← Product Code [MBH](/productcode/MBH) · K053340

# TRABECULAR METAL TIBIAL AND FEMORAL CONE AUGMENTS (K053340)

_Zimmer Trabecular · MBH · Feb 27, 2006 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K053340

## Device Facts

- **Applicant:** Zimmer Trabecular
- **Product Code:** [MBH](/productcode/MBH.md)
- **Decision Date:** Feb 27, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3565
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Trabecular Metal Tibial Cone Augments and Trabecular Metal Femoral Cone Augments are intended for use where severe degeneration, trauma, or other pathology of the knee joint indicates total knee arthroplasty. When used with the NexGen Complete Knee Solution - Rotating Hinge Knee (RHK) System, the Trabecular Metal Tibial Cone Augments and Trabecular Metal Femoral Cone Augments are for cemented use only. When used with the NexGen Complete Knee Solution - Legacy Constrained Condylar Knee System, the Trabecular Metal Tibial Cone Augments and Trabecular Metal Femoral Cone Augments are for cementless or cemented use.

## Device Story

Trabecular Metal Tibial and Femoral Cone Augments are orthopedic implants designed to address bone loss during total knee arthroplasty. These augments are used in conjunction with Zimmer's NexGen Complete Knee Solution, specifically the Rotating Hinge Knee (RHK) System or the Legacy Constrained Condylar Knee (LCCK) System. When paired with the RHK system, the augments are restricted to cemented use; when paired with the LCCK system, they may be used with or without cement. The device provides structural support to the knee joint, aiding in reconstruction for patients with severe joint pathology. Surgeons implant these components during knee replacement procedures to restore joint stability and function.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on the identical nature of the subject devices to the predicate devices, for which performance characteristics were previously established.

## Technological Characteristics

Trabecular Metal (porous metal) construction. Designed for use as knee system augments. Compatible with NexGen RHK and LCCK systems. Cemented or cementless fixation depending on the specific knee system used.

## Regulatory Identification

A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.

## Special Controls

*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.

## Predicate Devices

- Trabecular Metal Tibial Cone Augments ([K031962](/device/K031962.md))
- Trabecular Metal Femoral Cone Augments ([K051756](/device/K051756.md))
- Trabecular Metal Knee System Augments ([K040487](/device/K040487.md))
- NexGen Knee System Uncemented Components ([K031462](/device/K031462.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K05334o

Zimmer Trabecular Metal Technology, Inc.

FEB 2 1 2006

Trabecular Metal Tibial and Femoral Cone Augments Cemented or Cementless Use 510(k) Premarket Notification

#### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

## Trabecular Metal Knee System Augments

- Submitter Name: Zimmer Trabecular Metal Technology, Inc. Submitter Address: 80 Commerce Drive Allendale, New Jersey 07401-1600 Contact Person(s): Marci Halevi Phone Number: (201) 818-1800 ext 507 Fax Number: (973) 829-0825 Date Prepared: November 28, 2005 Device Trade Name: Trabecular Metal Tibial Cone Augments and Trabecular Metal Femoral Cone Augments Device Common Name: Knee System Augments Classification Name: knee, patello/femorotibial, semi-constrained. Prosthesis, uncemented, porous, coated, polymer/metal/polymer
Substantial Equivalence: The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence found in the Federal Food, Drug and Cosmetic Act, as amended and as applied under 21 CFR 807, Subpart E under which a device can be marketed without premarket approval or reclassification. determination of substantial র্ব equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of substantial equivalence herein shall be construed as an admission against interest under the US Patent Laws or their application by the courts.

- Predicate Devices: Trabecular Metal Tibial Cone Augments (K031962) Trabecular Metal Femoral Cone Augments (K051756), Trabecular Metal Knee System Augments (K040487) NexGen Knee System Uncemented Components (K031462).

{1}------------------------------------------------

053340

Zimmer Trabecular Metal Technology, Inc

510(K) Summary of Safety and Effectiveness - Continued...

Device Description: The Trabecular Metal Tibial Cone Augments and Trabecular Metal Femoral Cone Auaments are manufactured to be compatible for use with Zimmer's NexGen Complete Knee Solution - Rotating Hinge Knee (RHK) System and Zimmer's NexGen Complete Knee Solution - Legacy Knee Constrained Condylar Knee (LCCK) System. When used with the RHK System. the Trabecular Metal Cone Augments are for cemented use only. When used with the LCCK System, the Trabecular Metal Augments are for either cementless or cemented use.

Intended Use: Trabecular Metal Tibial Cone Augments and Trabecular Metal Femoral Cone Augments are intended for use where severe degeneration, trauma, or other pathology of the knee joint indicates total knee arthroplasty. When used with the NexGen Complete Knee Solution - Rotating Hinge Knee (RHK) System, the Trabecular Metal Tibial Cone Augments and Trabecular Metal Femoral Cone Augments are for cemented use only. When used with the NexGen Complete Knee Solution - Legacy Constrained Condylar Knee System, the Trabecular Metal Tibial Cone Augments and Trabecular Metal Femoral Cone Augments are for cementless or cemented use.

- Performance Data: The predicate and subject devices are identical; performance characteristics therefore remain as documented in the predicate submission (K031962 and K051756).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" are arranged around the seal's perimeter.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 2 7 2006

Zimmer Trabecular Metal Technology, Inc. c/o Ms. Marci Halevi Manager of Regulatory Affairs 80 Commerce Drive Allendale, New Jersey 07401-1600

Re: K053340

Trade/Device Name: Trabecular Metal Tibial Cone Augments and Trabecular Metal Femoral Cone Augments

Regulation Number: 21 CFR 888.3565 and 21 CFR 888.3560

Regulation Name: Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis, Knee joint patellofemorotibial polymer/metal/polymer semiconstrained cemented prosthesis

Regulatory Class: II Product Code: MBH, JWH Dated: November 30, 2005 Received: December 1, 2005

Dear Ms. Halevi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

## Page 2 - Ms. Halevi

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CMR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

fo

Mark N. Melkerson, M.S. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### Indications for Use KOS 3340 510(k) Number (if known):

# Device Name: Trabecular Metal Tibial Cone Augments and Trabecular Metal Femoral Cone Augments

Indications for Use:

Trabecular Metal Tibial Cone Augments and Trabecular Metal Femoral Cone Augments are intended for use where severe degeneration, trauma, or other pathology of the knee joint indicates total knee arthroplasty. When used with the NexGen Complete Knee Solution - Rotating Hinge Knee (RHK) System, the Trabecular Metal Tibial Cone Auqments and Trabecular Metal Femoral Cone Augments are for cemented use only. When used with the NexGen Complete Knee Solution - Legacy Constrained Condylar Knee System, the Trabecular Metal Tibial Cone Augments and Trabecular Metal Femoral Cone Augments are for cementless or cemented use.

× Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) |  |
|--------------------------------------------------------|--|
|--------------------------------------------------------|--|

Helbert Lumio

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

| 11<br>ﺮ ﺍﺯ ﺍﯾﻦ ﺍﺯ<br>œ | C                                                                                                                                                                              |  |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
|                        | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |  |

(Posted November 13, 200510(k) Number_

---

**Source:** [https://fda-staging.innolitics.com/device/K053340](https://fda-staging.innolitics.com/device/K053340)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
