Teco CX3 Reagent Set for SYNCHRON CX is intended for the quantitative determination of BUN, Calcium, Creatinine and Glucose in serum on Beckman CX3 System.
Device Story
Teco CX3 Reagent Set consists of chemical reagents for the quantitative determination of BUN, Calcium, Creatinine, and Glucose in human serum samples. Designed for use on the Beckman CX3 System; reagents are loaded into the analyzer. The system performs automated chemical reactions: BUN via urease hydrolysis (conductivity measurement); Calcium via Arsenazo III reaction (spectrophotometry at 650 nm); Creatinine via red color complex formation (spectrophotometry at 520/560 nm); Glucose via oxygen rate method (oxygen electrode). The analyzer processes samples and provides quantitative concentration outputs. Used in clinical laboratory settings by trained technicians. Results assist healthcare providers in diagnosing and monitoring renal, metabolic, parathyroid, and carbohydrate metabolism disorders. Benefits include standardized, automated diagnostic testing for patient management.
Clinical Evidence
Bench testing only. Performance evaluated using human serum samples. Precision (within-day and day-to-day) assessed with N=25 replicates per sample; CV% values were within acceptable limits (BUN <8%, Calcium <5%, Creatinine <10%, Glucose <10%). Linearity and accuracy (method comparison) were evaluated against predicate devices. Accuracy results showed high correlation (r=0.96 to 0.99) and slopes within 0.90-1.1 range for all analytes. No clinical prospective or retrospective studies were required.
Technological Characteristics
In vitro diagnostic reagent set. Sensing principles: conductivity (BUN), spectrophotometry (Calcium, Creatinine), and oxygen rate electrode (Glucose). Reagents are liquid-based. Designed for integration with Beckman CX3 System. Connectivity is via the host analyzer. Sterilization not applicable (chemical reagents). Software is embedded in the host analyzer; no specific algorithm architecture provided beyond standard clinical chemistry calculation methods.
Indications for Use
Indicated for quantitative determination of BUN, Calcium, Creatinine, and Glucose in serum. BUN used for diagnosis/treatment of renal and metabolic diseases. Calcium used for parathyroid, bone, chronic renal diseases, and tetany. Creatinine used for renal disease diagnosis, monitoring dialysis, and calculation of other urine analytes. Glucose used for carbohydrate metabolism disorders including diabetes mellitus, neonatal/idiopathic hypoglycemia, and pancreatic islet carcinoma. For in vitro diagnostic use only.
Regulatory Classification
Identification
A urea nitrogen test system is a device intended to measure urea nitrogen (an end-product of nitrogen metabolism) in whole blood, serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases.
{0}------------------------------------------------
MAR 2 7 2006
K060/20
Image /page/0/Picture/3 description: The image shows a black hexagon with the letters 'TC' in white inside. The letters are in a simple, sans-serif font and are stacked on top of each other. The hexagon is solid black, providing a strong contrast to the white letters.
1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 693-7788 Toll Free: 1-800-222-9880 Fax: (714) 693-3838 www.tecodiag.com Email:tecodiag@tecodiag.com
Updated 510K summary Teco CX3 Reagent Set for SYNCHRON CX Contact Name: Jian Vaeches Phone Number: 714-693-7788 ext.131 Prepared Date: 03/20/2006
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a black hexagon with the letters "TC" in white inside. The letters are in a bold, sans-serif font and are centered within the hexagon. The hexagon is oriented with one point facing upwards.
1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 693-7788 Toll Free: 1-800-222-9880 Fax: (714) 693-3838 www.tecodiag.com Email:tecodiag@tecodiag.com
#### Device Name
The device trade names and common/classifications name are:
| Device Trade Name | Common/Classification Name |
|---------------------------------------------|-------------------------------------------|
| Teco CX3 Reagent Set for SYNCHRON CX System | BUN, CALCIUM, CREATININE, GLUCOSE REGENTS |
#### Address and Registration
The address and registration number of the manufacturer site for Teco CX3 Reagent Set for SYNCHRON CX System:
TECO Diagnostics 1268 N. Lakeview Ave. Anaheim, CA 92807, U.S.A. FDA Registration # 1832216
#### Device Classs
ાં બે Teco CX3 Reagent Set has been classified as Class II with Product Code Lare, JPP, CGX, CGA "in vitro" diagnostics reagent set having the classification number: 21 CFR. 862.1770, 21 CFR 862.1145, 21 CFR 862.1225, 21 CFR 862.1345. This is the description available from the classification names listed in the " CDRH Home Page- Listing Database."
#### Predicate Device Information
The predicate device is Teco BUN Liquid, Teco Calcium Color, Teco Creatinine and Teco Glucose Liquid Reagent.
The 510 (K) approval letter is provided in Appendix I. 510(K) #: K981106, K864741, K880629, K863926. Approval Date: 04-16-1998, 01-09-1987, 03-23-1988, 11-13-1986
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a black hexagon with the letters 'TC' in white inside. The letters are in a simple, sans-serif font and are stacked on top of each other. The overall design is clean and minimalist, with a high contrast between the black background and the white letters.
1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 693-7788 Toll Free: 1-800-222-9880 Fax: (714) 693-3838 www.tecodiag.com Email:tecodiag@tecodiag.com
## Labeling and Intended Use
Draft labels and Instructions for use can be found in appendix A.
Intended Use
Teco CX3 Reagent Set for SYNCHRON CX is intended for the quantitative determination of BUN, Calcium, Creatinine and Glucose in serum on Beckman CX3 System.
## Device Description and Comparison
The Teco CX3 Reagent Set for SYSNCHRON CX designed for use on the Beckman CX System. The Reagent Set includes BUN, Calcium, Creatinine and Glucose.
Method Comparison of Teco CX3 Reagent Set for SYNCHRON CX to Teco BUN Liquid, Teco Calcium Color, Teco Creatinine and Teco Glucose Liquid following the guidelines of NCCLS Guideline EP9-A2 was conducted.
Comparison with predicate: Similarities Intended use. Sample Handling Subsystem. Operational Environment Performance.
| | Predicate Device | Candidate Device |
|----------------|------------------------------------------------------------------------------------------------------------|------------------|
| Packaging Size | | |
| BUN | 5 x 25 ml & 5 x 5 ml | 500 ml |
| Calcium | 4 x 120 ml | 500 ml |
| Creatinine | 4 x 120 ml | 2000 ml |
| Glucose | 4 x 120 ml | 500 ml |
| Analyzer | General Chemistry Analyzers | SYSCHRON CX3 |
| Components | | |
| BUN | Tris Buffer 100 mmol<br>2-Oxoglutarate 5 mmol<br>Urease >20,000 U/L<br>GLDH > 1500 U/L<br>NADH 0.25 mmol/L | Urease >592 U/mL |
#### Differences
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a black hexagon with the letters "TC" in white inside. The letters are in a simple, sans-serif font and are stacked on top of each other. The hexagon is solid black, providing a strong contrast to the white letters.
#### 1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 693-7788 Toll Free: 1-800-222-9880 Fax: (714) 693-3838 www.tecodiag.com Email:tecodiag@tecodiag.com
| Calcium | O-Cresolphthalein Complexone 0.14 mM | Arsenazo III 0.15 mml/L |
|------------|--------------------------------------|---------------------------|
| | 8-Hydroxyquinoline 13 mM | |
| | Diethylamide 363 mM | |
| Creatinine | Picric Acid 10mM | Picric Acid 50mM |
| | Sodium Hydroxide 240mM | Sodium Hydroxide 0.188 mM |
| Glucose | Glucose Oxidase 15 u/ml | Glucose Oxidase 590 u/ml |
| | Peroxidase 1.2 u/ml | Ethanol 10% |
| | 4-Aminoantipyrine 0.38 mM | Potassium Iodide 0.04 mM |
| | p-Hydroxybenzene Sulfonate 10 mM | |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a black hexagon with the letters 'TC' in white inside. The letters are stylized and bold. The background is plain white.
1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 693-7788 Toll Free: 1-800-222-9880 Fax: (714) 693-3838 www.tecodiag.com Email:tecodiag@tecodiag.com
## Standard/ Guidance Document Referenced (if applicable)
NCCLS EP5-A-Evaluation of Precision Performance of Clinical Chemistry Devices NCCLS EP6-A --- Evaluation of Linearity of Quantitative Analytical Methods NCCLS EP9-A---Method Comparison and Bias Estimation Using Patient Samples
### Test Principle
BUN (Urea Nitrogen) is hydrolyzed by urease to produce ammonium and bicarbonate which resulted in the increase of solution conductivity which is directly proportional to the concentration of BUN present in the test.
Calcium reacts with Arsenazo to form an bluish-purple chromophore which is measured spectrophotometrically at 650 nm. A reagent blank reading is taken just prior to sample injection, and a final absorbance reading is taken 21 seconds later. The differential absorbance, corrected for the reagent blank, is directly proportional to the calcium concentration.
Creatinine from the sample combines with the reagent to produce a red color complex. Absorbance reading is taken at both 520nm and 560nm at 25.6 seconds after picking up sample. The differential absorbance has been shown to be a direct measure of the concentration of creatinine in the sample.
SYSNCHRON CX3 determines glucose concentration by an oxygen rate method using a glucose Oxygen electrode. The rate of oxygen depletion is proportional to the glucose concentration in the samples.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a black hexagon with the letters "TC" in white inside. The letters are in a bold, sans-serif font and are centered within the hexagon. The hexagon is solid black, providing a strong contrast to the white letters.
1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 693-7788 Toll Free: 1-800-222-9880 Fax: (714) 693-3838 www.tecodiag.com Email:tecodiag@tecodiag.com
## TECO DIAGNOSTICS
Image /page/5/Picture/4 description: The image shows a black hexagon with the letters "TC" in white inside. The letters are in a sans-serif font and are slightly offset from each other. The hexagon is oriented with one point facing upwards.
1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 693-7788 Toll Free: 1-800-222-9880 Fax: (714) 693-3838 www.tecodiag.com Email:tecodiag@tecodiag.com
### Substantial Equivalence
The Teco CX3 Reagent Set for SYNCHRON CX System is substantially equivalent to other devices legally marketed in the United States. We have compared TECO Diagnostics Teco CX3 Reagent Set for SYNCHRON CX System to TECO Diagnostics Teco BUN Liquid, Teco Calcium Color, Teco Creatinine and Teco Glucose Liquid Reagent for Hitachi.The 510 (K) numbers are K981106, K864741, K880629, and K863926. Both devices are for the quantitative determination of the same analyst in serum.
### Summary of Design Control Activities
The risk analysis method used to assess the impact of modifications was a Failure Modes and Effects Analysis (FMEA). The design verification tests that were performed as a result of this risk analysis assessment are listed in table 1&2 below. The tests data can also be found in appendix C through F.
# 1. Table 1 -- Performance Characteristics (Human Serum)
| Feature | Predicate Device | | Candidate Device | | |
|-------------------------|------------------|-----------|------------------|----------|-----------|
| | Sample 1 | Sample 2 | Sample 1 | Sample 2 | Sample 3 |
| Precision<br>Within-Day | N=20 | N=20 | N=25 | N=25 | N=25 |
| | Mean=12.9 | Mean=51.8 | Mean=18 | Mean=49 | Mean=97.5 |
| | SD=0.33 | SD=0.74 | SD=0.46 | SD=0.91 | SD=1.3 |
| | CV%=2.62 | CV%=1.43 | CV%=2.5 | CV%=1.9 | CV%=1.3 |
| Day to Day | Sample 1 | Sample 2 | Sample 1 | Sample 2 | Sample 3 |
| | N=20 | N=20 | N=25 | N=25 | N=25 |
| | Mean=12.4 | Mean=44.6 | Mean=18 | Mean=49 | Mean=98 |
| | SD=0.15 | SD=0.75 | SD=0.37 | SD=1.04 | SD=1.14 |
| | CV%=1.2 | CV%=1.67 | CV%=2.0 | CV%=2.0 | CV%=1.1 |
Teco CX3 Reagent Set, the acceptability of CV% in BUN test is within 8%
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a black hexagon with the letters "TC" in white inside. The letters are stacked on top of each other, with "T" on top and "C" on the bottom. The letters are in a bold, sans-serif font. The hexagon is solid black, and the letters are solid white.
1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 693-7788 Toll Free: 1-800-222-9880 Fax: (714) 693-3838 www.tecodiag.com Email:tecodiag@tecodiag.com
## Calcium:
| Feature | Predicate Device | | | Candidate Device | | |
|-------------------------|-------------------------------------------------------|--------------------------------------------------------|-------------------------------------------------------|--------------------------------------------------------|------------------------------------------------------|--|
| Precision<br>Within-Day | Sample 1<br>N=20<br>Mean=9.1<br>SD=0.39<br>$CV\%=4.3$ | Sample 2<br>N=20<br>Mean=13.7<br>SD=0.02<br>$CV\%=0.2$ | Sample 1<br>N=25<br>Mean=9.1<br>SD=0.08<br>$CV\%=0.9$ | Sample 2<br>N=25<br>Mean=12.3<br>SD=0.11<br>$CV\%=0.9$ | Sample 3<br>N=25<br>Mean=14<br>SD=0.16<br>$CV\%=0.8$ | |
| Day to Day | Sample 1<br>N=20<br>Mean=9.2<br>SD=0.21<br>$CV\%=2.2$ | Sample 2<br>N=20<br>Mean=13.3<br>SD=0.32<br>$CV\%=2.4$ | Sample 1<br>N=25<br>Mean=9.1<br>SD=0.09<br>$CV\%=1.0$ | Sample 2<br>N=25<br>Mean=12.4<br>SD=0.14<br>$CV\%=1.1$ | Sample 3<br>N=25<br>Mean=14<br>SD=0.14<br>$CV\%=1.0$ | |
Teco CX3 Reagent Set, the acceptability of CV% in Calcium test is within 5%
# Creatinine:
| Feature | Predicate Device | | Candidate Device | | |
|-------------------------|----------------------------------------------------|---------------------------------------------------|----------------------------------------------------|----------------------------------------------------|-----------------------------------------------------|
| Precision<br>Within-Day | Sample 1<br>N=20<br>Mean=1.9<br>SD=0.05<br>CV%=2.6 | Sample 2<br>N=20<br>Mean=8.2<br>SD=0.6<br>CV%=7.3 | Sample 1<br>N=25<br>Mean=1.3<br>SD=0.04<br>CV%=2.7 | Sample 2<br>N=25<br>Mean=6.4<br>SD=0.11<br>CV%=1.6 | Sample 3<br>N=25<br>Mean=20<br>SD=0.35<br>CV%=1.7 |
| Day to Day | Sample 1<br>N=20<br>Mean=2.0<br>SD=0.2<br>CV%=10 | Sample 2<br>N=20<br>Mean=8.2<br>SD=0.4<br>CV%=4.6 | Sample 1<br>N=25<br>Mean=1.3<br>SD=0.03<br>CV%=2.2 | Sample 2<br>N=25<br>Mean=6.6<br>SD=0.24<br>CV%=3.7 | Sample 3<br>N=25<br>Mean=19.9<br>SD=0.33<br>CV%=1.6 |
Teco CX3 Reagent Set, the acceptability of CV% in Creatinine test is within 10%
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows a black hexagon with the letters "TC" in white inside. The letters are in a bold, sans-serif font. The hexagon is solid black, and the letters are centered within it. The image is simple and graphic.
1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 693-7788 Toll Free: 1-800-222-9880 Fax: (714) 693-3838 www.tecodiag.com_Email:tecodiag@tecodiag.com
## Glucose:
| Feature | Predicate Device | | Candidate Device | | |
|-------------------------|--------------------------------------------------|---------------------------------------------------|--------------------------------------------------|---------------------------------------------------|---------------------------------------------------|
| | Sample 1 | Sample 2 | Sample 1 | Sample 2 | Sample 3 |
| Precision<br>Within-Day | N=20<br>Mean=87<br>SD=4.2<br>CV%=4.8 | N=20<br>Mean=282<br>SD=5.4<br>CV%=1.9 | N=25<br>Mean=78<br>SD=2.0<br>CV%=2.5 | N=25<br>Mean=255<br>SD=3.4<br>CV%=1.3 | N=25<br>Mean=735<br>SD=9.8<br>CV%=1.5 |
| Day to Day | Sample 1<br>N=20<br>Mean=85<br>SD=3.7<br>CV%=4.3 | Sample 2<br>N=20<br>Mean=287<br>SD=9.6<br>CV%=3.3 | Sample 1<br>N=25<br>Mean=79<br>SD=2.6<br>CV%=3.3 | Sample 2<br>N=25<br>Mean=262<br>SD=5.9<br>CV%=2.2 | Sample 3<br>N=25<br>Mean=734<br>SD=8.7<br>CV%=1.2 |
Teco CX3 Reagent Set, the acceptability of CV% in Clucose test is within 10%
# Table 2-Performance Characteristics (Human Serum) BUN:
| Feature | Candidate Device | Predicate Device |
|-----------------------------------|---------------------------|--------------------------------|
| Linearity | 3 to 102 mg/dL | up to 80 mg/dL |
| Accuracy | R=0.99<br>Y= $1.00-0.911$ | R=0.99<br>Y= $0.962 X - 0.721$ |
| Expect values/<br>Reference range | 15-39 mg/dL | 8-23 mg/L |
Teco CX3 Reagent Set BUN test, Acceptance Criteria of Accuracy: r > 0.95; Slope: 0.97-1.1
## Calcium:
| Feature | Candidate Device | Predicate Device |
|-----------------------------------|-------------------------|--------------------------|
| Linearity | 1.0 to 15 mg/dL | up to 20 mg/dL |
| Accuracy | R=0.96<br>Y=1.07X +0.60 | R=0.97<br>Y=0.94 X+ 0.53 |
| Expect values/<br>Reference range | 8.4-10.2 mg/dL | 8.5-10.5 mg/dL |
Teco CX3 Reagent Set Calcium test, Acceptance Criteria of Accuracy: r > 0.90; Slope: 0.90-1.1
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows a black hexagon with the letters "TC" in white inside. The letters are in a bold, sans-serif font. The hexagon is solid black, and the white letters stand out against the dark background. The logo appears simple and modern.
1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 693-7788 Toll Free: 1-800-222-9880 Fax: (714) 693-3838 www.tecodiag.com Email:tecodiag@tecodiag.com
## Creatinine:
| Feature | Candidate Device | Predicate Device |
|-----------------------------------|----------------------------|-----------------------------------------------|
| Linearity | 0.2 to24g/dL | Up to 25 mg/dL |
| Accuracy | $R=0.99$<br>$Y=0.97 -0.11$ | $R=0.99$<br>$Y=0.96 X+ 0.06$ |
| Expect values/<br>Reference range | 0.6-1.3 mg/dL | Male: 0.9-1.5 mg/dL<br>Female: 0.7-1.37 mg/dL |
Teco CX3 Reagent Set Creatinine test, Acceptance Criteria of Accuracy: r > 0.96; Slope: 0.90-1.1
# Glucose:
| Feature | Candidate Device | Predicate Device |
|-----------------------------------|------------------------|------------------------|
| Linearity | 30 to 750 mg/dL | up to 500 mg/dL |
| Accuracy | R=0.99<br>Y=0.96X+0.36 | R=0.99<br>Y=1.02 X+3.1 |
| Expect values/<br>Reference range | 70-105 mg/dL | 70-106 mg/dL |
Teco CX3 Reagent Set Glucose test, Acceptance Criteria of Accuracy: r > 0.90; Slope: 0.85-1.1
{9}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and a caduceus intertwined with the eagle's body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is black and white.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAR 2 7 2006
Ms. Jian Vaeches Teco Diagnostics 1268 N. Lakeview Ave Anaheim, CA 92807
Re:
k060120 Trade/Device Name: Teco CX3 Reagent Set for SYNCHRON CX System Regulation Number: 21 CFR§ 862.1770 Regulation Name: Urea nitrogen test system Regulatory Class: Class II Product Code: CDQ, CIC, CGX, CGA Dated: March 16, 2006 Received: March 16, 2006
Dear Ms. Vaeches:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{10}------------------------------------------------
Page 2 –
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Alberto Gutierrez
Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
# Indications for Use
## 510(k) Number (if known): k060120
# Device Name: Teco CX3 Reagent Set for SYNCHRON CX System
Indications For Use: Teco CX3 Reagent Set for SYNCHRON CX System is intended for the quantitative determination of BUN, Calcium, Creatinine and Glucose in serum on Beckman CX3 System.
BUN measurements are used in the diagnosis and treatment of certain renal and metabolic diseases.
Calcium measurements are used in the diagnosis and treatment of parathyroid diseases, a variety of bone diseases, chronic renal and tetany.
Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and a calculation basis for measuring other urine analytes.
Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders, including diabetes mellitus, neonatal hypoglycemia and idiopathic hypoglycemia, and pancreatic islet carcinoma.
This reagent set is intended for in vitro diagnostic use only.
Prescription Use J
AND/OR
Over-The-Counter Use
(Part 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Page 1 of 1
Of In Vitro Diagnostic Device Evaluation and Safe
n Sign-Off
060120
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.