K060464 · Beckman Coulter, Inc. · JPK · Mar 28, 2006 · Hematology
Device Facts
Record ID
K060464
Device Name
COULTER 5C CELL CONTROL
Applicant
Beckman Coulter, Inc.
Product Code
JPK · Hematology
Decision Date
Mar 28, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.8625
Device Class
Class 2
Indications for Use
5C Cell control is a hematology quality control mixture used to monitor the performance of coulter hematology analyzers listed on the Table of Expected Results in conjunction with specific Coulter reagents. The assigned values and expected ranges on the Table of Expected Results can be used to monitor instrument performance.
Device Story
5C Cell Control is a stabilized hematology quality control mixture; composed of treated human erythrocytes, stabilized platelet-sized components, and fixed erythrocytes simulating leukocytes. Used in clinical laboratory settings to monitor daily performance of automated hematology analyzers. Provided in three levels (Normal, Abnormal I, Abnormal II). Healthcare providers analyze the control on the instrument; compare results against assigned values and expected ranges provided in the Table of Expected Results. Discrepancies in control values indicate potential instrument performance issues, prompting calibration or maintenance. Benefits include ensuring accuracy and reliability of patient hematology test results.
Clinical Evidence
Bench testing only. Stability studies were conducted to support claims of 13 days (open vial) and 95 days (closed vial) for the RDW-SD parameter.
Technological Characteristics
Reference material composed of treated, stabilized human erythrocytes in isotonic, bacteriostatic medium; includes platelet-sized components and fixed erythrocytes. Derived parameter (RDW-SD) calculated from RBC histogram data. No changes to formulation or manufacturing process.
Indications for Use
Indicated for use as a quality control material for hematology analyzers to monitor instrument performance for parameters including WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV, and differential leukocyte counts.
Regulatory Classification
Identification
A hematology quality control mixture is a device used to ascertain the accuracy and precision of manual, semiautomated, and automated determinations of cell parameters such as white cell count (WBC), red cell count (RBC), platelet count (PLT), hemoglobin, hematocrit (HCT), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC).
Special Controls
*Classification.* Class II (special controls). Except when intended for use in blood components, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE
A. 510(k) Number:
K060464
B. Purpose for Submission:
The parameter RDW-SD (Red Cell Distribution Width-Standard Deviation) is being added to the Coulter 5C Cell Control.
C. Manufacturer and Instrument Name:
Beckman Coulter, Inc., Coulter 5C Cell Control is intended to be used on a new automated hematology analyzer called the Coulter LH780 which will be capable of analyzing the RDW-SD parameter.
D. Type of Test or Tests Performed:
Coulter 5C Cell Control is a quality control material for the parameters WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV, Ly%, Mo%, Ne%, Eo%, Ba%, Ly#, Mo#, Ne#, Eo#, and Ba#.
E. System Descriptions:
1. Device Description:
5C Cell Control is a hematology quality control mixture intended to be used with automated cell counters and differential cell counters. It is prepared from stabilized human blood so that repeated measurements by an automated cell counter, or differential cell counter, can be made to monitor instrument daily performance. 5C Cell Control consists of treated stabilized human erythrocytes, a stabilized platelet sized component, and fixed erythrocytes that simulate leukocytes. Three levels of varying component concentrations (Abnormal II, Abnormal I, and Normal) are offered.
2. Principles of Operation:
N/A
3. Modes of Operation:
N/A
{1}
4. Specimen Identification:
N/A
5. Specimen Sampling and Handling:
N/A
6. Calibration:
N/A
7. Quality Control:
5C Cell Control has assigned values and expected ranges which can be used to monitor instrument performance.
8. Software: N/A
FDA has reviewed applicant's Hazard Analysis and Software Development processes for this line of product types:
Yes _______ or No _______
F. Regulatory Information:
1. Regulation section:
21 CFR 864.8625
2. Classification:
Class II
3. Product code:
JPK
4. Panel:
Hematology (81)
2
{2}
G. Intended Use:
1. Indication(s) for Use:
5C Cell control is a hematology quality control mixture used to monitor the performance of coulter hematology analyzers listed on the Table of Expected Results in conjunction with specific Coulter reagents. The assigned values and expected ranges on the Table of Expected Results can be used to monitor instrument performance.
2. Special Conditions for Use Statement(s):
N/A
H. Substantial Equivalence Information:
1. Predicate Device Name(s) and 510(k) numbers:
COULTER® 5C® Cell Control, K912133 (cleared as COULTER® PX Cell Control)
2. Comparison with Predicate Device:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Product Description | 5C Controls consist of treated, stabilized human erythrocytes in an isotonic, bacteriostatic medium. It also contains a stabilized platelet-sized component, and fixed erythrocytes to simulate leukocytes. | Same |
| Intended Use | 5C Cell Control is a hematology quality control mixture used to monitor the performance of Coulter hematology analyzers listed on the Table of Expected Results in conjunction with specific coulter reagents. The assigned values and expected | Same |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Cellular Populations | ranges on the Table of Expected Results can be used to monitor instrument performance. | Same |
| Manufacturing Process | Erythrocytes, platelets, lymphocytes, monocytes, neutrophils, and eosinophils. | Same |
| Sample Preparation | 5C Cell Control is prepared from pre-screened units of whole human blood as well as animal blood. Stabilized RBC and platelet pools and fixed lymphocyte, monocyte, neutrophil, and eosinophils pools are manufactured independently. These pools are combined to provide the desired target ranges. | Same |
| Final Product Form | No special sample preparation is required. The control is a ready to use product. The control is warmed to ambient temperature, mixed by hand and analyzed on specified hematology analyzers. | Same |
| Open Vial Stability | Three levels, liquid, ready to use. | Same |
| Closed Vial Stability | 13 events over 13 days when stored at 2-8° C. | Same |
| | 95 days when stored at 2-8° C. | Same |
{4}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Assayed Parameters | WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PCT, PDW, Ly%, Mo%, Ne%, Eo%, Ba%, Ly#, Mo#, Ne#, Eo#, Ba# | Same with the additional parameter RDW-SD. |
I. Special Control/Guidance Document Referenced (if applicable):
N/A
J. Performance Characteristics:
1. Analytical Performance:
a. Accuracy:
N/A
b. Precision/Reproducibility:
N/A
c. Linearity:
N/A
d. Carryover:
N/A
e. Interfering Substances:
N/A
{5}
2. Other Supportive Instrument Performance Data Not Covered Above:
For Open Vial Stability and Closed Vial Stability, a total of nine lots of 5C Controls were tested through expiration starting near the end of the product's shelf life. Two tubes per level (3) were evaluated. Open Vial Stability was determined to be 13 events over 13 days when stored at 2-8° C. Closed Vial Stability was determined to be 95 days when stored at 2-8° C.
K. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
L. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
6
{6}
7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.