ACRO BIOTECH LLC RAPID BENZOYLECGONINE URINE TEST; CATALOG# MET001000

K053035 · Acro Biotech, LLC · DIO · May 5, 2006 · Clinical Toxicology

Device Facts

Record IDK053035
Device NameACRO BIOTECH LLC RAPID BENZOYLECGONINE URINE TEST; CATALOG# MET001000
ApplicantAcro Biotech, LLC
Product CodeDIO · Clinical Toxicology
Decision DateMay 5, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3250
Device ClassClass 2

Indications for Use

Acro Rapid Benzoylecgonine (Cocaine) Urine test is a lateral flow, one-step immunoassay for the qualitative detection of benzoylecgonine in human urine at a cutoff of 300 ng/mL. The test is used to obtain a visual qualitative result and is intended for laboratory use only. This assay provides only preliminary result. Clinical consideration and professional judgment must be applied to a drug test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectroscopy (GC/MS) analysis is preferred.

Device Story

Lateral flow, one-step immunoassay for qualitative detection of Benzoylecgonine in human urine; 300 ng/mL cutoff. Device used in laboratory settings by professional personnel. Input is human urine sample; output is visual qualitative result (presence/absence of analyte). Preliminary result requires confirmation via alternate chemical method, preferably GC/MS. Clinical decision-making relies on professional judgment and confirmatory testing.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Lateral flow, one-step immunoassay. Qualitative visual readout. Cutoff concentration: 300 ng/mL. Designed for laboratory use.

Indications for Use

Indicated for the qualitative detection of Benzoylecgonine (cocaine metabolite) in human urine at a 300 ng/mL cutoff. Intended for laboratory use only to obtain preliminary results requiring confirmation by GC/MS.

Regulatory Classification

Identification

A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.

Special Controls

*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k053035 B. Purpose for Submission: New device C. Measurand: Benzoylecgonine D. Type of Test: Qualitative lateral flow immunochromatographic test E. Applicant: Acro Biotech, LLC F. Proprietary and Established Names: Acro Rapid Benzoylecgonine (Cocaine) Urine Test G. Regulatory Information: 1. Regulation section: 21 CFR 862.3250 Cocaine and Cocaine metabolite test system 2. Classification: Class II 3. Product code: DIO 4. Panel: 91, Toxicology H. Intended Use: 1. Intended use(s): See indications for use below 2. Indication(s) for use: Acro Rapid Benzoylecgonine (Cocaine) Urine test is a lateral flow, one-step immunoassay for the qualitative detection of benzoylecgonine in human urine at a cutoff of 300 ng/mL. The test is used to obtain a visual qualitative result and is intended for laboratory use only. {1} This assay provides only preliminary result. Clinical consideration and professional judgment must be applied to a drug test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectroscopy (GC/MS) analysis is preferred. 3. Special conditions for use statement(s): For prescription use 4. Special instrument requirements: Not applicable. The device is a visually read single-use device. I. Device Description: The Acro Rapid Benzoylecgonine (Cocaine) Urine Test is a one-step immunoassay in which a chemically labeled drug (benzoylecgonine-BSA conjugate) competes with Cocaine and its metabolite, benzoylecgonine, in urine for limited antibody binding sites. The test device contains a membrane strip, which is pre-coated with benzoylecgonine-BSA conjugate at the test band region of the membrane strip. A wicking pad containing anti-benzoylecgonine monoclonal antibody-colloidal gold conjugate is placed at one end of the membrane. The device contains a control region which has a different antigen/antibody from the test region. The device is for single-use and visually read. J. Substantial Equivalence Information: 1. Predicate device name(s): Syntron QuickPac II OneStep Benzoylecgonine test 2. Predicate 510(k) number(s): k992990 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Test Principle | Qualitative, immunochromatographic assay | Same | | Intended Use | Detection of benzoylecgonine in human urine | Same | {2} | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Cutoff concentration | 300 ng/mL | 150 ng/mL | | Read time | Read result after 5-10 minutes | Read result after 5 minutes | K. Standard/Guidance Document Referenced (if applicable): Draft Guidance for Industry and FDA Staff: Premarket Submission and Labeling Recommendations for Drugs of Abuse Screening Tests (December 2, 2003) L. Test Principle: The Acro Rapid Benzoylecgonine (Cocaine) test is based on the principle of competitive immunochemical reaction between an immobilized drug-protein conjugate and the drug or drug metabolites, which may be present in the urine sample for limited binding sites of a labeled drug antibody. When sample is applied to the test device, the sample migrates by capillary action through the device. When benzoylecgonine is absent in the urine or is present at a concentration below the cutoff level (300 ng/mL), the anti-benzoylecgonine antibodies in the colloidal gold conjugate will bind to the benzoylecgonine-BSA conjugate in the test region of the device to form a line, indicating a negative result. No line will form if the sample contains benzoylecgonine at the cutoff level or higher, because it will compete with the benzoylecgonine-BSA for antibody binding sites on the antibody colloidal gold conjugate, indicating a preliminary positive result. Each device contains a procedural control which generates a visible line to indicate that the correct amount of sample was added. If the device is functioning correctly, a line in the control region should always appear regardless of the presence or absence of drug or drug metabolite in the urine specimen. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: The reproducibility of the device was evaluated at four different sites. The samples were prepared from fresh drug-free normal human urine specimens collected from non-user volunteers and pooled. Test samples were prepared by spiking the pooled urine with benzoylecgonine to concentrations of 150 ng/mL (50% below cutoff), 225 ng/mL (25% below cutoff), 375 ng/mL (25% above cutoff), 450 ng/mL (50% above cutoff) and 600 ng/mL (100% above cutoff). The testing was performed in one day by one operator at each site. Each sample was tested n=15. {3} The study data is presented below: | Test sites | 0 ng/mL | | 150 ng/mL | | 225 ng/mL | | 375 ng/mL | | 450 ng/mL | | 600 ng/mL | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Neg. | Pos. | Neg. | Pos. | Neg. | Pos. | Neg. | Pos. | Neg. | Pos. | Neg. | Pos. | | 1 | 15 | 0 | 15 | 0 | 11 | 4 | 3 | 12 | 0 | 15 | 0 | 15 | | 2 | 15 | 0 | 15 | 0 | 11 | 4 | 4 | 11 | 0 | 15 | 0 | 15 | | 3 | 15 | 0 | 15 | 0 | 10 | 5 | 4 | 11 | 0 | 15 | 0 | 15 | | 4 | 15 | 0 | 15 | 0 | 12 | 3 | 4 | 11 | 0 | 15 | 0 | 15 | | Total | 60 | 0 | 60 | 0 | 44 | 16 | 15 | 45 | 0 | 60 | 0 | 60 | b. Linearity/assay reportable range: Not applicable. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Controls: A procedural control is included in the test strip of the device. A line appears in the control region (C) with every test as long as a sufficient specimen volume was added to the device. If the line does not form, the test is considered invalid. External control materials are not supplied with this test however, the labeling includes a recommendation that external controls be tested to ensure proper kit performance. Users are instructed to follow local, state and federal guidelines for testing QC material. **Stability:** The manufacturer claims an expiration date of 24 months when the devices are stored at 2-30 °C. Accelerated studies are being used by the sponsor to estimate the shelf-life of the device, however, on-going real time studies are being performed. Protocols and acceptance criteria were provided by the sponsor. d. Detection limit: See 1(f) Assay Cut-off e. Analytical specificity: Cross-reactivity was established by spiking various concentrations of similarly structured drug compounds into drug-free urine. By analyzing various concentrations of each compound, the sponsor determined the concentration {4} of the drug that produced a response approximately equivalent to the cutoff concentration of the assay. Results of the studies appear in the table below: | Compound | Response equivalent to cutoff | | --- | --- | | Benzoylecogonine | 300 ng/mL | | Cocaethylene | 600 ng/mL | | Cocaine | 800 ng/mL | The following list of structurally unrelated substances showed no interference at concentrations of 100 µg/mL in normal urine spiked with 150 and 450 ng/mL benzoylecgonine. Ascorbic acid, glucose, hemoglobin, ibuprofen, and sodium chloride were tested at higher concentrations (listed below) and each showed no interference as well. | Acetaminophen | Ibuprofen (200 µg/mL) | | --- | --- | | Acetone | (+/-)-Isoproterenol | | Albumin | Ketamine | | Ampicillin | Levorphanol | | Ascorbic Acid (500 µg/mL) | Lidocaine | | Aspartame | (+)-Naproxen | | Aspirin | Niacinamide | | Atropine | Nicotine | | Benzocaine | (+/-)-Norephedrine | | Bilirubin | Oxalic Acid | | Caffeine | Penicillin-G | | Chloroquine | Pheniramine | | (+)-Chlorpheniramine | Phenothiazine | | (+/-)-Chlorpheniramine | 1-Phenylephrine | | Creatine | β -Phenylethylamine | | Dexbrompheniramine | Procaine | | Dextromethorphan | Quinidine | | Diphenhydramine | Ranitidine | | Dopamine | Riboflavin | | (+/-)-Epinephrine | Sodium Chloride (10,000 µg/mL) | | Erythromycin | Sulindac | | Ethanol (0.2%) | Theophylline | | Furosemide | Tyramine | | Glucose (500 µg/mL) | 4-Dimethylaminoantipine | | Guaiacol Glyceryl Ether | (1R, 2S)-(-)-N-Methyl-Ephedrine | | Hemoglobin (500 µg/mL) | | Aliquots of test samples with concentrations of benzoylecgonine 50% below and 50% above the cutoff were adjusted to pH 4 to 9 in 1 pH unit increments and evaluated. Altering the pH of the sample did not affect the expected {5} results of the test. The effect of specific gravity on test results was evaluated. Aliquots of test samples with concentrations of benzoylecgonine 50% below and 50% above the cutoff were adjusted to specific gravities ranging from 1.003-1.04. Altering the specific gravity of the sample did not affect the expected results of the test. f. Assay cut-off: The cutoff of the device is stated to be 300 ng/mL. The cutoff was determined by repetitive assaying of six levels of benzoylecgonine controls. The study was performed as part of the reproducibility study in section 1.a. of this template. The results of the cut-off study are summarized in the following table: | Benzoylecgonine Conc. | # Tested | # Positive (+) | # Negative (-) | | --- | --- | --- | --- | | 0 ng/mL | 60 | 0 | 60 | | 150 ng/mL | 60 | 0 | 60 | | 225 ng/mL | 60 | 16 | 44 | | 300 ng/mL | 60 | 40 | 20 | | 375ng/mL | 60 | 45 | 15 | | 450 ng/mL | 60 | 60 | 0 | | 600 ng/mL | 60 | 60 | 0 | 2. Comparison studies: a. Method comparison with predicate device: Urine samples were collected from 60 self-declared non-drug users and were tested using the Acro Rapid Benzoylecgonine Urine Test and the predicate device. Note: The cutoff concentration of the candidate device is 300 ng/mL and the cutoff concentration of the predicate device is 150 ng/mL. Additionally, seven (7) drug-negative samples were tested on the candidate device and a reference method, Gas Chromatography/Mass Spectrometry GC/MS. The results are summarized in the table below. Sixty (n=60) clinical drug positive urine specimens were pre-screened by GC/MS and then assayed on the Acro Rapid Benzoylecgonine Urine Test. Approximately 10% of these study samples were evenly distributed between minus and plus 50% of the claimed cutoff concentration, 150 ng/mL and 450 ng/mL respectively, {6} The results are summarized in the table below: Candidate Device Results vs. stratified GC/MS Values (Cut-off 300 ng/mL) | Candidate Device Results | Negative by the predicate device* | Near Cutoff Negative (150-300 ng/mL) | Near Cutoff Positive (300-450 ng/mL) | High Positive (>450 ng/mL) | | --- | --- | --- | --- | --- | | Positive | 0 | 3 | 7 | 40 | | Negative | 60 | 7 | 3 | 0 | * Drug-negative samples were tested by predicate with 150 ng/mL cut-off. These samples were not confirmed by GC/MS ** Additionally, 7 drug-negative urines were tested on candidate device and GC/MS. All were negative with both methods. % Agreement to GC/MS among positives is 47/50 =94% % Agreement to GC/MS among negatives is 14/17 =82% b. Matrix comparison: Not applicable; this device is only for use with urine samples 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Not applicable N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...