← Product Code [LLL](/productcode/LLL) · K053653

# FIDIS CONNECTIVE 10, MODEL MX006 (K053653)

_Biomedical Diagnostics (Bmd) SA · LLL · Mar 13, 2006 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K053653

## Device Facts

- **Applicant:** Biomedical Diagnostics (Bmd) SA
- **Product Code:** [LLL](/productcode/LLL.md)
- **Decision Date:** Mar 13, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5100
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

The FIDIS™ Connective 10* kit is a fluorescent immunoassay for the semi-quantitative simultaneous detection of 10 autoantibody specificities directed against double stranded DNA (dsDNA), SSA (60 kDA and 52 kDA), SSB, Sm, Sm/RNP, Scl-70, Jo-1 ribosome and centromere in human serum. (*Antibodies to dsDNA, SSA, SSB, Sm, Sm/RNP, Scl-70, Jo-1, ribosome and centromere can be reported using this assay). The test system is used to screen serum samples and detect the presence of antinuclear antibodies associated with connective diseases systemic lupus erythematosus (SLE), Sjogren's syndrome, mixed connective tissue disease (MCTD), scleroderma, dermatomyositis, and CREST syndrome, in conjunction with clinical findings and other laboratory tests.

## Device Story

FIDIS™ Connective 10* is a multiplexed fluorescent immunoassay for semi-quantitative detection of 10 autoantibody specificities in human serum. Input: patient serum samples. Process: antigens (dsDNA, SSA, SSB, Sm, Sm/RNP, Scl-70, Jo-1, ribosome, centromere) are covalently coupled to color-coded microspheres; patient autoantibodies bind to immobilized antigens; phycoerythrin-labeled anti-human IgG conjugate binds to antibody-antigen complexes. Output: flow cytometric measurement of fluorescence intensity per bead set, interpolated against a calibration system to determine antibody titers. Used in clinical laboratories; operated by trained technicians. Results assist clinicians in screening for systemic autoimmune diseases in conjunction with other clinical findings.

## Clinical Evidence

Bench testing only. Performance supported by comparison studies against predicate devices using positive, equivocal, and negative serum samples. Data included analysis of samples from apparently healthy subjects and samples with potential biological cross-reactivity.

## Technological Characteristics

System includes FIDIS flow cytometer, XYP automatic sampling platform, SD pump, and MLX-BOOSTER software. Uses color-coded microspheres sensitized with specific antigens and phycoerythrin-conjugated anti-human IgG. Detection via flow cytometry fluorescence measurement. Fully integrated, automated immunodiagnostic system.

## Regulatory Identification

An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).

## Predicate Devices

- Varelisa dsDNA antibodies ([K950031](/device/K950031.md))
- Varelisa RO (SS-A) antibodies ([K944169](/device/K944169.md))
- Varelisa LA (SS-B) antibodies ([K944168](/device/K944168.md))
- Varelisa SM antibodies ([K042629](/device/K042629.md))
- Varelisa RNP antibodies ([K993589](/device/K993589.md))
- Varelisa JO1 antibodies ([K944169](/device/K944169.md))
- Varelisa SCL-70 antibodies ([K944172](/device/K944172.md))
- Varelisa CENTROMERE antibodies ([K944171](/device/K944171.md))
- QuantaLite RIBOSOME P ELISA ([K981237](/device/K981237.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k053653

B. Purpose for Submission:
This is a new device.

C. Measurands:
Anti-SS-A, anti-SS-B, anti-Sm, anti-Sm/RNP, anti-dsDNA, anti-Scl-70, anti-Jo1, ribosome and centromere

D. Type of Test:
Multiplex bead-based flow cytometric immunoassay

E. Applicant:
Biomedical Diagnostics (bmd) S.A.

F. Proprietary and Established Names:
FIDIS™ Connective 10*

G. Regulatory Information:
1. Regulation section:
21CFR§ 866.5100, Antinuclear Antibody Immunological Test System
2. Classification:
Class II
1. Product code:
LLL, Extractable Antinuclear Antibody, Antigen, and Control
LKJ, Antinuclear Antibody, Antigen, Control
LKO, Anti-RNP Antibody, Antigen, Control
LKP, Anti-Sm Antibody, Antigen, and Control
LSW, Anti-DNA Antibody, Antigen and Control
LJM, Antinuclear Antibody (Enzyme Labeled), Antigen, Controls
MQA, Anti-Ribosomal P Antibodies
4. Panel:
Immunology (82)

H. Intended Use:
1. Intended use(s):
The FIDIS™ Connective 10* kit is a fluorescent immunoassay for the semi-quantitative simultaneous detection of 10 autoantibody specificities directed against double stranded DNA (dsDNA), SSA (60 kDA and 52 kDA), SSB, Sm, Sm/RNP, Scl-70, Jo-1 ribosome and centromere in human serum. (*Antibodies to dsDNA, SSA, SSB, Sm, Sm/RNP, Scl-70, Jo-1, ribosome and centromere can be reported using this assay).
2. Indication(s) for use:
The FIDIS™ Connective 10* kit is a semi-quantitative homogeneous fluorescent based microparticles immunoassay using flow cytometry readings. It is designed for the simultaneous detection of autoantibody specificities: double stranded DNA (dsDNA), SSA (60 kDA and 52 kDA), SSB, Sm, Sm/RNP, Scl-70, Jo-1 ribosome and centromere in human serum. (*Antibodies to dsDNA, SSA, SSB, Sm, Sm/RNP, Scl-70, Jo-1 ribosome and centromere can be reported using this assay).

{1}

The test system is used to screen serum samples and detect the presence of antinuclear antibodies associated with connective diseases systemic lupus erythematosus (SLE), Sjogren's syndrome, mixed connective tissue disease (MCTD), scleroderma, dermatomyositis, and CREST syndrome, in conjunction with clinical findings and other laboratory tests.

3. Special conditions for use statement(s):

This device is for prescription use only.

4. Special instrument requirements:

FIDISTM Instrument (Luminex  $100^{\mathrm{TM}}$  plus FIDISTM MLX-Booster Software)

CARISTM (Optional diluting and dispensing device)

# I. Device Description:

The device consists of the following: color-coded sets of microspheres (ready-to-use). Each microsphere set is conjugated to one of the following antigens: dsDNA, SSa (60 kDA and 52 kDA), SSB, Sm, Sm/RNP, Scl-70, Jo-1, ribosome and centromere; calibrator (ready to use); positive control (to be diluted); negative control (to be diluted); goat anti-human IgG conjugate coupled phycoerythrin (to be diluted) and 10x concentrated PBS-Tween (to be diluted with distilled water)

# J. Substantial Equivalence Information:

1. Predicate device name(s):

Varelisa LA (SS-B) Antibodies, Varelisa RO (SS-A) Antibodies, Varelisa Jo-1 Antibodies, Varelisa Sm Antibodies, Varelisa RNP Antibodies, Varelisa dsDNA Antibodies, Varelisa Scl-70 Antibodies, Varelisa Centromere Antibodies, and QuantaLite Ribosome P ELISA.

2. Predicate 510(k) number(s):

k944168, k944169, k944173, k944170, k993589, k950031, k944172, k944171 and k981237.

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | FIDISTM Connective 10* | Individual Varelisa ELISA assays for autoantibodies to LA (SS-B), RO (SS-A), Jo-1, Sm, RNP, dsDNA, ScL-70, Centromere and QuantaLite Ribosome P  |
|  Intended Use | Individual determination of IgG antibodies to dsDNA, SSA 60 kDA and 52 kDA, SSB, Sm/RNP, ScL-70, Jo-1, ribosome and centromere | Same  |
|  Sample type | Serum | Same  |
|  Type of test | Semi-quantitative | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Assay type | Flow Cytometer based | ELISA  |

{2}

|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Assay Format | Multiplexed | Individual analytes  |
|  Sample Dilution | 1:200 | 1:101  |
|  Substrate solution | None | TMB  |
|  Instrument | Luminometer (Luminex 2.2) | Spectrophotometer  |
|  Detection method | Fluorescence | Colorimetric  |
|  Conjugate | Phycoerythrin | HR peroxidase  |
|  Solid Phase Capture | Color-coded microsphere | Microwells  |
|  Antigens | Recombinant dsDNA, SSA,SSB, Scl70, Jo1, and Centromere; Purified Sm, Sm/RNP and Ribosome P | Recombinant dsDNA, SSA, SSB, Scl70, Jo1, RNP, and Centromere; Synthetic SmD peptide, And Ribosome P  |

**K. Standard/Guidance Document Referenced (if applicable):**
None referenced.

**L. Test Principle:**
FIDISTM Connective 10* is a multiplexed semi-quantitative, fluorescent immunoassay performed on the FIDISTM Instrument (Luminex 100™) with the MLX Booster software. Each antigen (dsDNA, SSA [60 kDA and 52 kDA], SSB, Sm, Sm/RNP, Scl-70, Jo-1, ribosome and centromere) is covalently coupled to an individual set of microspheres through its surface functional groups. The different sets of antigen-coupled microspheres are mixed together to constitute the final microspheres reagent and put into wells of a microtiter plate. Prediluted controls and diluted patient sera are added to separate wells allowing autoantibodies to bind to the immobilized antigens on the beads. After incubation, a wash step through a filtration process will remove the unbound antibodies. Then a phycoerythrin labeled anti-human IgG is added to each well and binds to any patient autoantibodies/antigen complexes on the microspheres. The samples are subsequently measured in the FIDIS Instrument. The flow cytometer discriminates the different bead sets as well as measures the fluorescent intensity of the conjugate on each bead. For each sample, the antibody titer for each antigenic specificity is interpolated against a calibration system.

FIDISTM Connective 10* was optimized by flow cytometry for the average binding capacity at the given dilution (1:200) from the median fluorescence value using 200 microspheres per parameter.

**M. Performance Characteristics (if/when applicable):**
1. Analytical performance:
a. Precision/Reproducibility:
To evaluate intra-assay and inter assay reproducibility, five samples covering the reportable range of the assay were analyzed on the FIDISTM Connective 10*. For within-run, the five samples were assayed 10 times in one run and for between-run; the three samples were assayed 2 times per run for 6 runs. Results were as follows:

{3}

|   | Within-run |   | Between-run  |   |
| --- | --- | --- | --- | --- |
|  Antigenic Specificity | Mean value | %CV | Mean Value | %CV  |
|  dsDNA | 45 | 6.0 | 48 | 14.0  |
|   |  55 | 2.5 | 49 | 6.6  |
|   |  71 | 5.0 | 67 | 8.0  |
|   |  105 | 6.1 | 102 | 7.4  |
|   |  687 | 2.6 | 667 | 5.1  |
|  SSA 60 kDA / SSA52kDA | 28 | 11.0 | 27 | 9.0  |
|   |  50 | 12.0 | 46 | 13.0  |
|   |  66 | 13.8 | 52 | 8.5  |
|   |  81 | 3.5 | 76 | 3.9  |
|   |  248 | 3.6 | 241 | 4.9  |
|  SSB | 46 | 8.0 | 41 | 11.0  |
|   |  54 | 6.0 | 53 | 10.0  |
|   |  56 | 6.7 | 45 | 7.5  |
|   |  80 | 3.1 | 73 | 9.1  |
|   |  131 | 2.5 | 122 | 6.2  |
|  Sm | 25 | 11.0 | 26 | 11.0  |
|   |  28 | 9.0 | 29 | 8.0  |
|   |  45 | 9.7 | 35 | 7.1  |
|   |  90 | 4.0 | 87 | 6.4  |
|   |  115 | 2.5 | 134 | 6.7  |
|  Sm/RNP | 20 | 10.0 | 31 | 10.0  |
|   |  44 | 5.5 | 55 | 8.9  |
|   |  67 | 4.0 | 63 | 9.0  |
|   |  89 | 3.1 | 88 | 4.6  |
|   |  110 | 2.6 | 114 | 7.1  |
|  Scl-70 | 28 | 9.0 | 30 | 14.0  |
|   |  34 | 15.0 | 35 | 14.0  |
|   |  49 | 10.9 | 76 | 8.8  |
|   |  97 | 8.4 | 84 | 5.8  |
|   |  180 | 3.5 | 152 | 8.5  |
|  Jo-1 | 50 | 8.0 | 59 | 9.0  |
|   |  60 | 8.0 | 59 | 9.0  |
|   |  91 | 7.0 | 79 | 4.7  |
|   |  176 | 6.2 | 170 | 7.8  |
|   |  285 | 2.1 | 266 | 8.8  |
|  Centromere | 30 | 8.0 | 27 | 12.0  |
|   |  54 | 4.2 | 44 | 6.3  |
|   |  94 | 3.3 | 82 | 7.3  |
|   |  97 | 7.0 | 88 | 9.0  |
|   |  181 | 2.1 | 156 | 7.3  |
|  Ribosome | 51 | 5.0 | 49 | 8.0  |
|   |  68 | 3.0 | 52 | 8.6  |
|   |  69 | 7.0 | 64 | 11.0  |
|   |  87 | 2.0 | 79 | 8.5  |
|   |  136 | 2.5 | 132 | 5.5  |

Precision of the assay using the optional automated CARIS system was assessed. For within-run, seven samples were assayed 10 times in one run and for between-run; seven samples were assayed 4 times per run for 6 runs. Three antigens were chosen for this study. Results were as follows:

{4}

|   | Within-run |   | Between-run  |   |
| --- | --- | --- | --- | --- |
|  Antigenic Specificity | Mean value | %CV | Mean Value | %CV  |
|  dsDNA | 370 | 6.0 | 360 | 6  |
|   |  77 | 12.0 | 73 | 11.0  |
|  SSB | 193 | 3.0 | 187 | 7.0  |
|  SSA 60 kDA | 150 | 6.5 | 149 | 9.0  |
|   |  70 | 5.0 | 69 | 4.0  |
|   |  56 | 8.0 | 53 | 14.0  |
|   |  12 | 8.0 | 11 | 10.0  |

b. Linearity/assay reportable range:

Linearity is not claimed for this assay.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The dsDNA values in the calibrator are established using the WHO

International Standard for anti-double stranded DNA (dsDNA), human code:

WO/80. Other calibrator titers are expressed in arbitrary units per mL

(AU/mL).

d. Detection limit:

Not applicable

e. Analytical specificity:

Interfering substances

To evaluate the system for potential cross reactivity to other antibodies and interference from blood components, 52 samples were tested. High level of complement proteins were used but the specific kind of complement was not provided. A statement to avoid the use of abnormal concentration of these samples was added to the Limitations of the Procedure. The following results were obtained:

|   | Number of Positive Samples  |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  dsDNA | SSA 60 & 52kD | SSB | Sm | Sm/RNP | Scl70 | Jo1 | Centro mere | Ribo some  |
|  Cryoglobulinemia (6) |  |  |  |  |  |  |  | 1 |   |
|  Complement (8) | 2 | 1 |  |  | 2 |  |  |  | 1  |
|  Hypergammaglobulinemia (1) |  |  |  |  |  |  |  |  |   |
|  IgG monoclonal Ig (3) |  |  |  |  |  |  |  |  |   |
|  IgM monoclonal Ig (8) |  |  |  |  |  |  |  |  |   |
|  Rheumatoid Factor (8) |  |  |  |  |  |  |  |  |   |
|  Blood Plasma (6) |  |  |  |  |  |  |  |  |   |
|  Hemolyzed sera (6) |  |  |  |  |  |  |  |  |   |
|  Citrated Plasma (6) |  |  |  |  |  |  |  |  |   |

f. Assay cut-off:

Not applicable

2. Comparison studies:

a. Method comparison with predicate device:

The tables below show the comparison of serum samples  $(N = 434)$  that were

{5}

tested with the FIDIS™ Connective 10* and the predicate devices. No information about age, gender, and clinical status was provided.

- 307 positive samples for one or more parameters related to systemic autoimmune diseases
- 227 negative samples

All borderline results with the two devices were considered negative.

|   | Varelisa dsDNA  |   |   |
| --- | --- | --- | --- |
|   |   |  Pos | Neg  |
|  FIDIS dsDNA | Pos | 58 | 8  |
|   |  Neg | 3 | 365  |
|  Positive % agreement: |   | 95.08% (95% CI: 89.7% - 100%)  |   |
|  Negative % agreement: |   | 97.86% (95% CI: 96.4% n- 99.3%)  |   |
|  Overall % agreement: |   | 97.47% (95% CI: 96.0% - 98.9%)  |   |
|   | Varelisa SS-A  |   |   |
| --- | --- | --- | --- |
|   |   |  Pos | Neg  |
|  FIDIS SS-A | Pos | 116 | 30  |
|   |  Neg | 8 | 283  |
|  Positive % agreement: |   | 93.55% (95% CI: 89.2% - 97.9%)  |   |
|  Negative % agreement: |   | 90.42% (95% CI: 87.2% - 93.7%)  |   |
|  Overall % agreement: |   | 91.30% (95% CI: 88.7% - 93.9%)  |   |
|   | Varelisa SS-B  |   |   |
| --- | --- | --- | --- |
|   |   |  Pos | Neg  |
|  FIDIS SS-B | Pos | 49 | 19  |
|   |  Neg | 3 | 363  |

Positive % agreement: 94.23% (95% CI: 87.9%, -100.0%)

Negative % agreement: 95.03% (95% CI: 92.8 - 97.2%)

Overall % agreement: 94.93% (95% CI: 92.9% -97.0%)

|   | Varelisa Sm  |   |   |
| --- | --- | --- | --- |
|   |   |  Pos | Neg  |
|  FIDIS Sm | Pos | 51 | 13  |
|   |  Neg | 5 | 365  |

Positive % agreement: 91.07% (95% CI: 83.6% - 98.5%)

Negative % agreement: 96.56% (95%CI: 94.1% - 98.4%)

Overall % agreement: 95.85% (95% CI: 94.0% - 97.7%)

|   | Varelisa Sm/RNP  |   |   |
| --- | --- | --- | --- |
|   |   |  Pos | Neg  |
|  FIDIS Sm/RNP | Pos | 84 | 18  |
|   |  Neg | 14 | 318  |

Positive % agreement: 85.71% (95% CI: 78.8% - 92.6%)

{6}

Negative % agreement: 94.64% (95%CI: 92.2% - 97.1%)

Overall % agreement: 92.63% (95% CI: 90.2%- 95.1%)

|   | Varelisa Scl70  |   |   |
| --- | --- | --- | --- |
|   |   |  Pos | Neg  |
|  FIDIS Scl-70 | Pos | 32 | 5  |
|   |  Neg | 2 | 385  |

Positive % agreement: 94.12% (95% CI: 86.2% - 100%)

Negative % agreement: 98.75% (95%CI: 97.7% - 99.8%)

Overall % agreement: 98.4% (95% CI: 97.2% - 99.6%)

|   | Varelisa Jo1  |   |   |
| --- | --- | --- | --- |
|   |   |  Pos | Neg  |
|  FIDIS Jo-1 | Pos | 34 | 8  |
|   |  Neg | 0 | 392  |

Positive % agreement: 100.0%

Negative % agreement: 98.00% (95%CI: 96.6% - 99.4%)

Overall % agreement: 98.16% (95% CI: 96.9% - 99.4%)

|   | Varelisa Centromere  |   |   |
| --- | --- | --- | --- |
|   |   |  Pos | Neg  |
|  FIDIS Centromere | Pos | 45 | 4  |
|   |  Neg | 2 | 383  |

Positive % agreement: 95.74% (95%CI: 90.0% - 100.0%)

Negative % agreement: 98.97% (95%CI: 98.0% - 100.0%)

Overall % agreement: 98.62% (95% CI: 97.5% - 99.7%)

|   | QuantaLite Ribosome  |   |   |
| --- | --- | --- | --- |
|   |   |  Pos | Neg  |
|  FIDIS Ribosome | Pos | 24 | 4  |
|   |  Neg | 2 | 404  |

Positive % agreement: 92.31% (95%CI: 82.1% - 100.0%)

Negative % agreement: 99.02% (95%CI: 98.1% - 100.0%)

Overall % agreement: 98.62% (95% CI: 97.5% - 99.7%)

Comparison of the automated CARIS system and manual method

A comparison study between the manual method and the automated

CARIS™ system was also performed. The comparison was performed on 36 samples. All borderline results with the two methods were considered negative.

|  SSA (60kD+52kD) | Manual FIDIS  |   |   |
| --- | --- | --- | --- |
|   |   |  Pos | Neg  |
|  CARIS FIDIS | Pos | 18 | 1  |
|   |  Neg | 0 | 17  |

{7}

Positive % agreement: 100.0%
Negative % agreement: 94.4%
Overall % agreement: 97.2%

|  SSB | Manual FIDIS  |   |   |
| --- | --- | --- | --- |
|   |   |  Pos | Neg  |
|  Caris FIDIS | Pos | 4 | 0  |
|   |  Neg | 1 | 31  |

Positive % agreement: 80.0%
Negative % agreement: 100.0%
Overall % agreement: 97.2%

|  Sm | Manual FIDIS  |   |   |
| --- | --- | --- | --- |
|   |   |  Pos | Neg  |
|  Caris FIDIS | Pos | 2 | 0  |
|   |  Neg | 0 | 34  |

Positive % agreement: 100.0%
Negative % agreement: 100.0%
Overall % agreement: 100.0%

|  Sm/RNP | Manual FIDIS  |   |   |
| --- | --- | --- | --- |
|   |   |  Pos | Neg  |
|  Caris FIDIS | Pos | 6 | 0  |
|   |  Neg | 0 | 30  |

Positive % agreement: 100.0%
Negative % agreement: 100.0%
Overall % agreement: 100.0%

|  Scl70 | Manual FIDIS  |   |   |
| --- | --- | --- | --- |
|   |   |  Pos | Neg  |
|  Caris FIDIS | Pos | 4 | 0  |
|   |  Neg | 0 | 32  |

Positive % agreement: 100.0%
Negative % agreement: 100.0%
Overall % agreement: 100.0%

|  Jo1 | Manual FIDIS  |   |   |
| --- | --- | --- | --- |
|   |   |  Pos | Neg  |
|  Caris FIDIS | Pos | 5 | 0  |
|   |  Neg | 0 | 31  |

Positive % agreement: 100.0%
Negative % agreement: 100.0%
Overall % agreement: 100.0%

8

{8}

9

|  Centromere | Manual FIDIS  |   |   |
| --- | --- | --- | --- |
|   |   |  Pos | Neg  |
|  Caris FIDIS | Pos | 7 | 0  |
|   |  Neg | 0 | 29  |

Positive % agreement: 100.0%
Negative % agreement: 100.0%
Overall % agreement: 100.0%

|  Ribosome | Manual FIDIS  |   |   |
| --- | --- | --- | --- |
|   |   |  Pos | Neg  |
|  Caris FIDIS | Pos | 0 | 0  |
|   |  Neg | 0 | 39  |

Positive % agreement: 100.0%
Negative % agreement: 100.0%
Overall % agreement: 100.0%

|  dsDNA | Manual FIDIS  |   |   |
| --- | --- | --- | --- |
|   |   |  Pos | Neg  |
|  Caris FIDIS | Pos | 7 | 0  |
|   |  Neg | 0 | 29  |

Positive % agreement: 100.0%
Negative % agreement: 100.0%
Overall % agreement: 100.0%

b. Matrix comparison:
Serum is the only recommended matrix.

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
The reported expected ranges were estimated from 2 populations:
- 50 samples from blood donors
- 48 samples selected from their potential biological interferences and according to WHO standard for dsDNA specificity

|  Arbitrary units (AU/mL) | <30 AU/mL | 30-40 AU/mL | >40 AU/mL  |
| --- | --- | --- | --- |
|  International units (IU/mL) | <30 | 30-40 | >40  |
|  For dsDNA | IU/mL | IU/mL | IU/mL  |
|  Interpretation | Negative | Equivocal¹ | Positive  |

{9}

The negative thresholds (30 AU/mL or 30 IU/mL) correspond to the 97.9th percentile for dsDNA, SSA, Sm/RNP; 99.0% for centromere and ribosome, and 100% for SSB, Sm, Scl70 and Jo1 for the populations studied.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

10

---

**Source:** [https://fda-staging.innolitics.com/device/K053653](https://fda-staging.innolitics.com/device/K053653)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
