← Product Code [DPS](/productcode/DPS) · K060167

# ATRIA 3100, ATRIA 6100 (K060167)

_Cardiac Science Corporation · DPS · Apr 19, 2006 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K060167

## Device Facts

- **Applicant:** Cardiac Science Corporation
- **Product Code:** [DPS](/productcode/DPS.md)
- **Decision Date:** Apr 19, 2006
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 870.2340
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

Under the supervision of a qualified physician trained in ECG interpretation, Atria can be used to record the electrical activity of the heart for the purpose of correlating the resultant waveforms with the health of the heart muscle tissue structures. This equipment is not designed to produce a definitive interpretation nor exhaustive evaluation of the patient's heart but rather provide an effective beginning for evaluation of adult and pediatric patients with cardiac abnormalities.

## Device Story

Atria 3100 and 6100 are 12-lead electrocardiographs; record electrical heart activity; display, print, and store ECG waveforms. Atria 6100 includes color TFT VGA preview screen and alphanumeric keyboard; Atria 3100 uses character display. Devices store 50-150 ECG records; support manual/automatic rhythm strip printing; offer optional interpretation. Connectivity includes wired Ethernet, wireless 802.11, and modem for fax, email, and network upload. Operated by clinicians in clinical settings; provides waveforms for physician review to assist in evaluating cardiac abnormalities; facilitates early assessment of heart muscle health.

## Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and intended use compared to predicate devices.

## Technological Characteristics

12-lead electrocardiograph; Atria 6100 features 640x480 color TFT VGA display; Atria 3100 features 2x40 character display. Connectivity: wired Ethernet, wireless 802.11, modem. Storage: 50 (3100) or 150 (6100) records. Optional interpretation software included.

## Regulatory Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

## Predicate Devices

- PageWriter Trim Series Cardiograph - Phillips Medical Systems ([K031422](/device/K031422.md))
- CP 100 and CP 200 - Welch Allyn, Inc. ([K050074](/device/K050074.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K060167 Page 1

Cardiac Science Corporation Bothell, WA Abbreviated Premarket Notification January 20, 2006

APR 1 9 2006

### II. 510(k) Summary

#### A. Name of Device

| Trade name:          | Atria 3100         |
|----------------------|--------------------|
|                      | Atria 6100         |
| Common name:         | Electrocardiograph |
| Classification name: | Electrocardiograph |

#### B. Predicate Devices

| Device Name                                                      | Premarket<br>Notification |
|------------------------------------------------------------------|---------------------------|
| PageWriter Trim Series Cardiograph - Phillips Medical<br>Systems | K031422                   |
| CP 100 and CP 200 - Welch Allyn, Inc.                            | K050074                   |

#### C. Device description

The Atria 3100 and 6100 are 12 lead electrocardiographs designed to record the electrical activity of the heart. They can display, print, electronically send and save ECG recordings. The Atria 6100 features a full alpha numeric keyboard and color display for data entry, waveform review and editing machine settings. The unit provides the ability to store 150 ECGs, print rhythm strips manually or automatically, and provides an optional interpretation. Advanced communications options allow the unit to fax, email and upload results to shared network drives via wired Ethernet, wireless 802.11 and modem.

The differences between the Atria 3100 and 6100 are as follows:

- 1. Screen Size:
	- Atria 3100 = 2 x 40 character display for data entry only (no waveform . viewing)
	- Atria 6100 = 640 x 480 color TFT VGA preview screen for waveform . viewing
- 2. Storage:
	- . Atria 3100 = 50 records

Page 1 of 2
Page 4 of 39

Confidential

{1}------------------------------------------------

zcbuiot Page 2

Cardiac Science Corporation Bothell, WA Abbreviated Premarket Notification January 20, 2006

- Atria 6100 = 150 records .
- 3. The Atria 6100 has a slightly different keyboard to accommodate navigation in the User Interface.

#### D. Intended use

Under the supervision of a qualified physician trained in ECG interpretation, Atria can be used to record the electrical activity of the heart for the purpose of correlating the resultant waveforms with the health of the heart muscle tissue structures.

This equipment is not designed to produce a definitive interpretation nor exhaustive evaluation of the patient's heart but rather provide an effective beginning for evaluation of adult and pediatric patients with cardiac abnormalities.

#### E. Summary

The intended use, indication for use and principle of operation are substantially equivalent to the predicate devices, and do not raise any new types of safety and effectiveness questions.

Page 2 of 2
Page 5 of 39

{2}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 1 9 2006

Cardiac Science Corporation c/o Ms. Cheryl L. Shea Vice President. RA/QA 3303 Monte Villa Parkway Bothell, Washington 98021-8969

Re: K060167

Trade Name: Atria 3100 and Atria 6100 Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (two) Product Code: DPS Dated: March 22, 2006 Received: March 24, 2005

## Dear Ms. Shea:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). 11 may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register

{3}------------------------------------------------

Page 2 -- Ms. Cheryl L. Shea

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that vour device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely vours,

Bfumtemon for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number: KoloOlo 7

Device Name: Atria 3100 and Atria 6100

## Indications for Use:

Under the supervision of a qualified physician trained in ECG interpretation, Atria can be used to record the electrical activity of the heart for the purpose of correlating the resultant waveforms with the health of the heart muscle tissue structures.

This equipment is not designed to produce a definitive interpretation nor exhaustive evaluation of the patient's heart but rather provide an effective beginning for evaluation of adult and pediatric patients with cardiac abnormalities.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Bifumblona
Off

lovascular Devices
K060667

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K060167](https://fda-staging.innolitics.com/device/K060167)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
