VIRTUAL IMPLANT PLACEMENT (VIP) DENTAL IMPLANT SURGERY PLANNING SOFTWARE
Applicant
Implant Logic Systems, Ltd.
Product Code
LLZ · Radiology
Decision Date
Mar 17, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
This device employs previously scanned DICOM CT images in a software tool which serves as an aid to visualizing and pre-planning of dental implant surgery.
Device Story
Virtual Implant Placement (VIP) is a PC-based software tool for dental implant surgery pre-planning. Input consists of DICOM CT scan images from external scanners. Clinicians process images by removing non-relevant data and defining a curve to reformat axial CT data into panoramic and cross-sectional views. The software provides measurement tools (distance, angle, rectangular, elliptical) and allows placement of virtual generic or manufacturer-specific implant models to assess size and position relative to patient anatomy. Used by dentists in clinical settings to evaluate bone density and anatomy for surgical planning. Output assists clinicians in visualizing surgical sites and designing surgical appliances; does not perform automated diagnosis.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
PC-based software; compatible with Windows 98, 2000, and XP. Requires minimum 256 MB RAM and 10 GB hard drive. Processes DICOM CT images. Features include image reformatting (axial to panoramic/cross-sectional), virtual implant modeling, and measurement tools. No specific hardware integration; standalone software application.
Indications for Use
Indicated for dental implant surgery planning in patients requiring visualization of jaw anatomy and bone density using DICOM CT images. Intended for use by dental clinicians.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K060267
MAK 1 7 2006
## EXHIBIT 2 510(k) Summary Implant Logic Systems, Ltd. 76 Spruce St. Cedarhurst, NY 11516 Tel 516-295-1121 Fax 516-295-2992 www.implantlogic.com Contact: Michael Klein, DDS, President January 30, 2006
- 1. Identification of the Device: Proprietary-Trade Name: Virtual Implant Placement™ (VIP) Dental Implant Surgery Planning Software Classification Name/Product Code: Picture archiving and communications system, 90 LLZ Common/Usual Name: Picture archiving and communications system
- 2. Equivalent legally marketed devices: ImplantMaster K042212 (I-Dent Ltd.) and SimPlant system, K033849, (Materialise.)
- 3. Indications for Use (intended use): This device employs previously scanned DICOM CT images in a software tool which serves as an aid to visualizing and pre-planning of dental implant surgery.
- Description of the Device: Virtual Implant Placement, or simply VIP, is a software 4. program that will allow dental implant clinicians to pre-plan their implant surgeries and/or to design surgical appliances that will be used during surgery. The program will presents the clinician with various reformatted CT images of their patient's jaw(s), allow the placement and manipulation of virtual implants, and provide measurement and other tools to assist the clinician. In typical usage a dentist evaluating a patient for dental implant surgery will often refer the patient for a CT scan to better visualize the patient's anatomy, and check the amount and density of the bone for its suitability for placing implants. The CT scan site will return the axial images from the CT scan on a CD in industry-standard DICOM format. Upon receipt of the CD, the doctor will "process" the case using VIP. Axial images are well-known to radiologists, but foreign to dentists. Processing involves the removal of unnecessary images which are outside the region of interest, and drawing a curve which will be used for the later reformatting of the data to produce images more familiar to dentists. After opening a disk of images, VIP will display the axial images and thumbnails of these, along with a scout view and a checklist of stops to follow in processing the case. After the case has been processed, the axial data will be processed to make panoramic images, which are parallel to the curve that was drawn during processing, and cross-sectional images, which are perpendicular to the panoramic image. Both types of images are normally generated by the Panorex machines dentists are familiar with. Since the primary purpose of VIP is to aid in the planning of implant surgeries, VIP will allow the surgeon to place simulated implants on the image and to gauge their size and position
{1}------------------------------------------------
relative to the surrounding anatomy. The simulated implants will be generic models of standard dental implants, which range from cylindrical to conical. When the data becomes available from various implant manufacturers, VIP will allow the user to pick from specific, currently-manufactured implants to approximately model any of their favorite implants.
| Description | SimPlant system, K033849, (Materialise.) | ImplantMaster K042212 (I-Dent Ltd.) | Virtual Implant Placement™ (VIP) |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Image Source | CT Scanner | DICOM CT | DICOM CT |
| Main indication | Indicated for use as a is a medical<br>front-end software that can be used<br>by medically trained people for the<br>purpose of visualizing gray value<br>images. It is indicated as a software<br>interface and image segmentation<br>system for the transfer of imaging<br>information from a medical<br>scanner such as a CT scanner or a<br>Magnetic Resonance scanner. It is<br>also indicated for use as a planning<br>and simulation software for dental<br>implant placement and surgical<br>treatment | Uses DICOM CT data<br>for visualization,<br>diagnosis and treatment<br>planning for dental<br>implant surgery. | This device employs<br>previously scanned DICOM<br>CT images in a software tool<br>which serves as an aid to<br>visualizing and pre-planning<br>of dental implant surgery. |
| Tools | Visualization, Implant placement,<br>measurement of distances, angles<br>and density | Visualization, Implant<br>placement. | Visualization,<br>Implant placement,<br>Distance measurement,<br>Angle measurement,<br>Rectangular measurement,<br>Elliptical measurement |
| Host platform | PC | PC | PC |
| Operating system | Not specified | Windows XP SP2 or<br>Win 2000 SP4 | Windows 98, 2000 and XP® |
| Host RAM | Not specified | 512 Mb | 256 MB RAM |
| Host Magnetic<br>Storage | Not specified | 1 GB | 10 GB hard drive |
| CD ROM | Yes (for installation) or download | Yes (for installation) | Yes (for installation) |
| Host Processor<br>Speed | Not specified | Intel P4 1.6 GHz<br>(recommended: 2.8<br>GHz) | Pentium III, 500 MHz |
- 5. Safety and Effectiveness, comparison to predicate devices:
- 6. Conclusion: In all important respects, the VIP is substantially equivalent to one or more predicate systems, including the one named above. This is based on testing to verify compliance with product specifications.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its wings. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Implant Logic Systems Ltd. % Mr. Daniel Kamm, P.E. Principal Consultant Kamm & Associates P.O. Box 7007 DEERFIELD IL 60015
Re: K060267
MAR 1 7 2006
Trade/Device Name: Virtual Implant Placement™ (VIP) Dental Implant Surgery Planning Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 30, 2006 Received: February 1, 2006
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely vours.
Nancy C. Brigdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: Virtual Implant Placement™ (VIP) Dental Implant Surgery Planning Software
Indications For Use:
This device employs previously scanned DICOM CT images in a software tool which serves as an aid to visualizing and pre-planning of dental implant surgery.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Chisholm
cal Devices 510(k) Num
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.