← Product Code [KJJ](/productcode/KJJ) · K060335

# FILE-EZE (K060335)

_Ultradent Products, Inc. · KJJ · Feb 24, 2006 · DE · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K060335

## Device Facts

- **Applicant:** Ultradent Products, Inc.
- **Product Code:** [KJJ](/productcode/KJJ.md)
- **Decision Date:** Feb 24, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** DE

## Indications for Use

Use File-Eze to aid in instrumentation and debridement of root canals.

## Device Story

File-Eze is a 19% EDTA dentin chelating agent in an aqueous, water-soluble solution with a lubricating base. Used by dentists during root canal procedures to aid in instrumentation and debridement. The solution is irrigated and rinsed from the canal; often used in conjunction with sodium hypochlorite solutions (e.g., ChlorCid), which produce a slight effervescent action upon contact with residual File-Eze to assist in removing extraneous material. The device facilitates mechanical cleaning of the root canal system.

## Clinical Evidence

Bench testing only. SEM study comparing dentin smear layer removal in root canal systems showed File-Eze is similar in effectiveness to RC-Prep. Biocompatibility study using the Agar Diffusion Method demonstrated that File-Eze is not cytotoxic.

## Technological Characteristics

19% EDTA aqueous, water-soluble chelating agent with lubricating base. Chemical-based root canal cleanser. No electronic components, software, or specialized materials.

## Predicate Devices

- RC-PREP (Prc-1976)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

L 060335

#### FEB 2 4 2006

## 510K SUMMARY

#### SUBMITTED BY: ULTRADENT PRODUCTS, INC. 505 WEST 10200 SOUTH SOUTH JORDAN, UTAH 84095

CONTACT PERSON: Tammy Lavery RA Senior Manager Phone: 1-801-553-4323

DATE PREPARED: 12-05-05 TRADE NAME: File-Eze COMMON NAME: Root Canal Cleanser CLASSIFICATION: Unclassified PREDICATE DEVICE: RC-PREP - Prc-1976 Device

DEVICE DESCRIPTION: File-Eze is a 19% EDTA dentin chelating agent in an aqueous, water-soluble solution with a lubricating base. File-Eze is irrigated and rinsed from canal as with other root canal preparations; ChlorCid and ChlorCid V (more viscous) sodium hypochlorite solution works well for this. Also, a slight effervescent cleansing action occurs when ChiorCid contacts residual File-Eze which helps remove extraneous material from canal.

INTENDED USE: Use File-Eze to aid in instrumentation and debridement of root canals.

The intended use of File-Eze and the predicate device, RC-Prep are the same, i.e. the cleansing and preparation of the root canal.

The physical properties of File-Eze and the predicate device are similar, i.e. viscosity, appearance, color, and odor.

TECHNOLOGICAL CHARACTERISTICS: File-Eze and the predicate device have the same active ingredient and the same intended use,

We believe that File-Eze is similar to RC Prep and because of the long established safe and efficacious use of the predicate device in the same intended use, the same active ingredient, the short duration of contact within the oral cavity, and the thorough removal of the product from the canal space, that File-Eze is safe and effective for the intended use. Additionally, the results of the following two studies further demonstrate the safety and effectiveness of File Eze for its intended use:

A study was done utilizing SEM`s comparing the removal of the dentin smear layer in root canal systems, the results show that File-Eze is similar in effectiveness compared to RC Prep. A biocompatibility study was done to evaluate cytotoxicity of File-Eze using the Agar Diffusion Method. The results of this study show that File-Eze is not cytotoxic.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

Public Health Service

FEB 2 4 2006

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Tammy Lavery Regulatory Affairs Senior Manager Ultradent Products, Incorporated 505 West 10200 South South Jordan, Utah 84095

Re: K060335

Trade/Device Name: File-Eze Regulation Number: N/A Regulation Name: Root Canal Cleanser Regulatory Class: Unclassified Product Code: KJJ Dated: December 05, 2005 Received: February 09, 2006

Dear Ms. Lavery:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

### Page 2 - Ms. Tammy Lavery

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA 's issualice of a substition of the requirements of the Act or
that FDA has made a determination that your device complies . You must comply with that FDA has made a delectimation that your devices. You must comply with any Federal statures and regulations administered by outding (21 CFR Part 807);
all the Act's requirements, including, but not limited to: registration and listing (21 CFR Pa all the Act's requirements, including, but not immed in see forth in the quality
labeling (21 CFR Part 801); good manufacturer and security of the colunt rediction labeling (21 CFR Part 801), good manufacturing prosolicable, the electronic product radiation
systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic pro systems (QS) regulation (21-54-12-12-12-12-21-21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to ocgii maketing your antial equivalence of your device to a legally
premarket notification. The FDA finding of substantial equivales and thus, pe prematket notification. The FDA Inding of substantial equice and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your ac not the regulation of the regulation entitled, a contact the Office of Comphance at (210) 276 or over 807.97). You may obtain other
"Misbranding by reference to premarket notification" (21 CFR Part Riginian of Small "Misoranding of Terence to promation on the Act from the Division of Small
general information on your responsibilities under the Act from the 1990 for general information on your responsionnes and its toll-free number (800) 638-2041 or and index think Manufacturers, International and Colisal.http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chas S. Liem, Ph.D.

Chiu S. Lin, PhD Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

#### Enclosure

{3}------------------------------------------------

# INDICATIONS FOR USE

510(k) Number (if known):

Device Name: File-Eze

Indications for use:

Use File-Eze to aid in instrumentation and debridement of root canals.

Prescription Use:_ X (Pcr 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use:_ (21 CFR 807 Subpart C)

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

---

**Source:** [https://fda-staging.innolitics.com/device/K060335](https://fda-staging.innolitics.com/device/K060335)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com