System for maturing immature oocytes of infertile women undergoing in vitro fertilization who for medical reasons cannot or choose not to undergo conventional ovarian stimulation using drugs. Oocyte Washing Medium: This medium is used for washing immature oocytes contained within cumulus-oocyte complexes (COCs), collected from the follicles before maturation in culture. Oocyte Maturation Medium: This medium is used for maturation in culture of immature oocytes in an incubator. Embryo Maintenance: This medium is used for embryonic culture following insemination by intracytoplasmic sperm injection (ICSI).
Device Story
IVM System comprises three specialized reproductive media: Oocyte Washing Medium, Oocyte Maturation Medium, and Embryo Maintenance Medium. Used in clinical IVF laboratories by embryologists/clinicians. Oocyte Washing Medium maintains immature oocytes during collection/washing; Oocyte Maturation Medium supports in vitro maturation of immature oocytes in an incubator; Embryo Maintenance Medium supports embryo culture following ICSI. These media provide necessary nutrients, amino acids, and vitamins to support gamete and embryo viability outside the body. By enabling maturation of immature oocytes, the system provides an alternative for patients unable or unwilling to undergo conventional drug-induced ovarian stimulation.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and intended use comparisons to legally marketed predicate devices.
Technological Characteristics
Balanced salt solutions containing nutrients, amino acids, and vitamins. Formulated for specific stages of IVF: oocyte washing, maturation, and embryo maintenance. Class II reproductive media (Product Code: MQL).
Indications for Use
Indicated for infertile women undergoing in vitro fertilization who cannot or choose not to undergo conventional ovarian stimulation using drugs. Includes Oocyte Washing Medium for washing immature oocytes (COCs), Oocyte Maturation Medium for in vitro maturation, and Embryo Maintenance Medium for culture post-ICSI.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Media For Gamete Preparation And Embryo Culture (Blastocyst Medium) (K002836)
Submission Summary (Full Text)
{0}------------------------------------------------
## 5. 510(k) Summary
Image /page/0/Picture/2 description: The image shows the word "SAGE" inside of a circle. The circle is not fully complete, as there is a break in the lower right-hand side. The word "SAGE" is written in all capital letters and is centered inside of the circle. The font is a simple sans-serif font.
Assisted Reproduction Products™
SAGE In-Vitro Fertilization, Inc. a CooperSurgical Company CA. Vanniato Don a mispa (2) fine a (2015) 6:11-6 11. 2490 100 11.4 30
KOS 36 46
Submitter's name: Address:
Phone: Fax number: SAGE In-Vitro Fertilization 95 Corporate Drive Trumbull, CT 09911 203-601-5200 203-601-4737
JUL 1 4 2006
Name of contact person:
Grace Holland Regulatory Specialists, Inc 3722 Ave. Sausalito Irvine, CA 92606 Phone: 949-262-0411 Fax: 949-552-2821 Email: grace@regulatoryspecialists.com
Date the summary was prepared: December 26, 2005
Name of the device: Trade or proprietary name: Common or usual name: Classification name:
Name of the device: Trade or proprietary name: Common or usual name: Classification name:
Oocyte Washing Medium Oocyte Washing Medium Oocyte Washing Medium Reproductive media
Oocyte Maturation Medium Oocyte Maturation Medium Oocyte Maturation Medium Reproductive media
Name of the device: Trade or proprietary name: Common or usual name: Classification name:
Embryo Maintenance Medium Embryo Maintenance Medium Embryo Maintenance Medium Reproductive media
{1}------------------------------------------------
The legally marketed devices to which we are claiming equivalence [807.92(a)(3)]:
| Device | Ref# | Decided |
|------------------------------------------------------------------------|---------|------------|
| MediCult IVM System | K041284 | 11/22/2004 |
| Media For Gamete Preparation And<br>Embryo Culture (Blastocyst Medium) | K002836 | 10/03/2000 |
Description of the devices:
Qocyte Washing Medium is a balanced salt solution containing nutrients, amino acids and vitamins to maintain immature oocytes during their collection and washing.
Oocyte Maturation Medium is a balanced salt solution containing nutrients, amino acids and vitamins for the in vitro culture of immature oocytes.
Embryo Maintenance Medium is a balanced salt solution containing nutrients and amino acids for maintaining embryos in culture.
Indications:
Oocyte Washing Medium: This medium is used for washing immature oocytes contained within cumulus-oocyte complexes (COCs), collected from the follicles before maturation in culture.
Oocyte Maturation Medium: This medium is used for maturation in culture of immature oocytes in an incubator.
Embryo Maintenance Medium: This medium is used for embryonic culture following insemination by intracytoplasmic sperm injection (ICSI).
Summary of the technological characteristics of our device compared to the predicate device:
The predicates and these devices were compared in the following areas and found to have equivalent technological characteristics and therefore are equivalent.
{2}------------------------------------------------
Formula Indications for Use Performance Testing
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract caduceus or a representation of human services, featuring three curved lines that suggest movement or flow.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
## JUL 1 4 2006
SAGE In-Vitro Fertilization, Inc. % Ms. Grace Holland Regulatory Specialist Regulatory Specialists, Inc. 3722 Ave. Sausalito IRVINE CA 92606
Re: K053646
Trade/Device Name: IVM System Regulation Number: 21 CFR 884.6180 Regulation Name: Reproductive media and supplements Regulatory Class: II Product Code: MQL Dated: June 15, 2006 Received: June 20, 2006
## Dear Ms. Holland:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image shows a logo with the text "1906-2006" at the top. Below that is the text "FDA Centennial" in a stylized font. There are three stars below the word "Centennial". The logo is surrounded by a circular border with text.
{4}------------------------------------------------
Page 2 –
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 4. Indications for Use Statement Indications for Use
510(k) Number (if known): Ko5 36466
Device Name: _ IVM System_____________________________________________________________________________________________________________________________________________________ Indications for Use:
System for maturing immature oocytes of infertile women undergoing in vitro fertilization who for medical reasons cannot or choose not to undergo conventional ovarian stimulation using drugs.
Oocyte Washing Medium: This medium is used for washing immature oocytes contained within cumulus-oocyte complexes (COCs), collected from the follicles before maturation in culture.
Oocyte Maturation Medium: This medium is used for maturation in culture of immature oocytes in an incubator.
Embryo Maintenance: This medium is used for embryonic culture following insemination by intracytoplasmic sperm injection (ICSI).
AND/OR
x Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) ---------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Ingram
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Device
Page 1 of 1
REGULATORY SPECIALISTS, INC.
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.