← Product Code [NBC](/productcode/NBC) · K053597

# I-STAT B-TYPE NATRIURETIC PEPTIDE (BNP) (K053597)

_I-Stat Corporation · NBC · Jul 21, 2006 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K053597

## Device Facts

- **Applicant:** I-Stat Corporation
- **Product Code:** [NBC](/productcode/NBC.md)
- **Decision Date:** Jul 21, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1117
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The i-STAT BNP test is an in vitro diagnostic test for the quantitative measurement of B-type natriuretic peptide (BNP) in whole blood or plasma samples using EDTA as the anticoagulant. BNP measurements can be used as an aid in the diagnosis and assessment of severity of congestive heart failure. The i-STAT Controls are assayed liquid plasma used to verify the integrity of newly received i-STAT BNP cartridges. The i-STAT BNP Calibration Verification Controls are assayed liquid plasma used to verify the calibration of i-STAT BNP cartridges throughout the reportable range.

## Device Story

i-STAT BNP test is a single-use cartridge-based assay for quantitative BNP measurement in whole blood or plasma. Operates on i-STAT 1 Analyzer; requires EDTA anticoagulant. Cartridge contains electrochemical sensors on silicon chip; utilizes two-site ELISA principle. Sample introduced to cartridge; BNP captured by antibody-enzyme conjugate; enzyme cleaves substrate to produce electrochemically detectable product proportional to BNP concentration. Used in clinical settings; provides rapid quantitative results to assist physicians in diagnosing and assessing congestive heart failure severity. System includes automated self-checks for sensor performance, fluid integrity, and thermal/pressure transducers; results suppressed if deviations occur. Calibration is factory-set.

## Clinical Evidence

Clinical evidence derived from studies using Abbott AxSYM BNP assay to support reference range transfer. Study population included 890 non-heart failure individuals and 693 heart failure patients (NYHA classes I-IV). AUC for heart failure diagnosis is 0.90 (95% CI: 0.86-0.92) at 100 pg/mL threshold. Clinical sensitivity 74.2%, specificity 91.5%. Method comparison with ARCHITECT assay (n=433) showed correlation slope of 0.97 and r=0.961.

## Technological Characteristics

Two-site ELISA; electrochemical (amperometric) detection. Cartridge contains silicon chip with electrochemical sensors, antibody/alkaline phosphatase conjugate, buffer, and preservatives. Sample type: EDTA whole blood or plasma. Factory-set calibration. Dimensions/form factor: single-use cartridge for i-STAT 1 Analyzer. No optical interferents due to electrochemical detection.

## Regulatory Identification

The B-type natriuretic peptide (BNP) test system is an in vitro diagnostic device intended to measure BNP in whole blood and plasma. Measurements of BNP are used as an aid in the diagnosis of patients with congestive heart failure.

## Special Controls

*Classification.* Class II (special controls). The special control is “Class II Special Control Guidance Document for B-Type Natriuretic Peptide Premarket Notifications; Final Guidance for Industry and FDA Reviewers.”

## Predicate Devices

- Biosite Triage BNP Test (k021317)

## Reference Devices

- Abbott ARCHITECT BNP test

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY AND INSTRUMENT COMBINATION TEMPLATE

A. 510(k) Number:
k053597

B. Purpose for Submission:
New device

C. Measurand:
B-type natriuretic peptide

D. Type of Test:
Quantitative

E. Applicant:
i-STAT Corporation

F. Proprietary and Established Names:
i-STAT BNP test
i-STAT Control Level 1
i-STAT Control Level 2
i-STAT Control Level 3
i-STAT BNP Calibration Verification Control Set

G. Regulatory Information:
1. Regulation section:
862.1117, B-type natriuretic peptide test system
862.1660, Single (specified) analyte controls (assayed and unassayed)

{1}

2. Classification:
Class II, Class I

3. Product code:
NBC, JJX

4. Panel:
75 Chemistry

H. Intended Use:

1. Intended use(s):
The i-STAT BNP test is an in vitro diagnostic test for the quantitative measurement of B-type natriuretic peptide (BNP) in whole blood or plasma samples using EDTA as the anticoagulant. BNP measurements can be used as an aid in the diagnosis and assessment of severity of congestive heart failure.

The i-STAT Controls are assayed liquid plasma used to verify the integrity of newly received i-STAT BNP cartridges.

The i-STAT BNP Calibration Verification Controls are assayed liquid plasma used to verify the calibration of i-STAT BNP cartridges throughout the reportable range.

2. Indication(s) for use:
See Intended use(s) above.

3. Special conditions for use statement(s):
Prescription use only

4. Special instrument requirements:
i-STAT 1 Analyzer

I. Device Description:
Each i-STAT BNP cartridge provides a sample inlet, sensors to detect the BNP, and all the necessary reagents needed to perform the test. The cartridge contains a buffer and preservatives. A list of reactive ingredients is indicated below:

{2}

|  Reactive Ingredient | Biological Source  |
| --- | --- |
|  Antibody/Alkaline Phosphatase Conjugate | Murine IgG:Bovine Intestine  |
|  IgG | Caprine IgG: Murine IgG  |
|  Sodium Aminophenyl Phosphate | N/A  |
|  Heparin | Porcine Intestine  |

The i-STAT BNP Controls are supplied as assayed frozen liquid plasma at 3 levels, Control Level 1, Control Level 2 and Control Level 3. The human sera used in the preparation of this product has been tested by FDA approved test methods and found negative/non-reactive for HIV-1, HIV-2, HBsAg, HCV, HTLV-1 and HTLV-2.

The i-STAT Verification Control Set is supplied as 3 levels of assayed frozen liquid plasma at 3 levels, Level 1, Level 2 and Level 3. The human sera used in the preparation of this product has been tested by FDA approved test methods and found negative/non-reactive for HIV-1, HIV-2, HBsAg, HCV, HTLV-1 and HTLV-2.

# J. Substantial Equivalence Information:

1. Predicate device name(s):

Biosite Triage BNP Test

2. Predicate  $510(\mathrm{k})$  number(s):

k021317

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Assay methodology | Two-site ELISA | Two-site ELISA  |
|  Capture site | Heterogeneous | Heterogeneous  |
|  Capture antibodies | Monoclonal | Monoclonal  |
|  Enzyme label antibody | Monoclonal | Monoclonal  |
|  Sample type | Whole blood or plasma | Whole blood or plasma  |
|  Acceptable samples | EDTA anti-coagulated blood or plasma | EDTA anti-coagulated blood or plasma  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Enzyme label | Fluorescent dye | Alkaline phosphatase  |
|  Enzyme detection | Fluorescent | Electrochemical  |
|  Sample volume | 250 μL | 20 μL  |
|  Reportable range | 15-5000 pg/mL | 5-5000 pg/mL  |

{3}

# K. Standard/Guidance Document Referenced (if applicable):

CLSI Guideline EP7-A; CLSI Guideline EP9-A2; CLSI Guideline C-28-A2

# L. Test Principle:

The i-STAT BNP test cartridge uses a two-site enzyme-linked immunosorbant assay (ELISA) method. Antibodies specific for BNP are located on an electrochemical sensor fabricated on a silicon chip. Also deposited in another location on the sensor silicon chip is an antibody/alkaline phosphatase enzyme conjugate specific to a separate portion of the BNP molecule. The whole blood or plasma sample is brought into contact with the sensors allowing the enzyme conjugate to dissolve into the sample. The BNP within the sample becomes labeled with alkaline phosphatase and is captured onto the surface of the electrochemical sensor during an incubation period of approximately seven minutes. The sample is washed off the sensors, as well as excess enzyme conjugate. Within the wash fluid is a substrate for the alkaline phosphatase enzyme. The enzyme bound to the antibody/antigen/antibody sandwich cleaves the substrate releasing an electrochemically detectable product. The electrochemical (amperometric) sensor measures this enzyme product which is proportional to the concentration of BNP within the sample.

# M. Performance Characteristics (if/when applicable):

# 1. Analytical performance:

# a. Precision/Reproducibility:

Precision data were collected as follows: duplicates of each control were tested daily for a period of 20 days for each of 3 lots of cartridges, resulting in a total of 434 replicates. The average statistics are presented below.

|  Aqueous Control | Mean | % CV (within-run) | % CV (total)  |
| --- | --- | --- | --- |
|  Level 1 | 126 | 9.0 | 11.1  |
|  Level 2 | 1551 | 6.6 | 8.1  |
|  Level 3 | 3337 | 8.0 | 9.8  |

Whole blood imprecision data were collected as follows: whole blood samples from 5 healthy donors were spiked to low, intermediate and high BNP concentrations affording 15 samples, each of which was measured in 10 i-STAT BNP cartridges from a single cartridge lot; three lots of cartridges were employed. The mean within-sample BNP concentration ranged from  $84 - 3925\mathrm{pg / mL}$  and the within-sample imprecision  $(\% \mathrm{CV})$  ranged from 3.4 to  $9.4\%$ ; the average BNP concentration and imprecision were  $1464~\mathrm{pg / mL}$  and  $6.5\%$  respectively. The individual results are presented in the table below:

|  Donor | Mean BNP in pg/mL | % CV  |
| --- | --- | --- |
|  1 | 99 | 6.3  |
|  1 | 765 | 3.4  |

{4}

|  Donor | Mean BNP in pg/mL | % CV  |
| --- | --- | --- |
|  1 | 3803 | 4.1  |
|  2 | 107 | 7.8  |
|  2 | 1049 | 7.8  |
|  2 | 2638 | 5.7  |
|  3 | 108 | 9.2  |
|  3 | 1036 | 6.5  |
|  3 | 3805 | 5.4  |
|  4 | 84 | 9.1  |
|  4 | 783 | 9.4  |
|  4 | 3925 | 7.6  |
|  5 | 95 | 7.0  |
|  5 | 763 | 3.9  |
|  5 | 2900 | 4.8  |

# b. Linearity/assay reportable range:

The dilution linearity of the i-STAT BNP test was studied using EDTA whole blood and plasma samples derived from 3 separate donors. For each donor, the original BNP negative sample and a BNP spiked sample were prepared. This process yielded three BNP positive whole blood samples that were then assayed in duplicate for each of 3 separate i-STAT BNP cartridge lots. These whole blood samples were then diluted using an equal mass of the original unspiked whole blood and assayed in duplicate. From this whole blood data, the BNP recovery was calculated.

|  Whole blood | Concentration | Diluted concentration | % recovery  |
| --- | --- | --- | --- |
|  A | 590 | 312 | 106%  |
|  B | 2765 | 1429 | 103%  |
|  C | 5123 | 2803 | 109%  |

The plasma derived from these three donors was combined in all pair-wise combinations in equal volumes. These combinations were then assayed in duplicate for each of 3 separate i-STAT BNP cartridge lots. The BNP recovery for each pair was calculated using the average of the 6 results.

|  Plasma Blood Sample | Concentration pg/mL) | Diluted Concentration (pg/mL) | % Recovery  |
| --- | --- | --- | --- |
|  A | 590 | — | —  |
|  B | 2764 | — | —  |
|  C | 5123 | — | —  |
|  A+B | — | 1570 | 94%  |
|  B+C | — | 3992 | 101%  |
|  A+C | — | 2734 | 96%  |

{5}

A plasma sample was spiked with BNP to a value of approximately 5000 pg/mL. This sample was subjected to a series of dilutions with fresh, un-spiked plasma in order to prepare a range of concentrations. The concentration of each sample/dilution was calculated based on the measured concentration of the initial solution and the dilutions performed. The diluted samples were then measured in i-STAT BNP test cartridges (N = 6-10). The procedure was repeated with a whole blood sample. The results of these experiments are summarized in the following table:

|  Sample | Dilution | Calculated [BNP] (pg/mL) | Measured [BNP] (pg/mL) | %Recovery  |
| --- | --- | --- | --- | --- |
|  Plasma | 1 | 52 | 57 | 110%  |
|  Plasma | 2 | 104 | 114 | 110%  |
|  Plasma | 3 | 259 | 265 | 103%  |
|  Plasma | 4 | 518 | 560 | 108%  |
|  Plasma | 5 | 1036 | 1002 | 97%  |
|  Plasma | 6 | 2072 | 2277 | 110%  |
|  Plasma | 7 | 3107 | 3384 | 109%  |
|  Plasma | 8 | 4143 | 4222 | 102%  |
|  Whole Blood | 1 | 44 | 41 | 93%  |
|  Whole Blood | 2 | 88 | 88 | 100%  |
|  Whole Blood | 3 | 269 | 287 | 107%  |
|  Whole Blood | 4 | 537 | 554 | 103%  |
|  Whole Blood | 5 | 725 | 720 | 99%  |
|  Whole Blood | 6 | 1450 | 1367 | 94%  |
|  Whole Blood | 7 | 3042 | 2826 | 93%  |
|  Whole Blood | 8 | 4056 | 3856 | 95%  |

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The i-STAT BNP calibrators are traceable to an internal reference standard that has been prepared gravimetrically with synthetic BNP. The internal reference standard underwent a one-time value assignment to align with the ARCHITECT BNP assay with a decision threshold of 100 pg/mL. Manufacturers working calibrators are prepared by gravimetric manipulation of the standard and incorporate a one-time value assignment for alignment of methods. The i-STAT, AxSYM and ARCHITECT assays have been designed, by virtue of their calibration, to report comparable values. The i-STAT vs. ARCHITECT method comparison data exhibits a correlation slope of 0.97 (see method comparison section). Similar data for the ARCHITECT vs. AxSYM exhibited a slope of 1.03.

Stability studies were performed to evaluate the intended storage (open and closed vial) for the i-STAT controls and calibration verification materials. The real-time frozen stability of BNP control/calibration verification materials was established for 3 lots of material, each comprised of 3 levels. Stability was judged to be acceptable provided that the mean BNP concentration measured at each test event be within ± 20 % of the original mean concentration. The stability was acceptable over 5 months frozen storage. The stability studies are ongoing.

{6}

The labeling for the i-STAT controls and calibration verification materials states that, after thawing, the opened or unopened vial is stable for 4 hours when capped and stored at 2 – 8°. Stability studies performed support the 4 hour time limit.

d. Detection limit:

The limit of the blank for the BNP method is 15 pg/mL, which is the lowest BNP level that can be distinguished from zero. The value was estimated using a control material with &lt; 5 pg/mL BNP during a 20 day precision study in which 3 separate lots of BNP test cartridges were tested in duplicate using a pool of 6 i-STAT 1 analyzers for a total of 147 test results.

e. Analytical specificity:

The following muscle proteins were tested at both 1000 pg/mL and 20,000 pg/mL concentrations and found to have no detectable cross-reactivity for BNP: ANP, CNP, and N-terminal pro-BNP.

The i-STAT BNP assay employs electrochemical rather than optical detection. An electrogenic substrate is cleaved by an enzyme label giving rise to an electroactive product that can be oxidized at a sensor electrode generating a signal comprised of electrical current, therefore optical interferents, including hemoglobin, bilirubin, and chylomicrons, do not interfere with this mode of detection.

The following substances were found to have no significant effect (less than 10%) on the BNP method, when added to a plasma pool containing approximately 1000 pg/mL of B-type natriuretic peptide at the concentrations indicated:

7

{7}

|  Compound | Test Level (μmol/L unless otherwise indicated)  |
| --- | --- |
|  Acetaminophen | 1660  |
|  Allopurinol | 294  |
|  Ampicillin | 152  |
|  Ascorbic Acid | 227  |
|  Acetyl Salicylic Acid | 3333  |
|  Atenolol | 37.6  |
|  Caffeine | 308  |
|  Captopril | 23  |
|  Chloramphenicol | 155  |
|  Diclofenac | 169  |
|  Digoxin | 6.15  |
|  Dopamine | 5.87  |
|  Enalaprilat | 0.86  |
|  Erythromycin | 81.6  |
|  Furosemide | 181  |
|  Sodium Heparin | 90 U/mL  |
|  Ibuprofen | 2425  |
|  Isosorbide dinitrate | 636  |
|  Methyldopa | 71  |
|  Nicotine | 6.2  |
|  Nifedipine | 1.156  |
|  Phenytoin | 198  |
|  Propanolol | 7.71  |
|  Salicylic Acid | 4.34  |
|  Theophylline | 222  |
|  Verapamil | 4.4  |
|  Warfarin | 64.9  |

f. Assay cut-off:
BNP results less than or equal to 100 pg/mL are representative of normal values in patients without CHF. See Clinical cut-off section below.

2. Comparison studies:

a. Method comparison with predicate device:

Method comparison data were collected using CLSI guideline EP9-A2. Venous blood samples were collected in EDTA evacuated tubes and analyzed in duplicate on the i-STAT System. A portion of the specimen was centrifuged and the separated plasma was analyzed in duplicate on the i-STAT 1 System and on the comparative method, the Abbott ARCHITECT BNP assay, within 1 hour of collection. Deming regression analysis was performed on the first replicate of each sample. In the method comparison table, n is the number of specimens in the first data set, Sxx and Syy refer to estimates of imprecision based on the duplicates of the comparative and the i-STAT methods respectively. Sy.x is the standard error of the estimate, and r is the correlation coefficient. The samples had BNP values ranging from 5-5000 mg/dL.

{8}

9

Method Comparison

|  Abbott ARCHITECT  |   |
| --- | --- |
|  N | 433  |
|  Mean (pg/mL) | 482.1  |
|  Sxx (pg/mL) | 38.1  |
|  Syy (pg/mL) | 97.6  |
|  Slope | 0.971  |
|  Intercept | -14.4  |
|  Sy.x | 198.0  |
|  Xmin | 5  |
|  Xmax | 4797.7  |
|  Correlation, r | 0.961  |

b. Matrix comparison:

EDTA plasma is the only sample type indicated. The labeling states that performance characteristics have not been established for samples taken from capillary tubes and direct skin punctures (e.g. fingersticks) so these sample types should not be used with the BNP cartridge.

3. Clinical studies:

Clinical studies performed with the Abbott AxSYM BNP assay are included in the labeling for the i-STAT BNP assay. The applicant provided the following to support the transfer of reference ranges:

- The AxSYM, ARCHITECT and i-STAT BNP assays employ an identical antibody set. The average imprecision is similar for the 3 assays as follows: AxSYM average %CV = 7.9 %; ARCHITECT average %CV = 5.2 %; i-STAT average %CV = 9.7 %.
- The i-STAT, AxSYM and ARCHITECT assays have been designed, by virtue of their calibration, to report comparable values.
- The CLSI document C28-A2, How to Define and Determine Reference Intervals in the Clinical Laboratory, provides guidance concerning the transferability of reference ranges from one measurement system to another. The i-STAT vs. ARCHITECT method comparison data exhibits a correlation slope of 0.97 (see method comparison section above). Also, similar data for the ARCHITECT vs. AxSYM exhibited a slope of 1.03 (see k060964).

{9}

# a. Clinical Sensitivity:

In studies performed with the AxSYM BNP Assay, age-matched analysis of the heart failure and non-heart failure populations was performed based on the data published by the American Heart Association in the 2000 Heart and Stroke Statistical Update and according to the age structure of the United States population. The age distributions in the intended use population are approximately as follows: individuals less than 45 years old comprise  $9\%$ , individuals 45-54 years old comprise  $11\%$ , individuals 55-64 years old comprise  $22\%$ , individuals 65-74 years old comprise  $26\%$ , and individuals 75 years and older comprise  $32\%$ . The resulting combined AUC is 0.87 (0.85 to 0.90,  $95\%$  CI). The clinical sensitivity and specificity using a decision threshold of  $100~\mathrm{pg / mL}$  is presented in the table below.

|   | Males (Age Group)  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  All | <45 Years | 45-54 Years | 55-64 Years | 65-74 Years | 75+ Years  |
|  Sensitivity | 71.0% | 47.1% | 57.1% | 57.3% | 70.6% | 86.1%  |
|   |  (328/462) | (8/17) | (24/42) | (51/89) | (115/163) | (130/151)  |
|  95% Confidence Interval | 66.6 to 75.1% | 23.0 to 72.2% | 41.0 to 72.3% | 46.4 to 67.7% | 62.9 to 77.4% | 79.5 to 91.2%  |
|  Specificity | 94.8% | 97.2% | 100.0% | 97.9% | 88.7% | 89.5%  |
|   |  (403/425) | (104/107) | (71/71) | (92/94) | (102/115) | (34/38)  |
|  95% Confidence Interval | 92.3 to 96.7% | 92.0 to 99.4% | 94.9 to 100.0% | 92.5 to 99.7% | 81.5 to 93.8% | 75.2 to 97.1%  |
|   | Females (Age Group)  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  All | <45 Years | 45-54 Years | 55-64 Years | 65-74 Years | 75+ Years  |
|  Sensitivity | 80.5% | 44.4% | 73.3% | 50.0% | 80.6% | 91.7%  |
|   |  (186/231) | (4/9) | (11/15) | (13/26) | (58/72) | (100/109)  |
|  95% Confidence Interval | 74.8 to 85.4% | 13.7 to 78.8% | 44.9 to 92.2% | 29.9 to 70.1% | 69.5 to 88.9% | 84.9 to 96.2%  |
|  Specificity | 88.4% | 95.9% | 90.7% | 89.6% | 85.7% | 80.5%  |
|   |  (411/465) | (94/98) | (68/75) | (69/77) | (114/133) | (66/82)  |
|  95% Confidence Interval | 85.1 to 91.2% | 89.9 to 98.9% | 81.7 to 96.2% | 80.6 to 95.4% | 78.6 to 91.2% | 70.3 to 88.4%  |

# b. Clinical specificity:

See Clinical Sensitivity section above.

{10}

c. Other clinical supportive data (when a. and b. are not applicable):

# 4. Clinical cut-off:

Data from the clinical studies performed with the AxSYM BNP assay were used to generate The Receiver Operating Characteristic (ROC) curve of BNP decision thresholds versus clinical sensitivity and clinical specificity. At a decision threshold of  $100\mathrm{pg / mL}$ , the BNP assay demonstrated a clinical sensitivity and specificity of  $74.2\%$  and  $91.5\%$  respectively. The area under the curve is 0.90 (0.86 to 0.92,  $95\%$  CI).

# 5. Expected values/Reference range:

Plasma samples from 890 individuals (465 females, 425 males) who had not been diagnosed with heart failure were tested with the AxSYM BNP assay. This population included non-hospitalized patients with renal disease (not on dialysis), diabetes, hypertension and chronic obstructive pulmonary disease. BNP levels for these patients were not statistically different from the population of apparently healthy individuals. The data are summarized below.

Non-Heart Failure Population - All (Age Group)

|   | All | <45 Years | 45-54 Years | 55-64 Years | 65-74 Years | 75+ Years  |
| --- | --- | --- | --- | --- | --- | --- |
|  Sample Size (N=) | 890 | 205 | 146 | 171 | 248 | 120  |
|  Median (pg/mL) | 21 | 17 | 9 | 24 | 23 | 31  |
|  Mean (pg/mL) | 39 | 28 | 21 | 37 | 47 | 63  |
|  SD (pg/mL) | 66 | 36 | 30 | 48 | 80 | 109  |
|  95th Percentile | 135 | 85 | 87 | 119 | 160 | 254  |
|  Percentage < 100 pg/mL | 91.5% | 96.6% | 95.2% | 94.2% | 87.1% | 83.3%  |
|  Minimum (pg/mL) | 0 | 0 | 0 | 0 | 0 | 0  |
|  Maximum (pg/mL) | 907 | 263 | 142 | 380 | 907 | 837  |

Non-Heart Failure Population - Males (Age Group)

|   | All | <45 Years | 45-54 Years | 55-64 Years | 65-74 Years | 75+ Years  |
| --- | --- | --- | --- | --- | --- | --- |
|  Sample Size (N=) | 425 | 107 | 71 | 94 | 115 | 38  |
|  Median (pg/mL) | 14 | 12 | 1 | 17 | 21 | 37  |
|  Mean (pg/mL) | 30 | 23 | 9 | 26 | 47 | 49  |
|  SD (pg/mL) | 61 | 34 | 14 | 45 | 96 | 51  |
|  95th Percentile | 104 | 73 | 40 | 80 | 150 | 121  |
|  Percentage < 100 pg/mL | 94.8% | 97.2% | 100.0% | 97.9% | 88.7% | 89.5%  |
|  Minimum (pg/mL) | 0 | 0 | 0 | 0 | 0 | 0  |
|  Maximum (pg/mL) | 907 | 200 | 57 | 380 | 907 | 254  |

{11}

Non-Heart Failure Population - Females (Age Group)

|   | All | <45 Years | 45-54 Years | 55-64 Years | 65-74 Years | 75+ Years  |
| --- | --- | --- | --- | --- | --- | --- |
|  Sample Size (N=) | 465 | 98 | 75 | 77 | 133 | 82  |
|  Median (pg/mL) | 26 | 23 | 23 | 37 | 23 | 25  |
|  Mean (pg/mL) | 46 | 34 | 34 | 51 | 46 | 69  |
|  SD (pg/mL) | 70 | 37 | 36 | 48 | 63 | 126  |
|  95^{th} Percentile | 150 | 89 | 111 | 155 | 159 | 266  |
|  Percentage < 100 pg/mL | 88.4% | 95.9% | 90.7% | 89.6% | 85.7% | 80.5%  |
|  Minimum (pg/mL) | 0 | 0 | 0 | 0 | 0 | 0  |
|  Maximum (pg/mL) | 837 | 263 | 142 | 230 | 374 | 837  |

Plasma samples from 693 patients with diagnosed heart failure (231 females, 462 males) were tested with the AxSYM BNP assay. All patients in this population were categorized according to the functional classification system published by the New York Heart Association (NYHA). This system divides heart failure patients into one of four categories of increasing disease progression (classes I to IV) based upon a subjective assessment of the patient's clinical signs and symptoms. The data from this study are summarized below.

Heart Failure Population - All

|   | NYHA Functional Class  |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|   | All | I | II | III | IV  |
|  Sample Size (N=) | 693 | 124 | 319 | 190 | 60  |
|  Median (pg/mL) | 298 | 133 | 266 | 335 | 1531  |
|  Mean (pg/mL) | 578 | 320 | 432 | 656 | 1635  |
|  SD (pg/mL) | 771 | 388 | 574 | 841 | 1097  |
|  5th Percentile | 14 | 9 | 15 | 12 | 188  |
|  95th Percentile | 2154 | 1257 | 1534 | 2516 | >4000  |
|  Percentage ≥ 100 pg/mL | 74.2% | 58.1% | 73.0% | 79.0% | 98.3%  |
|  Minimum (pg/mL) | 0 | 3 | 0 | 0 | 14  |
|  Maximum (pg/mL) | >4000 | 1651 | >4000 | >4000 | >4000  |

{12}

Heart Failure Population – Males

|   | NYHA Functional Class  |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|   | All | I | II | III | IV  |
|  Sample Size (N=) | 462 | 94 | 215 | 121 | 32  |
|  Median (pg/mL) | 268 | 122 | 258 | 293 | 1645  |
|  Mean (pg/mL) | 524 | 314 | 409 | 597 | 1646  |
|  SD (pg/mL) | 719 | 390 | 539 | 821 | 1032  |
|  5th Percentile | 12 | 9 | 14 | 22 | 265  |
|  95th Percentile | 1976 | 1281 | 1356 | 2288 | 3654  |
|  Percentage ≥ 100 pg/mL | 71.0% | 56.4% | 70.7% | 76.0% | 96.9%  |
|  Minimum (pg/mL) | 0 | 3 | 0 | 0 | 14  |
|  Maximum (pg/mL) | >4000 | 1408 | 3782 | >4000 | >4000  |

Heart Failure Population - Females

|   | NYHA Functional Class  |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|   | All | I | II | III | IV  |
|  Sample Size (N=) | 231 | 30 | 104 | 69 | 28  |
|  Median (pg/mL) | 385 | 174 | 298 | 466 | 1408  |
|  Mean (pg/mL) | 685 | 341 | 481 | 760 | 1623  |
|  SD (pg/mL) | 858 | 388 | 641 | 870 | 1186  |
|  5th Percentile | 16 | 14 | 21 | 12 | 244  |
|  95th Percentile | 2593 | 1022 | 2031 | 2718 | >4000  |
|  Percentage ≥ 100 pg/mL | 80.5% | 63.3% | 77.9% | 84.1% | 100.0%  |
|  Minimum (pg/mL) | 0 | 10 | 0 | 0 | 173  |
|  Maximum (pg/mL) | >4000 | 1651 | >4000 | >4000 | >4000  |

N. Instrument Name:
i-STAT 1 Analyzer

O. System Descriptions:
1. Modes of Operation:

{13}

Single use cartridge

2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐

3. Specimen Identification:
Bar code reader is incorporated into the system

4. Specimen Sampling and Handling:
Whole blood samples are applied directly into the sample well of the cartridge

5. Calibration:
Factory set

6. Quality Control:
The reliability of the results is maintained through a combination of user testing and instrument self-checks. The self checks occur with every cartridge run and verify performance of the analyzer and cartridge sub-systems. This includes checks on the individual sensor’s performance, the integrity of the calibrant fluid, the response of the pressure and thermal transducers, and the flow of calibrant and sample within the cartridge. Any values that are statistically deviant from the factory established expectation values would cause the test results to be suppressed. Daily monitoring is through the use of internal and external electronic simulators. Liquid controls are provided for the verification of cartridge lot performance for all newly received cartridge lots.

P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

14

---

**Source:** [https://fda-staging.innolitics.com/device/K053597](https://fda-staging.innolitics.com/device/K053597)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
