GIVEN DIAGNOSTIC SYSTEM

K053639 · Given Imaging , Ltd. · NEZ · May 8, 2006 · Gastroenterology, Urology

Device Facts

Record IDK053639
Device NameGIVEN DIAGNOSTIC SYSTEM
ApplicantGiven Imaging , Ltd.
Product CodeNEZ · Gastroenterology, Urology
Decision DateMay 8, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1300
Device ClassClass 2

Indications for Use

The Given® AGILE Patency System is an accessory to the PillCam video capsule and is intended to verify adequate patency of the gastrointestinal tract prior to administration of the PillCam video capsule in patients with known or suspected strictures.

Device Story

AGILE Patency System serves as accessory to PillCam video capsule; verifies GI tract patency in patients with known/suspected strictures. System components: AGILE Patency capsule, AGILE Patency Scanner, TesTag. Patient ingests capsule; peristalsis moves it through GI tract. Scanner detects capsule presence in body. If capsule excreted whole or not detected by scanner within 30 hours post-ingestion, GI tract patency is indicated, allowing PillCam administration. If scanner detects capsule at 30 hours, patency not indicated. Capsule designed to dissolve 30-42 hours post-ingestion. Fluoroscopy may be used for localization if scanner unavailable. Benefits patient by preventing PillCam retention in strictured GI tracts.

Clinical Evidence

No clinical data provided; device relies on bench testing and design verification to demonstrate substantial equivalence.

Technological Characteristics

System includes ingestible capsule, handheld scanner, and interference tester (TesTag). Capsule is designed to dissolve 30-42 hours post-ingestion. Scanner detects capsule presence via telemetric signal. Standalone operation.

Indications for Use

Indicated for patients with known or suspected gastrointestinal strictures to verify GI tract patency prior to administration of the PillCam video capsule.

Regulatory Classification

Identification

An ingestible telemetric gastrointestinal capsule imaging system is used for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an antenna array, a receiving/recording unit, a data storage device, computer software to process the images, and accessories.

Special Controls

*Classification.* Class II (special controls). The special control is FDA's guidance, “Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging Systems; Final Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Page 1 of 2 MAY - 8 2003 # 510(K) SUMMARY Given® AGILE Patency System 510(k) Number K_ K053639 ## Applicant's Name: Given Imaging Ltd. Hermon Building (Shaar Yoqneam) P.O. Box 258 New Industrial Zone Yokneam 20692, Israel Tel.: 011-972-4-9097730 Fax: 011-972-4-9592466 #### Contact Person: Shosh Friedman, RAC Senior V.P. Regulatory & Clinical Affairs Tel: 011-972-4- 909 7784 Fax: 011-972-4-993 8060 Email: shosh@givenimaging.com #### Trade Name: Given® AGILE Patency System (an optional accessory to the Given® Diagnostic System with PillCam Capsules) ### Classification Name: Ingestible Telemetric Gastrointestinal Capsule Imaging System #### Classification: FDA has classified Ingestible Telemetric Gastrointestinal Capsule Imaging System as class II devices (product code 78NZE and 78NSI) and they are reviewed by the Gastroenterology Panel. #### Predicate Device: The Given AGILE Patency System is an accessory for the Given® Diagnostic System with PillCam™ Capsules (Given Imaging Ltd.) cleared under K010312, K020341, K022362, K022980, K031033, K032405, K040248, K041149, K042960, and K052184. {1}------------------------------------------------ Ko:53639 Page 2 of ### Intended Use: The Given® AGILE Patency System is an accessory to the PillCam video capsule and is intended to verify adequate patency of the gastrointestinal tract prior to administration of the PillCam video capsule in patients with known or suspected strictures. #### Device Description: The Given® AGILE Patency System is a simple and easy to use device for verifying the patency of the GI tract. It consists of the following components: - · Given AGILE Patency capsule - · Given AGILE Patency Scanner - · TesTag (interference tester) Once the patient ingests the Given AGILE Patency capsule it is propelled through the GI tract by normal peristalsis. If the AGILE Patency capsule is excreted structurally whole, then this indicates patency of the GI tract of the patient and a PillCam capsule can be administered. The capsule is designed to dissolve starting 30 hours following ingestion, during a period of approximately 12 hours. If anytime before 30 hours after capsule ingestion the AGILE Patency capsule cannot be detected in the patient's body by the Given AGILE Patency Scanner, it indicates that the GI tract is patent and that the capsule was excreted naturally. The Given AGILE Patency Scanner is used to detect the presence of the Given AGILE Patency capsule in the patient's body. If a scanner is not available, or if localization of a retained capsule is desired, fluoroscopy can be used. If the Given AGILE Patency Scanner detects the Given AGILE Patency capsule in the GI tract 30 hours post-ingestion, then patency of the GI tract is not indicated. Eventually, the capsule dissolves into small fragments and is naturally excreted. ## Substantial Equivalence: Given Imaging Ltd. believes that, based on the information provided in the submission, the Given® AGILE Patency System is substantially equivalent to its predicate devices without raising any new safety and/or efficacy issue. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, with its wings spread and head facing to the right. Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850 MAY -8 2006 Shoshana Friedman, RAC Senior VP, Regulatory and Clinical Affairs Given Imaging Ltd. Hermon Building (Shaar Yoqneam) PO Box 258 New Industrial Zone Yokneam 20692 ISRAEL Re: K053639 Trade/Device Name: Given® AGILE Patency System Regulation Number: 21 CFR §876.1300 Regulation Name: Ingestible telemetric gastrointestinal capsule imaging system Regulatory Class: II Product Code: NSI and NEZ Dated: April 26, 2006 Received: May 1, 2006 Dear Ms. Friedman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Image /page/2/Picture/10 description: The image shows the FDA Centennial logo. The logo is circular with the words "1906-2006" at the top and "For the science of food" at the bottom. In the center of the logo is the acronym "FDA" in large, bold letters. Below the acronym is the word "Centennial" in a smaller font. There are three stars below the word "Centennial". The image also contains the words "Protecting and Promoting Public Health". {3}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, Nancy C Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## INDICATIONS FOR USE STATEMENT 510(k) Number (if known): _K053639 . Device Name: Given® AGILE Patency System Indications for Use: The Given® AGILE Patency System is an accessory to the PillCam video capsule and is intended to verify adequate patency of the gastrointestinal tract prior to administration of the PillCam video capsule in patients with known or suspected strictures. (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED) 510(k) Number_K053639 Prescription Use (Per 21 CFR 801.109) OR Over the Counter Use_ David A. Seymann (Division Sign-Off) Division of Reproductive, and Radiological Devices **510(k) Number.** 7-6
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...