← Product Code [NDN](/productcode/NDN) · K053336

# STAXX FX SYSTEM (K053336)

_Spine Wave, Inc. · NDN · Jul 3, 2006 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K053336

## Device Facts

- **Applicant:** Spine Wave, Inc.
- **Product Code:** [NDN](/productcode/NDN.md)
- **Decision Date:** Jul 3, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3027
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The StaXx™ FX System is indicated for use in the reduction of spinal fractures. It is intended to be used in combination with Stryker Spineplex™ Radiopaque Bone Cement.

## Device Story

StaXx™ FX System is an internal fracture reduction device; consists of stackable wafers made of preformed PMMA and a PEEK-OPTIMA base wafer containing 6% Barium Sulfate. Wafers are inserted incrementally into the vertebral body to form a column, providing fracture reduction. Used in combination with Stryker Spineplex™ Radiopaque Bone Cement. Operated by surgeons in a clinical/surgical setting. Device provides mechanical support to reduce spinal fractures; output is the physical column formed by stacked wafers; clinical benefit is the restoration of vertebral height and stabilization of fractures.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

System consists of stackable wafers fabricated from preformed PMMA and a base wafer of PEEK-OPTIMA with 6% Barium Sulfate. Wafers are 8mm wide with lengths of 20mm, 25mm, and 30mm. Mechanical device; no energy source or software components.

## Regulatory Identification

Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”

## Predicate Devices

- KyphX® Xpander Inflatable Bone Tamps ([K041454](/device/K041454.md))
- SKy Bone Expander System ([K040612](/device/K040612.md))
- KyphX® HV-R™ Bone Cement ([K041584](/device/K041584.md))
- Stryker Spineplex™ Radiopaque Bone Cement ([K032945](/device/K032945.md))
- The Wafer System ([K033303](/device/K033303.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# SPINE WAVE

K05.3336
Page 1 of 2

JUL - 3 2006

## 510(k) Summary StaXxTM FX System

### I. Submitter Information

Spine Wave, Inc. Two Enterprise Drive Suite 302 Shelton, CT 06484 Telephone: 203-944-9494 203-944-9493 Telefax:

Contact: Ronald K. Smith Date Prepared: April 28, 2006

#### II. Device Information

| Trade name:          | StaXx™ FX System                                                                    |
|----------------------|-------------------------------------------------------------------------------------|
| Common name:         | Internal Fracture Reduction System                                                  |
| Classification:      | Class II per 21 CFR 888.4540; Class II per 888.3027                                 |
| Classification Name: | Orthopedic manual surgical instrument;<br>Polymethylmethacrylate (PMMA) bone cement |
| Product Code:        | LXH; NDN                                                                            |

## III. Device Information

The StaXx™ FX System is a vertebral fracture reduction device composed of stackable wafers fabricated from preformed PMMA. The System includes a base wafer fabricated from PEEK-OPTIMA with 6% Barium Sulfate. The wafers are designed to be inserted incrementally into the vertebral body to form a column that provides the desired fracture reduction. Twenty-four wafers are provided per package. The wafers are provided in one width (8mm) with three lengths (20mm, 25mm, 30mm)

### IV. Intended Use

The StaXx™ FX System is indicated for use in the reduction of spinal fractures. It is intended to be used in combination with Stryker Spineplex™ Radiopaque Bone Cement.

{1}------------------------------------------------

## US 3336

:

## V. Substantial equivalence¹

: . The StaXxTM FX System was demonstrated to be substantially equivalent to the following devices:

| Predicate Device                          | Manufacturer        | 510(k) No. |
|-------------------------------------------|---------------------|------------|
| KyphX® Xpander Inflatable Bone Tamps      | Kyphon, Inc.        | K041454    |
| SKy Bone Expander System                  | Disc-O-Tech, Ltd.   | K040612    |
| KyphX® HV-R™ Bone Cement                  | Kyphon, Inc.        | K041584    |
| Stryker Spineplex™ Radiopaque Bone Cement | Stryker Corporation | K032945    |
| The Wafer System                          | Spine Wave, Inc.    | K033303    |

In addition, mechanical testing demonstrated that the StaXx™ FX System meets the performance requirements for its intended use. Any differences between the StaXx™ FX and the predicate devices do no affect the safety or effectiveness of this device.

.

:

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows a stylized drawing of what appears to be a bird in flight. The bird is represented with simple, bold lines, giving it an abstract and minimalist appearance. The lines suggest the shape of the bird's body, wings, and tail, with a sense of movement and dynamism conveyed through the curved and flowing lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL - 3 2006

Spine Wave, Inc. c/o Mr. Ronald K. Smith Director. Quality Systems and Regulatory Affairs Two Enterprise Drive Shelton, CT 06484

Re: K053336

Trade/Device Name: StaXx™ FX System Regulation Number: 21 CFR 888.3027, 21 CFR 888.4540 Regulation Name: PMMA bone cement; orthopedic manual surgical instrument Regulatory Class: Class II Product Code: NDN, HXG Dated: April 28, 2006 Received: May 1, 2006

Dear Mr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federall Food, Dr w and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the 1.01 The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-1050.

{3}------------------------------------------------

Page 2 - Mr. Ronald K. Smith

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

C Barbare Buchup
for

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### A. Indications for Use

510(k) Number (if known): _K053336

Device Name: _ .______________________________________________________________________________________________________________________________________________________________

Indications for Use:

The StaXx™ FX System is indicated for use in the reduction of spinal fractures. It is intended to be used in combination with Stryker Spiner Fractions. " (11)

Prescription Use (Part 21 CFR 801 Subpart D)

. :

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

:

: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Barbara Buchum

. Restorative. Division of Gr and Neurological Devices

Page 1 of 1

**510(k) Number** K053336

---

**Source:** [https://fda-staging.innolitics.com/device/K053336](https://fda-staging.innolitics.com/device/K053336)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).