality CT and / or PET diagnostic imaging.

## Device Story

Combined PET/CT scanner; integrates Siemens LSO HiRez PET scanner and Siemens Somatom Sensation 64 or 40 CT. Inputs: positron-emitting radiopharmaceuticals (PET) and X-ray transmission data (CT). Operation: CT data provides attenuation correction for PET and anatomical reference; system performs registration and fusion of metabolic and anatomic images. Used in clinical settings by physicians/technicians for oncology, neurology, cardiology. Outputs: fused cross-sectional images. Clinical utility: enables precise anatomical localization of metabolic activity; facilitates diagnostic assessment of physiologic functions. System supports independent PET or CT modality operation.

## Clinical Evidence

No clinical data provided; substantial equivalence supported by bench testing and comparison to predicate devices.

## Technological Characteristics

Combined PET/CT system. PET component: LSO (Lutetium Oxyorthosilicate) crystal-based scanner. CT component: Somatom Sensation 64 or 40-slice spiral CT. Connectivity: supports image registration and fusion. Energy: X-ray and positron emission. Software: reconstruction algorithms for PET and CT data.

## Regulatory Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

## Predicate Devices

- Biograph 6 ([K060060](/device/K060060.md))
- LSO PET/CT HiRez Series ([K050509](/device/K050509.md))
- Somatom Project P30F ([K040665](/device/K040665.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K06D63!

#### MAR 1 7 2006

### 510(k) Summary

as required by 21 CFR Part 807.87(h)

## Identification of the Submitter

| Submitter:          | Alaine Medio, RAC |
|---------------------|-------------------|
| Telephone Number:   | (865)218-2703     |
| Fax Number:         | (865)218-3019     |
| Date of Submission: | 2/10/06           |

### ldentification of the product

| Device Proprietary Name: | Biograph 64 and Biograph 40                                                                                           |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Common Name:             | Combination Positron Emission Tomography (PET)<br>and Computed Tomography (CT) system                                 |
| Classification Name:     | Emission Computed Tomography System per 21<br>CFR 892.1200<br>Computed Tomography X-Ray System per 21 CFR<br>892.1750 |

# Marketed Devices to which Equivalence is claimed

| Device                  | Manufacturer                                                   | 510(k) Number |
|-------------------------|----------------------------------------------------------------|---------------|
| Biograph 6              | Siemens Medical<br>Systems USA Inc.                            | K060060       |
| LSO PET/CT HiRez Series | CTI PET Systems (now<br>Siemens Medical<br>Solutions USA Inc.) | K050509       |
| Somatom Project P30F    | Siemens Medical<br>Systems USA Inc.                            | K040665       |

{1}------------------------------------------------

### Device Description:

The Biograph 64 and Biograph 40 are combined Positron Emission Tomography and X-Ray Computed Tomography scanners. These systems are designed for whole-body oncology, neurology and cardiology examinations. The Biograph systems provide registration and fusion of high-resolution metabolic and anatomic information from the two major components of each system, the Siemens LSO HiRez PET scanner and the Siemens Somatom Sensation 64 or 40 CT.

The combined PET/CT scanner is intended for use as a clinical, whole-body machine with high-end spiral CT and PET performance. The CT component provides attenuation correction for PET studies as well as precise anatomical reference through fused PET and CT images. In addition, the PET / CT system maintains independent functionality of the PET and CT scanning systems, allowing for most standard stand-alone CT and PET clinical diagnostic protocols to be available as well.

The Biograph 64 and Biograph 40 systems which are the subject of this application are substantially equivalent to the commercially available Biograph 6 PET/CT system (K060060) and the Siemens Somatom Project P30F CT system (K040665).

### Indications for Use:

The Siemens PET/CT scanners are combined Positron Emission Tomography (PET) and X-Ray Computed Tomography (CT) scanners. The Biograph 64 and 40 are intended to be utilized by appropriately trained health care professionals to:

- Image and measure the distribution of injected positron emitting . radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body; and
- Produce cross-sectional images of the body by computer reconstruction of X-. Ray transmission data from either the same axial plan taken at different angles or spiral planes taken at different angles.

The Biograph 64 and 40 PET/CTs provide registration and fusion of highresolution metabolic and anatomic information. The PET component utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for fused PET and CT images. Additionally, the system maintains independent functionality of the PET and CT devices, allowing for single modality CT and / or PET diagnostic imaging.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.

Public Health Service

MAR 1 7 2006

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Siemens Medical Solutions USA, Inc. % Mr. Daniel Lehtonen Responsible Third Party Official Intertek Testing Services NA, Inc. 2307 East Aurora Road, Unit B7 TWINSBURG OH 44087

Re: K060631

Trade/Device Name: Biograph 64 and Biograph 40 Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: KPS and JAK Dated: March 8, 2006 Received: March 9, 2006

Dear Mr. Lehtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act 's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industrv/support/index.html.

Sincerely yours.

Nancy C. Brigdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known):

K060631

Biograph 64 and Biograph 40 Device Name:

Indications for Use:

The Siemens PET/CT scanners are combined Positron Emission Tomography (PET) and X-Ray Computed Tomography (CT) scanners. The Biograph 64 and 40 are intended to be utilized by appropriately trained health care professionals to:

- Image and measure the distribution of injected positron emitting . radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body; and
- . Produce cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plan taken at different angles or spiral planes taken at different angles.

The Biograph 64 and 40 PET/CTs provide registration and fusion of highresolution metabolic and anatomic information. The PET component utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for fused PET and CT images. Additionally, the system maintains independent functionality of the PET and CT devices, allowing for single modality CT and / or PET diagnostic imaging.

Prescription Use × (Part 21 CFR 801 Subpart D) Over the Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy C. Bragdon

(Division Sign-Off Division of Reproductive, Abd and Radiological Devices 510(k) Number

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K060631](https://fda-staging.innolitics.com/device/K060631)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
