K053159 · Brainlab AG · HAW · Jun 6, 2006 · Neurology
Device Facts
Record ID
K053159
Device Name
VECTORVISION SPINE
Applicant
Brainlab AG
Product Code
HAW · Neurology
Decision Date
Jun 6, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
BrainLAB's VectorVision spine is intended for use as an intraoperative image-quided localization system for minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative image data that is processed by a VectorVision workstation. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to a CT, x-ray or MR-based model of the anatomy, Example procedures include but are not limited to: - Navigated spinal procedures in support of standard approaches (e.g. anterior, lateral, oblique) - Spinal implant procedures such as - Pedicle screw placement - Anterior plating - Kyphoplasty and vertebroplasty procedures - Placement of other temporary or permanent devices such as k-wires, needles, catheters or electrodes - Thoracic spine surgerv - Tumor surgery on the spinal column and adjacent soft tissue - Placement of acetabular and SI screws on the pelvis
Device Story
VectorVision spine is an intraoperative image-guided localization system; assists surgeons in minimally invasive spinal and pelvic procedures. Inputs: preoperative 3D CT/MR datasets or intraoperative 2D fluoroscopic images. Operation: links tracked freehand surgical instruments to virtual image space via passive marker sensor system. Registration methods: paired point matching, surface matching, region matching, or CT-fluoro matching. Output: real-time navigation visualization on workstation display. Usage: OR environment; operated by surgeons. Benefit: provides spatial orientation for precise instrument placement during spinal implants, tumor surgery, or device insertion; improves accuracy of minimally invasive approaches.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness verified through product design and development validation procedures.
Technological Characteristics
Image-guided surgery system; passive marker sensor tracking; workstation-based processing. Supports 3D (CT/MR) and 2D (fluoroscopy) data integration. Registration algorithms include paired point, surface, region, and CT-fluoro matching. Connectivity via workstation interface. Sterilization requirements for surgical instruments apply.
Indications for Use
Indicated for patients requiring stereotactic surgery where rigid anatomical structures (skull, pelvis, long bone, vertebra) can be referenced to CT, X-ray, or MR-based models. Applicable for spinal navigation, implant placement (pedicle screws, anterior plating), kyphoplasty, vertebroplasty, and placement of temporary/permanent devices (k-wires, needles, catheters, electrodes) in spinal or pelvic procedures.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
# KOS 3159 510 (k) Summary of Safety and Effectiveness for VectorVision spine
| Manufacturer: | BrainLAB AG |
|-----------------------------|---------------------------------------------------------------------------------------|
| Address: | Ammerthalstrasse 8<br>85551 Heimstetten<br>Germany |
| Phone: | +49 89 99 15 68 0 |
| Fax: | +49 89 99 15 68 33 |
| Contact Person: | Mr. Rainer Birkenbach |
| Summary Date: | September 13, 2005 |
| Device Name: | |
| Trade name: | VectorVision spine |
| Common/Classification Name: | VectorVision spine, BrainLAB Image Guided Surgery System /<br>Instrument, Stereotaxic |
#### Predicate Device:
BrainLAB VectorVision CT / Fluoro (K010968) BrainLAB Kolibri spine (K042721)
Device Classification Name: Instrument, Stereotaxic Requlatory Class: Class II
## Intended Use:
BrainLAB's VectorVision spine is intended for use as an intraoperative image-quided localization system for minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative image data that is processed by a VectorVision workstation. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to a CT, x-ray or MR-based model of the anatomy, Example procedures include but are not limited to:
{1}------------------------------------------------
- · Navigated spinal procedures in support of standard approaches (e.g. anterior, lateral, oblique)
- · Spinal implant procedures such as
- · Pedicle screw placement
- · Anterior plating
- · Kyphoplasty and vertebroplasty procedures
- · Placement of other temporary or permanent devices such as k-wires, needles, catheters or electrodes
- · Thoracic spine surgerv
- · Tumor surgery on the spinal column and adjacent soft tissue
- Placement of acetabular and SI screws on the pelvis
# Device Description:
VectorVision spine is a device that allows surgical planning and navigation. It links a surgical instrument. (tracked by passive marker sensor system) to a location on a virtual computer image, which is either based ou patient's preoperative 3D information of a CT or MR dataset or based on patient's intraoperative acquired 2D fluoro image(s) of a c-arm.
The device enables the navigation based on 3D data and / or based on acquired fluoro images,
Based on 2D fluoro images, the registration is done automatically by using the exact spatial position informat. of the intra-operatively acquired fluoro images.
Based on 3D data, the procedure of linking the surgical instrument to the virtual computer image is achieved by performing registration methods as paired point matching, surface matching, region matching or CT fluoro matching. The last registration method uses 2D fluoro images to register the previously acquired 3D dataset. Thus, CT fluoro matching combines 2D fluoro imaging with 3D datasets,
After registration, the device assists the surgeon in performing certain surgical procedures as described in the indications for use.
## Substantial equivalence:
VectorVision spine has been verified and validated according to BrainLAB's procedures for product design and development. The validation proves the safety and effectiveness of the system. The information provided by BrainLAB in this 510 (k) application was found to be substantially equivalent with the predicate device BrainLAB VectorVision CT / Fluoro (K010968) and BrainLAB Kolibri spine (K042721).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo features a stylized caduceus, a symbol often associated with healthcare, with three abstract human figures intertwined within it. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 6 2006
BrainLAB AG % Mr. Martin Ringholz Project Engineer IGS Spine Ammerthalstrasse 8 Heimstetten, Germany 85551
Re: K053159
Trade/Device Name: Vector Vision spine Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulation Class: II Product Code: HAW Dated: May 23, 2006 Received: May 25, 2006
Dear Mr. Ringholz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Martin Ringholz
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Mark N. Melkerson Director
Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K053159
Device Name: Vector Vision spine
Indications For Use:
BrainLAB's VectorVision spine is intended for use as an intraoperative image-quided localization system for minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative image data that is processed by a VectorVision workstation. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to a CT, x-ray or MR-based model of the anatomy.
Example procedures include but are not limited to:
- · Navigated spinal procedures in support of standard approaches (e.g. anterior, lateral. oblique)
- · Spinal implant procedures such as
- · Pedicle screw placement
- · Anterior plating
- · Kyphoplasty and vertebroplasty procedures
- · Placement of other temporary or permanent devices such as k-wires, needles, catheters or electrodes
- · Thoracic spine surgery
- · Tumor surgery on the spinal column and adjacent soft tissue
- · Placement of acetabular and SI screws on the pelvis
Prescription Use × (Per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|---------------------------------------------------------------|---------|
| (Division Sign-Off) | |
| Division of General, Restorative,<br>and Neurological Devices | |
| 510(k) Number | K053159 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.