The SmartPill GI Monitoring System is indicated for use in evaluating patients with suspected delayed gastric emptying (gastroparesis). The SmartPill GI Monitoring System measures pH, pressure and temperature throughout the gastrointestinal tract. These physiological measurements are used to determine gastric emptying time (GET), total transit time (TTT), and combined small-large bowel transit time (SLBTT). In addition, pressure contraction patterns from the antrum and duodenum are used to calculate motility indices. Delayed gastric emptying is implicated in such disorders as idiopathic and diabetic gastroparesis and functional non-ulcer dyspepsia. Not for use in pediatric populations
Device Story
Ingestible capsule measures pH, pressure, and temperature throughout GI tract; transmits data via 434 MHz ASK modulated RF signal to patient-worn Data Receiver; data downloaded to PC for analysis. MotiliGI software performs automated analysis; generates printable report for physician review. Used in clinical settings; physician interprets output to assess gastric emptying time, total transit time, small-large bowel transit time, and motility indices. Assists in diagnosing gastroparesis and functional non-ulcer dyspepsia; provides objective physiological data to guide clinical management.
Clinical Evidence
Clinical testing conducted to validate indications for use; bench testing performed to verify accuracy and precision. Biocompatibility tested per ISO-10993 and USP <661>. Electrical safety verified per IEC 60601-1; EMC verified per IEC 60601-1-2. Results support intended use and clinical claims.
Technological Characteristics
Ingestible capsule with solid-state piezo-resistive pressure sensor, pH sensor, and temperature sensor. RF transmission at 434 MHz (ASK modulation). Patient-worn receiver; PC-based analysis software. Biocompatibility per ISO-10993 and USP <661>. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2.
Indications for Use
Indicated for patients with suspected delayed gastric emptying (gastroparesis), including idiopathic and diabetic gastroparesis and functional non-ulcer dyspepsia. Not for use in pediatric populations.
Regulatory Classification
Identification
A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.
Predicate Devices
Heidelberg pH Capsule and the Heidelberg Gold Medallion Gastro-intestinal pH Measurement System
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with a more abstract and flowing design.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Ms. Kathleen H. Selover Director, Regulatory Affairs and Quality The SmartPill Corporation 847 Main Street Buffalo, New York 14203-1109
JAN 1 0 2017
Re: K053547
Trade/Device Name: SmartPill GI Monitoring System Regulation Number: 21 CFR 876.1725 Regulation Name: Gastrointestinal motility monitoring system Regulatory Class: II Product Code: NYV Dated: May 18, 2006 Received: May 18, 2006
Dear Ms. Selover:
This letter corrects our substantially equivalent letter of July 18, 2006.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements
{1}------------------------------------------------
Page 2 - Ms. Kathleen H. Selover
as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the as set form in and radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809] ), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tou may of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Benjamin R. Fisher -S
Benjamin Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Response to FDA Request for Additional. Information FOR K053547 DATED March 24, 2006
5/17/2006
SECTION 16
Indications for Use Statement
510(k) Number, if known: K05354Z
Device Name: SmartPill GI Monitoring System
Indications for Use:
The SmartPill GI Monitoring System is indicated for use in evaluating patients with suspected delayed gastric emptying (gastroparesis).
The SmartPill GI Monitoring System measures pH, pressure and temperature The SmartPill Of Montoning System measurements are and masurements are throughout the gastromics time (GET), total transit time (GET), total transit time (TTT), and used to determine gastric emplying this (SLBTT). In addition, pressure Combined shaketarge bower cransit time (S) (S) (S) (S) (S) (S) (Calculate motility indices.
Delayed gastric emptying is implicated in such disorders as idiopathic and Detayed gastroparesis and functional non-ulcer dyspepsia.
Not for use in pediatric populations
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR
Over-the-Counter Use _
(Per 21CFR 801.109)
Doucuc hoadon
Division Sion-Off Division of Renmout ive. Ahdon and Radiological Devi
{3}------------------------------------------------
JUL 1 8 2006
Page 1 of 3
## 510(k) Summary
.
:
| DATE OF SUMMARY: | July 7, 2006 |
|------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MANUFACTURER: | The SmartPill Corporation<br>847 Main Street<br>Buffalo, NY 14203 |
| CONTACT INFORMATION: | Phone: 716.882.0701, ext. 106<br>Fax: 716.882.0706 Fax<br>Email: kselover@smartpillcorp.com |
| CONTACT PERSON: | Kathleen H. Selover<br>Director, Regulatory Affairs |
| DEVICE TRADE NAME: | SmartPill GI Monitoring System |
| DEVICE COMMON NAME: | Gastrointestinal pH and pressure monitoring<br>equipment |
| CLASSIFICATION NAME,<br>REGULATORY REFERENCE<br>AND PRODUCT CODE | Unclassified<br>Product Code: 78 NYV |
| PREDICATE DEVICE(S) | • Heidelberg pH Capsule and the Heidelberg Gold<br>Medallion Gastro-intestinal pH Measurement<br>System<br>• Unisensor UniTip Gastrointestinal Pressure<br>Catheter |
{4}------------------------------------------------
510(K) Summary K053547 Page 2 of 3 The SmartPill GI Monitoring System senses and Product records pH and pressure measurements from the DESCRIPTION entire length of the gastrointestinal tract for use by physicians to evaluate patients with delayed gastric emptying. Sensors on board an ingestible capsule measure pH and pressure as the capsule travels the length of the GI tract. Measurements are transmitted from the capsule within the GI tract via ASK modulated RF signal at 434 MHz to a patient-worn Data Receiver and subsequently downloaded to PC for analysis and review. MotiliGI™ Software performs data analyses automatically and provides the physician with a printable report containing gastric emptying time, motility index, and total transit time. The SmartPill GI Monitoring System is indicated for INTENDED USE INDICATIONS FOR USE use in evaluating patients with suspected delayed gastric emptying (gastroparesis). The SmartPill GI Monitoring System measures pH, pressure and temperature throughout the gastrointestinal tract. These physiological measurements are used to determine gastric emptying time (GET), total transit time (TTT), and combined small-large bowel transit time (SLBTT). In addition, pressure contraction patterns from the antrum and duodenum are used to calculate motility indices. Delaved gastric emptving is implicated in such disorders as idiopathic and diabetic gastroparesis and functional non-ulcer dyspepsia.
Not for use in pediatric populations
{5}------------------------------------------------
510(K) Summary
K053547
Page 3 of 3
:
| PHYSICAL AND<br>TECHNICAL COMPARISONS | The SmartPill Gl Monitoring System, as it applies to<br>pH measurements, is the same as the intended use of<br>the predicate, Heidelberg pH capsule and Monitoring<br>System. Both devices use an ingestible capsule<br>capable of measuring pH throughout the Gl tract to<br>provide a continuous data stream of measured pH<br>values.<br><br>The device is technically comparable to UniSensor<br>UniTop pressure catheter. Both use a solid state<br>piezo resistive pressure sensor for sensing pressure in<br>the gastrointestinal tract and transmit the sensed<br>signal to a receiving device. Display of contractile<br>patterns are similar. |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PERFORMANCE<br>SUMMARY<br>SAFETY TESTING | The SmartPill Gl Monitoring System was tested in<br>multiple bench tests to verify the accuracy and<br>precision of the device. In addition, clinical testing<br>was conducted to validate the device's indications<br>for use. Results of these tests support the product's<br>intended use, indications for use, performance and<br>clinical claims. |
| BIOCOMPATIBILITY | All patient contacting materials were tested for<br>biocompatibility in accordance with ISO-10993, Part I<br>for a surface device that contacts breached or<br>compromised surfaces for prolonged contact and<br>plastic materials in accordance with USP <661>. |
| ELECTRICAL SAFETY | Electrical safety was conducted and found to meet<br>the requirements of IEC 60601-1 |
| ELECTROMAGNETIC<br>COMPATIBILITY | EMC testing was tested and found to meet the<br>requirements of IEC 60601-1-2 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.