BD MULTITEST 6-COLOR TBNK REAGENT WITH BD TRUCOUNT TUBES WITH MODELS 337 AND 166
Applicant
Bd Biosciences
Product Code
GKZ · Hematology
Decision Date
Jun 6, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Indications for Use
BD Multitest 6-Color TBNK reagent with BD Trucount tubes is intended for in vitro diagnostic use with the BD FACSCanto system to identify and determine the percentages and absolute counts of T, B, and natural killer (NK) cells as well as the CD4 and CD8 subpopulations of T cells in peripheral blood.
Device Story
Reagent kit uses fluorochrome-labeled antibodies to bind leukocyte surface antigens in whole blood samples. Processed in BD Trucount tubes containing a lyophilized pellet that releases a known number of fluorescent beads upon dissolution. Samples treated with BD FACS lysing solution to remove erythrocytes. Flow cytometer detects scatter and fluorescence signals as cells and beads pass through laser beams. Software analyzes signals to determine cell size, complexity, and fluorescence intensity. Absolute counts (cells/µL) calculated by comparing cellular events to bead events. Used in clinical laboratory settings by trained personnel. Output provides clinicians with lymphocyte subset profiles to aid in patient assessment. Benefits include standardized, automated quantification of immune cell populations.
Clinical Evidence
Bench testing only. Method comparison study (n=117) against predicate device showed high correlation (R² > 0.99 for most subsets). Precision/reproducibility evaluated over 21 days; repeatability and within-device precision reported for percentages and absolute counts. Linearity established for CD4, CD8, CD3, CD19, and CD16+CD56 subsets. Reference intervals established using 123 healthy donor samples (ages 18-65).
Technological Characteristics
Six-color direct immunofluorescence reagent (CD4 PE-Cy7, CD8 APC-Cy7, CD3 FITC, CD19 APC, CD16/CD56 PE, CD45 PerCP-Cy5.5). Uses BD Trucount tubes with lyophilized beads for absolute counting. Compatible with BD FACSCanto flow cytometer. Software-based analysis of scatter and fluorescence signals. Sodium azide (0.1%) used as preservative.
Indications for Use
Indicated for use with the BD FACSCanto flow cytometer and K3 EDTA whole blood to identify and quantify T, B, and NK cells, including CD4 and CD8 T-cell subpopulations, in peripheral blood for in vitro diagnostic use.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
BD FACSCanto system with BD FACSCanto clinical software (K041074)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k060375
B. Purpose for Submission:
New Device
C. Measurand:
T, B, and natural killer (NK) cells as well as the CD4 and CD8 subpopulations of T cells
D. Type of Test:
Quantitative and Semi-quantitative flow cytometric assay
A. Applicant:
BD BIOSCIENCES
F. Proprietary and Established Names:
BD Multitest 6-Color TBNK Reagent with BD Trucount Tubes
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| GKZ | Class II | 864.5220
Automated Differential Cell Counter | Hematology (81) |
H. Intended Use:
1. Intended use(s):
BD Multitest 6-Color TBNK reagent with BD Trucount tubes is intended for in vitro diagnostic use with the BD FACSCanto system to identify and determine the percentages and absolute counts of T, B, and natural killer (NK) cells as well as the CD4 and CD8 subpopulations of T cells in peripheral blood.
{1}
2. Indication(s) for use:
For use with the BD FACSCanto flow cytometer.
For use with whole blood collected in K3 EDTA tubes.
For use in the identification and determination of percentages and absolute counts of T, B, and natural killer (NK) cells as well as the CD4 and CD8 subpopulations of T cells in peripheral blood.
For in vitro diagnostic use.
3. Special conditions for use statement(s):
Not Applicable
4. Special instrument requirements:
BD Multitest 6-Color TBNK reagent is for use with the BD FACSCanto system.
I. Device Description:
BD Multitest 6-color TBNK reagent with BD Trucount tubes is packaged in a kit consisting of one vial containing sufficient reagent to yield 50 tests, and two pouches, each containing 25 BD Trucount tubes. One BD Trucount tube is required for each test, and each tube is for single-use only.
The BD Multitest 6-color TBNK reagent is provided in 1 mL of buffered saline with 0.1% sodium azide. The following antibodies comprise the reagent:
| Antibody | Clone | Fluorochrome |
| --- | --- | --- |
| CD4 | SK3 | Phycoerythrin-Cyanine 7 (PE-Cy7) |
| CD8 | SK1 | Allophycocyanin-Cyanine 7 (APC-Cy7) |
| CD3 | SK7 | Fluorescein Isothiocyanate (FITC) |
| CD19 | SJ25C1 | Allophycocyanin (APC) |
| CD16+CD56 | B73.1 / NCAM16.2 | Phycoerythrin (PE) |
| CD45 | 2D1 (HLe-1) | Peridinin chlorophyll protein – Cyanine 5.5 (PerCP-Cy5.5) |
BD Trucount tubes each contain a freeze-dried pellet of fluorescent beads held by a metal retainer at the bottom of the tube.
{2}
J. Substantial Equivalence Information:
| Predicate | Item | Similarities | Differences |
| --- | --- | --- | --- |
| K980858 - BD Multitest IMK Kit with BD Trucount Tubes | Intended Use | Identification and determination of percentages and absolute counts of the following mature human lymphocyte subsets in erythrocyte-lysed whole blood: T lymphocytes (CD3+, CD3+CD4+, and CD3+CD8+), B lymphocytes (CD3-CD19+), and NK lymphocytes (CD3-CD16+CD56+) | Same |
| K980858 - BD Multitest IMK Kit with BD Trucount Tubes | Components (including Cluster Designation, Conjugation, and Clone) | BD Multitest CD3/CD16+CD56/CD45/CD19 (1 vial, 50 tests) including CD3 FITC (SK7), CD16 PE (B73.1) + CD56 PE (NCAM16.2), CD45 PerCP (2D1), CD19 APC (SJ25C1), and buffer with 0.1% sodium azide
BD Multitest CD3/CD8/CD45/CD4 (1 vial, 50 tests) including CD3 FITC (SK7), CD8 PE (SK1), CD45 PerCP (2D1), CD4 APC (SK3), and buffer with 0.1% sodium azide
BD FACS Lysing Solution
BD Trucount Tubes (100 tubes) | BD Multitest 6-Color TBNK Reagent (1 vial, 50 tests) including CD4 PE-Cy7 (SK3), CD8 APC-Cy7 (SK1), CD3 FITC (SK7), CD19 APC (SJ25C1), CD16 PE (B73.1) + CD56 PE (NCAM16.2), CD45 PerCP-Cy5.5 (2D1), and buffer with 0.1% sodium azide
BD FACS Lysing Solution (not included with the reagent)
BD Trucount Tubes (50 tubes) |
| K980858 - BD Multitest IMK Kit with BD Trucount Tubes | Specificity | Specificities of antibodies have been verified by the International Workshop on Human Leukocyte Differentiation Antigens. | Same. |
| K980858 - BD Multitest IMK Kit with BD Trucount Tubes | Method to Identify Populations of Interest | Lyse/no-wash method using a two-tube panel with four-color antibody reagents (one tube stained with CD3/CD16+CD56/CD45/CD19, the other with CD3/CD8/CD45/CD4) to identify lymphocytes with specific cell-surface antigens, fluorescence triggering, and CD45 vs. SSC for gating. Uses fluorescent beads to quantify absolute counts | Same, except uses a one-tube panel with six-color antibody reagent |
{3}
4
| Predicate | Item | Similarities | Differences |
| --- | --- | --- | --- |
| K980858 - BD Multitest IMK Kit with BD Trucount Tubes | Control | Recommend use of commercially available whole blood control with established values for subset percentages and absolute counts | Recommend use of two levels of commercially available whole blood controls with established values for subset percentages and absolute counts. BD specifically recommends the use of BD Multi-Check and BD Multi-Check CD4 Low controls |
| K041074 - BD FACSCanto System with BD FACSCanto Clinical Software | Instrument and Software | BD FACSCanto flow cytometer with BD FACSCanto clinical software version 1.0 | Same, except uses version 2.0 of BD FACSCanto clinical software |
| K041074 - BD FACSCanto System with BD FACSCanto Clinical Software | System Setup | BD FACSCanto flow cytometer with BD FACSCanto clinical software version 1.0 and BD FACS 7-color setup beads | Same, except uses version 2.0 of BD FACSCanto clinical software |
| K980858 - BD Multitest IMK Kit with BD Trucount Tubes | Sample and Stain Stability | Anticoagulated blood stored at room temperature (20–25°C) must be stained within 48 hours of draw and then analyzed within 24 hours of staining | Anticoagulated blood stored at room temperature (20–25°C) must be stained within 24 hours of draw and then analyzed within 6 hours of staining |
| K980858 - BD Multitest IMK Kit with BD Trucount Tubes | Results | CD3+, CD3+CD4+, and CD3+CD8+ T lymphocytes; CD3-CD19+ B lymphocytes, and CD3-CD16+CD56+ NK lymphocytes expressed as percentages of total lymphocytes or as absolute counts (cells/μL) in whole blood | Same |
| K980858 - BD Multitest IMK Kit with BD Trucount Tubes | Sample Type | Erythrocyte-lysed whole blood, collected in K3 EDTA blood collection tubes | Same |
## K. Standard/Guidance Document Referenced (if applicable):
| STANDARDS |
| --- |
| Title and Reference Number |
| Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline (EP09-A2) |
| Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline (EP5-A) |
| How to Define and Determine Reference Intervals in the Clinical Laboratory; Approved Guideline - Second Edition (C28-A2) |
{4}
5
# Other Standards
| GUIDANCE | | | |
| --- | --- | --- | --- |
| Document Title | Office | Division | Web Page |
| Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA | OIVD | DIHD | http://www.fda.gov/cdrh/ode/guidance/1184.html |
| Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff | ODE | | http://www.fda.gov/cdrh/ode/guidance/337.html |
| Guidance for Industry and FDA Staff; Replacement Reagent and Instrument Family Policy | OIVD | | http://www.fda.gov/cdrh/oivd/guidance/950.html |
## L. Test Principle:
The BD FACSCanto system for six-color immunophenotyping consists of a BD FACSCanto flow cytometer, BD Multitest 6-color TBNK reagent, BD Trucount tubes, a Windows-based PC workstation, BD FACSCanto clinical software version 2.0 for automated acquisition and analysis.
Determining lymphocyte subset percentages and absolute counts requires: (1) obtaining a whole blood sample; (2) cell-surface antigen staining with the six-color monoclonal antibody reagent in a BD Trucount tube; (3) erythrocyte lysis; and (4) flow cytometric acquisition and analysis of list mode data.
When whole blood is added to the reagent, the fluorochrome-labeled antibodies in the reagent bind specifically to leukocyte surface antigens. During acquisition, the cells travel past the laser beam and scatter the laser light. The stained cells fluoresce. These scatter and fluorescence signals, detected by the instrument, provide information about the cell's size, internal complexity, and relative fluorescence intensity. BD Multitest reagents employ fluorescence triggering, allowing direct fluorescence gating of the lymphocyte population to reduce contamination of unlysed or nucleated red blood cells in the gate. A precise volume of sample is stained directly in a BD Trucount tube. The lyophilized pellet in the tube dissolves, releasing a known number of fluorescent beads. During analysis, the absolute number (cells/μL) of positive cells in the sample can be determined by comparing cellular events to bead events. BD FACSCanto clinical software (version 2.0 or higher) determines absolute counts.
{5}
6
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision was determined by running duplicate measurements of commercially available control at different levels of analyte concentration twice a day for a period of 21 operating days on three different instruments.
Standard deviations with 95% confidence limits were calculated for the within-run precision and the total precision of the lymphocyte subset percentages.
Percent CVs with 95% confidence limits were calculated for the within-run precision and the total precision of the lymphocyte subset percentage.
Repeatability of Lymphocyte Subset Percentages
| Lymphocyte Subset | N | Low Sample SD | Normal Sample SD |
| --- | --- | --- | --- |
| CD4 | 42 | 0.64 | 0.95 |
| CD8 | 42 | 1.07 | 0.65 |
| CD3 | 42 | 1.17 | 0.86 |
| CD19 | 42 | 0.89 | 062 |
| CD16+CD56 | 42 | 0.90 | 0.61 |
Within-Device Precision for Lymphocyte Subset Percentages
| Lymphocyte Subset | n | Low Sample SD | Normal Sample SD |
| --- | --- | --- | --- |
| CD4 | 42 | 0.69 | 1.23 |
| CD8 | 42 | 1.29 | 0.81 |
| CD3 | 42 | 1.23 | 0.90 |
| CD19 | 42 | 0.89 | 0.62 |
| CD16+CD56 | 42 | 0.96 | 0.62 |
Repeatability of Lymphocyte Subset Absolute Counts
| Lymphocyte Subset | n | Low Sample %CV | Normal Sample %CV |
| --- | --- | --- | --- |
| CD4 | 42 | 7.6 | 4.7 |
| CD8 | 42 | 4.1 | 4.7 |
| CD3 | 42 | 4.0 | 4.2 |
| CD19 | 42 | 5.7 | 5.3 |
| CD16+CD56 | 42 | 7.0 | 7.9 |
{6}
Within-Device Precision of Lymphocyte Subset Absolute Counts
| Lymphocyte Subset | n | Low Sample %CV | Normal Sample %CV |
| --- | --- | --- | --- |
| CD4 | 42 | 8.0 | 4.8 |
| CD8 | 42 | 5.0 | 5.4 |
| CD3 | 42 | 4.4 | 4.2 |
| CD19 | 42 | 6.0 | 5.7 |
| CD16+CD56 | 42 | 8.0 | 7.9 |
b. Linearity/assay reportable range:
Three whole blood samples were diluted with autologous plasma to known dilutions. High and low pools were created to span the expected CD4+ range of 200 to 3,000 cells/μL. Three replicates at each level were stained and acquired for each specimen. All samples from each specimen were acquired on the same day. Two BD FACSCanto instruments were used in this study. See linear ranges below:
| Lymphocyte Subset | Linear Range (cells/μL) | R² |
| --- | --- | --- |
| CD4 | 4 – 2,234 | 0.998 |
| CD8 | 158 – 1,125 | 0.991 |
| CD3 | 498 – 3,356 | 0.996 |
| CD19 | 71 – 447 | 0.989 |
| CD16+CD56 | 0 – 1,559 | 0.999 |
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Not applicable.
d. Detection limit:
Analytical limits at low levels were validated by accuracy, precision, and linearity studies.
e. Analytical specificity:
Not applicable.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
{7}
a. Method comparison with predicate device:
Whole blood samples (117) were collected and analyzed at two clinical evaluation sites. Lymphocyte subset percentage and absolute counts were enumerated with BD Multitest 6-color TBNK reagent in BD Trucount tubes and analyzed on the BD FACSCanto flow cytometer using BD FACSCanto clinical software. The results were compared with results from the BD Multitest™ IMK kit with BD Trucount tubes, and also were analyzed on the BD FACSCanto flow cytometer using BD FACSCanto clinical software. The results are as follows:
| Lymphocyte Subset | n | Unit | R2 | Slope | Intercept | Range |
| --- | --- | --- | --- | --- | --- | --- |
| CD3+ CD4+ | 117 | cells/μL | .995 | 0.965 | 6.0 | 4–1,593 |
| | | % | .998 | 1 | 0.0423 | 1–67 |
| CD3+CD8+ | 117 | cells/μL | .994 | 0.956 | 7.01 | 51–2,146 |
| | | % | .996 | 0.983 | 0.00592 | 11–83 |
| Total CD3+ | 117 | cells/μL | .995 | 0.968 | 13.5 | 107–3,403 |
| | | % | .996 | 0.985 | 0.895 | 34–88 |
| CD3–CD19 | 117 | cells/μL | .992 | 0.973 | 6.97 | 1–1,207 |
| | | % | .993 | 0.999 | 0.33 | 0–36 |
| CD3–(CD16+ CD56)+ | 117 | cells/μL | .99 | 0.98 | -0.291 | 7–918 |
| | | % | .992 | 0.985 | 0.0603 | 2–51 |
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
{8}
A total of 123 donor samples were collected and evaluated. Samples with an even gender distribution and representative of a healthy adult population – between the ages of 18 and 65 years – were tested and used to establish the lymphocyte subset reference intervals. Each sample was stained in duplicate, and the averaged values of the duplicates were used in determining the reference interval.
Reference Intervals for Lymphocyte Subset Percentages
| Lymphocyte Subset | n | Mean (%) | 95% Reference Interval (%) |
| --- | --- | --- | --- |
| CD4 | 123 | 46.4 | 28.2 – 62.8 |
| CD8 | 123 | 24.0 | 10.2 – 40.1 |
| CD3 | 123 | 71.1 | 49.1 – 83.6 |
| CD19 | 123 | 14.9 | 6.5 – 27.0 |
| CD16+CD56 | 123 | 11.7 | 4.2 – 25.2 |
Reference Intervals for Subset Absolute Counts
| Lymphocyte Subset | n | Mean (cells/μL) | 95% Reference Interval (cells/μL) |
| --- | --- | --- | --- |
| CD4 | 123 | 1106 | 441 – 2156 |
| CD8 | 123 | 583 | 125 – 1312 |
| CD3 | 123 | 1705 | 603 – 2990 |
| CD19 | 123 | 354 | 107 – 698 |
| CD16+CD56 | 123 | 266 | 95 – 640 |
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.