← Product Code [JDK](/productcode/JDK) · K060132

# SPINAL USA CEMENT RESTRICTOR SYSTEM (K060132)

_Spinal USA · JDK · Mar 16, 2006 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K060132

## Device Facts

- **Applicant:** Spinal USA
- **Product Code:** [JDK](/productcode/JDK.md)
- **Decision Date:** Mar 16, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Spinal USA Cement Restrictor System is intended as a cement restrictor in orthopedic surgeries such as those involving the femoral canal and tibial plateau in hip stem and total knee replacement. This device is not intended for any spinal indications. The safety and effectiveness of this device when implanted in the spine have not been established.

## Device Story

Hollow, rounded rectangular frame; various sizes; straight and tapered styles; exterior teeth on opposing flat sides prevent migration; holes throughout geometry hold bone cement. Used in orthopedic surgery (hip/knee replacement) to restrict bone cement flow in femoral canal or tibial plateau. Implanted by orthopedic surgeons during joint replacement procedures. Device acts as physical barrier to contain cement; prevents migration via teeth. Benefits patient by ensuring proper cement placement and containment during prosthetic fixation.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Materials: Titanium alloy (ASTM F136) or PEEK (ASTM F2026-02). Geometry: Hollow rounded rectangular frame with migration-preventing teeth. Styles: Straight and tapered. Mechanical device; no software or energy source.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- Scient'X Cement Restrictor ([K052367](/device/K052367.md))
- Quantum Cement Restrictor ([K040276](/device/K040276.md))
- Fortitude Cement Restrictor ([K021719](/device/K021719.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K060132

### MAR 1 6 2006

## 510 (k) Summary of Safety and Effectiveness

| Date Summary Prepared: | January 16, 2006                    |
|------------------------|-------------------------------------|
| Submitter Information: | Spinal USA                          |
|                        | 213 Eastside Lane                   |
|                        | Brandon, MS 39047                   |
| Contact Name:          | Jeffrey Johnson                     |
| Phone:                 | 601-992-7668                        |
| Fax:                   | 601-992-0380                        |
| E-mail:                | jeff@spinalusa.com                  |
| Device Trade Name:     | Spinal USA Cement Restrictor System |
| Common Name:           | Prosthesis, Hip, Cement Restrictor  |
| Regulatory Number:     | 878.3300                            |
| Classification:        | Class II                            |
| Product Code:          | IDK                                 |

#### INTENDED USE:

The Spinal USA Cement Restrictor System is intended as a cement restrictor in orthopedic surgeries such as those involving the femoral canal and tibial plateau in hip stem and total knee replacement.

This device is not intended for any spinal indications. The safety and effectiveness of this device when implanted in the spine have not been established.

#### DEVICE DESCRIPTION:

The Spinal USA Cement Restrictor is a hollow rounded rectangular frame with various holes located throughout its geometry to hold bone cement. The exterior surface of the device has teeth on two opposing flat sides to prevent migration. The device comes in various sizes and is offered in straight and tapered styles. The device may be made from titanium alloy as specified in ASTM F136 or polyetheretherketone (PEEK) that conforms to ASTM F2026-02.

#### EQUIVALENT DEVICE:

The Spinal USA Cement Restrictor System was demonstrated to be substantially equivalent to previous cleared devices such as the Scient'X Cement Restrictor (K052367), Quantum Cement Restrictor (K040276), and Fortitude Cement Restrictor (K021719). The subject device and predicate devices have the following similarities:

-the same indication for use

-the same operating principle

-the same basic design

-the same materials

-implanted using the same surgical techniques and equipment

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three lines forming its body and head, and two curved lines representing its legs. The seal is encircled by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA".

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 6 2006

Spinal USA C/o Mr. Jeffrey Johnson Manager, Regulatory Affairs 213 Eastside Lane Brandon, Mississippi 39047

Re: K060132

K000192
Trade/Device Name: Spinal USA Cement Restrictor System Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: II Product Code: JDK Dated: January 16, 2006 Received: January 19, 2006

Dear Mr. Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Section 3 rots) presidentially equivalent (for the indications ferenced above and nave determined by marketed predicate devices marketed in interstate for use stated in the clerosure) to regally manage of the Medical Device Amendments, or to commerce provide to may 20, 1978, are excerdance with the provisions of the Federal Food, Drug, devices that have been reclassified in accerative while in a premarket approval application (PMA). and Cosmetic Act (Act) that to not require apt to the general controls provisions of the Act. The 1 ou may, mercere, manel are as not include requirements for annual registration, listing of general controls provisions of wactice, labeling, and prohibitions against misbranding and adulteration.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this The Office or Device Evanuation has avot identified in the proposed labeling and that such use device with be used for an mosiose cordance with Section 513(i)(1)(E) of the Act, the following court cause namin. "Therefore, namings section of the device's package insert and also as a Warning on the product label:

WARNING: THIS DEVICE IS NOT INTENDED FOR ANY SPINAL INDICATIONS.

THE SAFETY AND EFFECTIVENESS OF THIS DEVICE WHEN IMPLANTED IN THE SPINE HAVE NOT BEEN ESTABLISHED.

{2}------------------------------------------------

### Page 2 - Mr. Jeffrey Johnson

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. r teast note that the act to aconing before these limitations are modified in any way or removed from the device's labeling.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This Ietter will allow you to begin marketing your device as described in your Section 510(k) icter will anow you to began minutes on statement described above is added to your labeling.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elassified (800 are rols. Existing major regulations affecting your device can may be subject to subli additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I that FDA has made a determination that your device complies with other requirements of the Act that I Dr Has made a and regulations administered by other Federal agencies. You taust of any I catales and the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI IT art 007); idocing (21 (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product from to begin marketing your device as described in your Section 510(k) I mis letter will and w yours ough finding of substantial equivalence of your device to a legally prematice notification: "The PDF interessification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you attire of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small oner general minternational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Herbert Lemmer

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

{3}------------------------------------------------

Page 3 - Mr. Jeffrey Johnson

:" »

Enclosure

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

ו

:

、

{4}------------------------------------------------

# K 060132

# Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Spinal USA Cement Restrictor System

Indications for Use:

The Spinal USA Cement Restrictor System is intended as a cement restrictor in orthopedic surgeries such as those involving the femoral canal and tibial plateau in hip stem and total knee replacement.

This device is not intended for any spinal indications. The safety and effectiveness of this device when implanted in the spine have not been established.

Prescription Use
(Part 21 CFR 801 Subpart D)

X

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert Lemon

Division Division of General, Restorative. and Neurological Devices

**510(k) Number** K060132

---

**Source:** [https://fda-staging.innolitics.com/device/K060132](https://fda-staging.innolitics.com/device/K060132)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).