This device is used to monitor physiologic parameters, including SpO2, PR and NIBP, on adult, pediatric, and neonatal patients in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. This device is not intended for transport or home use.
Device Story
VS-800 Vital Signs Monitor is a battery or line-powered bedside monitor; acquires physiological signals (NIBP, SpO2); converts signals to digital data; processes data for alarm conditions; presents numeric values and waveforms on front panel; optional built-in recorder provides hard copies of trends and waveforms. Used in healthcare facilities by physicians or medical staff. Output assists clinical decision-making by providing real-time patient monitoring and alerts. Benefits include continuous physiological tracking and timely notification of abnormal patient conditions.
Clinical Evidence
Bench testing only. Validation included environmental testing per FDA guidance, hazard analysis, and software validation. Compliance verified against EN865, IEC60601-1, IEC60601-1-1, IEC60601-1-2, IEC60601-1-4, IEC60601-2-30, ISO14971, and ANSI/AAMI SP10 for NIBP accuracy.
Technological Characteristics
Bedside vital signs monitor; battery or line-powered. Parameters: NIBP, SpO2, PR. Connectivity: bedside or central monitoring system. Standards compliance: EN865, IEC60601-1 series, IEC60601-2-30, ISO14971, ANSI/AAMI SP10. Software-controlled measurement and alarm generation.
Indications for Use
Indicated for monitoring SpO2, pulse rate (PR), and non-invasive blood pressure (NIBP) in adult, pediatric, and neonatal patients within healthcare facilities. Not for transport or home use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
### 510(K) SUMMARY Exhibit #1
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: K060281
# Submitter:
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Tel: +86 755 2658 2888 Fax: +86 755 2658 2680
#### Contact Person: ●
- Li Dongling
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
#### Date Prepared: ●
Jan. 16, 2005
# Name of the device:
- Trade/Proprietary Name: VS-800 Vital Signs Monitor
- Common Name: Vital Signs Monitor .
- Classification
| 21 CFR 870.2300 Cardiac monitor (including cardiotachometer and rate alarm) | Class |
|-----------------------------------------------------------------------------|-------|
| 21 CFR 870.1130 Non-Invasive blood pressure measurement System | Class |
| 21 CFR 870.2700 Oximeter, Pulse | Class |
| 21 CFR 870.2710 Ear Oximeter, Pulse | Class |
# Legally Marketed Predicate Device:
K043348 PM-8000 Patient Monitor (by Mindray Co., Ltd.)
{1}------------------------------------------------
# Description:
The VS-800 Vital Signs Monitor is battery or line-powered monitor used on human patient. The Vital Signs Monitor acquires the physiological signals non-invasive blood pressure (NIBP) and pulse oxygen saturation of the blood (SpO2). These physiological signals are converted into digital data and processed. The VS-800 Vital Signs Monitor examines the data for alarm conditions and presents them on the front panel. The Vital Signs Monitor also provides user with the convenient operating control and human-machine interface (HMI),
The optional built-in recorder provides hard copies of all digital data and waveforms as well as Tabular and Graphic Trend Information, and storage the previous monitoring data information when power off accidentally.
## Statement of intended Use:
This device is used to monitor physiologic parameters, including SpO2, PR and NIBP, on adult, pediatric, and neonatal patients in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians.
This device is not intended for transport or home use.
# Comparison of Technological Characteristics:
The VS-800 Vital Signs Monitor is substantially equivalent to systems currently marketed predicate devices. The design, components, storage technology and energy source of the VS-800 Vital Signs Monitor are similar to the predicate device named PM-8000 Patient Monitor (K#043348). Both VS-800 Vital Signs Monitor and PM-8000 Patient Monitor provide a means for interfacing with a patient, collecting parameter specific physiological data, and processing the data for alarm generation and display of numeric values and waveforms on a bedside or central monitoring system. The monitoring parameters of the PM-8000 Patient Monitor cover those of the VS-800 Vital Signs Monitor. And the parameters' specification of the VS-800 Vital Signs Monitor, including SpO2, PR and NIBP, is similar to the predicate PM-8000 Patient Monitor,
The notable difference between the technical specifications of the VS-800 Vital Signs Monitor and PM-8000 Patient Monitor is shown as following:
While the monitoring parameters of the PM-8000 Patient Monitor cover those of the VS-800 Vital Signs Monitor, but the parameters' specification of the VS-800 Vital Signs Monitor, including SpO2, PR and NIBP, is similar to the predicate PM-8000 Patient Monitor, and all these parameters of the pending device and the predicate device comply with some
. . . . . . . . . . . . . . . . . . . . . . .
{2}------------------------------------------------
ANSI/AAMI standards, IEC standards, EN standards and ISO standards.
These technological differences do not affect the safety or efficacy of the device. Any safety issues that may be raised by a software controlled medical device are the same issues already addressed by the predicate device and are addressed in the systems hazard analysis and in the system validation.
# Testing:
Laboratory testing was conducted to validate and verify that the VS-800 Vital Signs Monitor met all design specifications and was substantially equivalent to the predicate devices. This testing consisted of all environmental testing identified in the FDA's DCRND November 1993 "Reviewer Guidance Document for Premarket notification Submissions" Draft Guidance Document. Finally, a hazard analysis of the system and its software was performed and testing was conducted to validate the systems overall operation. Some safety testing has been performed by third party agencies to ensure the device complies with applicable industry and safety standards. The VS-800 Vital Signs Monitor has also been tested to assure compliance with the requirements of various published standards, including EN865. IEC60601-1, IEC60601-1-1, IEC60601-1-2, IEC60601-1-4, IEC60601-2-30, and ISO14971.
Testing of the non-invasive blood pressure portion of the system was conducted according to the requirements outlined in the ANSI/AAMI Standards SP10 "Electronic automated sphygmomanometers."
Although the device is neither life supporting nor life sustaining, diagnostic information derived from the use of the device and alarms generated by the device may be critical to the proper management of the patient. So, the areas of risk for this device are the same as other devices in this class, and the following:
- Electrical shock
chassis Excessive electrical leakage current can disturb the normal electrophysiology of the heart.
- Misdiagnosis
- ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Inadequate design of the signal processing and measurement circuitry or program can lead generation of inaccurate diagnostic data.. If inaccurate diagnostic data are used in managing the patient, the physician may prescribe a course of treatment that places the patient at risk unnecessarily.
- --Inadequate design of the device's software, used to make various measurements, can lead to generation of inaccurate diagnostic data. If inaccurate diagnostic data are used in managing the patient, the physician may prescribe a course of treatment that places the patient at risk unnecessarily.
- Inadequate design of the systems ability to alert the users through audible and ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{3}------------------------------------------------
visual indicators, can lead to user mistrust and/or inadequate response to the patient's condition. If an inadequate response to the patient's condition should occur the patient may unnecessarily be placed at risk.
# Conclusion:
The conclusions drawn from clinical and laboratory testing of the VS-800 Vital Signs Monitor demonstrates that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate device, the PM-8000 Patient Monitor numbered K#043348(by Mindray Co., Ltd).
and the comments of
{4}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2 3 2006
Shenzhen Mindray Bio-Medical Electronics Co. Ltd. c/o Ms. Susan D. Goldstein-Falk mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, New York 11021
Re: K060281
> Trade Name: VS-800 Vital Signs Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Patient Physiological Monitor (with Arrhythmia Detection or Alarms) Regulatory Class: Class II (two) Product Code: MHX Dated: June 16, 2006 Received: June 19, 2006
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent {for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2 - Ms. Susan D. Goldstein-Falk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to I ogally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation enjitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours.
Bhemnia for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Heath
Enclosure
{6}------------------------------------------------
## Page
510(k) Number (if known):
## Device Name: VS-800 Vital Signs Monitor
### Indications For Use:
The VS-800 Vital Signs Monitor is used to monitor physiological parameters, including Sp02, PR and NIBP on adults, pediatric and neonatal patients in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians.
The device is not intended for transport or home use.
Prescription Use × (Per 21 CFR 801 Subpart D)
Over-The Counter Use OR (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
------------------------------------
B. Bairamova
Division Sign Off
(Division Sign-Off) Division of Cardlovascular Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.