← Product Code [HWC](/productcode/HWC) · K060478

# ARTHREX BIO-COMPRESSION SCREW (K060478)

_Arthrex, Inc. · HWC · May 19, 2006 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K060478

## Device Facts

- **Applicant:** Arthrex, Inc.
- **Product Code:** [HWC](/productcode/HWC.md)
- **Decision Date:** May 19, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Arthrex Bio-Compression Screw is intended for fixation of fractures, osteotomies and arthrodesis in: Carpal, metacarpal, and small hand bone . tarsal and metatarsals . phalanges . Intra-articular fractures ● ankle . proximal and distal humerus . proximal and distal radius . proximal and distal ulna . osteochondral fixation and fractures . Osteochondritis Dissecans . Fixation of fractures and osteotomies about the knee . Oblique fractures of the fibula . . Reconstructive surgeries of the foot . malleolar fixation . where size of offered implant is patient appropriate.

## Device Story

Arthrex Bio-Compression Family consists of PLLA (Poly(L-Lactide)) bone fixation screws; used for fracture, osteotomy, and arthrodesis fixation in various skeletal sites including hand, foot, arm, and knee. Implants provided sterile; intended for surgical use by orthopedic surgeons. Device provides mechanical compression to bone fragments to facilitate healing. Substantial equivalence based on material characterization and comparison to predicate devices with similar indications.

## Clinical Evidence

No clinical data provided; substantial equivalence supported by in-house bench testing and cited literature.

## Technological Characteristics

Materials: Poly(L-Lactide) (PLLA). Form factor: Bone fixation screws of various lengths and diameters. Sterilization: Provided sterile. Principle: Mechanical bone fixation/compression.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K060478 page 1 of 2

O

# 510(k) Summary

| 510(k) Number:     | K06XXXX                                     |
|--------------------|---------------------------------------------|
| Company:           | Arthrex, Inc.                               |
| Address:           | 1370 Creekside Blvd., Naples, FL 34108-1945 |
| Telephone:         | (239) 643-5553                              |
| Facsimile:         | (239) 598-5508                              |
| Contact:           | Ann Waterhouse                              |
| Device Name:       | Arthrex Bio-Compression Family              |
| Classification:    | Screw, Fixation, Bone                       |
| Product Code:      | HWC                                         |
| Predicate Devices: |                                             |

#### Description:

The Arthrex Bio-Compression Family is currently comprised of enhanced Poly(I-Lactide) or PLLA implants in various lengths and diameters. They are offered sterile.

#### Indications for Use:

The Arthrex Bio-Compression Screw is intended for fixation of fractures, osteotomies and arthrodesis in:

(

- Carpal, metacarpal, and small hand bone .
- tarsal and metatarsals .
- phalanges .
- Intra-articular fractures ●
- ankle
- . proximal and distal humerus
- proximal and distal radius
- . proximal and distal ulna
- . osteochondral fixation and fractures
- . Osteochondritis Dissecans
- Fixation of fractures and osteotomies about the knee .
- Oblique fractures of the fibula .
- . Reconstructive surgeries of the foot
- malleolar fixation .

where size of offered implant is patient appropriate.

{1}------------------------------------------------

Pge yf

#### Technical Differences:

The basis for substantial equivalence is in supporting the use of your device in the same manner as that of a competitor. The in house test data provided, as well as articles cited, support the use of the Arthrex Bio-Compression Family in the indications sought.

#### Substantial Equivalence:

By definition, substantial equivalence means that a device has the same intended use and technical characteristics as the predicate device, or has the same intended use and different technological characteristics, but can be demonstrated to be as safe and effective as the predicate device for the previously cleared indications for use. The Arthrex Bio-Compression Family of Implants does not raise any questions regarding the safety and effectiveness of the implant. Furthermore, the materials used in construction of these devices are well characterized and have been used in predicate devices with similar indications.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows a logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" is arranged around the symbol in a circular fashion. The image is in black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 19 2006

Arthrex, Inc. % Ms. Ann Waterhouse, RAC Regulatory Affairs Project Manager 1370 Creekside Boulevard Naples, Florida 34108-1945

Re: K060478

> Trade/Device Name: Arthrex Bio-Compression Screw Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC Dated: April 20, 2006 Received: April 21, 2006

Dear Ms. Waterhouse:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Herbert Lerner vo

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 3 - Ms. Ann Waterhouse, RAC

2019 11:43 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

## Indications for Use

510(k) Number (if known):

Device Name:

Indications For Use:

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED )

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert Lerner

(Division Sign-Oif) Division of General, Restorative, and Neurological Devices

Page 1 of

**510(k) Number** k060478

{5}------------------------------------------------

510(k) Number (if known):

## Device Name: Arthrex Bio-Compression Screw

### Indications for Use:

The Arthrex Bio-Compression Screw is intended for fixation of fractures, osteotomies and arthrodesis in:

- Carpal, metacarpal, and small hand bone
- tarsal and metatarsals
- phalanges
- Intra-articular fractures
- ankle
- proximal and distal humerus
- proximal and distal radius
- proximal and distal ulna
- osteochondral fixation and fractures
- Osteochondritis Dissecans
- Fixation of fractures and osteotomies about the knee
- Oblique fractures of the fibula
- Reconstructive surgeries of the foot:
- malleolar fixation

where size of offered implant is patient appropriate.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use
(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1_ of _1

---

**Source:** [https://fda-staging.innolitics.com/device/K060478](https://fda-staging.innolitics.com/device/K060478)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).