DRLOCK DISTAL RADIUS VOLAR SYSTEM

K053182 · Orthohelix Surgical Designs, Inc. · HRS · Feb 27, 2006 · Orthopedic

Device Facts

Record IDK053182
Device NameDRLOCK DISTAL RADIUS VOLAR SYSTEM
ApplicantOrthohelix Surgical Designs, Inc.
Product CodeHRS · Orthopedic
Decision DateFeb 27, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The system is used to stabilize and aid in the fusion of fractures and osteotomies of the distal radius.

Device Story

The DRLock Distal Radius Volar System is a set of metallic bone fixation implants used to stabilize fractures and osteotomies of the distal radius. The system comprises four fixation plates, twenty-five screws, and eleven pegs. These components are surgically implanted by a physician to provide internal fixation, aiding in bone fusion. The device is intended for clinical use in a surgical setting.

Clinical Evidence

Bench testing only. Finite element analysis and mechanical testing were performed to confirm substantial equivalence and performance requirements.

Technological Characteristics

Materials: Type 316L Stainless Steel conforming to ASTM F 138. Components: Fixation plates, screws, and pegs. Mechanical fixation principle.

Indications for Use

Indicated for the fixation of unstable distal radius fractures and osteotomies.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ FEB 2 7 2006 510(k) Premarket Notification DRLock™ Distal Radius Volar System ## 510(k) SUMMARY # Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared. OrthoHelix Surgical Designs, Inc. 1815 West Market Street Suite 205 Akron, Ohio 44313 Phone: (866) 904-3549 Fax: (352) 371-3932 Contact Person: Edward A. Kroll Representative Consultant for OrthoHelix Surgical Designs, Inc. January 23, 2006 Date Prepared: ### Name of Device DRLock ™ Distal Radius Volar System ## Common or Usual Name Fixation Plates and Screws #### Classification Name Single/Multiple Component Metallic Fixation Appliances and Accessories #### Predicate Devices Hand Innovations Distal Volar Radius Fracture Repair System (K0022775) Synthes (USA) Stainless Steel Modular Hand System (K030310) #### Intended Use The system is used to stabilize and aid in the fusion of fractures and osteotomies of the distal radius. {1}------------------------------------------------ 510(k) Premarket Notification DRLock™ Distal Radius Volar System # Device Description The DRLock ™ Distal Radius Volar System (DRLock) is a series of metallic (stainless steel), implantable, bone fixation plates, pegs and screws. Its' intended use is to stabilize and aid in the fusion of fractures and osteotomies involving the distal radius. The System includes four (4) fixation plates, twenty-five (25) screws and eleven (11) pegs. All screws and plates are made from type 316L Stainless Steel in conformance with ASTM F 138 Standard Specification for Wrought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants. ## Performance Data Finite Element Analysis in conjunction with mechanical testing confirms that the DRLock System is substantially equivalent to its' predicate devices, and that it meets specified requirements for its' intended use. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines forming the staff and a snake winding around it. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" are arranged in a circular pattern around the caduceus. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB 2 7 2006 Mr. Lee A. Strnad Senior Development Manager OrthoHelix Surgical Designs, Inc. 1815 West Market Street, Suite 205 Akron, Ohio 44313 K053182 Re: K055162 Trade/Device Name: DRLock Distal Radius Volar System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Codes: HRS Dated: January 25, 2006 Received: January 26, 2006 Dear Mr. Strnad: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becaon 9 rQxy press. In substantially equivalent (for the indications felerenced above and nave determined ly marketed predicate devices marketed in interstate for use stated in the encrosure) to regions and ment date of the Medical Device Amendments, or to commerce provision way 20, 1976, the eccordance with the provisions of the Federal Food, Drug, devices that have been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I bu may, incretore, market the device, books requirements for annual registration, listing of general controls provisions of the ling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (600 world). Existing major regulations affecting your device can may be subject to suen additions, Title 21, Parts 800 to 898. In addition, FDA may oc found in the Couvencements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please oe advised that I DTT 3 issualled of a babbailer. that FDA has made a determination that your device complies with other requirements of the Act that I DA has made a deceminations administered by other Federal agencies. You must or any I catal suttates and regulations, including, but not limited to: registration and listing (21 {3}------------------------------------------------ Page 2 - Mr. Lee A. Strnad CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set a CFR Part 807), labeling (21 CFR Part 800); good in and 820); and if applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Sections over device as described in your Section 510(k) I his letter will anow you to ocgin manketing your arrivalence of your device to a legally premarket nothleation: "The PDF mailing of castination for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please IT you desire specific arrice ior J ow (240) 276-0120. Also, please note the regulation entitled, Colliact the Office of Compilance as (E++ Part 807.97). You may obtain Misolallang by relected to promantonionshibilities under the Act from the Division of Small other general information on your response Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Vicker Leen 40 for Mark N. Melkerson, M.S. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K053182 Device Name: DRLock™ Distal Radius Volar System Indications for Use: The DRLock Distal Radius Volar System is indicated for the fixation of unstable distal radius fractures and osteotomies. Prescription Use 区 (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page | of | (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number_________________________________________________________________________________________________________________________________________________________________ Image /page/4/Picture/14 description: The image shows a scatter plot with a concentration of points in the lower right corner. The density of points decreases as you move towards the upper left corner. The points appear to be randomly distributed, with no clear patterns or clusters visible other than the overall concentration in the lower right.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...