K060194 · March Healthcare Corporation · DXN · Feb 13, 2006 · Cardiovascular
Device Facts
Record ID
K060194
Device Name
MARCH HEALTHCARE HEALTH MONITORING KIT
Applicant
March Healthcare Corporation
Product Code
DXN · Cardiovascular
Decision Date
Feb 13, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The Health Monitoring Kit (HMK) is a non-invasive, spot-measuring device intended to measure heart rate and systolic and diastolic blood pressure using the oscillometric method: functional oxygen saturation levels: pulse rate; and heart and lung sounds of both pediatric patients aged three years and older and adult patients under the care of a physician in either clinical or home enviroments.
Device Story
Portable HMK device performs spot-checks of vital signs; inputs include oscillometric blood pressure (cuff), SpO2 (photodetector/red-IR light), and heart/lung sounds (electronic stethoscope). Device transforms inputs into physiological data; displays results on built-in LCD. Operates as standalone unit or connects to host via Bluetooth/USB for remote telehealth. In telehealth, stethoscope audio and vital data transmit to remote computer for clinician review. Used by clinicians or patients in clinical/home environments. Facilitates remote monitoring and clinical decision-making by providing objective vital sign data to off-site healthcare professionals.
Clinical Evidence
Bench testing only. Comparative performance testing conducted between HMK and Welch Allyn 52000 VSM (NIBP and SpO2) and Cardionics E-Scope (frequency response). Results demonstrated equivalent performance. Risk analysis performed per ISO 14971. Device met all applicable electrical, mechanical, EMC, and environmental safety standards.
Technological Characteristics
Portable handheld unit; ABS injection-molded case. NIBP: oscillometric step-wise deflation. SpO2: red/IR photodetector. Stethoscope: microphone-based electronic signal conversion. Connectivity: Bluetooth wireless, USB wired. Power: Sealed Lead Acid (SLA) battery or IEC 60601-1 AC adapter. Patient-contact materials: 6-nylon/PVC cuffs, steel stethoscope chest piece, biocompatible oximeter sensors (ISO 10993).
Indications for Use
Indicated for non-invasive spot-checking of blood pressure, heart rate, functional oxygen saturation, pulse rate, and heart/lung sounds in pediatric patients (age 3+) and adults under physician care in clinical or home settings.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
CLINICAL VITAL SIGNS MONITOR, MODEL #52STP-E1 (52000) (K951193)
E-Scope Electronic Stethoscope Model 7187-120 (K961301)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for March Healthcare. The logo consists of a geometric shape on the left, followed by the word "MARCH" in large, bold letters. Below "MARCH" is the word "HEALTHCARE" in smaller letters. Underneath the logo is the text "A March Networks Company".
FEB 1 3 2006
# 510(k) Summary of Safety and Effectiveness
### Submitter:
March Healthcare Corporation 555 Legget Drive. Tower B Ottawa, Ontario. K2K 2X3 Canada
## Contact:
Timon LeDain, VP, Engineering Toll Free: (866) 700-0050 Local: (613) 591-6496 Fax: (613) 591-7337 Email: tledain@marchhealthcare.com
Date: November 4. 2005
Trade Name: March Healthcare HMK
Common Name: Health Monitoring Kit
Classification Name: Non-Invasive Blood Pressure Measurement System
#### Product Code: DXN
Legally Marketed Devices: The March Healthcare HMK is substantially equivalent to the legally marketed devices listed in the chart below.
| LEGALLY<br>MARKETED DEVICE | MANUFACTURE<br>NAME | REGULATORY<br>CLASS AND<br>PRODUCT CODE | 510(K)<br>REGISTRATION<br>NUMBER |
|----------------------------------------------------------------|---------------------|-----------------------------------------|----------------------------------|
| CLINICAL VITAL<br>SIGNS MONITOR.<br>MODEL #52STP-E1<br>(52000) | Welch Allyn Inc | Class II/DXN | K951193 |
| E-Scope Electronic<br>Stethoscope Model<br>7187-120 | Cardionics | Class II/DQD | K961301 |
March Healtheare Corporation · Tower B · 555 Legget Drive · Ottawa. ON · Canada · K2K 2X3 613-591-6496 · 1-866-700-0050 · www.marchhealtheare.com
{1}------------------------------------------------
The new March Healthcare HMK is substantially equivalent to the legally marketed devices listed based on the following rationale:
- Same indications for use: for non-invasive monitoring of blood pressure, heart > rate, functional oxygen saturation, and pulse rate of adult and pediatric patients in various health environments.
- Similar key design technical characteristics- The HMK, like the Welch Allyn > monitor, contains comparable modular-based blood pressure monitor and SpO2 components. These OEM modules are currently in use in devices that have already been FDA approved for market. The HMK incorporates Nonin's Ipod or Xpod pulse oximetry technology and CAS Medical's NE NIBP module. In addition, the HMK incorporates Cardionics` networked electronic stethoscope technology into its design.
- Same/similar components for measuring the vital signs of patient. >
- Similar size (hand held), weight, power source. and performance. >
#### Description:
The March Healthcare Health Monitoring Kit (HMK) is a portable medical device equipped with a built-in blood pressure monitor. pulse oximeter, and electronic stethoscope intended for spot-checking the basic vital signs of pediatric patients aged three and older and adult patients under the care of a physician in both clinical and home environments. The HMK can be used as a stand-alone unit or it can be connected to a host using a standard Bluetooth® wireless interface or wired universal serial bus (USB) interface to support remote control of the HMK and audio and data transmission in telehealth applications.
In both stand-alone and remote telehealth applications. the user can spot-check the following vital signs with the blood pressure monitor and pulse oximeter: heart rate and svstolic and diastolic blood pressure. and functional blood oxygen saturation and pulse rate. In remote telehealth applications when the HMK is interfaced with a host, the electronic stethoscope transmits heart and lung sounds via the host to a health care professional at a remote computer. The data from the HMK can be viewed in one of two wavs: on the built-in HMK liquid crystal display (LCD) or from a remote computer connected to a host.
March Healtheare Corporation · Tower B · 555 Legget Drive · Ottawa. ON · Canada · K2K 2N3 613-591-6496 · 1-866-700-0050 · www.marchhealtheare.com
{2}------------------------------------------------
The HMK blood pressure monitor uses the non-invasive, oscillometric, step-wise deflation method to measure heart rate and systolic and diastolic blood pressure. The noninvasive pulse oximeter measures functional blood oxygen saturation and pulse rate using a photodetector and a red and infrared light source to measure the absorption of red and infrared (IR) light passed through the tissuc and arterial hemoglobin. The electronic stethoscope uses a microphone within the chestpiece. bchind a diaphragm. The microphone converts the sound in the chestpiece to an clectronic signal for amplification and transmission to a health care professional at a remote computer via the host.
The HMK does not store physiological data in memory. When using the HMK as a standalone unit. the user must manually record the results displayed on the LCD. However. when the HMK is interfaced with a host, the host may store the audio and physiological data in a database for remote access by authorized clinicians.
The HMK is powered with either a Sealed Lead Acid (SLA) battery or a Medical Grade IEC 60601-1 compliant AC adapter. The AC adapter is also used to charge the SLA battery.
The HMK has only two external inputs: AC adapter input and universal serial bus (USB) input. The user connects the AC adapter to the DC power plug socket when the battery needs charging. In addition, the user can connect the HMK to the host using the Type B USB input as a wired alternative to the wireless Bluetooth interface.
The HMK is for sale by or on the order of a physician.
#### Intended for Use:
The March Healthcare Health Monitoring Kit (HMK) is a non-invasive, spot-measuring device intended to measure heart rate and systolic and diastolic blood pressure using the oscillometric method: functional oxygen saturation levels: pulse rate; and heart and lung sounds of both pediatric patients aged three vears and older and adult patients under the care of a physician in either clinical or home enviroments.
#### Technological Comparison to the Legally Marketed Devices:
Technologically, the March Healthcare HMK is substantially equivalent to the legally marketed devices listed above. The risks, safety and effectiveness, and benefits of the March Healthcare HMK are also comparable. The Table of Comparison in Section 4 will provide additional information illustrating that the new March Healthcare HMK is substantially equivalent to the Welch Allvn 52000 and the Cardionics E-Scope Stethoscone.
Viarch Healtheare Corporation · Tower B · 555 Legget Drive · Ottawa. ON · Canada · K2K 2X3 613-591-6496 = 1-866-700-0050 = www.marchhealtheare.com
{3}------------------------------------------------
To highlight:
- The Non-Invasive Blood Pressure measurement specifications and performance . are equivalent to the Welch Allyn 52000 Vital Signs Monitor (K951193). The HMK and the Welch Allyn 52000 Vital Signs Monitor use a CAS Medical OEM NIBP module that is similar in design.
- The HMK SpO2 measurement and performance specifications are equivalent to . the Welch Allyn 52000 Vital Signs Monitor (K951193) because both units use Nonin OEM pulse oximetry technology.
- The heart and lung sound specifications and performance are derived from the . Cardionics stethoscope (K961301). The networked clectronic stethoscope is an added feature on the HMK that is not available on the Welch Allyn 52000. However, the stethoscope has been approved for market by the FDA to be incorporated into systems like the HMK: consequently, the inclusion of the FDAapproved stethoscope technology raises no new issues of safety and effectiveness and does not affect a substantial equivalence determination.
#### Materials and Biocompatibility:
The blood pressure monitor, pulse oximeter and electronic stethoscope modules are encased in an UL-approved acrylonitrile-butadiene-styrcne (ABS) injection molded carrying case.
The following components contact the patient:
- A & D Medical LifeSource Pressure Cuffs: The pressure cuff is made from cloth . material (6-nylon) and the cuff hose is made from latex-free PVC plastic.
- Nonin Pulse Oximeters: All oximeter sensors are constructed using materials that . have been previously validated by Nonin for biocompatibility per ISO 10993.
- Cardionics Stethoscope: The stethoscope chest piece is made from steel. .
These patient contacting materials have a history of biocompatibility with respect to patient contact. These materials contact the patient in the same locations for less or the same amount of time. Moreover. March Healthcare Corporation has not modified these materials in any way. A&D Medical. Nonin. and Cardionics have tested all patient contacting equipment and accessories for biocompatibility. and these manufacturers have obtained adequate biocompatibility results. For these reasons. March Healthcare Corporation believes that no further biocompatibility testing is required.
{4}------------------------------------------------
#### Summary of Performance Testing:
The March Healthcare HMK was tested for comformance to the design specifications in accordance to the HMK Validation Test Plan, included with the submission, using accordance to the Intent units prior to release to market. The HMK passed all tests and productuates the functionality indicated in the design specifications for the unit.
A risk analysis identifying potential hazards and documenting mitigations of the hazards A Tisk and your released as part of the March Healthcare Corporation product nus been actively and risk analysis is based on ISO14971 - Risk Analysis for Medical Devices.
Comparative testing between the March Healthcare Health Monitoring Kit and the Welch Allyn 52000 Vital Signs Monitor (K951193) was performed to validate the functional Anyh 52000 Vita. Bigneh Bankthcare HMK. In particular, CAS NE NIBP and Nonin SpO2 performance tests were performed to show substantial equivalence to the Welch Allyn 52000 VSM. Comparative testing was performed between the HMK and the E-Scope Electronic Stethoscope Model 7187-120 for similar frequency response. The comparison testing was performed with each vital signs measurement device and demonstrated that the HMK and the Welch Allyn 52000 Vital Signs Monitor and the HMK and the Cardionics E-Scope perform equivalently.
The HMK has been subjected to performance testing to applicable safety. electrical. mechanical, EMC standards. and environmental standards. All specifications were met.
#### Conclusion:
The March Healthcare Health Monitoring Kit (HMK) is safe and effective. complies with the appropriate medical standards, and is substantially equivalent in performance, design and intended use to the legally marketed devices listed above.
> March Healtheare Corporation - Tower B - 555 Legget Drive - Ottawa. ON - Canada - K2K 2X3 613-591-6496 · 1-866-700-0050 · www.marchhealtheare.com
> > HMK S&F- v
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus.
ﻤﻨﺘﺴ
FEB 1 3 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
March Healthcare Corporation c/o Ms. Laura Danielson TÜV Product Service 1775 Old Highway 8 New Brighton, MN 55112-1891
Re: K060194
Trade Name: Health Monitoring Kit (HMK) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: January 24, 2006 Received: January 25, 2006
Dear Ms. Danielson:
We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section 510(x) premainter is substantially equivalent (for the indications
referenced above and have determined the device is substantial in interstate for use stated in the enclosure) the arrived the devices marketed in interstate for use stated in the enclosure) to regally mancede producal Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Federal Food. Drug commerce prior to May 28, 1770, the enablems and the Federal Food, Drug.
devices that have been reclassified in accordance with the provisions of the Federal PMAN devices that nave been recassified in accordance wanted approval application (PMA).)
and Cosmetic Act (Act) that do not require approval of a premarket approval application o and Cosment Act (Act) that do not require approval on the general controls provisions of the Act. The You may, therefore, market the device, see, see, see in of the of the supportuned in and general Controls provisions of the Procellers, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), in If your device is classified (see above) the Existing major regulations affecting your device can
may be subject to such additional controls. Existing major and literal EDA may be subject to such additional controller and the 21, Parts 800 to 898. In addition, FDA may be found in the Oods of Peacharting your device in the Federal Register.
{6}------------------------------------------------
Page 2 -- Ms. Laura Danielson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that IDA s issualled of a backan............................................................................................................................ that IDA has made a decommance that your and your Federal agencies. You must of any I ectifical statutes and regulations and limited to: registration and listing (21 comply with an the Act 3 requirements, netwarts, and manufacturing practice requirements as set CPR Part 807); labeling (21 CFR Part 801); good market 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Part 820); and 11 GFR 1880 form in the quality systems (QD) regardin (Sections 531-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-105 product radiation control provisions (Sections over device as described in your Section 510(k) I his letter will anow you to begin manteang your antial equivalence of your device to a legally premarket notheation: "The PDF miding of basification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your ac not 2012-01-20. Also, please note the regulation entitled, comaci the Office of Compullion as (217) 21 CFR Part 807.97). You may obtain " Misoralluming by reference to premanter noutheast the Act from the Division of Small other general information on your response in its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bhimmar for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
| 510 (k) Number | K060194 (To be assigned) |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Health Monitoring Kit (HMK) |
| Indications for Use: | The Health Monitoring Kit (HMK) is a non-invasive, spot- measuring device used to measure blood pressure, heart rate, oxygen saturation levels, pulse rate, and heart and lung sounds. |
| Target Population: | Pediatric patients over the age of three and adult patients who are under the care of a health care professional. |
| Environment of Use: | For use in clinical environments by health care professionals and/or in home environments by patients under the care of a physician. |
Prescription Use: ✔ AND/OR (21 CFR 801 Subpart D)
Over-the-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blymmma
On Sign-Off
invision of Cardiovascular Devices ושמחשות (O(K) וי 060142
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.