The range of Nodecrest Transcranial Doppler Ultrasound System is intended for use as a diagnostic ultrasound fluid flow analysis system: 1) For the measurement of cerebral artery blood velocities to determine the presence of hemodynamically significant deviations from normal values 2) To assess arterial cerebral blood flow for the occurrence of micro embolic signals. Vessels intended for observation include, but are not limited to the middle, anterior and posterior cerebral arteries, via the temporal windows, the vertebral mid basilar arteries via the foramen magnum and the ophthalmic artery and intracranial internal carotid artery via the eye. The Nodecrest Transcranial Doppler is intended for use during: a) Diagnostic exams b) Surgical interventions The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.
Device Story
EMS9U Transcranial Doppler (TCD) system; uses pulse wave (PWD) and continuous wave (CWD) Doppler transducers (2MHz, 4MHz, 8MHz) to measure blood velocity and detect micro-embolic signals. Operates free-hand or head-frame mounted. Supplementary M-mode display assists in locating blood flow signals and positioning sample gates. System hardware identical across three models; software features differentiate functionality. Used in diagnostic exams and surgical interventions by clinicians. Output provides real-time blood flow data for hemodynamic assessment. Benefits include non-invasive monitoring of cerebral blood flow and emboli detection.
Clinical Evidence
Bench testing only. No clinical data provided. Device performance verified through electrical safety testing (IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, UL 2601-1) and biocompatibility testing (cytotoxicity, sensitization, skin irritation, acute toxicity, intracutaneous reactivity, implantation).
Technological Characteristics
TCD ultrasound system; 2MHz, 4MHz, 8MHz transducers. PWD and CWD modes. FFT-based frequency spectrum analysis (256/512 dots). Depth measurement 5-136 mm. Gain 0-40dB. Electrical safety per IEC 60601-1 series. Biocompatibility per ISO 10993 and USP standards. Software-controlled feature set.
Indications for Use
Indicated for diagnostic ultrasound fluid flow analysis in adults. Used for measuring cerebral artery blood velocities and assessing arterial cerebral blood flow for micro embolic signals. Vessels include middle, anterior, and posterior cerebral arteries, vertebral mid basilar arteries, ophthalmic artery, and intracranial internal carotid artery. Not for fetal applications or use within the sterile field.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K060112
# Noderrest LLC MAY - 5 2006
## SUMMARY STATEMENT
| Contact Person: | David G. Pegler |
|-----------------------|-----------------------------------------------------------------------|
| Address: | Nodecrest LLC<br>1478 Powells Tavern Place<br>Herndon, Virginia 20170 |
| Phone: | 703-627-6636 |
| Classification Name: | System, Imaging, Pulsed Doppler, Ultrasonic |
| Common Usual Name: | Transcranial Doppler Ultrasound System |
| Classification: | Class II |
| Product Code: | 90 IYN |
| Classification Panel: | Radiology |
### Device Description
The EMS9U range is a Transcranial Doppler (TCD) diagnostic ultrasound system, with a pulse wave and / or continuous wave Doppler transducer that can be used free hand (or head frame mounted for longer term monitoring). A supplementary M-mode disala ( nay be used to help locate blood flow signals to position the sample gate for Doppler signal, and for detection of emboli signals. Each of the three models in the range is identical in terms of hardware the difference being features that the end user does not request are turned off in the software.
### Intended Use
The range of Nodecrest Transcranial Doppler Ultrasound System is intended for use as a diagnostic ultrasound fluid flow analysis system:
1) For the measurement of cerebral artery blood velocities to determine the presence of hemodynamically significant deviations from normal values
2) To assess arterial cerebral blood flow for the occurrence of micro embolic signals. Vessels intended for observation include, but are not limited to the middle, anterior and posterior cerebral arteries, via the temporal windows, the
{1}------------------------------------------------
vertebral mid basilar arteries via the foramen magnum and the ophthalmic artery and intracranial internal carotid artery via the eye.
The Nodecrest Transcranial Doppler is intended for use during:
- a) Diagnostic exams
b) Surgical interventions
The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.
| Parameters | Nodecrest | Rimed Ltd. |
|-----------------------------------|---------------------------------------------------------------|-------------------------------------|
| 510(k) Number | | K974588 |
| Transducer Frequency | 2MHz, 4MHz, 8MHz | 2MHz, 4MHz, 8MHz |
| Frequency Spectrum | FFT 256/512 dots | 64/512 dots |
| Frequency Ranges | 1 to 16 MHz | 2 to 32 MHz |
| Depth Measurement | 5 to 136 mm | 15-150 mm |
| Gain | 0 ~ 40db | Manually control 20 steps |
| Clinical Application<br>for: 2MHz | | |
| PWD | Ophthalmic, Adult Cephalic<br>& Peripheral Vascular | Ophthalmic, Adult Cephalic |
| CWD | Ophthalmic & Peripheral<br>Vascular | Ophthalmic & Peripheral<br>Vascular |
| Clinical Application<br>for: 4MHz | | |
| PWD | Peripheral Vascular | |
| CWD | Peripheral Vascular | Peripheral Vascular |
| Clinical Application<br>for: 8MHz | | |
| PWD | Peripheral Vascular | |
| CWD | Peripheral Vascular | Ophthalmic & Peripheral<br>Vascular |
| General Electrical<br>Testing | IEC 60601-1-1,<br>IEC 60601-1-2, IEC 60601-<br>1-4, UL 2601-1 | EN 60601-1-1 &<br>EN 60601-1-2 |
| Biocompatibility Tests | | |
| ISO 10993 | Cytotoxicity, Sensitization | Unknown |
### 12.1 Product Comparison Chart
22
{2}------------------------------------------------
| | skin irritation | |
|-----|------------------------------------------------------------------------------|---------|
| USP | Acute Toxicity,<br>Intracutaneous Reactivity,<br>Implantation & Cytotoxicity | Unknown |
# 12.2 Summary of Safety and Effectiveness:
The Nodecrest Transcranial Doppler Ultrasound System and the Rimed Smart-Lite™ Transcranial Doppler have similar intended use: both offer similar transducer frequency, frequency spectrum, frequency ranges, depth measurement, gain, and offer similar clinical applications. Both systems meet IEC 60601-1 & IEC 60601-1-2 Standards for general electrical safety of medical devices and electrical emissions.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICE
Public Health Se
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Nodecrest, LLC % Mr. E. J. Smith President Smith Associates 1676 Village Green, Suite A CROFTON MD 21114
MAY 5 2006
Re: K060112
Trade Name: Nodecrest Transcranial Doppler Ultrasound System
Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic pulsed doppler imaging system
Regulation Number: 21 CFR 892.1570
Regulation Name: Diagnostic ultrasonic transducer
Regulatory Class: II
Product Code: IYN and ITX
Dated: March 1, 2006
Received: March 2, 2006
Dear Mr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Nodecrest Transcranial Doppler Ultrasound System, as described in your premarket notification:
Transducer Model Number
| 2 MHz |
|-------|
| 4 MHz |
| 8 MHz |
Image /page/3/Picture/20 description: The image is a black and white circular logo. The logo commemorates the centennial of the FDA from 1906-2006. The letters FDA are in bold in the center of the logo.
Investing and Promoting Public Health
{4}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, vou submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G. (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
{5}------------------------------------------------
If you have any questions regarding the content of this letter, please contact Andrew Kang, M.D. at (301) 594-1212.
Sincerely yours,
David A. Lyman
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use
510(k) Number (If known): < 060112
Device Name: Nodecrest Transcranial Doppler Ultrasound System
Indications for Use:
The range of Nodecrest Transcranial Doppler Ultrasound System is intended for use as a diagnostic ultrasound fluid flow analysis system:
1) For the measurement of cerebral artery blood velocities to determine the presence of hemodynamically significant deviations from normal values
2) To assess arterial cerebral blood flow for the occurrence of micro embolic signals. Vessels intended for observation include, but are not limited to the middle, anterior and posterior cerebral arteries, via the temporal windows, the vertebral mid basilar arteries via the foramen magnum and the ophthalmic artery and intracranial internal carotid artery via the eve.
The Nodecrest Transcranial Doppler is intended for use during:
- a) Diagnostic exams
b) Surgical interventions
The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Depaon
ive. Abdominal
{7}------------------------------------------------
148 - 124 - 12
And and and the same of the states
#### Diagnostic Ultrasound Indications for Use Form
Friday of the commend to the states
#### Fill out one form for each ultrasound system and each transqucer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the hurnan body as follows:
| Clinical Application: | A | B | M | PWID | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
|-------------------------------|---|---|---|------|-----|---------------|-------------------|------------------------|--------------------|-----------------|
| Ophthalmic | | | | X | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | X | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | X | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: EMS 9U TRANSCRANIAL Doppler SYSTEM 2MHz PROBE
(PLEASE DO NOT WRITE BELOW THIS LINE . CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
(Division Sign-Oll)
Division of Reproductive, Abdominal, F-3 and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________
011
:
{8}------------------------------------------------
19, 2017
#### Disgnostic Uitrasound Indications for Use Form
Fill out one form for cech ultrasound system and each transduder.
Intended Use: Diagnostic ultrasound imaging or fivid flow andlysis of the human body as follows:
| | | Mode of Operation | | | | | | | | | | |
|-------------------------------|---|-------------------|---|-----|-----|---------------|-------------------|------------------------|--------------------|-----------------|--|--|
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) | | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Peripheral Vascular | | | | X | X | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: EMS9U TRANSCRANIAL DOPPLE
SYSTEM 4MHZ PROBE
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
F-3
Prescription Use (Per 21 CFR 801.109)
Daniel A. Larson
(Division Sign-Off) Division of Reproductive, Ab and Radiological Devices 5 புரு) Number
{9}------------------------------------------------
Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|-------------------------------|-------------------|---|---|-----|-----|---------------|-------------------|------------------------|--------------------|-----------------|
| | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | XX | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new Indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: EMS9U TRANSCRANIAL DOPPER SYSTEM
8MHZ PROBE
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
F-3
Daniel A. Segerson
(Division Sign-Off) Division of Reproductive, Abdominal,
and Radiological Devices A (Alol 510(k) Number _
:
19
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.