PARA-PRO FC50

K053257 · Alpha-Tec Systems, Inc. · LKS · May 12, 2006 · Microbiology

Device Facts

Record IDK053257
Device NamePARA-PRO FC50
ApplicantAlpha-Tec Systems, Inc.
Product CodeLKS · Microbiology
Decision DateMay 12, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.2900
Device ClassClass 1

Indications for Use

Para-Pro™ fc50 is a unique, patent pending device for the separation of fecal debris from the appropriate preserved specimen (10% buffered formalin; SAF sodium acetate acetic acid formaldehyde; and Proto-fix™ CLR) for the concentration of eggs, larvae, protozoa and juvenile nematodes associated with intestinal infections.

Device Story

Para-Pro™ fc50 is a filtration device for processing human fecal specimens. It consists of a cone-shaped filter with 1.6 mm² openings designed to attach to a 30ml specimen vial and a 50ml centrifuge collection tube. The device creates a closed system; gravity filtration separates large fecal debris from the preserved specimen, allowing parasitic forms to pass into the collection tube. The resulting sediment is then processed via standard formalin-ethyl acetate or CONSED methods and examined microscopically by laboratory personnel. The device improves recovery of parasitic stages compared to traditional methods.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Specimen collection and transport device for fecal matter. Designed for use with chemical preservatives including 10% buffered formalin, SAF, and Proto-fix™ CLR. Mechanical separation principle for debris removal and parasite concentration. Non-electronic, manual operation.

Indications for Use

Indicated for the concentration of eggs, larvae, protozoa, and juvenile nematodes from fecal specimens in patients suspected of having intestinal infections.

Regulatory Classification

Identification

A microbiological specimen collection and transport device is a specimen collecting chamber intended for medical purposes to preserve the viability or integrity of microorganisms in specimens during storage of specimens after their collection and during their transport from the collecting area to the laboratory. The device may be labeled or otherwise represented as sterile. The device aids in the diagnosis of disease caused by pathogenic microorganisms.

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: K053257 B. Purpose of Submission: For the detection of ova and parasites in human stool C. Measurand: Parasitic eggs, larvae, protozoa and juvenile nematodes D. Type of Test: Parasite Concentration device E. Applicant: Alpha Tec Systems, Inc. F. Proprietary and Established Names: Para-Pro™ fc50 G. Regulatory Information: 1. Regulation section: 21 CFR Part 866.2900 Microbiological Specimen Collection and Transport Device 2. Classification: I 3. Product Code: LKS – Device, Parasite Concentration 4. Panel: 83 Microbiology H. Intended Use: 1. Intended use(s): Para-Pro™ fc50 is a unique, patent pending device for the separation of fecal debris from feces preserved in appropriate fixative reagents: 10% buffered formalin; SAF sodium acetate acetic acid formaldehyde; and Proto-fix™ CLR from the appropriate fixed/preserved specimen for the concentration of eggs, larvae, protozoa and juvenile nematodes associated with intestinal infections. Indication(s) for use: Para-Pro™ fc50 is a unique, patent pending device for the separation of fecal debris from the appropriate preserved specimen (10% buffered formalin; SAF sodium acetate acetic acid formaldehyde; and Proto-fix™ CLR) for the concentration of eggs, larvae, protozoa and juvenile nematodes associated with intestinal infections. 2. Special condition for use statement(s): For Prescription Use Only 3. Special instrument Requirements: Not applicable I. Device Description: {1} Page 2 of 4 Para-Pro™ fc50 is a large surface-area filtration device utilizing a 1.6 mm² opening in a cone shape to maximize the filtration area. When a fixed/preserved specimen is filtered, by gravity, through the filter device, the larger fecal debris is separated from the preserved fecal specimen. This device traps the fecal debris allowing large parasite forms to pass through the device into a collection centrifuge tube. The device is designed to accommodate a standard 30ml vial containing a suitable fixative for intestinal parasitic concentration procedures (10% buffered formalin; SAF sodium acetate acetic acid formaldehyde; and Proto-fix™ CLR), creating a completely closed system, and on the opposite end the device is attached to a collection tube consisting of a 50ml centrifuge tube and inverted to initiate the gravity filtration process. J. Substantial Equivalence Information: 1. Predicate device name(s): Para-Pac® Macro-CON Predicate K number(s): K860034 2. Comparison with predicate(s): | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended use | Device for filtering, concentrating and recovering helminth eggs, larvae, protozoa and juvenile nematodes from fixed/preserved feces. | same | | Specimen type | Fecal specimens | same | | Technology | Filtration, reagent separation and concentration via centrifugation | same | | Level of skill | Moderately complex | same | | | Differences | | | Item | Device | Predicate | | Procedure | 8ml Proto-fix CLR | 10% formalin | | Procedure | 4 ml ethyl acetate | 5 ml ethyl acetate | K. Standard/Guidance Document referenced (if applicable): Not applicable L. Test Principle: Para-Pro™ fc50 is a large surface-area filtration device utilizing a 1.6 mm² opening in a cone shape to maximize the filtration area. When a fixed/preserved specimen is filtered, by gravity, through the filter device, the larger fecal debris is separated from the preserved fecal specimen. This device traps the fecal debris allowing large parasite forms to pass through the device into a collection centrifuge tube. The device is designed to accommodate a standard 30ml vial containing a suitable fixative for intestinal parasitic concentration procedures (10% buffered formalin; SAF sodium {2} Page 3 of 4 acetate acetic acid formaldehyde; and Proto-fix™ CLR), creating a completely closed system. On the opposite end the device is attached to a collection tube consisting of a 50ml centrifuge tube and inverted to initiate the gravity filtration process. The air transfer between the 30 ml collection vial and the receiver centrifuge tube is accomplished through the opening at the top of the cone. The filtered specimen, collected in the centrifuge tube, can be concentrated by the formalin ethyl acetate method or the CONSED method, resulting in a pellet or sediment that contains the parasitic material, if present in the sample, and viewed microscopically on an appropriate cover-slipped glass slide using standard wet mount procedures. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: a. Precision/Reproducibility: Not applicable for this type device. Recovery studies were performed. The samples were tested with a conventional analytical method, formalin/ethyl-acetate concentration procedure, using human fecal samples. Each specimen recovered duplicate types of parasitic stages, showing only a slight difference in the recovered parasitic load within each specimen. The parasitic load, per 50 uL aliquot of the concentrated pellet, resulted in 1,849 ova and parasites for the test device when samples were fixed and tested using 10% formalin and 1,211 ova and parasites for the Predicate device when tested using duplicate samples fixed in 10% formalin. This represents 34.3% more parasitic forms recovered in the test device as compared to the Predicate device. b. Linearity/assay reportable range: Not applicable c. Traceability (controls, calibrators, or method): Not applicable d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off Not applicable ### 2. Comparison studies: a. Method comparison with gold standard: The data for SAF fixative samples showed an increase of 287 parasitic stages for the test device, when compared to the Predicate device. There was a (28% increase in parasitic stages). The data for Proto-fix CLR showed an increase of 1,043 parasitic stages for the test device when compared to the predicate device. (39% increase in parasitic stages) ### 3. Clinical studies: a. Clinical sensitivity: {3} Page 4 of 4 b. Clinical specificity: Refer to (a.) above c. Other clinical supportive data (when a and b is not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: (Interpretive Criteria) Not applicable N. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...