← Product Code [LLZ](/productcode/LLZ) · K052995

# ADVANTAGE WORKSTATION 4.3. (K052995)

_GE Healthcare · LLZ · Nov 8, 2005 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K052995

## Device Facts

- **Applicant:** GE Healthcare
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Nov 8, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

Advantage Workstation 4.3 is a review station, which allows easy selection, review, processing and filming of multi-modality DICOM images from a variety of diagnostic imaging systems. When interpreted by a trained physician, filmed or displayed images on the AW monitor may be used as a basis for diagnosis, except in the case of mammography images.

## Device Story

AW 4.3 is a multi-modality review workstation for DICOM images; operates on HP Linux-based hardware with 19" LCD monitors, keyboard, and mouse. Device integrates features from AW 4.1 and Volume Viewer Plus software. Used in clinical settings by physicians to review, post-process, and film diagnostic images. Output displayed on monitors serves as a basis for clinical diagnosis. System facilitates workflow by enabling image selection and manipulation from various imaging modalities. Does not support mammography interpretation.

## Clinical Evidence

No clinical data provided. Safety and effectiveness are supported by software development, validation, and verification processes, and adherence to industry and international standards.

## Technological Characteristics

HP Linux-based workstation (xw4000, xw8000, or xw8200); 19" LCD monitors; DICOM-compliant image processing; multi-modality review software; standard QWERTY keyboard and three-button mouse input.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Advantage Workstation 4.1 ([K020483](/device/K020483.md))
- Volume Viewer Plus ([K041521](/device/K041521.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular border. There are small, decorative flourishes or swirls around the outer edge of the circle. The logo is black and white.

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h).

- Larry A. Kroger, Ph.D. Submitter Senior Regulatory Programs Manager Tel: (262) 544-3894 Fax: (262) 544-4768 GE Healthcare W-400 3000 North Grandview Blvd. Waukesha, WI 53188 USA Date Prepared: July 25, 2005.
## PRODUCT IDENTIFICATION

| Name:                   | Advantage Workstation 4.3                                                             |
|-------------------------|---------------------------------------------------------------------------------------|
| Classification Name:    | PACS per 21 CFR 892-2050                                                              |
| Manufacturer :          | General Electric Medical Systems<br>283, rue de la Minière<br>78533 Buc Cedex, FRANCE |
| Distributor:            | GE Healthcare, P.O. Box 414, Milwaukee, WI 53210                                      |
| Marketed Devices below: | The Advantage Workstation is substantially equivalent to the devices listed           |
| • Model:                | Advantage Workstation 4.1, 510(k) # K020483                                           |
| • Manufacturer:         | General Electric Medical Systems, Buc, France                                         |
| • Model:                | Volume Viewer Plus, 510(k) #K041521                                                   |

- General Electric Medical Systems, Buc, France Manufacturer: .
### Device Description:

AW 4.3 is a multi-modality review workstation. It includes one or two color flat panel monitors for image review. The workstation allows for easy review, post processing and filming of DICOM images from a variety of imaging systems. AW 4.3 combines AW 4.1 features with the 2D image display features of the Volume Viewer Plus software option.

{1}------------------------------------------------

The hardware configuration supported by AW 4.3 includes:

- HP Linux based workstation (xw4000, xw8000 or xw8200) ●
- One or two 19" LCD monitors .
- 1 US QWERTY Keyboard
- l three button mouse .

#### Indications for Use:

Advantage Workstation 4.3 is a review station, which allows easy selection, review, processing and filming of multi-modality DICOM images from a variety of diagnostic imaging systems. When interpreted by a trained physician, filmed or displayed images on the AW monitor may be used as a basis for diagnosis, except in the case of mammography images.

#### Comparison with Predicate:

A W4.3 is substantially equivalent to the predicate devices listed above :

| Device Name               | FDA Clearance Number |
|---------------------------|----------------------|
| Advantage Workstation 4.1 | K020483              |
| Volume Viewer Plus        | K041521              |

#### Adverse Effects on Health:

The potential hazards are identified in a risk management summary (hazard analysis) and are controlled by:

- · Software Development, Validation and Verification Process to ensure performance to specifications, Federal Regulations and user requirements.
- · Adherence to industry and international standards.

#### Conclusions:

Advantage Workstation 4.3 does not result in any new potential safety risks and performs as well as devices currently on the market. GE considers features of the Advantage Workstation 4.3 to be equivalent to those of Advantage Workstation 4.1 and Volume Viewer Plus.

{2}------------------------------------------------

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird with its wings spread, rendered in a thick, black line. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circular pattern around the bird symbol.

NOV - 8 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

GE Healthcare % Mr. Daniel W. Lehtonen Responsible Third Party Official Intertek Testing Service NA., Inc. 70 Codman Hill Road BOXBOROUGH MA 01719

#### K052995 Re.:

Trade/Device Name: Advantage Workstation 4.3 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and Communication system Regulatory Class: II Product Code: LLZ Dated: October 27, 2005 Received: October 28, 2005

Dear Mr. Lehtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registrations, listing of devices, good manufacturing practice, labeling and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advices that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of he Act or any Federal statues and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Nancy C. Hogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are enclosed within a circular border, giving the logo a rounded appearance. The logo is black and white.

# Indications for Use

510(k) Number (if known): حكومك مك Device Name: Advantage Workstation 4.3 Indications for Use:

Advantage Workstation 4.3 is a review station, which allows easy selection, review, processing and filming of multi-modality DICOM images from a variety of diagnostic imaging systems. When interpreted by a trained physician, filmed or displayed images on the AW monitor may be used as a basis for diagnosis, except in the case of mammography images.

Prescription Use _ Over-The-Counter Use × AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy C Brogdon

(Division Sign-Off) Division of Reproductive, Abdominal. and Radiological Devices 510(k) Number

Page __ of _

---

**Source:** [https://fda-staging.innolitics.com/device/K052995](https://fda-staging.innolitics.com/device/K052995)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).