K053139 · Fotona D.D. · GEX · Feb 16, 2006 · General, Plastic Surgery
Device Facts
Record ID
K053139
Device Name
FOTONA QX ND:YAG/KTP LASER SYSTEM
Applicant
Fotona D.D.
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 16, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Nd:YAG (1064 nm) laser is capable of delivering up to 1.0 J of laser energy. Its spot sizes available range from 2 to 8 mm. This nominal pulse is intended for incision/excision, ablation, vaporization of soft tissue in general dermatology, removal of dark ink (black, blue and brown) tattoos, removal or lightening of unwanted hair, treatment of Common Nevi. The frequency doubled Nd: YAG (532 nm) KTP laser delivers up to 0.5J of laser energy in 4-9 nanosecond pulses. Spot sizes available range from 2 to 7 mm. This laser wavelength is intended to be used for incision/excision, ablation, vaporization of soft tissue, removal of light ink (red, tan, purple, and orange) tattoos, removal of pigmented lesions, removal of vascular lesions, treatment of Lentigines, treatment of Cafe-Au-Lait, treatment of Common Nevi, treatment of Seborrheic Kearotoses, and treatment of Post Inflammatory Hyperpigmentation.
Device Story
Fotona QX is a Q-switched Nd:YAG/KTP laser system for dermatological procedures. System utilizes Nd:YAG crystal (1064 nm) and frequency-doubling KTP crystal (532 nm) activated by flashlamps. Physician selects wavelength via control panel; therapeutic beam combined with red diode aiming beam via beamsplitter; articulated arm delivers energy to focusing handpiece for non-contact application. System includes high-voltage power supply, water-to-air cooling, and microprocessor-based controller for parameter regulation. Used in clinical settings by physicians to perform tissue ablation, tattoo removal, and lesion treatment. Output energy affects tissue via photothermal/photomechanical interaction; clinical benefit includes precise tissue removal and lesion clearance.
Indicated for patients requiring soft tissue incision, excision, ablation, or vaporization in general dermatology; removal of tattoos (dark ink at 1064nm; light ink at 532nm); removal/lightening of unwanted hair; treatment of pigmented/vascular lesions, Lentigines, Cafe-Au-Lait, Common Nevi, Seborrheic Keratoses, Post Inflammatory Hyperpigmentation, and Nevus of Ota.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Spectra -VRM Q-switched Nd: YAG Laser System (K000317)
{0}------------------------------------------------
## Appendix E : Summary of Safety and Effectiveness Data
| I. General Information | FEB 1 6 2006 | K053139 |
|------------------------|--------------------------------------------------------|---------|
| Company : | Fotona d.d. | |
| | Stegne 7, 1210 Ljubljana | |
| | SLOVENIA | |
| Contact Person : | Stojan Trošt | |
| Preparation Date : | 11-04-05 | |
| Device Trade Names : | Fotona QX Nd:YAG/KTP Laser System | |
| Common Name : | Combination of Q-switched Nd:YAG and KTP Nd:YAG Lasers | |
| Classification Name : | Instrument, Surgical, Powered, Laser | |
| | 79-GEX | |
| | 21 CFR 878-48 | |
#### II. Description
The Fotona QX laser system is based on the Nd:YAG (1064 nm) and frequency doubled KTP Nd:YAG (532 nm) laser technology. There is one optical cavity containing the Nd:YAG crystal. The frequency doubled KTP Nd:YAG wavelength is achieved by directing the Nd: YAG laser beam through a frequency doubling non-linear crystal. The Nd:YAG laser is activated by means of the use of flashlamps. After the cavity, a red diode aiming beam is reflected onto a coaxial beam path using a beamsplitter assembly. The combined therapeutic and aiming beams are guided by articulated arm to a focusing handpiece. Both lasers are used in non-contact mode.
Both laser wavelengths share a common power supply, control system, and cooling system. The internal computer can be directed to select either the Nd: YAG head or the KTP Nd: YAG laser wavelength. When the laser is first turned on the physician is able to select the desired wavelength via control panel.
The Fotona QX laser system consists of:
a) A high voltage power supply, which converts and rectifies the a.c. mains current to provide regulated power for the flashlamp simmer current and main triggering pulse.
b) A cooling system consisting of an internal water flow circuit together with water-to-air heat exchanger.
c) The microprocessor based controller, which regulates the functions of both lasers and allows parameter selection by the user.
d) The Nd: YAG laser head.
{1}------------------------------------------------
e) The frequency doubling crystal.
g) The optical delivery system, interfacing the energy from the laser to the patient via an articulated arm and focusing handpiece.
#### III. Intended Use
The Nd:YAG (1064 nm) laser is capable of delivering up to 1.0 J of laser energy. Its The Nd. ITAS (100+ mir) tabel 16 capaces on or sizes available range from 2 to 8 mm. This nominal pulsewithin is 1 7 minoooooonaed for incision/excision, ablation, vaporization of soft tissue in general dermatology, removal of dark ink (black, blue and brown) tattoos, Son lissue in general definatorial or lightening of unwanted hair, treatment of Common Nevi.
The frequency doubled Nd: YAG (532 nm) KTP laser delivers up to 0.5J of laser energy in 4-9 nanosecond pulses. Spot sizes available range from 2 to 7 mm. This laser wavelength is intended to be used for incision/excision, ablation, vaporization of soft wavelength is intended to be assureal of light ink (red, tan, purple, and orange) tatoos, removal of pigmented lesions, removal of vascular lesions, treatment of Lentigines, treatment of Cafe-Au-Lait, treatment of Common Nevi, treatment of Seborrheic Kearotoses, and treatment of Post Inflammatory Hyperpigmentation.
### IV. Summary of Substantial Equivalence
Fotona believes that its Fotona QX laser system is substantially equivalent to the
a) Spectra -VRM Q-switched Nd: YAG Laser System (K000317) previously cleared for:
- Incision, excision, ablation, vaporization of soft tissue for general dermatology
- 532 nm wavelength:
- removal of light ink (red, tan, purple, and orange) tatoos,
- removal of pigmented lesions, removal of vascular lesions,
- treatment of Lentigines, treatment of Cafe-Au-Lait,
- treatment of Common Nevi, ।
- treatment of Seborrheic Kearotoses, ।
- treatment of Post Inflammatory Hyperpigmentation. ।
- 1064 nm wavelength: =
- removal of dark ink (black, blue and brown) tattoos, -
- treatment of Nevus of Ota, -
- removal or lightening of unwanted hair, -
- treatment of Common Nevi. -
b) Medlite C6 Q-Switched Nd:YAG Laser (K 014234) previously cleared for incision, excision, ablation, vaporization of soft tissue for general dermatology, Tatoo removal, treatment of Vascular Lesions and treatment of Pigmented Lesions.
c) Palomar Q-YAG Nd:YAG Laser System (K023967) previously cleared for skin resurfacing with or without adjuvant preparation at the 1064 nm wavelength.
510(k) Submission: Fotona QX Nd: YAG/KTP Laser System
{2}------------------------------------------------
They therefore have the same Intended Use as the Fotona QX laser system.
The Fotona QX laser system shares the same design features (wavelength, active medium, cooling system, power supply, beam deliveries, controls, housing) as the medicate devices. The output characteristics are the same as those of the predicate devices.
The risk and benefits for the Fotona QX laser system are comparable to the predicate devices when used for similar clinical applications.
It is therefore believed that there are no new questions of Safety or Effectiveness raised by the introduction of the Fotona QX Nd: YAG/KTP laser system.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three wavy lines extending from its head, representing the department's commitment to health, human services, and well-being. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" are arranged in a circular pattern around the eagle.
FEB 1 6 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Stojan Trost Manager Quality Assurance & Regulatory Affairs Fotona D.D. Stegne 7 1210 Ljubljana SLOVENIA
Re: K053139
Trade/Device Name: Fotona QX Nd: Y AG/KTP Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: 21 CFR 078.1070
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: January 5, 2006 Received: January 9, 2006
Dear Mr. Trost:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becaon 910(t) premained is substantially equivalent (for the indications ferenced above and nave determined ly marketed predicate devices marketed in interstate for use stated in the encrosule) it regars and the Medical Device Amendments, or to commerce prior to May 20, 1976, the excordance with the provisions of the Federal Food. Drug, devices that have been recuire approval of a premarket approval application (PMA). and Cosmetic Act (Act) that Go not require appen o the general controls provisions of the Act. The r ou may, therefore, market the act include requirements for annual registration, listing of general controls provisions of the line, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device it may of Subject to Saon additional Regulations, Title 21, Parts 800 to 898. In addition, FDA can be found in the Oous of eneming your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that I DA s issuance or a cassince complies with other requirements of the Act that I DA has made a decertimation administered by other Federal agencies. You must of any I cocraf statutes and regaraments, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Kahan
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic
s (2001) 100 (2006) (2006) 1985 (2006) 1985-1994 (21 CFR 1000-1050 forth in the quality systems (QS) regulation (Sections 531-542 of the Act); 21 CFR 1000-1050.
Product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-
This letter will allow you to begin marketing your device as described in your Section 510(k)
The success of the support of any and consumeration of your device to a legal This letter will allow you to begin marketing your avvalence of your device to a legally
premarket notification. The FDA finding of substantial end the permits vour premarket notification. The FDA Inding of basisance of the results your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
rs the latest of the support of 115 - All and assesses the regulation entit If you desire specific advice for your device on our an Also, please note the regulation entitled.
contact the Office of Compliance at (240) 276-0115. Also, please note the r contact the Office of Compliance at (240) 270 - 121 - 121 Part 807.97). You may obtain
"Misbranding by reference to premarket notification" (21 - 1 - 1 - 1 - 1 - 1 - 1 - 1 "Misbranding by reference to prematici notification (c) * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * other general information on your responsionities and its toll-free number (800) 638-2041 or Manufacturers, International and Consumer Assistance at to vote a
(301) 443-6597 or at its Internet address http://www.fda.cov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson
Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Appendix F : Indications for Use Statement
510(k) Number (if known): K053139
Device Name: Fotona QX Nd:YAG/KTP Laser System
Indications For Use:
- 1. Incision, excision, ablation, vaporization of soft tissue for general dermatology.
- 2. 532 nm wavelength:
- removal of light ink (red, tan, purple, and orange) tatoos, "
- removal of pigmented lesions, removal of vascular lesions,
- treatment of Lentigines, treatment of Cafe-Au-Lait,
- treatment of Common Nevi,
- treatment of Seborrheic Kearotoses,
- treatment of Post Inflammatory Hyperpigmentation.
- 3. 1064 nm wavelength:
- removal of dark ink (black, blue and brown) tattoos, -
- treatment of Nevus of Ota, -
- removal or lightening of unwanted hair, -
- treatment of Common Nevi. -
- skin resurfacing with or without adjuvant preparation --
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
ØR
(Division Sig
Division of General, Restorative,
and Neurolog Devices
510(k) Number
Over-The-Counter Use
510(k) Submission: Fotona QX Nd:YAG/KTP Laser System
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.