K052999 · ETHICON, Inc. · MEB · Mar 22, 2006 · General Hospital
Device Facts
Record ID
K052999
Device Name
SEMPERFLO INFUSION SYSTEM
Applicant
ETHICON, Inc.
Product Code
MEB · General Hospital
Decision Date
Mar 22, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SEMPERFLO Infusion System is intended to provide continuous percutaneous infusion of prescribed solutions at a pre-set rate. The device is intended for intraoperative placement in hospital or alternate medical site environments. Once implanted, the device and pump may be used in homecare settings. The device is not intended for intravascular, intrathecal, or epidural use, nor is it indicated for use with blood, blood products, lipids or fat emulsions. When packaged in the kit configuration for pain management (to be known as the SEMPERFLO* Pain Management System), the intended use will be as follows: The SEMPERFLO Pain Management System is intended to provide continuous percutaneous infusion of prescribed solutions at a pre-set rate for post-operative pain management. The SEMPERFLO Pain Management System is also intended to significantly decrease narcotic use and pain when used to deliver local anesthetics to surgical wound sites or in close proximity to nerves when compared with narcotic only pain management.
Device Story
Single-use disposable elastomeric infusion pump; includes microcatheter for fluid delivery. Elastomeric reservoir maintains constant pressure; forces solution through 0.2μ filter into microcatheter. Provides continuous percutaneous infusion of prescribed solutions at pre-set rate. Used in hospital or alternate medical sites for intraoperative placement; suitable for homecare settings post-implantation. Reduces narcotic use and pain by delivering local anesthetics to surgical wound sites or near nerves. Operated by clinicians; provides passive, constant-rate medication delivery.
Clinical Evidence
Bench and animal functional performance testing conducted. Biological reactivity assessed per ISO 10993-1. No clinical trial data presented.
Technological Characteristics
Single-use disposable elastomeric infusion pump; 0.2μ filter; microcatheter. Materials assessed per ISO 10993-1. Passive mechanical energy source (elastomeric reservoir pressure). Sterile convenience kit configuration.
Indications for Use
Indicated for patients requiring continuous percutaneous infusion of prescribed solutions for post-operative pain management via local anesthetic delivery to surgical wound sites or near nerves. Contraindicated for intravascular, intrathecal, or epidural use; not for use with blood, blood products, lipids, or fat emulsions.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
On-Q Pain Pump (I-Flow Corp)
VICRYL* Suture
BAXTER Pain Management System
Submission Summary (Full Text)
{0}------------------------------------------------
K 052999
JMP-2005-014
## MAR 2 2 2006
# Summary of Safety and Effectiveness
Information supporting claims of substantial equivalence, as defined under the Federal Food, Drug and Cosmetic Act, respecting safety and effectiveness is summarized below.
#### Submitted by:
Jennifer M. Paine Manager, Regulatory Affairs Ethicon, Inc., A Johnson & Johnson Company Route 22 West, PO Box 151 Somerville, NJ 08876
#### Name/Classification of Device:
Class II in 21 CFR § 880.5725 (MEB) and 21 CFR § 868.5120 (BSO)
#### Trade Name:
SEMPERFLO Infusion System
#### Predicate Devices:
On-Q Pain Pump (I-Flow Corp) VICRYL* Suture BAXTER Pain Management System
#### Statement of Intended Use:
The SEMPERFLO Infusion System is intended to provide continuous percutaneous infusion of prescribed solutions at a pre-set rate.
The device is intended for intraoperative placement in hospital or alternate medical site environments. Once implanted, the device and pump may be used in homecare settings.
The device is not intended for intravascular, intrathecal, or epidural use, nor is it indicated for use with blood, blood products, lipids or fat emulsions.
When packaged in the kit configuration for pain management (to be known as the SEMPERFLO* Pain Management System), the intended use will be as follows:
> The SEMPERFLO Pain Management System is intended to provide continuous percutaneous infusion of prescribed solutions at a pre-set rate for post-operative pain management.
The SEMPERFLO Pain Management System is also intended to significantly decrease narcotic use and pain when used to deliver local anesthetics to surgical wound sites or in close proximity to nerves when compared with narcotic only pain management.
{1}------------------------------------------------
#### Device Description:
The proposed device consists of a single-use disposable elastomeric infusion pump, designed to deliver medication/fluids at a constant flow rate, and a microcatheter for fluid delivery. The device is intended for use in tissue, and is not intended for use in intravascular, epidural or intrathecal modes of delivery. Delivery of medication is achieved via an elastomeric reservoir that maintains relatively constant pressure, forcing solution through a 0.2μ filter and into the associated microcatheter assembly. The device will be provided as a convenience kit.
### Summary of Technological Characteristics of New Device to Predicate Devices:
The modified device has similar technological characteristics as the predicate devices. Like currently marketed devices, it is a kit providing a sterile infusion system. The proposed device uses the same or similar materials as those used in other currently marketed implantable or non-implantable medical devices. The mechanism of action is the same as for predicate devices.
#### Performance Data:
Biological reactivity of the materials has been assessed using methods specified in ISO Standard 10993-1, and the material was found to be acceptable for its intended use. Results of functional performance testing (bench and animal testing) have been submitted. These results indicate that the proposed device meets or exceeds all functional requirements.
#### Conclusions:
Based on the similarities to the predicate devices identified in this submission, we conclude that the modified device is substantially equivalent to the predicate devices under the Federal Food, Drug, and Cosmetic Act.
*Trademark of Ethicon, Inc.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with healthcare. The words "U.S. Department of Health and Human Services, USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 2 2006
Ms. Jennifer M. Paine Ethicon, Incorporated Route 22 West P.O. Box 151 Somerville, New Jersey 08876
Re: K052999 Trade/Device Name: SEMPERFLO Infusion System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: II Product Code: MEB Dated: February 27, 2006 Received: February 28, 2006
Dear Ms. Paine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{3}------------------------------------------------
### Page 2 - Ms. Jennifer M. Paine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements mean that i DA may Federal statutes and regulations administered by other Federal agencies. 01 the Act of ally rederal baker and segments, including, but not limited to: registration ? 1 ou indisting (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and and listing (21 CFR Part 007), labeling systems (QS) regulation (21 CFR Part 820); and if requirements as set form in the quarty agon control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to objak finding of substantial equivalence of your device to a premarket notification. - The PDF results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), if you desire specific ad rios for Journe at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may overn built be generational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sujette y. Michau Oms.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
# 510(k) Number (if known): K052999
Device Name: SEMPERFLO Infusion System
Indications for Use:
The SEMPERFLO Infusion System is intended to provide continuous percutaneous infusion of prescribed solutions at a pre-set rate.
The device is intended for intraoperative placement in hospital or alternate medical site environments. Once implanted, the device and pump may be used in homecare settings.
The device is not intended for intravascular, intrathecal, or epidural use, nor is it indicated for use with blood, blood products, lipids or fat emulsions.
When packaged in the kit configuration for pain management (to be known as the SEMPERFLO* Pain Management System), the intended use will be as follows:
The SEMPERFLO Pain Management System is intended to provide continuous percutaneous infusion of prescribed solutions at a pre-set rate for post-operative pain management.
The SEMPERFLO Pain Management System is also intended to significantly decrease narcotic use and pain when used to deliver local anesthetics to surgical wound sites or in close proximity to nerves when compared with narcotic only pain management.
*Trademark of Ethicon, Inc.
(21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Cinta Ong
on Man Mogy, General Huspic II 1150 Md Dental Davices
INDICATIONS FOR USE
- 54529999 -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.