ACON 30 SECOND RELIABLE DIGITAL THERMOMETER, ACON DIGITAL THERMOMETER PROBE COVERS
K060173 · ACON Laboratories, Inc. · FLL · Mar 17, 2006 · General Hospital
Device Facts
Record ID
K060173
Device Name
ACON 30 SECOND RELIABLE DIGITAL THERMOMETER, ACON DIGITAL THERMOMETER PROBE COVERS
Applicant
ACON Laboratories, Inc.
Product Code
FLL · General Hospital
Decision Date
Mar 17, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The ACON™ 30 Second Reliable Digital Thermometer and the ACON™ Digital Thermometer Probe Covers are intended for the quantitative detection of body temperature, which can be obtained orally, under the arm, and/or rectally. It is intended to be sold over-the-counter for home use by lay persons without any professional training.
Device Story
Electronic thermometer; pen-like housing; metal probe tip; LCD display. Input: thermal energy detected by sensor in probe. Operation: converts thermal input to digital temperature reading; displays result on LCD; provides audible beep upon measurement completion. Used in home or clinical settings by laypeople or clinicians. Output: quantitative body temperature (89.6–109.2°F). Benefits: rapid, accurate temperature monitoring; alternative to mercury thermometers. Includes optional disposable probe covers for hygiene.
Clinical Evidence
Controlled human clinical studies evaluated clinical bias, uncertainty, repeatability, and offset. Study compared ACON thermometer performance against the BD Flexible Digital Thermometer (K935267) using non-professional, inexperienced lay users. Results confirmed lay users obtained comparable temperature data and found the device easy to operate. Bench testing verified accuracy (±0.2°F), measurement range, and mechanical safety.
Technological Characteristics
Electronic clinical thermometer; metal probe sensor; 3-digit LCD. Powered by 1.5V LR41 battery. Dimensions: 5.3" x 1.0" x 0.6"; weight: 0.4 oz. Complies with ASTM E1112, EN 60601-1-2, EN 12470-3, and ISO 10993 biocompatibility standards. Features automatic shut-off (8 min) and last memory recall. Waterproof tip area.
Indications for Use
Indicated for quantitative body temperature measurement in lay persons (all ages) via oral, axillary, or rectal routes. Intended for OTC home use; no contraindications specified.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
BD Flexible Digital Thermometer, Model #524034 (K935267)
{0}------------------------------------------------
# . 17 . . .
#### ક. 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The Assigned 510(k) number is _ K060173
# Submitter's Identification:
ACON Laboratories, Inc. 4108 Sorrento Valley Boulevard San Diego, California 92121
Tel.: 858-535-2030 Fax: 858-535-2038
Date Prepared: January 19, 2006
## Contact Person:
Edward Tung, Ph.D. V.P., Regulatory Affairs
## Proprietary Name of the Device:
ACON™ 30 Second Reliable Digital Thermometer/ACON Digital Thermometer Probe Covers
### Common Name:
Thermometer, Electronic Thermometer, Predictive Thermometer
### Classification Name:
Class II §880.2910 Clinical Electronic Thermometer (To be manufactured and marketed for consumer home use)
### Predicate Device:
BD Flexible Digital Thermometer, Model #524034 Becton, Dickinson and Company 510(k) Number: K935267
{1}------------------------------------------------
# Description:
The ACON™ 30 Second Reliable Digital Thermometer consists of a temperature sensor within a metal probe and a Liquid Crystal Display all situated within a pen-like housing. The device is intended for the quantitative detection of body temperature, which can be obtained orally, under the arm, and / or rectally. It is intended to be sold over-the-counter for home use by laypeople without any professional training. And it could be used in hospitals as a substitute for the traditional mercury thermometer.
This device is intended for repeated use. Although the digital thermometer contains no sterile fittings, the user is recommended to clean the device prior to each use. A storage case or probe cover is provided to keep the metal probe tip covered when the thermometer is not in use. The ACONTM 30 Second Reliable Digital Thermometer can be used in conjunction with or withou a disposable probe cover, when preferred.
## Intended Use:
The ACON™ 30 Second Reliable Digital Thermometer and the ACON™ Digital Thermometer Probe Covers are intended for the quantitative detection of body temperature, which one be obtained orally, under the arm, and / or rectally. It is intended to be sold over-the-counter for home use by laypeople without any professional training. And it could be used in hosputals a a substitute for the traditional mercury thermometer.
| Feature | Technical Specification |
|-------------------------------------|-------------------------------------------------------------------|
| Measuring Time | 30 seconds (Oral), 38 seconds (Underarm), 30 seconds (Rectal) |
| Beep | When thermometer is turned on and after completion of measurement |
| Measurement Range | 89.6 - 109.2°F |
| Measurement Accuracy | $\pm$ 0.2°F |
| Resolution | 0.1°F |
| Memory | Last memory recall |
| Automatic Shut Off | After about 8 minutes |
| Display | 3 digit LCD display |
| Operating Temperature | 50 - 109.2°F |
| Storage Temperature | 14 - 140°F |
| Operating/Storage Relative Humidity | 10 - 95% |
| Waterproof | At tip area |
| Battery and Battery Life | 1.5V LR41 battery; about 1,000 uses or 2 years battery life |
| Thermometer Dimensions | 5.3" x 1.0" x 0.6" |
| Thermometer Weight | 0.4 oz |
# Technological Characteristics:
{2}------------------------------------------------
# Comparison to Predicate Devices:
The ACON™ 30 Second Reliable Digital Thermometer is substantially equivalent to the BD Flexible Digital Thermometer, Model #524034, K935267.
The ACON™ Digital Thermometer Probe Covers are similar to FDA-cleared Sanitherm Oral Disposable Thermometer Sheaths for Banta Healthcare Group, LTD (K983406) (Rite Aid Brand).
# Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Compliance to applicable voluntary standards includes ASTM E1112, as well as EN 60601-1-2 and EN 12470-3 requirements and as well as ISO 10993 biocompatibility testing.
Guidance documents included the "FDA Guidance on the Content of Premarket Notification 510(k) Submissions for Clinical Electronic Thermometers".
# Discussion of Clinical Tests Performed:
Controlled human clinical studies were conducted using the ACON™ 30 Second Reliable Digital Thermometer and the ACON™ Digital Thermometer Probe Covers. Clinical were presented evaluating clinical bias, clinical uncertainty, clinical repeatability and clinical offset per ACON clinical test protocol for Digital Thermometer. Clinical study results indicate that the non-professional, inexperienced lay persons were able to obtain comparable body temperature measurement data when using the ACONTM 30 Second Reliable Digital Thermoment and a legally marketed digital thermometer, BD Flexible Digital Thermometer, Model #524034 (K935267). In addition, the lay persons also satisfied to the ease of the operation following the Instructions for Use in the Instruction Manual.
## Conclusion:
The performance characteristics of the ACON™ 30 Second Reliable Digital Thermometer were verified by temperature range test, accuracy test, resolution test, cleaning test, battery condition test, operation environment test, storage environment test, indicating unit test, function safient, mechanical safety test and Rite Aid disposable probe cover validation test. Testing results, indicate that the ACON™ Digital Thermometer is robust and can perform satisfactorily when used according to the "Indication for Use" statement specified in the Instruction Manuals.
The laboratory testing results and lay person studies demonstrated a substantial equivalency on performance between the ACON™ 30 Second Reliable Digital Thermometer and a FDA-cleared Digital Thermometer, BD Flexible Digital Thermometer, Model #524034 (K935267), with the same intended use and product features. The lay person study results also demonstrated that the ACON™ 30 Second Reliable Digital Thermometer and ACON™ Digital Thermoment that he Covers are safe, accurate and easy-to-use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and head. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 17 2006
Dr. Edward Tung, Ph.D. Regulatory Affairs Acon Laboratories, Incorporated 4108 Sorrento Valley Boulevard San Diego, California 92121
Re: K060173
Trade/Device Name: ACON™ 30 Second Reliable Digital Thermometer ACONTM 30 Digital Thermometer Probe Covers Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: February 24, 2006 Received: March 1, 2006
Dear Dr. Tung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Dr. Edward Tung
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CF/R Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### 4. INDICATIONS FOR USE
K660173 510(k) Number (if known):
ACONTM 30 Second Reliable Digital Thermometer Device Name: ACONTM Digital Thermometer Probe Covers
### Indications for Use:
The ACON™ 30 Second Reliable Digital Thermometer and the ACON™ Digital Thermometer Probe Covers are intended for the quantitative detection of body temperature, which can be obtained orally, under the arm, and/or rectally. It is intended to be sold over-the-counter for home use by lay persons without any professional training.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anthony V. Grant
and only General Husain 1. . Jental Dovices
Page 1 of _ 1 ________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.