ARCHITECT B12 CALIBRATORS

K053330 · Abbott Laboratories · JIT · Dec 16, 2005 · Clinical Chemistry

Device Facts

Record IDK053330
Device NameARCHITECT B12 CALIBRATORS
ApplicantAbbott Laboratories
Product CodeJIT · Clinical Chemistry
Decision DateDec 16, 2005
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1150
Device ClassClass 2

Indications for Use

The Abbott ARCHITECT B12 Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of vitamin B12 in human serum and plasma.

Device Story

Abbott ARCHITECT B12 Calibrators; used in clinical laboratory settings by trained personnel. Modification involves increasing calibrator materials from two to six levels (A-F). Calibrator A: borate buffer; Calibrators B-F: cyanocobalamine in borate buffer at 100, 250, 500, 1000, and 2000 pg/mL. Used to establish calibration curves for quantitative Vitamin B12 assays on ARCHITECT immunoassay systems. Output enables accurate quantification of patient Vitamin B12 levels, supporting clinical diagnosis and monitoring of B12 deficiency or related metabolic conditions.

Clinical Evidence

Bench testing only. Verification and validation activities performed per design control procedures; results met predetermined acceptance criteria. No clinical data provided.

Technological Characteristics

Liquid, ready-to-use buffered aqueous solution. Calibrator concentrations span the dynamic range of the ARCHITECT i System vitamin B12 assay. No specific materials or software algorithms described beyond standard clinical chemistry calibrator formulation.

Indications for Use

Indicated for use as calibrators for the Abbott ARCHITECT B12 assay to establish the calibration curve for the quantitative determination of Vitamin B12 in human serum or plasma.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification ODE Review Memorandum To: THE FILE RE: DOCUMENT NUMBER k053330 Abbott ARCHITECT® B12 Calibrators This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary): 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) k984108 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials. 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was to modify the number of calibrator materials provided in the device. The modified device consists of six calibrator materials. Calibrator A contains borate buffer. Calibrators B through F contain cyanocobalamine in borate buffer at 100, 250, 500, 1000, and 2000 pg/mL. The predicate device consists of two calibrator materials containing cyanocobalamine in borate buffer at 100 and 500 pg/mL. 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, immunoassay technology, test protocol, performance, assay principle, cutoff level and reagents. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied c) A declaration of conformity with design controls. The declaration of conformity should include: i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review. 6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices). The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...