IPUMP PAIN MANAGEMENT SYSTEM, MODEL 2L3107, 2L3107K
K052973 · Baxter Healthcare Corp · FRN · Jan 25, 2007 · General Hospital
Device Facts
Record ID
K052973
Device Name
IPUMP PAIN MANAGEMENT SYSTEM, MODEL 2L3107, 2L3107K
Applicant
Baxter Healthcare Corp
Product Code
FRN · General Hospital
Decision Date
Jan 25, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ipump™ Pain Management System is indicated for the controlled delivery (continuous, intermittent, and continuous plus intermittent) of analgesic, sedative, and anesthetic solutions through clinically acceptable routes of administration including intravenous, subcutaneous, and epidural, and for regional (local) analgesia applications.
Device Story
External infusion pump for delivery of analgesic, sedative, and anesthetic solutions. Modalities: PCA, PCA plus basal, and continuous. Input: user-programmed delivery parameters via keypad. Mechanism: linear peristaltic pumping with DC motor; powered by 9V battery or AC adapter. Features: air sensor, upstream/downstream occlusion detectors, real-time clock, and data logging. Used in clinical settings; operated by healthcare professionals. Output: controlled medication infusion. Benefits: precise, programmable medication delivery for pain management. System includes custom tubing sets and reservoir bags.
Clinical Evidence
Bench testing only. Evidence includes electrical safety, EMC, environmental, performance, and software verification/validation. Biocompatibility confirmed via comparison to predicate materials and manufacturing methods.
Technological Characteristics
Linear peristaltic pump; DC motor; 9V battery or AC power. Features: air sensor, occlusion detectors, LCD, keypad. Connectivity: data transfer capability. Software: system, unit, and integration level testing performed. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-24.
Indications for Use
Indicated for patients requiring controlled delivery of analgesic, sedative, and anesthetic solutions via intravenous, subcutaneous, epidural, or regional routes.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
KOS2973
.
## 510(k) SUMMARY
:
JAN 2 5 2007
| Date | 19 October 2005 |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Owner | Baxter Healthcare Corporation |
| Contact Person | John Manarik<br>1620 Waukegan Road<br>McGaw Park, IL 60085<br>Manager, Global Regulatory Affairs<br>Telephone: (847) 473-6294<br>Fax: (847) 785-5116<br>E-mail: john_manarik@baxter.com |
| Device name | IpumpTM Pain Management System |
| Common name | External Infusion Pump |
| Classification name | Infusion Pump (21 CFR 880.5725, Product Code 80 FRN) |
| Predicate Device | IpumpTM Pain Management System, K993387 cleared on<br>01 November 1999 |
:
:
{1}------------------------------------------------
### Device Description
The Ipump™ Pain Management System is an external infusion pump offering three different modalities for the delivery of medications to a patient. The modalities include Patient Controlled Analgesia (PCA) mode, PCA plus basal (continuous) mode, and continuous mode.
PCA mode delivers a bolus of a medication at a fixed rate of 90 mL/hr at patient-controlled time intervals by pressing the PCA button. PCA plus basal (continuous) mode delivers medication at 90 mL/hr during patient-controlled time intervals in conjunction with the desired basal delivery of between 0.1 mL/hr and 19.9 mL/hr between PCA doses. In continuous mode, the pump delivers medication continuously at a rate adjustable between 0.1 mL/hr and 90.0 mL/hr.
The Ipump operates using a linear peristaltic pumping mechanism and a DC motor with either a 9 volt alkaline battery or an AC adapter power source. The pump's features include an air sensor, an upstream occlusion detector and a down stream occlusion detector. The disposables for the Ipump™ Pain Management System consist of numerous custom tubing sets and flexible non-vented reservoir bags of up to 500 ml.
#### Statement of Intended Use
The Ipump™ Pain Management System is indicated for the controlled delivery (continuous, intermittent, and continuous plus intermittent) of analgesic, sedative, and anesthetic solutions through clinically acceptable routes of administration including intravenous, subcutaneous, and epidural, and for regional (local) analgesia applications.
#### Technological Characteristics
The subject and predicate devices are similar in operating principle, material composition, energy sources, environmental specifications, performance, dimensional specifications and packaging. A summary of the technological characteristics for the subject and predicate devices follows in Table 1.
KoS2973
{2}------------------------------------------------
| Functional Features | Predicate Ipump | Subject Ipump |
|--------------------------------------------------|-----------------|---------------|
| Power Sources | | |
| 9 volt battery | included | included |
| 100-120 VAC, 60 Hz | included | included |
| 220-240 VAC, 50 Hz | | |
| Key Features: | | |
| Air sensor | included | included |
| Preventative maintenance alert | included | included |
| Multilanguage interface | included | included |
| Detailed history display and printout capability | included | included |
| Ability to transfer data to another Ipump | included | included |
| Upstream and downstream occlusion detectors | included | included |
| Programmable limits for PCA doses | included | included |
| Record Management: | | |
| Tracking of programming, time and history | included | included |
| Data retained in memory after pump is off | included | included |
| Real time clock | included | included |
| Security | | |
| Key only | included | included |
| Code only | included | included |
| Key plus code | included | included |
| Setting up of the Pump | | |
| PCA cord | included | included |
| May be IV pole mounted | included | included |
| Pumping Mechanism: | | |
| DC motor | included | included |
| Linear peristaltic motion | included | included |
| Display Technology | | |
| Liquid crystal display | included | included |
| Keypad | | |
| Keypad with action keys | included | included |
Table 1. Ipump™ Pain Management System Technological Characteristics Comparison
The technological characteristics of the subject Ipump™ Pain Management System do not raise new questions of safety or effectiveness.
Baxter Confidential
20
Ko52973
{3}------------------------------------------------
| Assessment of<br>Non-clinical Data | A summary and assessment of the non-clinical performance data<br>relied upon to support substantial equivalence follows: |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Electrical safety: Ipump was evaluated against the applicable<br>requirements contained within IEC 60601-1 with successful results. |
| | Electromagnetic Compatibility (EMC): The Ipump characteristics<br>were evaluated against the applicable requirements contained within<br>IEC 60601-1-2 second edition, augmented by the IEC 60601-2-24<br>clause 36 requirements, with sucessful results. |
| | The Ipump performance during environmental testing, such as the |
| | The performance characteristics of the Ipump were evaluated<br>against the applicable requirements contained within<br>IEC 60601-2-24 with successful results. |
| | Software testing was performed at the system level, unit level and<br>integration level to demonstrate that the design changes satisfy the<br>various Ipump™ Pain Management System requirements.<br>Regression testing was performed to demonstrate that the design<br>changes have not adversely affected those parts of the system not<br>changed. The data recorded from the software testing has been<br>determined to meet its acceptance criteria. |
| | A comparison of the predicate device with the subject Ipump shows<br>that the same components come into direct and indirect contact with<br>the patient. These components are processed by the same<br>manufacturing methods, and are of the same patient contact duration<br>and material type. Biocompatibility requirements are therefore<br>satisfied. |
| | The data supports the subject Ipump has shown equivalent<br>characteristics in its electrical safety, EMC, environmental,<br>performance, software and biocompatibility testing. |
| Conclusion | Baxter Healthcare Corporation concludes that the results obtained<br>during non-clinical performance testing for Ipump™ Pain<br>Management System supports a determination of substantial<br>equivalence. |
:
21
Kos 2973
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its body and wings. The eagle is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES, U.S.A." in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Baxter Healthcare Corporation C/O Ms. Barbara K. Barbeau Senior Director, Global Regulatory Affairs Global Regulatory Affairs 1620 Waukegan Road McGaw Park, Illinois 60085-6730
14 2 5 1007
Re: K052973
Trade/Device Name: Ipump™ Pain Management System Regulation Number: 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: November 13, 2006 Received: November 15, 2006
Dear Ms. Barbeau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements
{5}------------------------------------------------
Page 2 -- Ms. Barbeau
of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chih-Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): ____ K 052973
Device Name: Jpump™ Pain Management System
Indications For Use:
The Ipump™ Pain Management system is indicated for the controlled delivery (continuous, intermittent, and continuous plus intermittent) of analgesic, sedative, and anesthetic solutions through clinically acceptable routes of administration including intravenous, subcutaneous, and epidural, and for regional (local) analgesia applications.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND / OR
Over-the-Counter Use (21 CFR 807 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
K452973
KoS2973
Baxter Confidential
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.