INTELLIVUE GAS MODULES, MODELS G5 M1019A AND G1 M1013A
K053204 · Draeger Medical AG & Co. Kgaa · CCK · Feb 3, 2006 · Anesthesiology
Device Facts
Record ID
K053204
Device Name
INTELLIVUE GAS MODULES, MODELS G5 M1019A AND G1 M1013A
Applicant
Draeger Medical AG & Co. Kgaa
Product Code
CCK · Anesthesiology
Decision Date
Feb 3, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.1400
Device Class
Class 2
Indications for Use
The IntelliVue gas modules are indicated for measuring and monitoring CO2 concentration and the concentrations of N2O, O2, Halothane, Enflurane, Sevoflurane and Desflurane. Federal Law restricts this device to sale by or on the order of a physician.
Device Story
IntelliVue G5 M1019A/G1 M1013A are anesthetic gas modules for ventilated patients; used in clinical settings during anesthesia induction, maintenance, and emergence. Inputs: airway gas samples. Principle: nondispersive infrared measurement for respiratory/anesthetic gases; paramagnetic measurement for O2 (Fast O2). Outputs: display waves for O2, CO2, N2O, anesthetic agents; numerics for inspired/end-tidal values and airway respiration rate. Features: automatic identification of up to two anesthetic agents; optional sample gas return to breathing circuit. Operation: connects to Philips IntelliVue monitors via RS232 digital interface. Automatic zero calibration maintains accuracy. Clinicians use output to monitor patient gas concentrations and adjust anesthesia delivery, potentially improving patient safety during ventilation.
Clinical Evidence
Bench testing only. Verification and validation performed in accordance with FDA regulations and company internal standards to confirm device meets specifications and is safe and effective for intended use.
Technological Characteristics
Nondispersive infrared (NDIR) sensor for respiratory/anesthetic gases; paramagnetic sensor for O2. Digital RS232 interface for connectivity with Philips IntelliVue monitors. Automatic zero calibration. Optional sample gas return. Form factor: gas module.
Indications for Use
Indicated for measuring and monitoring CO2, N2O, O2, and anesthetic agents (Halothane, Enflurane, Isoflurane, Sevoflurane, Desflurane) in ventilated patients during anesthesia induction, maintenance, and emergence. Prescription use only.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{0}------------------------------------------------
**Dräger**medical
A Dräger and Siemens Company
K053204
510(k) Summary (Section 10)
ਗੁਰੂਰ re Bour
## Summary of Safety and Effectiveness
## Applicants Name and Address
Draeger Medical AG & Co. KGaA Moislinger Allee 53-55 D-23542 Luebeck Germany
## Applicants Contact Person
Dr. Karin Luebbers Senior Manager Regulatory Affairs
| Tel. No.: | 011 49 (451) 882-5367 |
|-----------|-------------------------|
| Fax No.: | 011 49 (451) 882-7-5367 |
#### Applicants US Contact Person
Ms Monica Ferrante Director of Regulatory Affairs
| Tel. No .: | (215) 721-5400 |
|------------|----------------|
| Fax No.: | (215) 721-5412 |
### Date the Summary was prepared
October 31, 2005
#### Device Name
a
| Trade Name: | IntelliVue G5 M1019A / IntelliVue G1 M1013A |
|--------------|---------------------------------------------|
| Common Name: | Anesthetic Multi Gas Monitor |
#### Classification
| Regulation No. | Device | Product Code |
|----------------|----------------------------------------------|--------------|
| 868.1400 | Analyzer, Gas, Carbon Dioxide, Gaseous Phase | (73CCK) |
| 868.1700 | Analyzer, Gas, Nitrous Oxide, Gaseous Phase | (73CBR) |
| 868.1500 | Analyzer, Gas, Enflurane, Gaseous Phase | (73CBQ) |
| 868.1620 | Analyzer, Gas, Halothane, Gaseous Phase | (73CBS) |
| 868.1500 | Analyzer, Gas, Desflurane, Gaseous Phase | (73NHO) |
| 868.1500 | Analyzer, Gas, Sevoflurane, Gaseous Phase | (73NHP) |
| 868.1500 | Analyzer, Gas, Isoflurane, Gaseous Phase | (73NHQ) |
| 868.1720 | Analyzer, Gas, Oxygen, Gaseous-Phase | (73CCL) |
{1}------------------------------------------------
## Legally marketed device to which Substantial Equivalence is claimed
VAMOS (K012139, K040847) Manufactured by Draeger Medical AG & Co. KGaA; Germany Distributed in the United States by Draeger Medical Inc.
SCIO (with Infinity Patient Monitors) (K031340, K051628) Manufactured by Draeger Medical AG & Co. KGaA; Germany Distributed in the United States by Draeger Medical Inc.
Philips M1026B Anesthetic Gas Monitor (with IntelliVue Patient Monitors) (K040917) Distributed in the United States by Philips Medical Systems
EGM - Essential Gas Monitor (K041956) Manufactured by Draeger Medical AG & Co. KGaA; Germany Distributed in the United States by Draeger Medical Inc.
## Description of the Device
The IntelliVue G5 M1019A / G1 M1013A gas modules provide a nondispersive infrared measurement of respiratory and anesthetic gases and a paramagnetic measurement of oxygen (Fast O2).
They are designed to work with Philips IntelliVue through a digital interface (RS232). They are intended for measuring the airway gases of ventilated patients during the induction of, maintenance of, and emergence from anesthesia.
The modules produce display waves for O2, CO2, N2O, and anesthetic agents, together with numerics for inspired and end-tidal values for O2, CO2, N2O, anesthetic agents, and airway respiration rate. Automatic identification of up to two anesthetic aqents and measurement of mixtures of up to two anesthetic agents are available as options. Return of the sample gas into the breathing circuit is available optional.
An automatic zero calibration is performed by the IntelliVue gas module as required to maintain measurement accuracy.
## Intended Use
The IntelliVue gas modules are indicated for measuring and monitoring CO2 concentration and the concentrations of N2O, O2, Halothane, Enflurane, Sevoflurane and Desflurane.
Federal Law restricts this device to sale by or on the order of a physician.
{2}------------------------------------------------
## Substantial Equivalence
The intended use of IntelliVue gas modules is comparable by the referenced predicate devices
- Draeger Medical VAMOS Anesthetic Gas Monitor
- Draeger INFINITY Monitors with SCIO
- Philips M1026B Anesthetic Gas Monitor
- EGM Essential Gas Monitor
The technical characteristics of the IntelliVue gas modules do not raise new questions regarding safety or effectiveness. Furthermore the labeling of the IntelliVue gas module provides similar information as the predicate devices except for the subject of this submission.
Information provided in the 510(k) Premarket Notification supports the determination of substantial equivalence. Design, development, verification and validation of the device was performed in accordance with FDA regulations and guidance and company internal standards. The testing and analysis of results provide assurance that the devices meet their specifications and are safe and effective for their intended use.
In summary Draeger Medical AG & Co. KGaA has demonstrated that the IntelliVue gas modules are safe and effective. They are considered to be substantial equivalent to currently marketed predicate devices which have been previously cleared by the FDA.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name encircling a stylized eagle emblem. The text reads "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" in a circular arrangement around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Monica Ferrante Director, of Regulatory Affairs Draeger Medical AG & Company KG 3155 Quarry Road
Telford, Pennsylvania 18969
Re: K053204
Trade/Device Name: IntelliVue G5 M1019A / IntelliVue G1 M1013A Regulation Number: 868.1440 Regulation Name: Carbon Dioxide gas analyzer Regulatory Class: II Product Code: CCK Dated: January 24, 2006 Received: January 27, 2006
FEB 3
2006
Dear Mr. Ferrante:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 - Ms. Ferrante
Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-firee number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K O53 このキ
Device Name: IntelliVue G5 M1019A / IntelliVue G1 M1013A
Indications for Use: The IntelliVue gas modules are indicated for measuring and monitoring CO2 concentration and the concentrations of N2O, O2, Halothane, Enflurane, Isoflurane, Sevoflurane and Desflurane.
Federal law restricts this device to sale by or on the order of a physician.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Auj Sabom
Division of Microbiology, General Hospital,
Infection Control, Dental Devices
K.05.3.2011
510(is) Number
Page 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.