CODONICS HORIZON XL DIGITAL DIRECT LONG FILM IMAGER; CODONICS HORIZON XL-X LONG FILM IMAGER
K060440 · Codonics, Inc. · LMC · Mar 16, 2006 · Radiology
Device Facts
Record ID
K060440
Device Name
CODONICS HORIZON XL DIGITAL DIRECT LONG FILM IMAGER; CODONICS HORIZON XL-X LONG FILM IMAGER
Applicant
Codonics, Inc.
Product Code
LMC · Radiology
Decision Date
Mar 16, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2040
Device Class
Class 2
Indications for Use
The intended uses of the Horizon XL are identical to the Codonics Horizon® Series Medical Hardcopy Dry Imagers (premarket notification K021054); high resolution hard copy imaging of digital image source material combining color, film, reflective media in one imager. The hardcopy output includes however is not limited to, digital radiography, nuclear medicine, ultrasound, CT, MRI, CR and Radiation Therapy planning. Images are suitable for medical image diagnosis use and referral. The system is intended for use by medical radiologists, imaging modality specialists, and communications to referring physicians. The addition of 14" x 36" and 14" x 51" true size "long" film and (TBD) reflective Direct Vista Paper media permits digital direct orthopaedic application hardcopy including diagnosis and analysis of scoliosis, weight bearing spine/hip/knee, and long bone/hip Prosthetic and orthopedic appliances work-up and surgical planning. Horizon XL is applicable to true-size hardcopy of whole body CT, MRI and Angiographic and Venous flow imaging procedures.
Device Story
Horizon XL is a dry, thermal, grey-scale and color direct thermal printer/imager. It accepts digital image inputs via LAN (FTP, TCP/IP, DICOM) from modalities like CR, DR, CT, and MRI. The device uses image rendering algorithms and format conversion software to process various digital formats (TIFF, JPEG, DICOM, etc.) without lossy compression. Imaging is performed via a directly-modulated discrete-element thin-layer linear thermal print head. It produces diagnostic-quality hardcopy on transmissive film or reflective media using heat-sensitive silver or dye-diffusion technology. Used in clinical settings by radiologists and imaging specialists, the device provides true-size long-film hardcopies for orthopedic diagnosis, surgical planning, and patient referral. It eliminates the need for wet chemistry processing.
Clinical Evidence
Clinical tests documented effective application and expected results consistent with predicate devices. Bench testing included electrical compliance (UL 2601-1, CAN/CSA-C22.2 No 601.1-M90, IEC EN-60601-1, TUV/EN-60950), electromagnetic compatibility (EN-60601-1-2), and printer resolution pattern testing.
Technological Characteristics
Direct thermal printer/imager; grey-scale (heat-sensitive silver) and color (dye-diffusion) technology. Directly-modulated discrete-element thin-layer linear thermal print head. Connectivity via LAN (FTP, TCP/IP, DICOM). Supports 14" x 36" and 14" x 51" film. Software-controlled densitometry compensation and image rendering. No lossy compression.
Indications for Use
Indicated for medical radiologists, imaging modality specialists, and referring physicians to produce diagnostic-quality hardcopy images from digital sources (CR, DR, CT, MRI, nuclear medicine, ultrasound, radiation therapy). Supports true-size long-film imaging (14"x36", 14"x51") for orthopedic applications including scoliosis, weight-bearing spine/hip/knee, long bone/hip prosthetic work-up, and surgical planning.
Regulatory Classification
Identification
A medical image hardcopy device is a device that produces a visible printed record of a medical image and associated identification information. Examples include multiformat cameras and laser printers.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Horizon® Series Medical Hardcopy Dry Imagers (K021054)
Submission Summary (Full Text)
{0}------------------------------------------------
Kocoyyo
:
MAR 16 - 30
# Attachment 1 Summary of Safety and Effectiveness
* Attachments labeled "CONFIDENTIAL" as follows: Codonics, Inc. regards the information defined as part of this Attachment to be a trade secret and confidential in nature.
{1}------------------------------------------------
Kok 0440
Image /page/1/Picture/1 description: The image shows the logo for Codonics. The logo consists of a geometric shape on the left and the word "CODONICS" in bold, sans-serif font on the right. Below the word "CODONICS" is the tagline "We bring the future into focus" in a smaller, italicized font. The logo is black and white.
510(k) Summary-Special 807.92(c) Submitter Information: 807.92(a)(1) 1 1.1 Submitter: Codonics, Inc. 17991 Englewood Drive Middleburg Heights, Ohio 44130 1.2 Manufacturing Facility: Same as above 1.3 Representative: Not applicable at this time 1.4 Contact: Alan DeSantis, Phone:(440) 243-1198 / Fax : (440) 243-1334 17991 Englewood Drive Middleburg Heights, Ohio 44130 1.5 Date: February 17, 2006 Device Name 807.92(a)(2) & 807.92(a)(3) N 2.1 Camera, Multiformat, Radiological 2.2 Classification Name: Medical Image Hardcopy Devices Classification Number: 892.2040 2.3 Classification Code: LMC Horizon® Series MEDICAL IMAGE HARDCOPY MULTIMEDIA 2.4 Trade/Proprietary Name: PRINTERS Device Modification to Cleared Device K021054 Model Horizon XL- Long Film Imager Model Horizon XL-X Various Models TBD Horizon® Series Medical Hardcopy Dry Imagers (premarket 2.5 Predicate Devices: notification K021054 3 Device Description 807.92(a)(4)
## 3.1 Function
The Horizon® Series Imagers are dry, thermal, grey scale and grev scale/color (Ci model) direct thermal printer/imagers. The devices produce continuous tone, diagnostic quality BM images on transmissive film and reflective incident light viewed media. The color images produced via dye-diffusion technology are photographic medical color matched quality.
{2}------------------------------------------------
The Horizon® XL provides the capability to image True Size Long films used in Orthopedic applications, particularly in spine and long bone studies. These have been traditionally performed utilizing contact-screen film radiography utilizing special long, light-sensitive x-ray films in special long film/screen cassettes. As more imaging departments and centers become committed to PACS implementation and the application of digital radiography (CR and DR) acquisition, Horizon® XL brings long films into the all-digital world.
## 3.2 Scientific Concepts:
Digital images input directly or via Local Area Network is managed via communication standards including however not limited to FTP, TCP/IP, and DICOM. Images of a variety of digital formats are managed via industry standard format conversion software and image rendering algorithms including however not limited to TIFF, GIF, PCX, BMP, PBM, PGM, PPM, XWD, JPEG, SunRaster, SGI, Targa, DICOM, DEFF, and Postscript. Interpolation and scaling of images without Lossy data compression is employed in this device to maintain data integrity. Validated digital linear and visual linear routines and verified industry/modality specific Look Up Tables (LUTs) are applied to optimize color and CRT image hardcopy display results.
Imaging is accomplished via directly-modulated discrete-element thin-laver linear thermal print head technology. The recording medium is either heat sensitive silver in the case of DV grey scale film and DV DirectView reflective record imaging, or thermal heat activated dvediffusion of color in the case of CV DirectView and Transparency/film record imaging. The action of heat on the grey scale media produces a black dye in the medium. The action of heat on the dye-diffusion media produces a precision mixing of colors, which diffuse the medium top layer. The image formation is accomplished without wet chemistry processing common to many laser film imaging systems in use today.
## 3.3 Physical And Performance Characteristics:
The Horizon® XLSeries Imagers are high resolution hard copy imagers of digital image source material substantially equivalent to the Codonics Horizon® Series Medical Hardcopy Dry Imagers (premarket notification K021054) in system function and intended uses. The technology and applications are substantially equivalent to models of printers already cleared to market by the FDA, with the added capability to image 14" x 36" and 14" x 51" true size film and (TBD) reflective media. The Horizon® Series Imagers output diagnostic quality hardcopy images from digital sources over LAN: indirect or direct digital capture devices including Computed Radiography (CR) and Digital Radiography (DR). A multitude of digital formats include TIFF, GIF, PCX, BMP, PBM, PGM, PPM, XWD, JPEG, SunRaster, SGI, Targa, DICOM, DEFF, and Postscript. Lossy data compression is not employed in this device.
The system software is used to control the image management and machine functions. including densitometry compensation of printing power with film & media characteristics.
#### 4 Device Intended Use:
## 807.92(a)(5)
- 4.1 The intended uses of the Horizon XL is identical to the Codonics Horizon® Series Medical Hardcopy Dry Imagers (premarket notification K021054) as a combined color, film, reflective media imager. The hardcopy output includes however is not limited to, digital radiography. nuclear medicine, ultrasound, CT, MRI, CR and Radiation Therapy planning. Images are suitable for medical image diagnosis use and referral. The system is intended for use by medical radiologists, imaging modality specialists, and communications to referring physicians. The addition of 14" x 36" and 14" x 51" true size "long" film " film permits digital direct orthopaedic application hardcopy including diagnosis and analysis of scoliosis, weight
{3}------------------------------------------------
bearing spine/hip/knee, and long bone/hip Prosthetic and orthopedic appliances work-up and surgical planning. Horizon XL is applicable to true-size hardcopy of whole body CT, MRI and Angiographic and Venous flow imaging procedures.
### Device Technological Characteristics: 807.92(a)(6) 5
- 5.1 The characteristics of the Horizon XL model imagers compare substantially to the Horizon® Series Medical Hardcopy Dry Imagers (premarket notification K021054) in system function and intended uses. The technology and applications are substantially equivalent to models of printers already cleared to market by the FDA, with the added convenience of combining multi-sized color and grey scale film and reflective media with capability to image 14" x 36" and 14" x 51" true size "long" film for digital direct orthopaedic application hardcopy. Additional image memory, film transport software to permit longer film handling, modified film and film feed trays, mechanical extender arms to direct long film output to the external catch bin have been added to the standard Horizon Series imagers. Differences of note do not affect safety and effectiveness of the device, general radiographic imaging intended uses, or application methods. The device operates in a manner substantially equivalent to other cleared 892.2040 devices in this category.
#### e Testing and Equivalence: 807.92(b)(1), 807.92(b)(2) & 807.92(b)(3)
6.1 In the code implementation, electrical compliance tests, simulation, printer resolution pattern testing, and clinical studies, results and outcomes have been thoroughly reviewed with proper operation and intended functions verified. The device passed a series of electrical safety tests including UL 2601-1, CAN/CSA-C22.2 No 601.1-M90, IEC EN-60601-1. TUV/EN-60950:1992 and EN 60950/A1:1993. The devices comply with electromagnetic standards defined in EN-60601-1-2. Clinical tests have documented effective application and expected results consistent with predicate devices currently in commercial distribution.
Codonics believes the Horizon XL Model Imagers to be substantially equivalent to Medical Image Hardcopy Devices currently in commercial distribution in the U.S. We have selected Horizon® Series Medical Hardcopy Dry Imagers (premarket notification K021054) as the predicate device for our claim of substantial equivalence, Attachment 6 contains information describing these predicate devices and provides a comparison of the Horizon XL Imagers to the predicate device and describes how any differences of note are substantially equivalent.
## 7 Hazard Analysis and Safety Concerns
- 7.1 Hazard analysis on this product has been performed throughout the definition, design, and testing phases of the product development and implementation. This process has emphasized:
- Identification of potential hazards, their causes and their effects .
- Development of methodologies to control the occurrence of hazards and to constrain their . effects:
- . Determine any effect on patient safety and system effectiveness
The potential hazards associated with this software product are not different than those of the Horizon® Series Medical Hardcopy Dry Imagers (premarket notification K021054) image components. These are primarily related to the failure of computer system components, and may be variously obviated by decisions taken by the end users of the product. None of the failures are expected to materially contribute to patient death or injury.
It is our conclusion that no hardware or software component, operating in a properly configured environment, whose latent design defect would be expected to result in death or injury of the patient. Thus the "level of Concern" is "Minor".
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains an abstract image of an eagle. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are written around the edge of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 6 1303
Mr. Alan DeSantis Director, Quality and Regulatory Affairs Codonics. Inc. 17991 Englewood Drive MIDDLEBURG HEIGHTS OH 44130
Re: K060440
Trade/Device Name: Horizon® XL Medical Long Film Imager, Image Hardcopy Multimedia Printers Regulation Number: 21 CFR 892.2040 Regulation Name: Medical image hardcopy device Regulatory Class: II Product Code: LMC Dated: February 17, 2006 Received: February 21, 2006
Dear Mr. DeSantis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
T. Manayle. Brigeton
Naney C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Inclosure
{6}------------------------------------------------
510(k) Number (if known): _
Horizon® XL MEDICAL Long Film Imager IMAGE HARDCOPY Device Name: MULTIMEDIA PRINTERS
Model Horizon XL- Long Film Imager Models:
Indications For Use:
The intended uses of the Horizon XL are identical to the Codonics Horizon® Series Medical Hardcopy Dry Imagers (premarket notification K021054); high resolution hard copy imaging of digital image source material combining color, film, reflective media in one imager. The hardcopy output includes however is not limited to, digital radiography, nuclear medicine, ultrasound, CT, MRI, CR and Radiation Therapy planning. Images are suitable for medical image diagnosis use and referral. The system is intended for use by medical radiologists, imaging modality specialists, and communications to referring physicians. The addition of 14" x 36" and 14" x 51" true size "long" film and (TBD) reflective Direct Vista Paper media permits digital direct orthopaedic application hardcopy including diagnosis and analysis of scoliosis, weight bearing spine/hip/knee, and long bone/hip Prosthetic and orthopedic appliances work-up and surgical planning. Horizon XL is applicable to true-size hardcopy of whole body CT, MRI and Angiographic and Venous flow imaging procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use _ V (Per 21 CFR 801.109) OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
David A. Severson
Division Sian-Off Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.