K053266 · Thought Technology , Ltd. · IPF · Dec 29, 2005 · Physical Medicine
Device Facts
Record ID
K053266
Device Name
MYOTRAC INFINITI
Applicant
Thought Technology , Ltd.
Product Code
IPF · Physical Medicine
Decision Date
Dec 29, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The MyoTrac Infiniti system is indicated for the ongoing treatment of the following conditions: Relaxation of Muscle Spasms, Prevention of disuse atrophy, increasing local blood circulation, immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, Maintaining or increasing range of motion and Stroke Rehab by Muscle re-education. It is also used for Biofeedback, Relaxation & Muscle Re-Education purposes.
Device Story
MyoTrac Infiniti is a powered muscle stimulator and EMG biofeedback device. It inputs bioelectric signals via surface electrodes; processes signals using RMS (Root Mean Square) detection; and outputs electrical stimulation to peripheral nerves/muscles. Used in clinical settings by physicians or patients for muscle re-education, relaxation, and biofeedback. The device displays EMG activity via line graphs, bar graphs, and digital readouts. Healthcare providers use this visual/auditory feedback to guide patient muscle control and stimulation parameters, facilitating rehabilitation, spasm reduction, and atrophy prevention.
Clinical Evidence
Bench testing only. Evidence consists of verification of product specifications, system validation, safety, and EMC testing. Biocompatibility of electrodes was confirmed via laboratory testing against established standards.
Technological Characteristics
Powered muscle stimulator and EMG biofeedback device. Asymmetrical balanced pulsed current waveform. Frequency: 2–100 Hz; Pulse width: 50–400μs; Max phase charge: 60μC. EMG bandwidth: 20–500 Hz. Uses bipolar EMG detection. Connectivity via surface electrodes (Axelgaard, Uni-Gel, Triode). Powered by electrical stimulation output (0-100mA at 500ohms).
Indications for Use
Indicated for patients requiring muscle re-education, relaxation, biofeedback, spasm relaxation, atrophy prevention, circulation improvement, post-surgical venous thrombosis prevention, or range of motion maintenance. Applicable for stroke rehabilitation.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
DEC 2 9 2005
K053266
# Premarket Notification 510(k) Summary
| Date Prepared: | 20-10-2005 |
|----------------|-------------------------------------------------------------------------------------|
| Applicant: | Thought Technology Ltd<br>2180 Belgrave Avenue<br>Montreal Quebec<br>Canada H4A 2L8 |
| Contact: | Suresh Sugirtharaja<br>Design Coordinator |
| Tel: | 489-8251 x127 |
| Fax: | 489-8255 |
{1}------------------------------------------------
| Device Name | | |
|----------------------|------------------------------------------------------------------------------|------------------------------------|
| Trade Name: | MyoTrac Infiniti System | |
| Common Name: | Powered muscle stimulator and biofeedback device | |
| Classification Name: | 89IPF, 84HCC | (882.5050 and 890.5850)<br>ClassII |
| Predicate Devices | | |
| Trade Name: | Danmeter AutoMove AM800 EMG Triggered Stimulator<br>(K972997) | |
| Classification Name: | 89IPF, 84HCC, ClassII | |
| Trade Name: | Mettler Electronics Corp Sys STIM 208 powered muscle<br>stimulator (K031017) | |
| Classification Name: | 89IPF, ClassII | |
#### Description of Device:
The MyoTrac Infiniti device is an electrical muscle stimulator for contraction of muscles as indicated below.
The MyoTrac Infiniti is also an electromyography device. It is intended for medical purposes, such as to monitor and display the bioelectric signals produced by muscles, to stimulate peripheral nerves and to monitor and display the electrical activity produced by nerves. The indications for use are muscle re-education, relaxation and biofeedback.
#### Intended Use
Indications for use:
- Biofeedback, Relaxation and Muscle Reeducation. ●
- Relaxation of Muscle Spasms. .
- Prevention or retardation of disuse atrophy. .
- Increasing local blood circulation. .
- Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis. .
- . Maintaining or increasing range of motion.
- Stroke Rehab by Muscle re-education. .
{2}------------------------------------------------
## Technical Characteristics Comparison to Predicate Device
MyoTrac Infiniti is compared to AutoMove AM800 (K972997) and Sys STM 208 (K031017). The technical specifications of the MyoTrac Infiniti substantially fall within the range of the AutoMove AM800. The differences are that the MyoTrac Infiniti is able to deliver more charge per pulse and more power density than the AutoMove AM800 but still with in the range of Sys STIM 280 and power density limits specified by 21 CFR 890 5850 (Powered Muscle Stimulator) and outputs EMG RMS vs. EMG peak from the AutoMove AM800. These differences should not affect the safety or effectiveness of the device.
| | PROPOSED DEVICE<br>MyoTrac Infiniti | Danmeter AutoMove AM800<br>EMG Triggered Stimulation<br>K972997 | SYS STIM 208<br>K031017 |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Intended Use | Muscle Stimulation and<br>EMG Biofeedback | Muscle Stimulation and<br>EMG Biofeedback | Powered Muscle<br>Stimulation |
| Primary Functions | Delivery of stimulation and<br>reading of Electromyography | Delivery of stimulation and<br>reading of Electromyography | Delivery of stimulation |
| Stimulator Output | 0 - 100mA at 500ohms | 0 -60mA at 2.5Kohms | 0-184mA at 500 ohms |
| Waveform | Asymmetrical Balanced Pulsed<br>Current | Biphasic, alternating and<br>monophasic Current | Asymmetrical Biphasic<br>Rectangle with zero net<br>DC |
| Maximum Phase<br>Charge | 60μC | 31μC | 56μC |
| Frequency | 2 – 100 Hz | 10 - 100 Hz | 1-80Hz |
| Peak pulse intensity | 100mA | 60mA | 184mA |
| Pulse Width | 50 - 400μs | 100 – 400 μs | 200-800μS |
| Ramps | 0 - 10sec on and off ramp | 0.5 sec to 10 sec | |
| Duty Cycle | On(sec): 2 -20<br>Off(sec): 2-50 | On(sec): 2-20<br>Off(sec): 2-50 | ON 0.375 to 3.75sec<br>OFF 0.375 to 3.75sec |
| Session Duration<br>(min) | 1 to 120 minutes | 10 to 800 minutes or continue | |
| Programmable<br>features | Frequency, Current intensity,<br>pulse width, ramp up and down,<br>session length, threshold<br>trigger by the patient and the<br>physician. | Frequency, Current, Pulse<br>width, ramp up and down,<br>session duration, threshold for<br>trigger and duty cycle. By<br>patient or physician. | |
| Surface Electrodes | K874469A AxelGaard<br>electrodes for EMG or Stim.<br>K935213 Electrodes Uni-Gel<br>for EMG only<br>K903497A Triode Electrodes<br>for EMG only<br>K903497A Single Electrodes<br>for EMG only | PALS FLEX electrodes -<br>AxelGaard - K8764469A | |
| Current Density (full<br>output) | Axelgaard Model 895220<br>0.24mA/cm²<br>Axelgaard Model 895340<br>0.08mA/cm² | 0.5 mA/ cm² | 0.132mA/ cm² |
| Power Density (full<br>output @ 500ohms) | Axelgaard Model 895220<br>16mW/cm² | 13mW/cm² | 12mW/ cm² |
{3}------------------------------------------------
| | Axelgaard Model 895340<br>5.3mW/cm² | | |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------|------------------------------|------------------|
| EMG Ranges in μV | 0-5, 0-10, 5-10, 0-20, 5-20, 10-<br>20, 0-50, 10-50, 0-100, 50-100,<br>0-200, 50-200, 100-200, 0-500,<br>100-500, 0-1000, 0-2000 | 2 - 2000μV | N/A |
| EMG Bandwidth | 20 – 500 Hz | 50 – 1000 Hz | N/A |
| EMG Signal<br>Processing | Root Mean Square (RMS) | Peak | N/A |
| EMG Detection | Bipolar | Bipolar | N/A |
| Work Period (sec) | 2 – 20 seconds | 2 - 20 seconds | 0.375 to 3.75sec |
| Rest Period (sec) | 2 - 50 seconds | 2 - 50 seconds | 0.375 to 3.75sec |
| Session Duration<br>(min) | 1-120 minutes | 1 – 30 minutes | 0-60minutes |
| Feedback Modes | Line Graph, Bar Graphs, Digital<br>Display | Bar Graphs, Digital Displays | Digital |
。
{4}------------------------------------------------
#### Performance Data
Non-clinical tests were performed consisted of verification of the product specification, system validation, safety and EMC testing. Device equivalency is determined by a direct comparison of the device functional and hardware specifications of MyoTrac Infiniti system with the legally marketed predicate devices, Automove AM800 EMG triggered stimulator and Sys STIM 208 powered muscle stimulator. Such a comparison table is present in the above section
#### Biocompatibility:
The Axelgaard EMG/Stimulation electrodes (K874469A), Thought Technology single, triode and Uni-gel Electrodes for EMG (K903497A & K935213) have been laboratory tested for the safety of the materials and were found to be safe under the standard required for each test.
#### Conclusion:
The MyoTrac Infiniti system is safe and effective for its intended use. The MyoTrac Infiniti system is substantially equivalent to the predicate devices.
End of 510(k) Summary
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, in the center. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" are arranged in a circular pattern around the caduceus.
Public Health Service
DEC 2 9 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Thought Technology Limited c/o Mr. Robert Mosenkis CITECH 5200 Butler Pike Plymouth Meeting, PA 19462
Re: K053266/S1
Trade/Device Name: MyoTrac Infiniti System Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: II Product Code: IPF, HCC Dated: December 14, 2005 Received: December 16, 2005
Dear Mr. Mosenkis:
We have reviewed your Section 510(k) premarket notification of intent to narket the device indication indical we nave revewed your Section 910(t) promantee is substantially equivalent (for the indications for referenced above and nave decemined are as a see as a see a marketed in interstate winners the use stated in the enclosure) to regally manufacture Medical Device Amendments, or to devices that prior to May 26, 1970, the chacinon auto of the Federal Food, Drug, and Cosmetic
have been reclassified in accordance with the provisions of the Federal Food, Drug, May may have been reclassified in accordance with the premarket approval application (PMA). You may, Act (Act) that do not require apploval of a provisions of the Act. The general therefore, market the device, subject to the genirements for annual registration, listing of devices, good controls provisions of the necesses institutions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), in h If your device is classified (see ao re) into Existing major regulations affecting your device can be may be subject to such additional volts of Environs. Title 21, Parts 800 to 898. In addition, FDA may publish found in the Sous ents concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination as fithe Assurance Flease be advised that 1 177 - 15 usualled of aevice complies with other requirements of the Act or any FDA has made a decommation administered by other Federal agencies. You must comply with all
Federal statutes and regulations administered by other Federal (1) CED Box 807) Federal statutes and regulations administered by on the segistration and listing (21 CFR Part 807); the rive s requirements) in the manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2- Mr. Robert Mosenkis
This letter will allow you to begin marketing your device as described in your section 510(k) I his letter will anow you to begin marketing your arrial equivalence of your deversits vour develop premarket notification: The PDA midnig of substanted vice and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your do not on one one as not the regulation entitled, contact the Office of Comphalled at (210) 210 - 120 - 120 .97). You may obtain other "Misbranding by relerence to preharter nothedron" (2) Cfrom the Division of Small Manufacturers, general information on your responsionnies and its tool-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Вавда Виелино
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications for Use:
The MyoTrac Infiniti system is indicated for the ongoing treatment of the The Myo Frac Immit System is Indiation of Muscle, Spasms, Prevention of Musical rollowing conditions: Notaxation of more of culation, immediate post-surgical disuse atropity, increasing lood blood on us thrombosis, Maintaining or increasing stimulation of call muselou to problem to re-education. It is also used for range of motion anxation & Muscle Re-Ec ucation purposes.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/ OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS: LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark H. Mulkerson
: Vision Sign-Off) Orvision of General, Restorative and Neurological Devices
Number K053266
Page 12 of 27
.... !
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.