EPISPIN, EPISPIN II SAFETY, EPISPIN II, EPISPIN I, EPISPIN IV
Applicant
Pajunk GmbH
Product Code
CAZ · Anesthesiology
Decision Date
Jun 2, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5140
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Pajunk EpiSpin anesthesia conduction kit is indicated for administration of regional Combined Spinal-Epidural (CSE) anesthesia. The anesthesia conduction needle is intended for insertion into the spinal and epidural space, as well as catheter placement into the epidural space to facilitate a longer anesthetic effect. After the anesthesia conduction needle has been withdrawn from the patient, the cather tip can remain in the epidural space for as long as determined by the professional anesthetist and the instructions for use (up to 72 hours).
Device Story
EpiSpin is a single-use, sterile, latex-free Combined Spinal-Epidural (CSE) anesthesia kit. It utilizes needle-through-needle technology: a specialized Tuohy cannula with a 'backeye' aperture in the curved tip allows a Sprotte cannula to pass through into the spinal space without bending or metal-on-metal friction. The kit includes the Tuohy cannula, Sprotte cannula, LOR-syringe for verifying epidural space placement, Tuohy-Borst adapter, flat filter for particulate minimization, and Fixolong fixation device for catheter stabilization. Used in clinical settings by anesthesiologists to administer regional anesthesia. The Sprotte needle is removed after spinal injection, while the epidural catheter remains for up to 72 hours to provide prolonged anesthetic effect. The system facilitates precise needle placement and secure catheter management, benefiting patients by enabling both immediate spinal and extended epidural anesthesia.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence is based on design, material, and technological comparison to legally marketed predicate devices.
Indicated for patients requiring regional Combined Spinal-Epidural (CSE) anesthesia; intended for use by professional anesthetists for needle insertion into spinal/epidural spaces and epidural catheter placement for anesthetic administration up to 72 hours.
Regulatory Classification
Identification
An anesthesia conduction kit is a device used to administer to a patient conduction, regional, or local anesthesia. The device may contain syringes, needles, and drugs.
Predicate Devices
Pajunks Stimulong Plus Plexus Catheter (K043130, K033018)
{0}------------------------------------------------
EpiSpin Premarket Notification Submission
Image /page/0/Picture/2 description: The image shows the word "PAJUNK" in a bold, sans-serif font, set against a black background. Below this, the word "MEDIZINTECHNOLOGIE" is printed in a smaller, regular font. The text appears to be a logo or brand name, possibly for a medical technology company.
# 510(k) Premarket Notification Submission:
JUN - 2 2006
# Summary of Safety and Effectiveness
Date of Preparation: December 1, 2005
# Submitter Information/ production site:
Sterilizer:
Pajunk GmbH Medizintechnologie Karl-Hall-Straße 01 78187 Geisingen Germany Fon: +49(0)7704-9291-533 Fax: +49(0)7704-9291-605
# SteriPro Lab & EO Facility Dreieichstrasse. 7 64546 Mörfelden Germany Tel +49 6105 23091 or +49 (0) 6105 93470 Fax +49 6105 24760
## Contact:
Christian Quaß, Regulatory Affairs Fon: +49(0)7704-9291-533 Fax: +49(0)7704-9291-605 E-Mail: christian.quass@pajunk.com Establishment Registration Number: 9611612
# Device Information:
| Trade Names: | EpiSpin, EpiSpin Set, EpiSpin II Safety Kit |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Anesthesia Conduction Kit |
| Classification Name: | Anesthesia Conduction Kit |
| Classification Reference: | 21 CFR § 868.5140, April 1, 2005, |
| Proposed Classification: | Regulatory Class II |
| Proposed Product Classification Code: | CAZ, Kit, Conduction Anesthesia |
| Panel: | Anesthesiology |
| Predicate Devices: | 1. Pajunks Stimulong Plus Plexus Catheter K043130, K033018<br>2. Pajunks Plexolong Sets K042979, K023218, K013041<br>3. Pajunk Sprotte and various Anesthesia conduction cannulas K040965, K911202, K911260, K911221<br>4. BBrauns Espocan K932400 (including BBrauns Perifix K813186) |
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is in large, bold, white letters against a black background. Below this, the word "MEDIZINTECHNOLOGIE" is written in smaller, white letters.
### Device Description:
Pajunk's EpiSpin set is a single use, sterile, non-pyrogenic and latex free medical device kit. It is a Combined Spinal-Epidural (CSE) anesthesia kit consisting of two needles, one for insertion into the epidural space (a specialized Tuohy Cannula) and one for insertion into the spinal space (a Sprotte Cannula). This needle through-needle technology is accomplished using an aperture in the curved tip of the specialized Tuohy Cannula (called the "backeye"), for advancement of the Sprotte Cannula into the spinal space. The Spinal needle does not bend at all, but passes easily thru the Tuohy needle. There is no significant friction of metal against metal involved at all.
#### Predicate Devices:
The devices we claim substantial equivalence with are Pajunks Stimulong Plus Plexus Catheter cleared under K043130, K033018, Pajunks Plexolong Sets cleared under K042979, K023218, K013041 for the whole Kit and cleared kit components and BBrauns Espocan K932400 (including BBrauns Perifix K813186) especially for the Kit components Tuohy Back-Eye and the LOR-Syringe as well as the intended use of the whole Kit.
The detailed discussion of substantial eqivalence can be found in Section 12 of this submission.
#### Sterilization
The contract sterilizer and the sterilizing process other than a company name change (was IBA Griffith Micro Science, and now is Sterigenics) is the same as that used for anesth esia conduction cannulae cleared for market by FDA under 510(k) number K040965 and Kit Products Pajunks Stimulong Plus Plexus Catheter K043130, K033018 Pajunks Plexolong Sets K1042879, K023218, K013041.
#### Technology Characteristics:
The EpiSpin kit provides needle-through-needle technology, employing the use of a Tuohy Cannula featuring a backeye, through which a Sprotte Cannula can be inserted in the spinal space for immediate administration of spinal anesthesia. The Sprotte Cannula is then removed, and the epidural catheter is then placed for prolonged (up to 72 hours) administration of epidural anesthesia. Safety is optimized by the use of the catheters fixation adapter, Tuohy-Borst adapter, flat filter, Fixolong catheter fixation device, and optional Introducer Aid provided in the kit.
#### Conclusion:
The comparison between the predicate devices and the proposed devices in section 12 of this submission demonstrates that the proposed devices are safe and effective, as well as substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle-like symbol with three curved lines representing the bird's body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 2 2006
Mr. Christian Quass Regulatory Affairs Pajunk GmbH Medizintechnologie Karl-Hall-Strasse 01 78187 Geisingen GERMANY
Re: K060563
Trade/Device Name: EpiSpin Regulation Number: 21 CFR 868.5140 Regulation Name: Anesthesia Conduction Kit Regulatory Class: II Product Code: CAZ Dated: May 19, 2006 Received: May 22, 2006
Dear Mr. Quass:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Mr. Quass
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and inoments as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is written in large, white, bold letters on a black background. Below the black background, the word "MEDIZINTECHNOLOGIE" is written in smaller, white letters.
K060563 510(k) Number:
Device Name: LOR-Syringe
#### Indications for Use:
The Pajunk Loss of Resistance (LOR)-Syringe is intended for use in conjunction with an epidural needle, to verify the needle tip placement in the epidural space by the Loss of Resistance technique as explained in standard medical textbooks.
These syringes are not intended for injection or aspiration.
Over-The-Counter Use AND/OR Prescription Use (21 CFR 801 Subpart C) (Per 21 CFR 801.109)
(Please Do Not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Auy safiom
(off)
Anesthesiology, General Hospital,
on Control, Dental Devices
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is written in large, bold, white letters on a black background. Below the word "PAJUNK" is the word "MEDIZINTECHNOLOGIE" written in smaller, white letters.
K060563 510(k) Number:
Fixolong Fixation device Device Name:
#### Indications for Use:
The Pajunk Fixolong is an adhesive disk with catheter clips and flat filter attachment point on top. It is employed for fixing the catheter during long term anesthesia conduction.
The adhesive is hypo-allergenic and prevents accidental removal of epidural catheter. The catheter fastening clips allows stabilization of epidural catheter and prevents accidental movement. On the top there is an attachment site for flat filter to protect fixation
Prescription Use (Per 21 CFR 801.109)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ainy Dalton
nesthesiology, General H
: Number: K0601
Page 1 of 1
01 :
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for PAJUNK. The word "PAJUNK" is in large, bold, white letters on a black background. Below the word "PAJUNK" is the word "MEDIZINTECHNOLOGIE" in smaller, white letters.
510(k) Number: K060563
Device Name: Flat Filter
Indications for Use:
The Pajunk anesthesia conduction flat filter is a microporous filter used while administering to a patient injections of local anesthetics to minimize particulate (foreign material) contamination of the injected fluid.
| Prescription Use | X |
|----------------------|---|
| (Per 21 CFR 801.109) | |
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
sa த்துட்-Cit) on Signeral | General Hospital, . ان ان , Dental Devices
Amy Saloom
of Anas lesiology, General Hospital,
John Control, Lental De
Page 1 of 1
CQ
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is in large, bold, white letters against a black background. Below the word "PAJUNK" is the word "MEDIZINTECHNOLOGIE" in smaller, white letters.
# Indications for use K060563 510(k) Number: Tuohy Borst Adaptor Device Name: Indications for Use: The Pajunk Tuohy Borst Adaptor is connected to the needle via Luer connector. It ensures The I agains Tuony Doroth laster is common of a Luer Lock syringe or infusion device. Over-The-Counter Use AND/OR Prescription Use (21 CFR 801 Subpart C) (Per 21 CFR 801.109) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Aury Suliom
g. Off
ion of Anesthesiology, General Hospital,
ion Control, Dental Devices
Number: K060563
Page 1 of 1
{8}------------------------------------------------
K060563
Image /page/8/Picture/1 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is in large, bold, white letters against a black background. Below that, in smaller letters, is the word "MEDIZINTECHNOLOGIE".
#### Indications for use
510(k) Number:
Device Name:
EpiSpin Kits for Anesthesiology
#### Indications for Use:
The Pajunk EpiSpin anesthesia conduction kit is indicated for administration of regional The Pagain Epidural (CSE) anesthesia. The anesthesia conduction needle is intended for Combined opinal Lpharrir (002) and epidural space, as well as catheter placement into the epidural space to facilitate a longer anesthetic effect.
After the anesthesia conduction needle has been withdrawn from the patient, the cather tip can remain in the epidural space for as long as determined by the professional anesthetist and the instructions for use (up to 72 hours).
| Prescription Use | X |
|----------------------|---|
| (Per 21 CFR 801.109) | |
AND/OR
Over-The-Counter Use_ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Amy Solomon
1 .. 45-011) n of Anesthesiology, I .on Control. Dental D
Number: K060563
Page 1 of 1
Image /page/8/Picture/21 description: The image shows the text 'CQ 2006-05-18' in a simple, sans-serif font. The text appears to be a date or code, with 'CQ' possibly indicating a location or category. The date is formatted as year-month-day. The image quality is slightly degraded, but the text is still legible.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is in large, bold, white letters against a black background. Below this, the word "MEDIZINTECHNOLOGIE" is written in smaller, white letters.
510(k) Number: K060563
Device Name:
Anesthesia conducting Catheter
#### Indications for Use:
The Pajunk catheter is placed in the epidural space to facilitate a longer anesthetic effect.
After the anesthesia conduction needle has been withdrawn from the patient, the catheter tip can remain in the epidural space for as long as determined by the professional anesthetist and the instructions for use (up to 72 hours).
Prescription Use **X**
(Per 21 CFR 801.109)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**Page 1 of 1**
Auu Suicom
i Anesthesiology, General Hospital on Control. De
K060563
CQ
PMN ADD
..
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.